SG8 RF Signal Generator : Datasheet. Overview

SG8 RF Signal Generator : Datasheet Overview Key Features of SG8 RF Signal Generator Output frequency range: 10 MHz – 8 GHz Frequency resolution: < 0...
Author: Melvyn Thompson
20 downloads 0 Views 504KB Size
SG8 RF Signal Generator : Datasheet Overview

Key Features of SG8 RF Signal Generator Output frequency range: 10 MHz – 8 GHz Frequency resolution: < 0.001 Hz Spectral purity o non-harmonics (typ.):HPSS (auxiliary option): −65 dBc HP (default option): −50 dBc o harmonics: −30 dBc o phase noise: −120 dBc/Hz @ 1GHz, 10 kHz offset Output level: up to +28 dBm Absolute level accuracy (typ.): ±0.1 dB Phase shift control o adjusting range: 360° o min. step: 0.15° @ 1GHz External input reference frequency o range: 20 – 150 MHz Remote control: RS-232, USB Dimensions: H x W x D: 2U x 42HP(½ 19”) x 315 mm

Brief Description Advantex SG8 10 MHz - 8 GHz signal generator is based on PLL frequency synthesizer and supports five operation modes:Continuous Wave – CW Sweep Frequency – SWF Sweep Level – SWL Frequency Modulation – FM Phase Modulation – φM. In CW mode one can adjust frequency, level and phase offset. There are four adjustable parameters in sweep modes: Start and Stop Frequency (in SWF) or Start and Stop Level (in SWL) Dwell Time Shape: Saw or Triangle Triggering: internal, external Triggering regime: auto, single and manual The following parameters are available in modulation modes: Frequency Level Deviation frequency (in FM) or phase deviation (in φM), or sensitivity when external modulating source is used Modulating source: internal or external Modulating signal shape: sine, square Frequency of internal modulating signal Other settings refer to REF In and REF Out signals, as well as to AUX In / TRIG and Mic In inputs. The instrument has USB and RS-232 interfaces for SCPI remote control and firmware updating capability.

1

SG8 RF Signal Generator : Datasheet For more information about instrument functions see Operating Manual.

Applications of Advantex SG8 RF Signal Generator General Purpose RF Signal Generator Instrument frequency range covers the common bands for RF applications such as radio and television, mobile, wireless networks, and radio-relay communications. This makes it possible to use the instrument in testing and debugging applications for all systems listed above. High output power compensates for additional level loss on the way to the DUT. Convenient user interface, large control elements and high-contrast display make it easy to obtain the desired signal at the RF output. Small dimensions and lightweight save working space and maintain easy portability. Arbitrary Reference Frequency Generator Small frequency step combined with low phase noise in low bands allows to use SG8 synthesized signal generator as an arbitrary reference frequency source while testing the deviation tolerance of symbol clock rates and local oscillators of communication systems. Low phase noise at low output frequencies is possible due to frequency dividers used up to the lower bound of the frequency range instead of frequency conversion from high to low range. One-decibel Compression Point P1dB Measurement for High-power Output Stages High output level of SG8 wideband signal generator enables one-decibel compression point and saturated power PSAT measurements without additional pre-amplifiers. Usually high-power output stages have relatively small gain, so availability of high power at the instrument output sufficiently simplifies the testing process. Second and Third Order Intercept Point IP 2 and IP3measurements Low nonlinear distortion of output stage, and high output level compensating for additional losses in attenuators following by power combiner, enable intermodulation parameters measurements of almost any RF path. RF-path Frequency Response Measurements Good frequency response flatness of SG8 microwave signal generator in wide frequency band due to APC (Automatic Power Control) and digital correction system allows to run scalar network analysis of DUT. Debugging of the RF Transceiver blocks for Telecommunication and Test & Measurement Equipment High output level and relatively low phase noise enables to use the instrument as a local oscillator for passive balanced mixers, which usually require high power for ensuring better linearity of the path. When debugging some systems, it is required that several LO sources in the same path are bound to some reference by natural factor. And reference frequency may significantly differ from standard series of references offered by most signal generators. In this case SG8 can use any frequency as a reference, making no demands to its stability or precision. Test & Measurement Automation External control interfaces (RS-232, USB) combined withSCPI protocol allow to use low-cost RF signal generator in ATE applications in volume production as well as in education laboratories where equipment is usually combined to one complex controlled by PC.

2

SG8 RF Signal Generator : Datasheet

Specifications

Frequency Parameters of Advantex SG8 RF Signal Generator Frequency parameters of RF signal Parameter

Conditions

Range

min. 3.90625

typ.

max. 8000

Resolution

over entire range

0.28·10

Settling time

from rising edge at AUX In input 2.5 (frequency switching over entire range)

4.5

Unit MHz

–3

Hz ms

Phase shift resolution* fOUT = fREF 0.022° 14 * Phase shift resolution equals (360°/2 )·fout/fref , where fout – output frequency, fref – reference frequency Reference frequency parameters Parameter

Conditions

min. typ. max. Unit

Internal reference frequency source Frequency

147

MHz

Accuracy

at 25°C

±5

ppm

Temperature effect

0..+50°C

±5

ppm

50

Ω

Reference frequency output (REF Out) Output impedance Signal level (internal source)

SG8-HP

−3.5 0

+4.5 dBm

SG8-HPSS

−5

+5

dBm

−1

+1

dB

Gain (when repeating external source fed to REF SG8-HP In input) External reference frequency input (REF In) Input impedance

50

Ω

Level

−10

+10

dBm

Frequency

20

150

MHz

−70

dBc

−60

dBc

|foffset|≤1MHz, Spur level (allowable at the REF In input for the PLL fref=20MHz to lock) |foffset|≤1MHz, fref=100MHz

3

SG8 RF Signal Generator : Datasheet

Spectral parameters of Advantex SG8 RF Signal Generator Spectral characteristics of RF output signal Parameter 2-nd Harmonic (10MHz≤fOUT≤8GHz)

3-rd Harmonic (10MHz≤fOUT≤8GHz)

Non-harmonics*, 4GHz