Feature Article
JY Division Raman Instrumentation for the Information Raman Pharmaceutical Industry Gwenaelle Le Bourdon, Fran Adar
Abstract Over the past 5 years our Raman business has seen continually accelerating growth. The biggest component to this growth is in the pharmaceutical industry. Several factors contribute to this success. It is generally now known that it is easy to record a Raman spectrum, and that the information derived complements that of XRD (X-ray diffraction) and FTIR (Fourier Transform Infrared Spectroscopy). The spectral resolution at 1 µm is far greater than either of these techniques. And spectra can be recorded inside of glass and plastic containers. Of greatest interest is the differentiation between the various solid state forms, and distribution of the APIs (Active Pharmaceutical Ingredients) in final products. All of these issues are discussed in this article, with illustrative examples.
Introduction With the revolution and evolution of Raman instrumentation over the past 10-12 years, the Raman market has been growing at an accelerated pace. Consequently we get more and more requests to provide systems specialized for particular applications, and some of our “special developments” evolve into becoming standardized products. The largest area of growth that we are seeing is in the pharmaceutical industry. This is a result of several factors. • It has become recognized that measuring a Raman spectrum is an “easy” experiment. • There is information in the Raman spectrum that complements FTIR and XRD. • There are several types of information of interest to the pharmaceutical industry that can be derived from Raman: Differentiating Excipient vs. Active Ingredient, Differentiating polymorphic forms of the Active Ingredient, Monitoring Chemical changes of intermediate forms, and Measuring Composition of Solution. Of particular interest to the pharmaceutical industry is information on polymorphy *1 – for 2 reasons. Clearly the polymorph of the active ingredient will effect things like solubility and bioavailability. But, because of these properties, the pharmaceutical companies have been able to patent not only the chemical composition, but the polymorph as well.
56
What is important to recognize is that the polymorphy of small molecules is determined by chemical interactions. That means that intra- and/or inter-molecular interactions between specific molecular functional groups can determine the conformation of the molecules as well as the intermolecular packing in crystals. While XRD can identify different polymorphs, extraction of detailed information requires a full fitting of the pattern. On the other hand, many molecular functional groups exhibit identifiable vibrational modes in the Raman and/or IR spectra, and these modes will shift due to chemical associations. This article will show examples of results of measurements made on these types of materials that will demonstrate the power of the Raman instrumentation. *1 By polymorphy, we include related phenomena of pseudomorphy in which waters of hydration are included in the crystal, and salt forms in which positive or negative functional groups are associated with anions or cations.
Excipients vs. Active Ingredients – Identification and Mapping Pharmaceuticals are rarely sold as pure substances. APIs are mixed with excipients in order to provide bulk, extend shelf life, control solid state chemical reactions, and control the time of release (by wrapping APIs in polymer), etc. An important characteristic of solid mixtures is the distribution of the species. In fact, the operation of mixing English Edition No.7
Technical Reports
itself becomes critical. It has been shown that vibration of a dry powder mixture of several species of particles of equal size but different densities will result in completely separating the species [1]! Thus, significant engineering and quality control has to be exerted in guaranteeing tablet composition.
20
Length X (µm)
25
30
35
40
45
5µm
50
20
25
30
35
40
45
50
Length X (µm)
Intensity (a.u.)
(a)
1200 1000
1086.7
800 600 400
Raman
20
Length Y (µm)
(b)
Emission
10
(a)
Grating & OEM
Raman spectroscopy is an ideal tool for the differentiation of API from excipient. In fact, one could say it is custom tailored. Normally the API is a minor ingredient. But because it almost always has at least one aromatic group, its Raman signal is quite high compared to the other components. The first example that we will present represents a study of a pharmaceutical tablet. Fig. 1 shows a microscopic image of a tablet. While it is possible to “see” particles, there is no chemical information in the picture.
(c) 15
200
30 400
600
800
1000
1200
Wavenumber (cm-1)
40
(b) 2000
10
20
30
40
50
60
70
Length X (µm) Fig. 1 Optical Microscopic Image of a Tablet
1500
477.3
1000 500
400
600
800 1000 Wavenumber (cm-1)
1200
Thin Film
(c) 14000 12000
1003.0
10000 8000
Optical Spectroscopy
6000 4000 2000 0 400
600
800 1000 Wavenumber (cm-1)
1200
Fig. 2 Raman Map of a Tablet
Forensic
Spectral examination indicated the presence of at least three chemical species. It was decided to “map” the region indicated by the blue outline. In this operation, the sample is rastered in the X and Y directions. At each position, a full spectrum is acquired. The collection of spectra is used to map the position of each of the chemical species; in the map each color represents material whose spectrum is printed in the same color (Fig. 2). The two excipients have been identified as lactose (Fig. 2(a)) and starch (Fig. 2(b)) following a search of the spectra in a Raman spectral library. The third component, component C, has not been identified, but the presence of a strong line at ca. 1000 cm-1 indicates that it is an aromatic (Fig. 2(c)). Note also the intensity axis of component C is an order of magnitude higher than that of the other two components which confirms the expectation that aromatic API will produce more intense Raman spectra.
Intensity (a.u.)
0
Intensity (a.u.)
60
Fluorescence
50
(a) Raman Spectrum of Lactose (b) Raman Spectrum of Wheat Starch + Component C (c) Raman Spectrum of Component C (assumed as aromatic series)
57
Feature Article
Raman Instrumentation for the Pharmaceutical Industry
It was mentioned above that the identification of the excipients was done using Spectral ID, library-searching software. Fig. 3 shows the Raman spectra of lactose and starch, the two excipients mentioned above. Fig. 4 is a reproduction of the Spectral ID screen showing the overlay of the top spectrum with that of a library spectrum of starch.
Polymorphy Control of polymorphy is a major issue in the pharmaceutical industry, as mentioned above. To reiterate, solubility and bio-availability depend on polymorphy. Consequently, pharmaceutical companies want to control polymorphy, and patent-protect their products with polymorph spefications. Classically, polymorphy, and its variants as described above, were determined by XRD.
943.5
Intensity (a.u.)
478.3
2907.1
1086.2 3363.0
1462.0
Wheat Starch
hand, vibrational spectroscopy will indicate chemical
2887.4
359.1
interactions between identifiable functional groups. And the spatial resolution of vibrational techniques is determined
2946.5
478.3 852.9 1086.7 1326.5
2977.7 3382.9
Lactose 500
1000
1500
2000
2500
However, XRD requires rather large samples, and does not provide chemical information directly. On the other
3000
Wavenumber (cm-1)
Fig. 3 Raman Spectra of Lactose and Wheat Starch
3500
by the physical diffraction limit and is quite good compared to XRD. For Raman, of course, the spatial resolution is about 1 µm. For infrared spectroscopy (FTIR) the spatial resolution is about 20 µm. In the pharmaceutical industry, not only is the identification of the polymorph important, but its control during formulation of product is of great concern. Under certain circumstances it can be quite easy to convert from one polymorph to another. For instance, if relative humidity is not controlled, the degree of hydration of a crystal form can change, which changes its pseudopolymorphic phase. A different set of conditions that could require control might be the mixing of APIs with excipient and/or pressing into tablets. Under such operations one could also imagine the possibility of changing hydration or even salt forms. The following spectra(Fig. 5) document the ability to
Fig. 4 Spectral ID Screen
identify polymorphy of Tylenol by Raman spectroscopy. The spectra represent the A and B forms. FTIR spectra of the same species have been recorded on the LabRam IR and are shown in Fig. 6 – first the fingerprint part, and then the CH/NH part.
58
English Edition No.7
Technical Reports
(a)
(a) 3000
1731
0.3
Absorbance
Intensity (a.u.)
1729
0.4
2000
Form B
0.2
Form B 0.1
Form A
Form A 0
800
100
200
300
400
1000
500
(b)
1735
1650
Intensity (a.u.)
