User Manual

Please Read this Instruction manual before use

Clinical IR thermometer

DX6635

Contents INTRODUCTION……………………………….………….....……….... 3 IMPORTANT SAFETY INFORMATIONS………………….....…….... 3 BODY TEMPERATURE………………………………..................….. 5 PACKAGE COMPONENTS…………………………….........………... 5 PRESENTATION………………………………..........................….... 6 BEFORE USE…………………………….........................………….. 8 HOW TO MEASURE FOREHEAD TEMPERATURE………..……..... 9 HOW TO MEASURE EAR TEMPERATURE………………………..... 9 TEMPERATURE TAKING HINTS…………………………………......10 BACKLIGHT AND TORCH LIGHT FUNCTION….......……..…….... 11 CHECK THE MEMORY……………………………………………....... 11 DATE AND TIME SETTING……………………………………...........12 VOICE PROMPT……………………….........................………...... 13 CELSIUS/FAHRENHEIT OPTION AND RESET FUNCTION......... 13 REPLACING THE BATTERY……………………………………........ 13 SPECIFICATIONS…………………………………………………...... 14 CARE AND CLEANING………………………………………….......... 15 CALIBRATION…………………………………………………............ 15 TROUBLESHOOTING…………………………………………........... 16 NORMALIZED SYMBOLS……………………………………............ 17 DISPOSAL…………………………………………………..................23 WARRANTY…………………………………………………................24

INTRODUCTION Thank you for purchasing clinical IR Thermometer DX6635 from Shenzhen Dongdixin Technology Co., LTD. The thermometer has been carefully developed for accurate, safe and fast human body temperature measurements in the ear and forehead.

The full stop before "so" will be changed to the comma.

Research indicates that the ear is an ideal site for taking body temperature because body temperature is regulated by the hypothalamus, which shares the same blood supply as the tympanic membrane. So please take the ear temperature as much as possible.

IMPORTANT SAFETY INFORMATIONS However, as with any thermometer, proper technique is critical to obtaining accurate temperatures. Therefore, in order to obtain an accurate measurement, please read this manual carefully before using. Use of this thermometer is not intended as a substitution for consultation with your physician. Measurement results are for reference only. Contact your physician if you have or suspect any health concerns. The Infrared sensor must be kept clean, dry, and

KEEP INFRARED SENSOR CLEAN

GOOD

NO GOOD

3

undamaged at all times to ensure accurate measurements. Earwax in the ear canal or lipid on the forehead may cause an inaccurate measurement, so make sure the ear canal or forehead is clean. If you clean the ear or forehead, please waiting for 5~10 minutes before measuring. Don't touch or blow the infrared sensor. A polluted infrared sensor may cause inaccuracy. Do not use this thermometer for measurements other than human body temperature. Avoid measuring the ear temperature if the ear is inflamed or infected. The temperature of the storage area differs greatly from that of the measuring area, please wait the thermometer temperature to equalize to the room temperature about 30 minutes before use. Keep the thermometer out of the reach of children. Don't throw battery into fire. Battery is likely to explode in fire. Don't store the thermometer to temperature extremes (below -20°C/-4°F or over 55°C/131°F ) or humidity extremes (below 10%RH or over 90%RH). Failing to do so may cause inaccuracy. Don't use the product in a wet ear hole after swimming or bathing. It is possible to damage the ear hole. Don't use the thermometer if the main body is damaged (for example, the infrared sensor is broken). The continuous use of a damaged unit may cause injury, improper results, or serious danger. Don't take apart, repair or change any parts of the unit at any time. Don't use a mobile phone nearby when the thermometer is working. If your thermometer will not be used on a regular basis, remove battery to prevent possible damage due to chemical leakage. If battery does leak, remove carefully. Do not allow bare skin to touch leaking fluid. This product needs special precautions regrading EMC and needs to be installed and put into service according to the EMC information provided, and this unit can be affected by portable and mobile RF communications equipment.

4

Caution: this machine should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, this machine should be observed to verify normal operation in the configuration in which is will be used.

BODY TEMPERATURE Normal body temperature is a range. A person's normal temperature range tends to decrease with age. The following table shows normal temperature ranges by age: Normal temperature ranges by age: 0 – 2 years

36.4 – 38.0 °C

97.5 – 100.4 °F

3 – 10 years

36.1 – 37.8 °C

97.0 – 100.0 °F

11 – 65 years

35.9 – 37.6 °C

96.6 – 99.7 °F

> 65 years

35.8 – 37.5 °C

96.4 – 99.5 °F

The range of normal varies from person to person and can be influenced by many factors such as time of day, measure from different sits, level of activity, medications, emotion and so on. So we recommend that you practice with the Thermometer on yourself and family members when you are healthy. This way you know how the thermometer works and can feel more confident of the measurements you take when a family member is ill.

