EC-type examination certificate Nederlands Meetinstituut Number T10071 revision 1 Project number 710466 Page 1 of 2

Issued by

NMi Certin B.V. Hugo de Grootplein 1 3314 EG Dordrecht The Netherlands Notified Body number 0122

In accordance with

The “Metrologiewet” (Stb. 2006, 137) as Dutch implementation of Directive 2004/22/EC on measuring instruments (MID).

Manufacturer

Emerson Process Management Flow B.V. Neonstraat 1 6718 WX Ede The Netherlands

In respect of

An interruptible or non-interruptible measuring instrument (measuring system for liquids other than water) Designation : MMI-MID 001 Sensor : Micro Motion CMFXXX (see clause 1.2), DS600, F100 Electronics : Micro Motion model 700; 800 ; 1700 ; 2500; 2700; 3500 and 3700 Accuracy class : 0.3, resp. 0.5, resp. 1.0, resp. 2.5; Environment class : Depending on the composition of the measuring instrument. See the belonging Evaluation Certificate of each component for the classification of the environmental classes. Temperature range liquid : measuring mass: -200 °C / +200 °C for CMF200/300/400/HC3 and DS600; -10 °C / +50 °C for CMF025/050/100; measuring volume and density: -10 °C / +50 °C Temperature range ambient : Depending on the composition of the measuring instrument. See the belonging Evaluation Certificate of each component for the approved temperature range. Liquids : Oil and oil products, alcohol, chemicals, potable liquids, liquefied gasses under pressure and cryogenic liquids with densities between 400 and 2000 kg/m3

Characteristics

In the description number T10071 revision 1 further characteristics are described. Valid until

25 April 2017

NMi Certin BV Hugo de Grootplein 1 3314 EG Dordrecht PO Box 394 3300 AJ Dordrecht, NL T +31 78 6332332 F +31 78 6332309 [email protected] www.nmi.nl

Parties concerned can lodge objection against this decision, within six weeks after the date of submission, to the general manager of NMi (see “Regulation objection and appeal against decisions of NMi”)

This document is issued under the provision that no liability is accepted and that the applicant shall indemnify third-party liability. Reproduction of the complete document is permitted.

Description Nederlands Meetinstituut

1

Number T10071 revision 1 Project number 710466 Page 1 of 5

General information of the measuring system All properties of the measuring system, whether mentioned or not, shall not be in conflict with the legislation. This revision 1 is issued because: - The addition of the Model 800 Enhanced Core Processor. - The addition of the model 3500 and 3700 flow transmitter. - The addition of the CMFHC3 sensor. - The addition of the reference to the verification procedure for measuring system and meter. - The addition of the different flow computers. - The addition of pressure and temperature transmitters. - Changes in the construction of the measuring system, such as two meters in series and by-pass over the meter under restricted conditions.

1.1

Essential parts Manufacturer

Type(s)

Evaluation Certificate

Remarks

Measurement sensor Micro Motion

Rosemount

CMF025, CMF050, CMF100, CMF200, CMF300 ,CMF400, CMFHC3 and DS600

TC7056

Flow sensor

F100 3051S series 3144P series

TC7050 TC7457 TC7458

Flow sensor Pressure transmitter Temperature transmitter

Calculator/conversion device Micro Motion

700; 800; 1700; 2500;2700; 3500 and 3700

TC7057

Flow computer Emerson Process Management

OMNI Flow Computers, Inc.

Contrec Enraf

FloBoss S600

TC7470

OMNI 6000 / 3000

TC7375

Contrec 1010

TC 7348

For printing and data storage purposes. For more functionalities see the Evaluation Certificate of the involved flow computer.

Description Nederlands Meetinstituut

1.2

Number T10071 revision 1 Project number 710466 Page 2 of 5

Essential characteristics Beside the characteristics as mentioned in EC-type examination T10071 the measuring system has the following essential characteristics: Mass measurement

Sensor type CMFxxxy (xxx represents a number relating to the diameter of the sensor) and DS600 CMF025 y)

CMF050 y)

CMF100 y)

CMF200 y)

CMF300 y)

CMF400 M

CMF400 y)

1)

2)

DS600 S

CMFHC3

F100

Maximum Qmax [kg/min] for all classes

36

110

450

1450

4500

6800

6800

10.800

22.000

272

Minimium Qmin [kg/min], for accuracy class 0.5, 1.0 and 2.5

0,23

1,36

5,7

18

57

680

340

570

1134

18

Minimum Qmin [kg/min], for accuracy class 0.3

0,46

2,72

11,4

36

114

680

680

1140

2268

36

Minimum measured quantity sensor [kg]

