Automated Pensky Martens Closed Cup Flash Point Analyzer HFP 339

Standard Quotation No. Automated Pensky Martens Closed Cup Flash Point Analyzer HFP 339 Fuels have different components from light to heavy hydrocarb...
Author: Belinda Sanders
58 downloads 4 Views 521KB Size
Standard Quotation No.

Automated Pensky Martens Closed Cup Flash Point Analyzer HFP 339 Fuels have different components from light to heavy hydrocarbons. The flashpoint method gives information about the light components by their tendency to flash. It further gives important information on safety and shipping regulations. The HFP 339 is a fully automated instrument which perfectly mimics the manual process as described in the following methods: ASTM D 93 A&B, ISO 2719 A&B, EN ISO 2719 A&B, DIN EN ISO 2719 A&B, NF EN ISO 2719 A&B, IP 34 A&B, JIS K 2265, DIN 51758 The Instrument has a built in microprocessor system which controls all functions of the unit. Operation of the instrument is done via easy to read VF-Display and a solvent resistant alpha numeric keypad. This easy to use man machine interface lets the operator quickly initiate standard flash point testing, change test parameters to run customized protocols, run the systems diagnostics, and recall or print test results. Main Features ¾ Flash point determinations may be run using standard test method or customized test protocols ¾ 6 Standard programs and 3 user-defined test programs. ¾ Tests for samples with unknown flash points. ¾ Automatically corrects results to barometric pressure. ¾ Memorizes the last 9 Flash Point results ¾ Built-in safety functions ¾ Built in fire detection system with audible alarm and external alarm relay contact. ¾ Thermal flash point detection eliminates any interference from water or silicone containing samples. ¾ Electric ignition (standard) and Gas ignition (option)

HFP339 Technical Offer

22.8.2006

Page 1 of 15

Technical details Standards

ASTM D 93 Methods A&B, ISO 2719 A&B,

EN ISO 2719 A&B,

DIN EN ISO 2719 A&B, NF EN ISO 2719 A&B, IP 34 A&B, JIS K 2265, DIN 51758 Configuration

Stand alone automated Pensky Martens flashpoint analyzer HFP 339

Sample Temperature Measurement

Glass Pt 100 probe, Metal Pt 100 available as a plug in accessory. Temperature range:

+0 … +400°C, resolution 0.1°C +32 … +752°F, resolution 0.1°F

Each measurement program can be individually programmed in °C or °F. The system perfectly mimics ASTM 9C/9F, 88C/88F and 10C/10F-standardthermometers. Probe can be calibrated on the display Automatic calibrating of the electronic circuit with integrated, high precision calibration resistors eliminates temperature and long term drift of the temperature measuring unit. Cabinet/Chassis

All steel, baked epoxy resin coated: electromagnetic isolation, solvent resistant. Width only 25 cm (9.85”)

Sample Heating System

950 W, conforms to CE safety specification. Air cooled at the end of the test.

Heating Rate Control

Automatic proven algorithm controls temperature of heater block and sample.

Heating Rate

Per appropriate method or user-defined programmable from 0.5 to 14 °C/min for individual measurement programs.

Sample Stirrer

Stepper motor operated stirring. Per appropriate method or user-defined programmable from 20 to 255 revolutions/min for individual measurement programs.

Igniter

Reliable electric igniter comes as a standard. The unit monitors the correct function of the electric igniter. Gas ignition is optional available. The Gas flame is automatically ignited by an auxiliary electric igniter.

Ignition Interval

Per appropriate method or user-defined programmable in 0.5 to 5 °C steps

Barometric Pressure

Built in barometric pressure sensor makes automatic pressure correction of the flash point temperature data. Indicated as mbar, hPa, kPa, mmHg or Torr.

Flash Detection System

Thermal detection system to eliminate interference from water or silicone containing samples.

Display

A 2 line 20 character VF-Display gives information about the status of the HFP 339. The HFP 339 lets you quickly initiate standard flashpoint point tests. You are also able to easily calibrate the instrument, generate methods and review past test results.

Keyboard

Operation of the instrument is done via a 25 button solvent resistant foil keypad.

Result Documentation

Test results are displayed on the VF-Display and can also be output automatically or on demand to an optional printer or computer system (LIM).

Printer

HFP339 Technical Offer

A typical report includes the Sample ID, Date and Time, Measurement Program, expected Flash Point, measured and corrected Flash Point and barometric pressure. Any printer with CENTRONICS® interface and ASCII character set may be

22.8.2006

Page 2 of 15

Printer and LIMS crosslinking of several units

used. To save space a small footprint needle ticket printer is recommended (see accessories). In order to connect several units to one PC with a Laboratory Information System (LIMS) or to a printer, the units can be linked together as a Herzog Token Ring network (HTR). It is possible to link up to 16 units. Connect with cable (626-082) to the socket „HTR“ at the back of the instrument.

Communications

Serial transmission of all data via RS232 and HTR (Herzog Token Ring) protocol.

Results memory

The instrument stores the last 9 test results with Sample number and flash point temperature.

Test Programs

6 standard programs and 3 user defined programs

Safety Features Fire Extinguisher Fire Safety — A built-in fire sensor located next to the lid cover alerts instantly to flames outside the flash cup. A potential-free alarm relay contact is also available to link the HFP to a fire suppression system or alarm bell. Safety pre test By doing safety pre dip ins the unit can detect high volatile contamination in normally high flash point samples and avoid a risk of fire Alarm Functions Automatic detection of method or safety violations. User selects test termination or alarm message Password Protection

Password protection is available to prevent inadvertent changing of programmed test parameters and calibration.

Calibration and Diagnostics

The instrument incorporates a comprehensive diagnostic program which allows you to check every key component and assembly. In addition, it features a selfcalibrating sample temperature measuring system that automatically compensates for drift. Any probe deviation can be corrected on screen calibration of resistance thermometer.

Operating Requirements Electrical 115V or 230VAC ±10% switchable; 50-60Hz ; 1100 W Ambient Conditions Temperature: max. 10° to 35°C; recomm. 15° to 25°C Humidity: 80% rel. at 35°C Storage temperature: -15 to +55°C Dimensions

25 cm (9.85”) wide, 56 cm (22”) deep, 51 cm (20.1”) tall

Weight

25kg, (55 lb)

HFP339 Technical Offer

22.8.2006

Page 3 of 15

Ordering information: Order No.

Description

Picture

03391-000-01

HFP 339 Pensky Martens 230V, el. igniter HERZOG - automatic flash point tester acc. to Pensky Martens. Compact single unit with VFDisplay and alphanumeric keypad:

Price

Electric ignition 230 V, 50-60 Hz. 1 x 101-052

Power cord with Continental European IEC/DIN plug.

Accessories included:

03391-000-02

1 x 820-268

Instruction manual

1 x 302-026

Test cup Pensky Martens

1 x 636-069

Pt 100 for sample temperatures up to +400°C

1 x 636-095

Flashpoint sensor with plug

1 x 103-060

Test cover, PM, for electric igniter and thermal flash detection device.

2 x 620-113

Electric igniters

1 x 628-041

Thermo fuses for fire sensor

1 x 101-029

Cover-handle PM

1 x 326-070

Gauge as igniter simulator

1 x 326-072

Igniter dip in level adjustment tool

1 x 402-120

Gauge for adjustment of dip in level

1 x 101-029

Cover handle for HFP 339

1 x 307-006

Drip cup, Pensky Martens

1 x 303-001

Coupling for stirrer, HFP 339

1 x 412-221

Set of hexagonal wrenches: 1.5; 2; 2.5; 3; 4; 5mm.