PACKAGE COMPONENTS NO. 1

DESCRIPTION

QUANTITY

Clinical IR thermometer

1 piece

2

Table stand holder

1 piece

3

1.5V Battery(AAA)

2 pieces

4

User manual

1 piece 5

PRESENTATION Operation panel ① ⑤ ② ③





⑩ ⑧

⑦ ⑨

1) Infrared sensor 2) Ear/Forehead mode detector 3) TALK/OK button 4) Memory/Set (MEM/SET) button 5) Forehead cap 6) LED torch 7) Reset button 8) Celsius/Fahrenheit (℃/℉) button 9) Battery compartment 10) Torch button 11) Table stand holder 12) Measuring button 6

LCD display

① ② ③ ④

⑩ ⑨ ⑧







1) Temperature and time display 2) AM /PM display 3) Sleep mode indicator 4) Week indicator 5) Low battery indicator 6) Memory indicator 7) Year display 8) Month and day display 9) Voice indicator 10) Temperature measuring indicator 11) Ear temperature measuring indicator 12) Forehead temperature measuring indicator

7

add" is "

BEFORE USE 1. To achieve accurate measurements, it is very important to check that the infrared sensor clean before using. To clean the sensor, gently wipe its surface with a cotton swab slightly moistened with alcohol and immediately wipe dry with a clean cotton swab. After cleaning, allow at least 20 minutes drying time before taking temperatures. 2. Earwax in the ear canal or lipid on the forehead may cause an inaccurate measurement, so make sure the ear canal or forehead is clean. If you clean the ear or forehead, please wait for 5~10 minutes before measuring. 3. If the temperature of the storage area differs greatly from that of the measuring area, please wait for the thermometer temperature to equalize to the room temperature about 30 minutes before measuring. 4. Please open the battery cover and move the insulating piece away when you use the thermometer for the first time, and then the thermometer turns on automatically, and the LCD display will be as shown on the left. When it is turned off, you can press any button to turn it on. Caution: 1) In measuring mode and memory mode, the LCD is normal display; but in waiting mode and setting mode, the LCD is reverse display. 8

2) In waiting mode, the LCD circularly displays time, and room temperature.

HOW TO MEASURE FOREHEAD TEMPERATURE 1. Put on forehead cap before use. Place the thermometer in your hand as shown with your thumb on the [Measuring] button. Do not press the button yet. Make sure the probe is clean before measuring. 2. Hold the thermometer so that the forehead cap is flush on center of the forehead, midway between eyebrow and hairline. Touch the forehead cap to the forehead and press the [Measuring] button to start the measurement. 3. Keep the thermometer flat until you hear temperature voice announce the temperature or a beep. The measurement is complete and the thermometer may be removed from the head. 4. Read the temperature on the display.

HOW TO MEASURE EAR TEMPERATURE 1. Remove forehead cap before use. Place the thermometer in your hand as shown with your finger on the [Measuring] button. Do not press the button yet. Make sure the probe is clean before measuring.

9

2. Insert the probe very gently and slowly into the ear canal. Hold the thermometer so that the probe faces straight in the direction of the eardrum. It is important to point the probe window of thermometer toward the eardrum and its periphery inside the ear. Hold the outer ear and gently pull it toward the rear of the head to straighten the ear canal Always insert the thermometer into the same ear with the same direction and depth. 3. Press the [Measuring] button to start the measurement. Keep the probe flat until you hear temperature broadcast or suggestive sound. 4. Remove the thermometer from the ear canal. The LCD displays the measured temperature. You can measure again after the symbol "000" disappears.

TEMPERATURE TAKING HINTS 1. A temperature measurement taken in the right ear may differ from a measurement taken in the left ear. Therefore, always take the temperature in the same ear. 2. External factors may influence ear and forehead temperatures, particularly when an individual has: been lying on one ear or the other had their ears covered been exposed to very hot or very cold temperatures, or 10

been recently swimming or bathing. In these cases, remove the individual from the situation and waiting for 30 minutes prior to taking a temperature. 3. For persons wearing hearing aids or ear plugs, remove the device and wait 30 minutes prior to taking a temperature. 4. Use the untreated ear if ear drops or other ear medications have been placed in the ear canal.

BACKLIGHT AND TORCH LIGHT FUNCTION Press [TORCH] button to turn on the backlight or press and hold [TORCH] button for two seconds to turn on the torch light. Then press [TORCH] button again to turn off the backlight or torch light. Caution: The backlight or torch light will turn off automatically after 5 seconds or in temperature measuring mode or voice prompt mode.

