Transmitter Cond 7100 Technical Data Contents
Technical Data Cond 7100
2
Technical Data Cond 7100/2(XH)
4
Terminal Assignments
6
Dimension Drawings
7
– 1–
Cond 7100 Cond 7100
Conductivity input
input for 2- or 4-electrode sensors
Measuring range Conductivity
0,2 µS * c ... 1000 mS * c 1 Ω/c ... 5 MΩ/c
Resistivity Measurement error 1)
< 1 % v. M.
Measuring range Conductivity
0,000 ... 9,999 µS/cm 00,00 ... 99,99 µS/cm 000,0 ... 999,9 µS/cm 0,000 ... 9,999 mS/cm 00,00 ... 99,99 mS/cm 000,0 ... 999,9 mS/cm
Resistivity
0,000 ... 9,999 MΩ cm 00,00 ... 99,99 MΩ cm 000,0 ... 999,9 MΩ cm
Salinity
0,0 ... 45,0 ‰ (0 ... 35 °C)
Temperature input *)
Pt100 / Pt1000 / NTC 30 kΩ / NTC 100 kΩ 2-wire connection, adjustable within ± 5 °C
Measuring range
Pt 100/Pt 1000: NTC 30 kΩ/NTC 100 kΩ:
Resolution Measurement error
0,1 °C / 1 °F < 0,5 K 2)3)
Temperature compensation (Reference temperature 25 °C)
(LIN) (NLF) –01– –02– –03–
*) configurable 1) ± 1 digit 2) for Pt 100 < 1 K 3) for NTC: temp. > 100 ºC < 1K 4) for USP fixed 0,05 µS/cm, for SAL fixed 0,2 ‰
(c=cell constant)
–20,0 ... +150,0 °C / –4 ... +302 °F –20,0 ... +130,0 °C / –4 ... +266 °F
00,00 ... 19,99 %/K (-20 … 130 °C) natural water to EN 27888 (0 … 36 °C) ultra pure water with traces of NaCl (0 ... 120 °C) ultra pure water with traces of HCl (0 ... 120 °C) ultra pure water with traces of NH3 (0 ... 120 °C)
USP function
Monitoring of conductivity of water used in the pharmaceuticals industry according to USP 23, fixed set points for limit value and alarm contacts. Adjustable reduced limit value (10 … 100 %) of the USP 23 value.
Display
LC alarm LED
Current output
0 ... 20 mA or 4 ... 20 mA, max. 10 V, floating 22 mA for error message, can be disabled
Current characteristic *)
linear or logarithmic
Override *)
22 mA for error message, cut-off switch
Current error
< 0,3 % of current value + 0,05 mA
Beginning/end of scale *)
selectable within the measurement range for µS, mS, MΩ, SAL
Min. measurement span
LIN LOG
Current source mode
0,00 mA ... 22,00 mA
–2–
5 % of selected measurement range 1 decade
Cell matching
• entry of cell constant at simultaneous display of the conductivity value and the temperature • manual determination of the cell constant using a calibration solution • temperature compensation
Permissable cell constants
0,0050 … 1,9999 cm -1
Monitoring of cells
SensoCheck®: recognition of polarization and monitoring of the cable capacity (can be disabled)
Switching contacts *)
4 switching contacts, floating limit contact min N/O hysterisis 0,2% of MB 4) limit contact max N/O hysterisis 0,2% of MB 4) alarm contakt N/C wash contakt N/O (periodity + wash contact, configurable)
Loadability
AC DC
Data retention
> 10 years (EEPROM)
RFI suppression
according to EN 50 081-1 and EN 50 081-2
Immunity to ESD
according to EN 50 082-1 and EN 50 082-2
Power supply
20 ... 253 V AC/DC; approx. 2 VA AC: 45 ... 65 Hz
Overvoltage category
II
Protection class
II
Ambient conditions
operating/ambient temperature –20 ... +55 °C transport-and storage temperature –20 ... +70 °C
