Ritter® Excellence Chair-mounted Dental Unit
Instruction Manual User Manual Service Manual
J1180, V03.00.02
z
Thank you for choosing our product.
z
It is designed for users to safely and effectively use this equipment. Please carefully read the product instruction manual before installation and debugging as well as operation and maintenance, and properly keep it for future use.
z
Properly keep the product quality maintenance card, product instruction manual as well as installation and maintenance accessories provided with product.
z
Please strictly follow the operation instruction when using and properly maintain it.
z
Any problems arise during using, please contact with local distributor or our company, and we will provide you with excellent service and help.
z
Commitment: when handling failure, if necessary, we commit to provide users with necessary and more specific technical data.
J1180_Excellence_V03.00.02
1
03.12.2010
Number
Name
1 1.1 1.2 1.2.1 1.2.2 1.2.3 1.2.4 1.2.5 1.2.6 1.2.7 1.2.8 1.2.8.1 1.2.8.2 1.2.8.3 1.2.8.4 1.2.8.5 1.2.8.6 1.2.8.7
Chair Unit Part Dental Chair Part(main body) Light Arm Part Light Arm(main body) Bowl Rinse Spout Cup Fill spout Spittoon Cover Protective cushion cover Water Bottle Operation manual Plastic Membrane Saliva ejector tip Suction tip Suction Fitter Syringe tip Connector Valve Fuse
1.2.8.8
Pipe
1.3 1.3.1 1.3.2 1.3.3 1.4 1.4.1 1.4.2 1.4.3 1.4.4
Operation Light part Operation Light (main body) Bulb Tools Dentist Operating Stool Seat Cushion Backrest Gas Spring Furcation Truckle
J1180_Excellence_V03.00.02
Order Number
JPC8-G1/2” G1/2” RF1-20:(ø5X20, 0.2A 1A 2A 3.15A 5A 10A glass tube) ø8X1(Blue Black),L=1.5m
50W 12V halogen bulb
2
Quantity
Unit
1 1 1 1 1 1 1 1 1 1 1 1
Sets Sets Sets Sets p.c. p.c. p.c. p.c. p.c. p.c. p.c. p.c.
Remark
1 2 1 2
p.c. p.c. p.c. p.c.
Spare Part Spare Part Spare Part Spare Part Spare Part Spare Part Spare Part
1
p.c.
Spare Part
1 1 1 1 1 1 1 1 1
Sets Sets p.c. p.c. Sets p.c. p.c. p.c. p.c.
Spare Part
03.12.2010
Contents I. II. III.
General ....................................................................................................................................4 Transportation and storage....................................................................................................9
Environmental safety..............................................................................................................9 1. Disposal of packing material .................................................................................................9 2. Disposal of equipment ...........................................................................................................9 SECTION ONE (For end user).........................................................................................................11 IV. Illustrations ...........................................................................................................................11 V. Profile.....................................................................................................................................12 VI. Structure of complete machine ...........................................................................................12 VII. Working Conditions..............................................................................................................13 VIII. Main technical parameters ...................................................................................................13 IX. External dimension of equipment .......................................................................................14 X. External dimension of equipment .......................................................................................15 XI. Invocate power switch................................................................................................15 XII. Heating of rinsing water .............................................................................................15 XIII. Water supply of rinsing ..............................................................................................15 XIV. Water supply of spittoon washing .............................................................................15 XVI. Handpiece operation...................................................................................................16 XVII. Air and water pressure adjustment of handpiece ....................................................16 XIX. Three-way syringe.......................................................................................................17 XX. Saliva ejector and high vacuum suction ...................................................................17 XXII. Operation of electric chair..........................................................................................18 XXIII. Reset button ............................................................................................................19 XXIV. Head rest..................................................................................................................19 XXV. Dentist Stool ................................................................................................................19 1. Handpiece ........................................................................................................................20 2. Other instruments ...........................................................................................................20 3. Exterior accessories .......................................................................................................20 SECTION TWO (for technicians) ....................................................................................................21 XXVII. Equipment installation .........................................................................................................21 • Preparation before installation.....................................................................................................21 1. Adjustment of filter relief valve ......................................................................................26 2. Main controlling board....................................................................................................26 3. Setup of the automatic control ......................................................................................28 4. Setting of instrument arm...............................................................................................29 XXVIII. Maintenance ....................................................................................................................30 1. Handpiece ........................................................................................................................30 2. Water filter........................................................................................................................30 4. Saliva ejector and suction ..............................................................................................31 5. Suction filter ....................................................................................................................31 6. Operating light.................................................................................................................31 XXIX. Failure diagnosis ..................................................................................................................32 XXXI. Diagram of the air and water pipes of the system .............................................................36 XXXII. Electric wiring diagram ........................................................................................................37
J1180_Excellence_V03.00.02
3
03.12.2010
I.
General
The manual is divided two sections: the first one is addressed to the end-user, the second section one is addressed to the technicians. The meaning of signs used in the instruction manual is as follows: Attention, Caution and Warning. Attention: important content is included that deserves attention. Caution: it comes before some particular operations. The equipment can be damaged once these provisions are violated. Warning: it comes before some particular operations. The equipment can be damaged and operator can be injured once these provisions are violated.
Warning: To guarantee the safe and normal operation, this machine shall not be placed in the environment of mixture of the inflammable anesthesia gas with air or oxygen or nitrous oxide. This unit is IPX0, Class I, Type B APPLIED PART. Operation method: Continuous operation with intermittent loading. Operation/ interval time: 2min/18min. The equipment does not cause electromagnetic or other influences at other equipment. It is not susceptible of such influences by other equipment, as it complies with Electro-Magnetic Compatibility Directive 89/336/EEC, emended by 92/68/EEC and 93/68/EEC directives, and it satisfies the requirements of the EN60601-1-2 harmonized standard. The units is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage supply network that supplies buildings used for domestic purposes. Floors should be wood, concrete or ceramic tile. If floor are covered with synthetic material, the relative humidity should be at lease 30%. Mains power quality should be that of a typical commercial or hospital environment.