0,5

5

10

20

200

500

500

1000

1000

10

6

12

25

50

80

100

100

150

200

25

CMFHC3

F100

Diameter in- and outlet [mm]

Volume measurement

Sensor type CMFxxxy (xxx represents a number relating to the diameter of the sensor) and DS600 CMF025 y)

CMF050 y)

CMF100 y)

CMF200 y)

CMF300 y)

CMF400 y)

CMF400 y)

1)

2)

DS600 S

Maximum Qmax [L/min], for all accuracy classes

Qmax/

min

--

Minimium Qmin [L/min], for accuracy class 0.5, 1.0 and 2.5

Qmin/

max

--

Minimum measured quantity sensor [L]

mmq/

max

--

Remarks: For the essential characteristics of the flow sensor see belonging Evaluation certificate. The Qmax and Qmin of the measuring system in kg/min have to be defined from the design of the measuring system (pump capacity and pressure losses) and shall not be in conflict with the characteristics as stated in the Evaluation Certificate belonging to the sensor. The minimum ratio of Qmax and Qmin shall be at least 4.

Description Nederlands Meetinstituut

Number T10071 revision 1 Project number 710466 Page 3 of 5

The minimum measured quantity (in kg and l) is the biggest of the following three: 1. minimum measured quantity as mentioned in table above; 2. one minute delivery at a flow rate, equal to Qmin of the measuring system; 3. 100 * the additional effect of the pipework between the meter and transfer point due to variations in temperature, equal to 10°C for exposed pipes and 2°C for insulated or underground pipes.

1.3

Essential shapes

1.3.1 Configuration Typical arrangement of the interruptible measuring system see drawing no. T10071/1-1. Typical arrangement of the non-interruptible measuring system see drawing no. T10071/1-2. 1.3.2 Inscriptions - Nameplate The name plate contains at least, clearly visible, the following information: The type approval mark no. T10071; Manufacturer’s identification mark or trademark Designation Serial number and year of manufacture Qmin - Qmax Pmax Liquid(s) to be measured Temperature range ambient CE marking Accuracy class of the flow sensor Environmental classes - See for the inscriptions of the sensor, flow transmitter, pressure transmitter, temperature transmitter and flow computer the belonging Evaluation Certificates. Note: In general it is preferred that the name plate layout is such that Safety, Metrological and common markings are logically arranged to cover also other relevant information, coming from other legislation. An example of the nameplate is presented in document no T10071/1-3. - The minimum measured quantity, in kg or L. 1.3.3

Sealings See chapter 3 of this Description.

Description Nederlands Meetinstituut

1.4

Number T10071 revision 1 Project number 710466 Page 4 of 5

Conditional parts The measuring system contains also the following conditional parts: - Temperature sensor (see Evaluation Certificate of the sensor) - Pressure sensor (see Evaluation Certificate of the sensor) - Back-up power supply (for non-interruptible systems)

1.5

Conditional characteristics -

1.6

Conditional shapes -

1.7

The construction shall be such that no air pockets remain after the air is released by the vent-off valves. Diameter of the valves and piping.

Non essential parts -

2

Before a delivery is started, the system shall be free of air. By-pass of a sensor (optional) See T10071/1-4 for prescribed conditions for by-passing a sensor.

Pump, pipe work and connections. Block-in valve(s) Vent-off valve(s) Filter/strainer

Conditions for approval - Verification procedure of the system For the initial verifications the NMi procedure C-SP-HW-281 is applied with the title 'Procedure C-SP-HW-281 for the MID conformity assessment for the Micro Motion flow meter when used for custody transfer gas applications (annex MI-002) and liquid applications (annex MI-005).’ - Verification procedure of the meter For the verification of the meter itself, the NMi procedure C-SP-HW-280 is applied with the title 'Procedure C-SP-HW-280 for the MID conformity assessment for the Micro Motion Flow meter when used for custody transfer in gas applications (annex MI-002) and liquid applications (annex MI-005)'. - The initial verification of the meter is based on: a water calibration, which includes: - a mass flow test - a zero mass flow verification - if applicable a density test In the field a zero mass flow verification and if applicable a density verification.

Description Nederlands Meetinstituut

Number T10071 revision 1 Project number 710466 Page 5 of 5

This procedure is justified because of the fact that tests have proven that the mass and density accuracy on water is representative for mass and density accuracy on other liquids. Note: The initial verification method can also be applied for the subsequent verifications. 3

Sealings The nameplate of the measuring system is sealed. For the sealing of the sensor, flow transmitter and flow computer the belonging Evaluation Certificates. The securing component has to bear either: - A mark of the manufacturer laid down in an approved quality system by a Notified Body, or; - A mark of a Notified Body.