Standard Quotation No.
Automated Pensky Martens Closed Cup Flash Point Analyzer HFP 339 Fuels have different components from light to heavy hydrocarbons. The flashpoint method gives information about the light components by their tendency to flash. It further gives important information on safety and shipping regulations. The HFP 339 is a fully automated instrument which perfectly mimics the manual process as described in the following methods: ASTM D 93 A&B, ISO 2719 A&B, EN ISO 2719 A&B, DIN EN ISO 2719 A&B, NF EN ISO 2719 A&B, IP 34 A&B, JIS K 2265, DIN 51758 The Instrument has a built in microprocessor system which controls all functions of the unit. Operation of the instrument is done via easy to read VF-Display and a solvent resistant alpha numeric keypad. This easy to use man machine interface lets the operator quickly initiate standard flash point testing, change test parameters to run customized protocols, run the systems diagnostics, and recall or print test results. Main Features ¾ Flash point determinations may be run using standard test method or customized test protocols ¾ 6 Standard programs and 3 user-defined test programs. ¾ Tests for samples with unknown flash points. ¾ Automatically corrects results to barometric pressure. ¾ Memorizes the last 9 Flash Point results ¾ Built-in safety functions ¾ Built in fire detection system with audible alarm and external alarm relay contact. ¾ Thermal flash point detection eliminates any interference from water or silicone containing samples. ¾ Electric ignition (standard) and Gas ignition (option)
HFP339 Technical Offer
22.8.2006
Page 1 of 15
Technical details Standards
ASTM D 93 Methods A&B, ISO 2719 A&B,
EN ISO 2719 A&B,
DIN EN ISO 2719 A&B, NF EN ISO 2719 A&B, IP 34 A&B, JIS K 2265, DIN 51758 Configuration
Stand alone automated Pensky Martens flashpoint analyzer HFP 339
Sample Temperature Measurement
Glass Pt 100 probe, Metal Pt 100 available as a plug in accessory. Temperature range:
+0 … +400°C, resolution 0.1°C +32 … +752°F, resolution 0.1°F
Each measurement program can be individually programmed in °C or °F. The system perfectly mimics ASTM 9C/9F, 88C/88F and 10C/10F-standardthermometers. Probe can be calibrated on the display Automatic calibrating of the electronic circuit with integrated, high precision calibration resistors eliminates temperature and long term drift of the temperature measuring unit. Cabinet/Chassis
All steel, baked epoxy resin coated: electromagnetic isolation, solvent resistant. Width only 25 cm (9.85”)
Sample Heating System
950 W, conforms to CE safety specification. Air cooled at the end of the test.
Heating Rate Control
Automatic proven algorithm controls temperature of heater block and sample.
Heating Rate
Per appropriate method or user-defined programmable from 0.5 to 14 °C/min for individual measurement programs.
Sample Stirrer
Stepper motor operated stirring. Per appropriate method or user-defined programmable from 20 to 255 revolutions/min for individual measurement programs.
Igniter
Reliable electric igniter comes as a standard. The unit monitors the correct function of the electric igniter. Gas ignition is optional available. The Gas flame is automatically ignited by an auxiliary electric igniter.
Ignition Interval
Per appropriate method or user-defined programmable in 0.5 to 5 °C steps
Barometric Pressure
Built in barometric pressure sensor makes automatic pressure correction of the flash point temperature data. Indicated as mbar, hPa, kPa, mmHg or Torr.
Flash Detection System
Thermal detection system to eliminate interference from water or silicone containing samples.
Display
A 2 line 20 character VF-Display gives information about the status of the HFP 339. The HFP 339 lets you quickly initiate standard flashpoint point tests. You are also able to easily calibrate the instrument, generate methods and review past test results.
Keyboard
Operation of the instrument is done via a 25 button solvent resistant foil keypad.
Result Documentation
Test results are displayed on the VF-Display and can also be output automatically or on demand to an optional printer or computer system (LIM).
Printer
HFP339 Technical Offer
A typical report includes the Sample ID, Date and Time, Measurement Program, expected Flash Point, measured and corrected Flash Point and barometric pressure. Any printer with CENTRONICS® interface and ASCII character set may be
22.8.2006
Page 2 of 15
Printer and LIMS crosslinking of several units
used. To save space a small footprint needle ticket printer is recommended (see accessories). In order to connect several units to one PC with a Laboratory Information System (LIMS) or to a printer, the units can be linked together as a Herzog Token Ring network (HTR). It is possible to link up to 16 units. Connect with cable (626-082) to the socket „HTR“ at the back of the instrument.
Communications
Serial transmission of all data via RS232 and HTR (Herzog Token Ring) protocol.
Results memory
The instrument stores the last 9 test results with Sample number and flash point temperature.
Test Programs
6 standard programs and 3 user defined programs
Safety Features Fire Extinguisher Fire Safety — A built-in fire sensor located next to the lid cover alerts instantly to flames outside the flash cup. A potential-free alarm relay contact is also available to link the HFP to a fire suppression system or alarm bell. Safety pre test By doing safety pre dip ins the unit can detect high volatile contamination in normally high flash point samples and avoid a risk of fire Alarm Functions Automatic detection of method or safety violations. User selects test termination or alarm message Password Protection
Password protection is available to prevent inadvertent changing of programmed test parameters and calibration.
Calibration and Diagnostics
The instrument incorporates a comprehensive diagnostic program which allows you to check every key component and assembly. In addition, it features a selfcalibrating sample temperature measuring system that automatically compensates for drift. Any probe deviation can be corrected on screen calibration of resistance thermometer.
Operating Requirements Electrical 115V or 230VAC ±10% switchable; 50-60Hz ; 1100 W Ambient Conditions Temperature: max. 10° to 35°C; recomm. 15° to 25°C Humidity: 80% rel. at 35°C Storage temperature: -15 to +55°C Dimensions
25 cm (9.85”) wide, 56 cm (22”) deep, 51 cm (20.1”) tall
Weight
25kg, (55 lb)
HFP339 Technical Offer
22.8.2006
Page 3 of 15
Ordering information: Order No.
Description
Picture
03391-000-01
HFP 339 Pensky Martens 230V, el. igniter HERZOG - automatic flash point tester acc. to Pensky Martens. Compact single unit with VFDisplay and alphanumeric keypad:
Price
Electric ignition 230 V, 50-60 Hz. 1 x 101-052
Power cord with Continental European IEC/DIN plug.
Accessories included:
03391-000-02
1 x 820-268
Instruction manual
1 x 302-026
Test cup Pensky Martens
1 x 636-069
Pt 100 for sample temperatures up to +400°C
1 x 636-095
Flashpoint sensor with plug
1 x 103-060
Test cover, PM, for electric igniter and thermal flash detection device.
2 x 620-113
Electric igniters
1 x 628-041
Thermo fuses for fire sensor
1 x 101-029
Cover-handle PM
1 x 326-070
Gauge as igniter simulator
1 x 326-072
Igniter dip in level adjustment tool
1 x 402-120
Gauge for adjustment of dip in level
1 x 101-029
Cover handle for HFP 339
1 x 307-006
Drip cup, Pensky Martens
1 x 303-001
Coupling for stirrer, HFP 339
1 x 412-221
Set of hexagonal wrenches: 1.5; 2; 2.5; 3; 4; 5mm.
HFP 339 Pensky Martens 115V, el. igniter HERZOG - automatic flash point tester acc. to Pensky Martens. Compact single unit with VFDisplay and alphanumeric keypad: Electric ignition 115 V, 50-60 Hz 1 x 411-021 Power cord with US NEMA 5-15 plug. Accessories included as for 03391-000-01
HFP339 Technical Offer
22.8.2006
Page 4 of 15
Order No.
Description
Picture
03391-000-03
HFP 339 Pensky Martens 230V, gas igniter HERZOG - automatic flash point tester acc. to Pensky Martens. Compact single unit with VFDisplay and alphanumeric keypad:
Price
Gas ignition 230 V, 50-60 Hz 1 x 101-052
Power cord with Continental European IEC/DIN plug
Accessories included: 1 x 820-268
Instruction manual
1 x 302-026
Test cup Pensky Martens
1 x 636-069
Pt 100 for sample temperatures up to +400°C
1 x 636-095
Flashpoint sensor with plug
1 x 103-061
Test cover, PM, for gas igniter
2 x 620-114
Electric igniters with closed filament
1 x 120-052
Straight gas connection
1 x 628-041
Thermo fuses for fire sensor
1 x 101-029
Cover handle for HFP 339
1 x 326-070
Gauge as igniter simulator
1 x 326-072
Igniter dip in level adjustment tool
1 x 402-120
Gauge for adjustment of dip in level
1 x 307-006
Drip cup, Pensky Martens
1 x 303-001
Coupling for stirrer, HFP 339
1 x 412-221 Set of hexagonal wrenches: 1.5; 2; 2.5; 3; 4; 5mm. 03391-000-04
HFP 339 Pensky Martens 115V, gas igniter HERZOG - automatic flash point tester acc. to Pensky Martens. Compact single unit with VFDisplay and alphanumeric keypad: Gas ignition 115 V, 50-60Hz 1 x 411-021
Power cord with US NEMA 5-15 plug
Accessories included as 03391-000-03
HFP339 Technical Offer
22.8.2006
Page 5 of 15
Accessories: Order No.
Description
Picture
110-191
Gas supply accessory set for HFP 339
Price
Ready to connect an external low pressure gas source and supply the gas ignition test cover with gas. In addition need following Parts! 103-061
Test cover, PM, for gas igniter
620-114
Electric igniters with closed filament
626-260
MV-connection cable
636-103
Stainless steel sample thermometer probe up to +400°C HFP 339
101-214
Matrix dot printer, 40 columns, using endless paper roll of 76mm width, Centronics interface, Dimensions: 156 mm (W) x 248 mm (D) x 132 mm (H), 100-230 VAC, 24W, Power cord with Continental European IEC/DIN plug. Centronics® Printer cable not included
101-212
Ink ribbon for Matrix dot printer 101-214
101-185
Paper rolls (10 pcs.) for Matrix dot printer printer 101-214 76 mm x ∅60 x ∅12mm
101-137
Ink jet printer with English and German manual, ASCII compatible, for 100 – 230VAC, Power cord with Continental European IEC/DIN plug.