J1180_Excellence_V03.00.02
4
03.12.2010
Guidance and manufacture’s declaration – electromagnetic emission The Ritter Excellence Chair-mounted Dental Unit is intended for use in the electromagnetic environment specified below. The customer of the user of the Ritter Excellence Chair-mounted Dental Unit should assure that it is used in such and environment.
Emission test
Compliance
Electromagnetic environment – guidance
RF emissions
The Ritter Excellence Chair-mounted Dental Unit
CISPR 11
uses RF energy only for its internal function. Therefore, its RF emissions are
Group 1
very low and are not likely to cause any interference in nearby electronic equipment. RF emission
The Ritter Excellence Chair-mounted Dental Unit
Class A
CISPR 11
is suitable for use in all establishments, other than domestic and those directly
Harmonic emissions
connected to the public low-voltage power
IEC 61000-3-2
supply network that supplies buildings used
Voltage fluctuations/
for domestic purposes.
Not applicable
flicker emissions IEC 61000-3-3
Guidance and manufacture’s declaration – electromagnetic immunity The Ritter Excellence Chair-mounted Dental Unit is intended for use in the electromagnetic environment specified below. The customer or the user of Ritter Excellence Chair-mounted Dental Unit should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment -
Electrostatic
±6 kV contact
±6 kV contact
guidance Floors should be wood, concrete
discharge (ESD)
±8 kV air
±8 kV air
or ceramic tile. If floor are covered with synthetic material, the relative
IEC 61000-4-2
humidity should be at least 30%. Electrical fast
±2 kV for power supply
±2kV for power
Mains power quality should be that
transient/burst
lines
supply lines
of a typical commercial or hospital
IEC 61000-4-4
environment.
Surge
±1 kV differential mode
±1 kV differential
Mains power quality should be that
IEC 61000-4-5
±2 kV common mode
mode
of a typical commercial or hospital
±2 kV common mode
environment.
Voltage dips, short
95% dip in UT)
of a typical commercial or hospital
voltage variations on
for 0.5 cycle
for 0.5 cycle
environment. If the user of the
40% UT
40% UT
Ritter Excellence Chair-mounted Dental Unit
(60% dip in UT)
(60% dip in UT)
requires continued operation
for 5 cycles
for 5 cycles
during power mains interruptions,
70% UT
70% UT
(30% dip in UT)
(30% dip in UT)
for 25 cycles
for 25 cycles
95% dip in UT)
power supply input lines IEC 61000-4-11
it is recommended that the
Ritter Excellence Chair-mounted Dental Unit be powered from an uninterruptible power supply or a battery.
for 5 sec
NOTE
for 5 sec UT is the a.c. mains voltage prior to application of the test level.
J1180_Excellence_V03.00.02
5
03.12.2010
Guidance and manufacture’s declaration – electromagnetic immunity The Ritter Excellence Chair-mounted Dental Unit is intended for use in the electromagnetic environment specified below. The customer or the user of Ritter Excellence Chair-mounted Dental Unit should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment - guidance Portable and mobile RF communications equipment should be used no closer to any part of the
Ritter Excellence Chair-mounted Dental Unit, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance Conducted RF
3 Vr m s
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
3 Vr m s
3 V/m
Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m). Field strengths from fixed RF transmitters, as a determined by an electromagnetic site survey, should be less than the compliance level in each frequency b range. Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a ) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Ritter Excellence Chair Mounted Dental Unit is used exceeds the applicable RF compliance level above, the Ritter Excellence Chair Mounted Dental Unit should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the Ritter Excellence Chair Mounted Dental Unit. b ) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
J1180_Excellence_V03.00.02
6
03.12.2010
Recommended separation distances between portable and mobile RF communications equipment and the Ritter Excellence Chair-mounted Dental Unit The Ritter Excellence Chair-mounted Dental Unit is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Ritter Excellence Chair-mounted Dental Unit can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Ritter Excellence Chair-mounted Dental Unit as recommended below, according to the maximum output power of the communications equipment. Separation distance according to frequency of transmitter (m) Rated maximum output power of
80 MHz to 800 MHz
150 kHz=to 80 MHz
800 MHz to 2.5 GHz
transmitter (W) 0.01
0.1167
0.1167
0.2334
0.1
0.3689
0.3689
0.7378
1
1.1667
1.1667
2.3334
10
3.6893
3.6893
7.3786
100
11.6667
11.6667
23.3334
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
J1180_Excellence_V03.00.02
7
03.12.2010
Attention: z Air source provided prohibited to use.
by
industrial
compressor
is
strongly
z No load running and over-pressure start of the handpiece is strongly forbidden. z
Clean or replace the core of the water filter timely.
z No touch of the reflecting mirror is allowed when the operating light is running, in case of the scald. z The dental chair is the continuous running of interval loading, and its running/interval time is 2min/18min. z
The operation of heater under dry condition is strictly prohibited.
z Adjust the position of the head rest and make sure that it has been locked before using. z
Cut off the power before replacing the safety fuse.
z No tangible thing shall be placed within the operational stroke of this machine. z The plastic membrane used for packing shall be disposed properly due to its harm to human body and environment.
J1180_Excellence_V03.00.02
8
03.12.2010
II.
Transportation and storage
The product can be transported or stored since it has been properly packed. Under the following ambient conditions; it can be placed fifteen weeks at most: a) Environment temperature: - 40°C~+55°C. b) Relative humidity: