Steens Physical Brukerveiledning for utstyr Instruction manual for equipment Introduction: This user manual for your new universal table is ment to guide you to the best use of the table , and that the patients safety is taken care of.

This user manual is for : 1. 372000 universal table std. artificial leather. 2. 372001 universal table std. leather. 3. 372002 universal table w/2 drops. Artificial leather. 4. 372003 universal table w/2 drops. Leather.

Read the user manual carefully, BEFORE using the table. Content in this manual: Cap. a) b) c) d) e) f) g) h) i) j)

dato/date: 21.10.1997

Text Declaration of conformity. User manual. Maintenance. Precautions. Limitations. Adjustments/settings. Area of use. Repairs. Drawings Technical specifications

Steens Industrier a/s

Side 2 3 4 4 5 5 5 5 6-8 9>>

side /page 1

Steens Physical Brukerveiledning for utstyr Instruction manual for equipment A _________________________________________________________________________________________ EC DECLARATION OF CONFORMITY FOR MACHINERY _________________________________________________________________________________________

Manufacturer: Steens Industrier a/s Address: Verkstedveien 25, 1400 SKI, NORWAY

Herewith declares that: Tables with number 37200,372001,372002,372003, is in conformity with the provisions of the Machinery Directive (directive 89/392/EEC), as amended, and with national implementing legislation.

Ski, Norway

21th of Oktober 1997

Terje Steen Terje Steen ass. managing director and CE-responsible

CE - labeling is done in accordance with the machine directive 89/392/eec

dato/date: 21.10.1997

Steens Industrier a/s

side /page 2

Steens Physical Brukerveiledning for utstyr Instruction manual for equipment

B. User manual: 1. Installation and unpacking: Inspect the wrapping carefully when receiving the table, and if there are signs of damage, you must contact your transporter immediately.

2. Do not use sharp objects or a knife when unpacking in order to avoid damage to upholstery or paint.

3. When receiving electrical tables, check specifications on motor, to be sure that your country has the correct voltage/outlet. 4. Test all functions on the table you have received. If there is something that is not working correctly, take contact with your dealer and describe the problem.

5. When using this table, follow the instructions picture map. 7. Armrests can be completely removed by loosening the locking wheels. 8. If the electrical motor is not working, check the voltage. If there is damage on the conductor, it must be replaced with the original. You must inspect if there are any visible damages on wires from the transformer box, if this is the case, it must be repaired or replaced. This must only be done by authorized personnel, or a service responsible that you have contact with.

dato/date: 21.10.1997

Steens Industrier a/s

side /page 3

Steens Physical Brukerveiledning for utstyr Instruction manual for equipment C.

Maintenance: 1.

If release of gas spring does not function properly, or that the spring is moving, it has to be adjusted. See page 6 for how to do this.

2.

You have to clean the table with mild soap water. This increase is the best for the table, and it is hygienic correct. On tables with artificial leather, you can use alcohol liquids to remove stains, but use mild soap water afterwards. You must NOT use alcohol liquids on tables with leather. We also recommend that you use covers on the table when persons are using it.

3.

Tables that include the drop function can be adjusted in hardness by unscrewing the locking wheel under the table. Screw with the clock to harden the drop, and screw against the clock to loosen it. Here there also is a handle to get cushion up after using the drop.

4.

If there are ”noises” on parts of the table, you can use silicon spray on this items.

5.

If any part has to be replaced for any reason, please always refer to part number on the item in question (see part drawing on page 7). Note, before changing parts you must make sure that you can do it without help, and if any questions, please contact your dealer for assistance.

6.

D.

If you have special requests regarding service and maintenance, take contact with your dealer.

Precautions: 1. The table is ONLY ment to be used for examination and treatment of people.

dato/date: 21.10.1997

Steens Industrier a/s

side /page 4

Steens Physical Brukerveiledning for utstyr Instruction manual for equipment

E.

Limitation: 1.

The table can NOT lift more than 130 kilo.

2.

Carrying capacity 250 kg.

3.

You must NOT sit on armrests, headrest or the adjustable leg rest.

4.

When the table is lowered in the lowest position, please remove the footswitch, or else they can be damaged.

5.

The table can NOT be machine washed.

6.

The equipment must NOT be stored or placed in areas corresponding to swimming pools, saunas or other areas of high humidity. This can cause corrosion and rust on the product.

7.

All moving parts MUST be locked when moving table up or down!.

F. Adjustments/settings: Look at the picture map and the drawings on page.6, for adjustments and settings.

G.

Area of use: The table is ment for use by chiropractors and manual therapist technicians.

H.

Repair: 1. If any problem does occur during use, try to fix it by using this manual. 2. If you still have problems with your table, take contact with your dealer. Remember to take note of the production year/month and the article number.

dato/date: 21.10.1997

Steens Industrier a/s

side /page 5

Steens Physical Brukerveiledning for utstyr Instruction manual for equipment I. Drawings:

Release head

ADJUSTMENT OF RELEASE AND LOCKING OF GAS SPRING.

dato/date: 21.10.1997

1.

Release nut.

2.

Turn rod on gasspring (uje appropriate tool with the clock for tightening, against the clock for loosening.

3.

If spring is slow, tighten spring.

4.

If spring is not releasing, it must be looser.

Steens Industrier a/s

side /page 6

Steens Physical Brukerveiledning for utstyr Instruction manual for equipment

I. Drawings:

dato/date: 21.10.1997

Steens Industrier a/s

side /page 7

Steens Physical Brukerveiledning for utstyr Instruction manual for equipment

CE LABLE

MAX LIFTING CAPABILITY 130 KG

CRUSH DANGER

I. Drawings:

dato/date: 21.10.1997

Steens Industrier a/s

side /page 8

Steens Physical Brukerveiledning for utstyr Instruction manual for equipment J. Technical specifications: 1. Length from 170cm adjustable up to 190cm. 2. Width 56cm. 3. Max height 93cm. 5. Min height 50cm. 6. Weight of the table 65 kg. 7. Lifting capacity 130 kg. 8. Carrying capacity 250 kg.

Electrical motor used: Magnetic Elektromotoren AG

SWISS MADE CH 4410 Liestal TYPE: MAX30-A180395A25100-000 24 V = 5 A 5.5 m.m /s Max 8000 N DRUCK / PUSH 6000 N ZUG/PULL 180 m.m int. 1min/9min IP66 CE

Transformer : Type:KOM31-1AA-000 U inn 230v Output 24v/ 6A =

Gas springs: 1. Bansbach ARNE SYVERSEN A/S type: K0B1PB3-080-283/150N for headrest. 2. Bansbach ARNE SYVERSEN A/S type: K0B1P03-040-225/750N for midpart. 3. Bansbach ARNE SYVERSEN A/S type: K0B1PB3-080-283/650N for legpart.

dato/date: 21.10.1997

Steens Industrier a/s

side /page 9

Universal Treatment Table For 25 years, STEENS-PHYSICAL has been providing safe, reliable and accurate rehabilitation

technology to the most demanding markets in the world. During this time, we have worked closely with

leading chiro praetors, medical doctors, physic~1 and manual therapists, and clinical researchers on

positioning and stabilization of the human body to safely isolate or incorporate joints and muscle groups.

Those years of intensive work have resulted in a body of knowledge concerning development of clinically

proven product applications available only for Steens-Physical. Our ability to apply that expertise has

enabled us to create tools for clinical rehabilitation that are unparalleled in the industry.

HSteens-Physical the most asked-for name in active rehabilitation H

Steens Physical

's therapY

.. 1

The Steens-Physical Universal Treatment Table (UTT) is no exception addressing the need for a highly adaptable format that doesn't sacrifice technicalquaHty or di~ical expertise. Knut Assjer and Vilhelm Scheel introduced the flexion distraction concept in 1983. They were pioneers in developing the first Flexion IDistraction Universal Treatment .. Table with the concept of combining all basic manual techniques.

-

­ -

-

-

-

-

-

Flexion with lateral flexion; multi functional pelvic section in both flexion and lateral positions.

-

-

Drop function can be built into the head section, upper and lower thoracic section and pelvic section.

-

-

.. L

The Steens-Physical team stands behind every Steens-Physical unit. UYou can trust the reliability, rely on the safety, and depend on your proven clinical· resultsU

Length adjustable ankle support permits proper body positioning and stability. Ergonomically designed handle allowing easy operation of the flexion focility.