CHECK THE MEMORY There are a total of 20 memory data for recording ear and forehead temperature measurements. The current measurement is always stored in the last storage space.When all storage spaces have been occupied, the oldest measurement is always deleted from the memory.

11

Press [MEM/SET] button to check the temperature and date. See figure on left. Press [MEM/SET] button repeatedly to cycle through the temperature in the memory. Press and hold [MEM/SET] button to check the memory data. If there is no press any button in 5 seconds, the device back to waiting mode.

DATE AND TIME SETTING Press and hold [MEM/SET] button about 2 seconds to enter into setting mode. Press [TALK/OK] button to switch the settings of hour mode (12hr or 24hr, AM or PM will display on LCD when you select 12hr mode), hour, minute, year (2001~2099), month, day and sleep mode (ON or OFF). When the sleep mode is active, the thermometer will turn off automatically about 1min or 3min(which is factory default); when the sleep mode is deactivated, the thermometer remains power on. For save power, we recommend to set sleep mode is set to on. Press [MEM/SET] button to increase the parameter which you selected.

12

VOICE PROMPT Press [TALK/OK] button to report measuring temperature, current time or room temperature. In waiting mode, press and hold [TALK/OK] button about 2 seconds to select the talking language which you want, it will change to anther language or non-talking function automatically.

CELSIUS/FAHRENHEIT OPTION AND RESET FUNCTION Move the battery cover, there are two keys on the two side of battery compartment. The left one is [ C/ F] key, you can press it with a pen to change between Celsius ( C) and Fahrenheit ( F). The right one is a [Reset] key. If you press it, the device will restart, and then the device restore factory parameter settings and empty all memory data.

[ C/ F] key

REPLACING THE BATTERY When the low battery indicator “ ” appears on the LCD, or the thermometer does not function at all, you should replace new batteries as soon as possible. The thermometer is supplied with two batteries (Type AAA). Move the battery cover, and pull up the batteries, then insert new AAA batteries into the battery compartment. Make sure you are installing batteries properly, be sure to match the positive and negative ends of the battery to the marking in the battery compartment of the thermometer. Replace the battery compartment cover and press to close.

13

 Caution: To protect the environment, dispose of exhausted battery according to national or local regulations. Keep the battery out of the reach of children.

SPECIFICATIONS Power supply: DC 3V (2*AAA) Body measuring range: 32.0°C ~ 42.9°C (89.6°F ~109.3°F) Ear and forehead measure accuracy: ±0.2°C (±0.4°F) in range of 35.5°C~42.0°C (95.9°F~107.6°F) ±0.3°C (±0.5°F) for other range Displayed room temperature range: 5.0°C ~ 59.9°C (41.0°F~139.8°F) Room temperature accuracy: ±1°C (±2°F) Dimension: 130(L)x41(W)x57(D)mm Weight: 56g (without battery) Forehead operating condition: 15.0°C~40.0°C (59.0°F~104.0°F) with a relative humidity of 20%~85% Ear operating condition: 10.0°C~40.0°C (50.0°F~104.0°F) with a relative humidity of 20%~85% Atmospheric pressure: 700 ~ 1060 hPa Storage environment: -20°C~55°C (-4°F~131°F) with a relative humidity of 20%~90% Atmospheric pressure: 700 ~ 1060 hPa Service life of the device: 2 years. Service life of the battery: With a new battery approx. 6 months based on the use frequency at 5 times/ day. Note: This thermometer has been thoroughly tested and inspected to assure proper performance and operation!

14

CARE AND CLEANING To ensure accurate measurements, it is very important to keep the probe tip clean and free of scratches. Finger prints, earwax or dirt will affect the accuracy of the thermometer. In order to get an accurate measurement, please clean the probe tip. Gently wipe its surface with a cotton swab slightly moistened with alcohol and immediately wipe dry with a clean cotton swab. Use a soft dry cloth to clean the body of the thermometer. Never clean the thermometer with an abrasive cleanser, thinner, benzene or submerge the thermometer into water or other liquids. After cleaning, place the thermometer in the protective case or Table stand holder. Store it in a clean, dry place at room temperature. Never expose the thermometer to extreme temperatures, humidity, direct sunlight or shock.

CALIBRATION The thermometer is initially calibrated at the time of manufacture. If the thermometer is used according to the instruction, periodic recalibration is not required. If at any time you question the accuracy of the measurement, please contact the retailer immediately. Don't attempt to modify or reassemble the thermometer.

