CLOUD POINT PROCESS ANALYZER
PROCESS ANALYZERS
CLOUD POINT PROCESS ANALYZER Norms and standards • • •
ASTM D-2500 DIN 51597 (old national) IP 219
• •
ISO 3015 (new international) EN 23015
These norms and standards set the operating limits for cloud point measurement for mineral oil products which are clear at a path length of 40 mm (for IP and ASTM 1 ½" = 38 mm) and have a cloud point below 49°C (120°F).
Application The cloud point process analyzer (CPA) is an analyzer for the fully automatic determination of the cloud point (CP) of liquid, transparent mineral oil products. The CPA operates online. It serves to monitor / maintain product quality for the in-spec production of mixtures such as diesel fuel and heating oil.
Measurement principle The product sample is cooled under defined conditions and its turbidity is observed. The temperature at which a cloud of paraffin crystals first appears, is known as the CP. The low- temperature behavior of liquid mineral oil products is very important for their usage, as it affects their ability to be pumped, transported and filtered. The CP is used as an internal process control parameter, or to test standard products offered for sale. The CPA uses a photometric measurement principle. The CP is determined by light scattering (diffuse reflection) caused by finely dispersed colloids which are formed when the product sample in the measuring cell is cooled. The degree of crystallization of the paraffins in the product correlates with the "cold energy" delivered to the sample. The crystals are finely divided particles which scatter and absorb light or other electromagnetic radiation. The scattering depends on the number of particles, the difference in refractive index between the liquid and solids, and the wavelength of the light. The measured CP is the temperature at which a set limit for the scattering is exceeded. The complete analysis procedure is controlled, monitored and visualized by the PACS (Process Analyzer Control System) software. It offers a user interface for local analyzer operation. At the end of the analysis cycle, in addition to the standard analog 4 - 20 mA CP signal, a digital output signal can be generated (programmable). During the analysis, an optional, galvanically separated 4 - 20 mA signal can also be used. As an alternative, a MODBUS interface is available as an option, which also allows a direct control of the anlalyzer from the DCS. In case remote access to the CPA is demanded, it can be equipped with a remote access interface (e.g. Modem, ISDN) for service and maintenance purposes.
PROCESS ANALYZERS Status / 10.07.2003
Seite 2 / 7
CLOUD POINT PROCESS ANALYZER Cloud Point Process Analyzer Analyzer type
CPA - 4
Method
DIN 3015, ASTM D-2500, IP 219 / 82, DIN ISO 3015
Measurement range
Please specify, ranges on request
Reproducibility
≤ DIN EN / ASTM
Repeatability
≤ DIN EN / ASTM
Measurement cycle
Discontinuous 4-8 min. depending on measuring point
Ambient temperature
5 - 40°C
Ambient humidity
Max. 70%, non - corrosive
Sample at analyzer inlet General conditions
Liquid (≤ 50 cSt), cooled, filtered (≤ 10 µm), dry (humidity max. 2000 ppm)
Flow rate
20 - 40 l/h
Pressure
100 - 300 kPa
Inlet temperature
At least 15°C above expected CP
Outlet / vent
Open to atmosphere
Utilities Cooling water
Clean cold water, temperature depending on measuring point 20 – 60 l/h / 100 – 300 kPa
Power supply
See Specification of purged electronic housing
Instrument air (inlet)
See Specification of purged electronic housing
Electrical signals Ouput signals
1 x 4–20 mA, 800 Ω out; (standard),additional outputs on request
System alarm
Digital output / potential-free
Ready contact
Digital output / potential-free
PROCESS ANALYZERS Status / 10.07.2003
Seite 3 / 7
CLOUD POINT PROCESS ANALYZER Options Validation request
Digital input
Analyzer reset
Digital input
Input signal
1 x 4–20 mA, 200 Ω in; (galvanically separated)
MODBUS
RS485 / RS422 or fibre optical interface
Remote access interface
Modem (V 90 analog) or ISDN
Closed loop chiller
If no suitable cooling water is available
Chiller alarm
Digital output / potential-free
Communication Display
Colour LC display 800 x 600 pixel
Keyboard
Virtual keyboard, operated via mouse in front door
Software
Windows 2000, PACS software, optional Remote Control Software
Explosion protection Protection type
II 2G EEx - pd II B T4 or optional: II 2 G EEx – pd II B +H2 T4
Test certificates
TÜV 02 ATEX 1846
Standard-connections Tube fittings
6 mm / 12 mm metric SWAGELOK
Cable entries
M20 x 1,5 / M25 x 1,5
Weight
Approx. 250 kg
Dimensions (L x D x H) in mm
1140 x 710 x 1900 (see dimensional drawing)
Important Notice: Analyzer is subject to continuous product improvement, specifications may be subject to change without notice.
PROCESS ANALYZERS Status / 10.07.2003
Seite 4 / 7
CLOUD POINT PROCESS ANALYZER Specification of the air purged electronic housing System type
PAGS 96-1 (Process Analyzer Housing System)
Version number
96 0200
Protection type
II 2 G EEx p II T4
Test certification number
TÜV 96 ATEX 1132X
EC – identification number
0032
Housing volume
Approx. 200 dm
Dimensions (L x D x H) in mm
515 x 470 x 800
Protection class
IP65 (with a Vortex chiller IP54)
Power supply
To be specified by user
Power consumption
Approx. 2 kW (analyzer approx. 600 W / chiller approx. 1000W)
Maximum power loss
515 W
Ignition prevention gas
Instrument air
3
100 - 150 kPa, constant (for EEx p system) Instrument air (inlet)
300 - 350 kPa, constant (for valves) Humidity class 1 / dew point ≤ - 30°C / oil-free
Housing overpressure
300 - 400 Pa (operating pressure)
Shut down pressure
≤ 80 Pa
Instrument air consumption
Min.1,4 Nm per flushing cycle (7 x housing volume)
3
Important Notice: Analyzer is subject to continuous product improvement, specifications may be subject to change without notice.
PROCESS ANALYZERS Status / 10.07.2003
Seite 5 / 7
CLOUD POINT PROCESS ANALYZER
Figure 1:
Main panel of process software
Figure 2:
Status of analog and digital In- and Outputs
PROCESS ANALYZERS Status / 10.07.2003
Seite 6 / 7
CLOUD POINT PROCESS ANALYZER
Figure 3:
Layout drawing of the analyzer (Example; incl. optional sample conditioning system and chiller)
BARTEC BENKE GmbH Borsigstrasse 10 D - 21465 Reinbek / Hamburg Status / 10.07.2003
PROCESS ANALYZERS
Phone +49 (0)40 727 03 - 0 Fax +49 (0)40 727 03 - 228 Internet www.bartec-benke.de Seite 7 / 7