Enclosure Color
polymer case, PBTP (polybutyleneterephthalate) blue-grey RAL 7031
Installation
• wall-mounting • pipe/bracket: ∅ 40 ... 60 mm, 30 ... 45 mm • control panel, cutout according to DIN 43 700 seal for case/panel
Dimensions Protection Cable entry Weight
144 x 144 x 105 mm IP 65 3 ports for PG 13,5 2 ports for NPT 1/2" or rigid metallic conduit approx. 1 kg
Connections Cross section
plug-in terminals, metal bracket ground for cell shielding max 2,5 mm2
Terminal assignments
[1] [2] [3] [4] [5] [6] [7] [8]
cell: current electrode + cell: voltage electrode + cell: voltage electrode – cell: current electrode – temperature sensor temperature sensor output
[9] [10] [11] [12] [13] [14] [15] [16] [17] [18] [19]
limit value contact min limit value contact max alarm contact wash contact free power supply
–3–
< 250 V / < 3 A / < 750 VA < 30 V / < 3 A / < 90 W
Cond 7100/2(XH) Cond 7100/2(XH)
Conductivity input
Input for 2- or 4-electrode sensors
Measuring range Conductivity
0,2 µS * c ... 1000 mS * c 1 Ω/c … 5 MΩ/c
Resistivity Measurement error 1)
< 1 % v. M.
Measuring range Conductivity
Resistivity
0,000 ... 9,999 µS/cm 00,00 ... 99,99 µS/cm 000,0 ... 999,9 µS/cm 0,000 ... 9,999 mS/cm 00,00 ... 99,99 mS/cm 000,0 ... 999,9 mS/cm 0,000 ... 9,999 MΩ cm 00,00 ... 99,99 MΩ cm 000,0 ... 999,9 MΩ cm
Salinity
0,0 ... 45,0 ‰ (0 ... 35 °C)
Temperature input *)
Pt100 / Pt1000 / NTC 30 kΩ / NTC 100 kΩ 2-wire connection, adjustable within ± 5 °C
Measuring range
Pt 100/Pt 1000: NTC 30 kΩ/NTC 100 kΩ:
Resolution Measurement error
0,1 °C / 1 °F < 0,5 K 2)3)
Temperature compensation (Reference temperature 25 °C)
(LIN) (NLF) –01– –02– –03–
*) configurable 1) ± 1 digit 2) for Pt 100 < 1 K 3) for NTC: temp. > 100 ºC < 1K
(c= cell constant)
–20,0 ... +150,0 °C / –4 ... +302 °F –20,0 ... +130,0 °C / –4 ... +266 °F
00,00 ... 19,99 %/K (–20 … 130 °C) natural water according to EN 27888 (0 … 36 °C) ultra pure water with traces of NaCl (0 ... 120 °C) ultra pure water with traces of HCl (0 ... 120 °C) ultra pure water with traces of NH3 (0 ... 120 °C)
Display
LC alarm LED
Power supply/output
loop current 4 ... 20 mA (22 mA), floating
Characteristic *)
linear or logarithmic
Feed voltage
14 … 30 V; Imax = 100 mA; Pmax = 0,8 W
Override *)
22 mA for error message, can be disabled
Current error
< 0,3 % of current value + 0,05 mA
Beginning/end of scale *)
configurable within the measurement range for µS, mS, MΩ, SAL
Min. measurement span
LIN LOG
Current source mode
3,8 ... 22,00 mA
–4–
5 % of selected measurement range 1 decade
Cell matching
• entry of the cell constant at simultaneous display of the conductivity value and the temperature • manual determination of the cell constant using a calibration solution • temperature compensation
Permissable cell constant
0,0050 … 1,9999 cm-1
Monitoring of cells
SensoCheck®: recognition of polarization and monitoring of the cable capacitance (can be disablet)
HART®-Communication Cond 7100/2H, 7100/2XH