1500 1000
4000
1617
500 0
3000
1600
1650
1700
1750
1600
1800
3426
0.20
3460
0.15
3506
0.10
Emission
5000
1730
Absorbance
1724
2500 2000
1400
0.25
1668 1670
3000
1200
Wavenumber (cm-1)
Wavenumber (cm-1)
(b)
Grating & OEM
1000
Form B
0.05
Form A 2000 2600
Form B
2800
3000
3200
3400
3600
3800
Wavenumber (cm-1)
Raman
1000
Form A 0 600
800
1000
1200
Wavenumber
1400
Fig. 6 Infrared Absorption Spectrum of Tylenol Polymorphism
1600
(a) Fingerprint Region
(cm-1)
(b) CH/NH Region
(c) 7000
Intensity (a.u.)
how the interactions between functional groups can determine the way the molecule crystallizes. Now we
5000
Form B
can see how vibrational spectra can provide information
4000 3000 2000
3434
3033
3460
on molecular associations. Table 1 tabulates the frequencies of >NH/-OH (3300-3600 cm-1) and >C=O
3050
3502
(1710-1740 cm-1) bands.
1000
Form A 0 2800
3000
3200
Wavenumber
3400
(cm-1)
Table 1 Raman and FTIR Peak Wavenumber for Tylenol Polymorphism
Raman FTIR
Form A 3460/3502 cm-1 1730 (sh 1735) cm-1 3460/3506 cm-1 1731 cm-1
Form B 3434 cm-1 1724 cm-1 3426 cm-1 1729 cm-1
Optical Spectroscopy
Fig. 5 Raman Spectrum of Tylenol Polymorphism (a) Wavenumber Approx. 50 to 500 cm-1 (b) Wavenumber Approx. 500 to 1800 cm-1 (c) Wavenumber Approx. 2600 to 3700 cm-1
3600
Thin Film
3025
Fluorescence
Recall that in the introductory section, we described
6000
Forensic 59
Feature Article
Raman Instrumentation for the Pharmaceutical Industry
There is overall agreement between the FTIR and Raman results. Form A has a doublet in the NH/OH region but a singlet in Form B. The carbonyl band has a frequency
(a)
slightly higher in form A than form B. However, the difference in frequency is more pronounced in the Raman spectrum rather than the FTIR spectrum. This is somewhat ironic because the IR is believed to be more sensitive to carbonyl than Raman, which is, in fact, supported by the relatively low Raman intensities of this band. However, the insensitivity of the FTIR frequency could possibly be a result of the larger width of IR band. These differences in frequencies can be assigned to phenomena such as hydrogen-bonding. A lower frequency
(b)
means more H-bonding, which is also often reflected in a wider band. In this case form B is exhibiting stronger Hbonding.
High Throughput Screening and the Multiwell LabRam
(c)
High Throughput Screening (HTS) is the latest approach in analytical technology. Used particularly in the pharmaceutical industry and in life science laboratories, this method enables systematic screening of drug design and high throughput analysis of batch production. The LabRam software has been adapted for automated study of Multiwell Plates. All of the phenomena discussed above are amenable to automated analysis for High Throughput Screening. Using a user-defined preprogrammed protocol for analysis, the software will output information in the form of an Excel spreadsheet. For example, the spreadsheet can report concentrations of API vs. excipient or different polymorphic forms as a function of well position. Fig. 7 shows the Raman Multiwell
Fig. 7 Raman Multiwell Analyzer (a) Multiwell LabRam (b) Multiwell Plate (c) Spectrum Screen
Analyzer.
60
English Edition No.7
Technical Reports
Conclusion Developments of the last 75 years have paved the way for the implementation of sophisticated, automated Raman instruments in the analytical laboratory. These developments are in fact, two-pronged. On the one hand,
Grating & OEM
the interpretation of spectra has to enable use of the phenomenon to extract useful information. On the other hand, the equipment has to produce the information with a minimum of effort. As this has converged, the pharmaceutical industry is now poised to take advantage of these developments to aid in the development of their
Emission
products at all stages.
Reference
Raman
[1] Tom Mullin, “Mixing and De-mixing,” Science 295, p.1851 (8 March, 2002)
Fluorescence Thin Film Optical Spectroscopy
Gwenaelle Le Bourdon, PhD Jobin Yvon S.A.S Raman Division Application Scientist
Jobin Yvon Inc. Raman Worldwide Senior Application Scientist
61
Forensic
Fran Adar, PhD