HFP 339 Pensky Martens 115V, el. igniter HERZOG - automatic flash point tester acc. to Pensky Martens. Compact single unit with VFDisplay and alphanumeric keypad: Electric ignition 115 V, 50-60 Hz 1 x 411-021 Power cord with US NEMA 5-15 plug. Accessories included as for 03391-000-01

HFP339 Technical Offer

22.8.2006

Page 4 of 15

Order No.

Description

Picture

03391-000-03

HFP 339 Pensky Martens 230V, gas igniter HERZOG - automatic flash point tester acc. to Pensky Martens. Compact single unit with VFDisplay and alphanumeric keypad:

Price

Gas ignition 230 V, 50-60 Hz 1 x 101-052

Power cord with Continental European IEC/DIN plug

Accessories included: 1 x 820-268

Instruction manual

1 x 302-026

Test cup Pensky Martens

1 x 636-069

Pt 100 for sample temperatures up to +400°C

1 x 636-095

Flashpoint sensor with plug

1 x 103-061

Test cover, PM, for gas igniter

2 x 620-114

Electric igniters with closed filament

1 x 120-052

Straight gas connection

1 x 628-041

Thermo fuses for fire sensor

1 x 101-029

Cover handle for HFP 339

1 x 326-070

Gauge as igniter simulator

1 x 326-072

Igniter dip in level adjustment tool

1 x 402-120

Gauge for adjustment of dip in level

1 x 307-006

Drip cup, Pensky Martens

1 x 303-001

Coupling for stirrer, HFP 339

1 x 412-221 Set of hexagonal wrenches: 1.5; 2; 2.5; 3; 4; 5mm. 03391-000-04

HFP 339 Pensky Martens 115V, gas igniter HERZOG - automatic flash point tester acc. to Pensky Martens. Compact single unit with VFDisplay and alphanumeric keypad: Gas ignition 115 V, 50-60Hz 1 x 411-021

Power cord with US NEMA 5-15 plug

Accessories included as 03391-000-03

HFP339 Technical Offer

22.8.2006

Page 5 of 15

Accessories: Order No.

Description

Picture

110-191

Gas supply accessory set for HFP 339

Price

Ready to connect an external low pressure gas source and supply the gas ignition test cover with gas. In addition need following Parts! 103-061

Test cover, PM, for gas igniter

620-114

Electric igniters with closed filament

626-260

MV-connection cable

636-103

Stainless steel sample thermometer probe up to +400°C HFP 339

101-214

Matrix dot printer, 40 columns, using endless paper roll of 76mm width, Centronics interface, Dimensions: 156 mm (W) x 248 mm (D) x 132 mm (H), 100-230 VAC, 24W, Power cord with Continental European IEC/DIN plug. Centronics® Printer cable not included

101-212

Ink ribbon for Matrix dot printer 101-214

101-185

Paper rolls (10 pcs.) for Matrix dot printer printer 101-214 76 mm x ∅60 x ∅12mm

101-137

Ink jet printer with English and German manual, ASCII compatible, for 100 – 230VAC, Power cord with Continental European IEC/DIN plug.

No Picture

Centronics® Printer cable not included 626-029

Printer connection cable (Centronics)

HFP339 Technical Offer

22.8.2006

Page 6 of 15

Order No.

Description

101-145

Pressure calibration instrument. U-tube system that creates a gauge pressure of 0 – 40 mbar

302-029

Special low volume (20 ml) sample test cup for Pensky Martens, HFP 339

103-066

Special cover for low sample volume (20 ml) for Pensky Martens, HFP 339 electric ignition

302-027

Cooling test cup Pensky Martens for flash points below +40°C (cooled with dry-ice)

302-028

Cooling test cup Pensky Martens, for flash points below +40°C ( with built on cooling coil for cooling with external chiller)

HFP339 Technical Offer

Picture

22.8.2006

Price

Page 7 of 15

Order No.

Description

Picture

101-193

Pt100 probe simulator box 0 to +400°C with 6 fixed temperatures, accuracy better 0,2%, including 4-wire connecting cable 626-122 with connector used for Herzog Pt100 probes

01000-329-51

Certified Reference Material, ASTM D93 Flash point Pensky Martens appr. +65°C, with certificate, 255 ml diesel fuel

01000-329-52

Certified Reference Material (CRM), Flash point Pensky Martens appr. +95,3°C, approx. 255 ml

01000-329-53

Certified Reference Material, ASTM D93 Flash point Pensky Martens appr. +215°C, with certificate, 255 ml base oil

634-198

Plug, 6-pol. mating halve for fire alarm outlet HFP 339

626-261

Cable for alarm outlet.

326-098

Angled gas connection without threat for

Price

HFP 339

226-205

Cup holder Additional Holder for storage the cup on the left side of the unit.

HFP339 Technical Offer

22.8.2006

Page 8 of 15

Recommended spare parts for appr. 2 years of operation Order No. Description

Qty

302-026

Pensky Martens Test cup, brass

2

303-001

Coupling for stirrer HFP 339

2

618-036

Stepper motor f. stirring HFP 339

1

618-133

Stepper motor f. dip in HFP 339

1

636-125

Temperature sensor for heating block HFP339

1

645-070

Bottom heater 450 W/ HFP 339

2

645-032

Wall heater 500W/MP HFP 339

1

620-113

Electric igniter with open filament for HFP 36x family 10 and HFP 339

HFP339 Technical Offer

22.8.2006

Picture

Price

Page 9 of 15

Order No. Description

Qty

628-007

Fuse 10 A – MT

10

628-021

Fuse 2 A – MT

10

628-023

Fuse 6.3 A – MT

10

628-025

Fuse 4 A – MT

10

628-041

Fire sensor

5

636-069

Sample temperature probe up to 400 °C HFP 339

4

636-095

Flash sensor for water containing samples, with plug 1

613-055

Solid state relay 4A 230V

HFP339 Technical Offer

Picture

Price

1

22.8.2006

Page 10 of 15

Other spare parts: Order No. Description

Picture

101-029

Cover-handle PM HFP 339

103-060

Test cover PM for HFP 339 with provisions for thermal flash detection and for electric ignition

103-061

Test cover PM for HFP 339 with provisions for thermal flash detection and for gas ignition

326-070

Gauge as igniter simulator

HFP339 Technical Offer

22.8.2006

Price

Page 11 of 15

Order No. Description

Picture

326-072

Igniter dip in level adjustment tool

307-006

Drip cup, Pensky Martens

402-120

Gauge for adjustment of dip in level

627-098

Blower

645-070

Bottom-Heater

108-266

Push-pull rod to open the cover shutter

412-209

Cardanic suspension for 108-266

326-210

Driving device with suspension for 108-266

HFP339 Technical Offer

22.8.2006

Price

Page 12 of 15

Order No. Description

Picture

620-114

Electric igniter with closed filament to re-ignite the gas ignition

126-050

Foil keypad

650-551

Temperature measurement P.C.B., with software

607-068

Motor control PCB for dip in system, without EPROM

650-140

EPROM with latest HFP339 software for 607-068

607-215

Motor control PCB for stirrer, without EPROM

650-140

EPROM wit latest HFP339 software for 607-215

607-082

Display

607-257

Keyboard Adapter

HFP339 Technical Offer

22.8.2006

Price

Page 13 of 15

Order No. Description

Picture

607-258

Adapter for Measuring board

607-259

Interface board

607-260

Power supply board

607-252

CPU board

607-211

Pressure sensor and flash detection P.C.B.

617-151

Main switch

619-013

Varistor metallic oxide for mains protection

625-084

Transformer for electric ignition

HFP339 Technical Offer

22.8.2006

Price

Page 14 of 15

Order No. Description 625-126 Transformer 230/115V-24V/8V/HFP339

HFP339 Technical Offer

22.8.2006

Picture

Price

Page 15 of 15