digital communication through FSK modulation of the loop current, point-to-point connection, multidrop, instrument identification, reading of measured values, status and messages reading and printing of parameters
Explosion-proof Cond 7100/2, 7100/2H
Zone 2 (USA/Kanada) NI, Class I, Div 2, Group A, B, C, D T4
Cond 7100/2X, 7100/2XH
Zone 1 (Europa/USA/Kanada) II 2 G EEx ib [ia] IIC T6 IS, Class I, Div 1 Group A, B, C, D T4
Data retention
> 10 years (EEPROM)
RFI
according to EN 50 081-1 and EN 50 081-2
Immunity to ESD
according to EN 50 082-1 and EN 50 082-2
Ambient conditions
operating/ambient temperature –20 ... +55 °C transport and storage temperature –20 ... +70 °C
Enclosure Color
polymer case of PBTP (polybutyleneterephthalate) blue-grey RAL 7031
Installation
• wall-mounting • pipe/bracket: ∅ 40 ... 60 mm, 30 ... 45 mm • control panel, cutout according to DIN 43 700 seal for case/panel
Dimensions Protection Cable entry Weight
144 x 144 x 105 mm IP 65 3 ports for PG 13,5 2 ports for NPT 1/2" or rigid metallic conduit approx. 1 kg
Connections Cross-section
plug-in terminals, metal bracket ground for cell shielding max 2,5 mm2
Terminal assignments
[1] [2] [3] [4] [5] [6] [7] [8]
cell: current electrode + cell: voltage electrode + cell: voltage electrode – cell: current electrode – shield free temperature sensor
[9] [10] [11]
free / PA when Ex output/power supply two clamp pins for HART®
–5–
Terminal Assignments
Cond 7100
Cond 7100
n.c. wash
power
alarm
min
max
8 —
7
contacts
6
5
4
3
2
1
T1 T2
shield
20 to 253 V
9
0(4) to 20 mA +
19 18 17 16 15 14 13 12 11 10
temp.
MAX. RELAY CONTACT RATING; 1/8 HP, 125 V AC/250 V AC or 3 A, 30 V DC/250 V AC, RESISTIVE
output
2-/4-electrode conductivity sensor
Cond 7100/2(H)
7 T2
6
n.c.
do not connect
4 to 20 mA
HART
8 T1
5
4
3
2
1
2
1
shield
9
+
+ 11 10 —
temp. 2-/4-electrode conductivity sensor
supply/ output
8
7
T1
T2
supply/ output
6
5 shield
9
4 to 20 mA
+
+ 11 10 — HART
II2G EEx ib [ia] IIC T6 Tamb –20 to +55 °C 0032 PTB No. Ex-98.E.xxxx Made in Germany/Berlin
n.c.
Cond 7100/2X(H)
4
3
temp. 2-/4-electrode conductivity sensor
–6–
Dimension Drawings Fig. 1
Cond 7100
Assembling the case
Plastic plug (for sealing in case of wall mounting)
Pg threaded cable gland Hinge pin (can be inserted from either side)
Rubber reducer Plug
Fig. 2
Dimension drawing for Transmitter Cond 7100, mounting diagram and ZU 0275 panel mounting kit 105 144
ca. 14
144
27
42
Pg 13,5 (3 x)
43 max. 25
not included in scope of delivery 78
27
72
32
80
Pg 13,5 or conduit 1/2"
21
84
6,2
Control panel cutout 138 x 138 mm (DIN 43700)
Seal Control panel 1- 22 mm Note: All dimensions in millimeters
–7–
Fig. 3 ZU 0274 bracket kit Ø40 ... Ø60
Hose clamp with worm drive to DIN 3017
for vertical or horizontal posts or pipes
Fig. 4
ZU 0276 protective hood for wall and pipe mounting 165
173
132
Note: All dimensions in millimeters
ISO
ISO
9001
14001
certified
certified
Management System certified according to ISO 9001 / ISO 14001
Sales and service:
Technical subject to changes. © Mettler-Toledo GmbH 3/00 Printed in Switzerland. 52 120 925
–8–