No Picture
Centronics® Printer cable not included 626-029
Printer connection cable (Centronics)
HFP339 Technical Offer
22.8.2006
Page 6 of 15
Order No.
Description
101-145
Pressure calibration instrument. U-tube system that creates a gauge pressure of 0 – 40 mbar
302-029
Special low volume (20 ml) sample test cup for Pensky Martens, HFP 339
103-066
Special cover for low sample volume (20 ml) for Pensky Martens, HFP 339 electric ignition
302-027
Cooling test cup Pensky Martens for flash points below +40°C (cooled with dry-ice)
302-028
Cooling test cup Pensky Martens, for flash points below +40°C ( with built on cooling coil for cooling with external chiller)
HFP339 Technical Offer
Picture
22.8.2006
Price
Page 7 of 15
Order No.
Description
Picture
101-193
Pt100 probe simulator box 0 to +400°C with 6 fixed temperatures, accuracy better 0,2%, including 4-wire connecting cable 626-122 with connector used for Herzog Pt100 probes
01000-329-51
Certified Reference Material, ASTM D93 Flash point Pensky Martens appr. +65°C, with certificate, 255 ml diesel fuel
01000-329-52
Certified Reference Material (CRM), Flash point Pensky Martens appr. +95,3°C, approx. 255 ml
01000-329-53
Certified Reference Material, ASTM D93 Flash point Pensky Martens appr. +215°C, with certificate, 255 ml base oil
634-198
Plug, 6-pol. mating halve for fire alarm outlet HFP 339
626-261
Cable for alarm outlet.
326-098
Angled gas connection without threat for
Price
HFP 339
226-205
Cup holder Additional Holder for storage the cup on the left side of the unit.
HFP339 Technical Offer
22.8.2006
Page 8 of 15
Recommended spare parts for appr. 2 years of operation Order No. Description
Qty
302-026
Pensky Martens Test cup, brass
2
303-001
Coupling for stirrer HFP 339
2
618-036
Stepper motor f. stirring HFP 339
1
618-133
Stepper motor f. dip in HFP 339
1
636-125
Temperature sensor for heating block HFP339
1
645-070
Bottom heater 450 W/ HFP 339
2
645-032
Wall heater 500W/MP HFP 339
1
620-113
Electric igniter with open filament for HFP 36x family 10 and HFP 339
HFP339 Technical Offer
22.8.2006
Picture
Price
Page 9 of 15
Order No. Description
Qty
628-007
Fuse 10 A – MT
10
628-021
Fuse 2 A – MT
10
628-023
Fuse 6.3 A – MT
10
628-025
Fuse 4 A – MT
10
628-041
Fire sensor
5
636-069
Sample temperature probe up to 400 °C HFP 339
4
636-095
Flash sensor for water containing samples, with plug 1
613-055
Solid state relay 4A 230V
HFP339 Technical Offer
Picture
Price
1
22.8.2006
Page 10 of 15
Other spare parts: Order No. Description
Picture
101-029
Cover-handle PM HFP 339
103-060
Test cover PM for HFP 339 with provisions for thermal flash detection and for electric ignition
103-061
Test cover PM for HFP 339 with provisions for thermal flash detection and for gas ignition
326-070
Gauge as igniter simulator
HFP339 Technical Offer
22.8.2006
Price
Page 11 of 15
Order No. Description
Picture
326-072
Igniter dip in level adjustment tool
307-006
Drip cup, Pensky Martens
402-120
Gauge for adjustment of dip in level
627-098
Blower
645-070
Bottom-Heater
108-266
Push-pull rod to open the cover shutter
412-209
Cardanic suspension for 108-266
326-210
Driving device with suspension for 108-266
HFP339 Technical Offer
22.8.2006
Price
Page 12 of 15
Order No. Description
Picture
620-114
Electric igniter with closed filament to re-ignite the gas ignition
126-050
Foil keypad
650-551
Temperature measurement P.C.B., with software
607-068
Motor control PCB for dip in system, without EPROM
650-140
EPROM with latest HFP339 software for 607-068
607-215
Motor control PCB for stirrer, without EPROM
650-140
EPROM wit latest HFP339 software for 607-215
607-082
Display
607-257
Keyboard Adapter
HFP339 Technical Offer
22.8.2006
Price
Page 13 of 15
Order No. Description
Picture
607-258
Adapter for Measuring board
607-259
Interface board
607-260
Power supply board
607-252
CPU board
607-211
Pressure sensor and flash detection P.C.B.
617-151
Main switch
619-013
Varistor metallic oxide for mains protection
625-084
Transformer for electric ignition
HFP339 Technical Offer
22.8.2006
Price
Page 14 of 15
Order No. Description 625-126 Transformer 230/115V-24V/8V/HFP339
HFP339 Technical Offer
22.8.2006
Picture
Price
Page 15 of 15