15

TROUBLESHOOTING Troubles No response / Automatic reset when pull out insulator

Checklists Battery used up? Battery in wrong polarity? Poor battery contact

Countermeasures Change new battery Take out battery, reinsert battery correctly

The measured temperature is lower than 32.0°C/ Follow steps in User's 89.6°F or higher than Manual for proper 42.9°C/109.3°F measurement. Please check the operation method

The measurement is not accurate or if there is any doubt of the measure result

16

Hardware problem

Contact the manufacturer or distributor

Please check if the infrared sensor is clean or not

Clean the infrared sensor with cotton swab according to user manual

Please check the measuring way is correct or not

Ensure you have read the manual and know how to use the thermometer properly.

Please check if you have let the thermometer and patient gets stabilized in the room for 30 minutes at least

Please keep the thermometer and patient in the measuring room at least 30 minutes before using

Are you using the thermometer indoor The measurement is not accurate or if there is any doubt of the measure result

Please check if you have hold the thermometer in your hand too long and affect the accuracy Please check if there is low battery icon on LCD

Please take the measurement indoor Put the thermometer on the table in the room where the measurement is taking place and let it cool down first Replace with new battery

IMPORANT INFORMATION REGARDING ELECTROMAGNETIC COMPATIBILITY(EMC) Important information regarding Electro Magnetic Compatibility (EMC) With the increased number of electronic devices such as PC’s and mobile (cellular) telephones, medical devices in use may be susceptible to electromagnetic interference from other devices. Electromagnetic interference may result in incorrect operation of the medical device and create a potentially unsafe situation. Medical devices should also not interfere with other devices. In order to regulate the requirements for EMC (Electro Magnetic Compatibility) with the aim to prevent unsafe product situations, the EN60601-1-2 standard has been implemented. This standard defines the levels of immunity to electromagnetic interferences as well as maximum levels of electromagnetic emissions for medical devices.

17

Medical devices manufactured by Shenzhen Dongdixin Technology Co.,ltd. conform to this EN60601-1-2:2007 standard for both immunity and emissions. Nevertheless, special precautions need to be observed: The use of accessories and cables other than those specified by Caretalk, with the exception of cables sold by Caretalk as replacement parts for internal components, may result in increased emission or decreased immunity of the device. • The medical devices should not be used adjacent to or stacked with other equipment. In case adjacent or stacked use is necessary, the medical device should be observed to verify normal operation in the configuration in which it will be used. • Refer to further guidance below regarding the EMC environment in which the device should be used. (Table 1) Guidance and manufacturer’s declaration – electromagnetic emissions Caretalk EIectrical stimulators are intended for use in the eIectromagnetic enviroment specified below. The customer or the user of these EIectrical stimulator should assure that it is used in such environment. Electromagnetic environment Emissions test Compliance - guidance The Caretalk EIectrical stimulator use RF energy only for its internal RF emissions function. Therefore, its RF emissions Group 1 CISPR 11 are very low and are not likely to cause any interference in nearby electronic equipment.

18

RF emissions CISPR 11

Class B

The EIectrical stimulators are suitable Harmonic Not applicable for use in all establishments, including emissions domestic establishments and those IEC 61000-3-2 directly connected to the public low-vo Voltage ltage power supply network that fluctuations/ flicker emissions Not applicable supplies buildings used for domestic purposes. IEC61000-3-3 (Table 2) Guidance and manufacturer’s declaration – electromagnetic immunity Caretalk EIectrical stimulators are intended for use in the eIectromagnetic enviroment specified below. The customer or the user of these EIectrical stimulator should assure that it is used in such environment. IEC 60601 Compliance Electromagnetic Immunity test Test level level environment – guidance Floor should be wood, concrete, or ceramic tile. If Electrostatic ±6 kV ±6 kV floors are covered with discharge contact ±8 contact ±8 synthetic material, the (ESD) kV air kV air relative humidity should be IEC 61000-4-2 at least 30 %. Electrical fast Not Not Not applicable transient/burst applicable applicable IEC 61000-4-4 Not Not Surge Not applicable IEC 61000-4-5 applicable applicable

19

Voltage dips, short interruptions Not Not and voltage variations on applicable applicable power supply IEC 61000-4-11

Not applicable

Power frequency (50/ 60 Hz) 3A/m magnetic field IEC 61000-4-8

Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.

3A/m

(Table 4) Guidance and manufacturer’s declaration – electromagnetic immunity Eiectrical Stimulators are intended for use in the electromagnetic environment specified below. The customers or the users of these EIectrical Stimulators should assure that it is used in such environment. Immunity IEC 60601 Compliance Electromagnetic test Test level level environment – guidance

Conducted RF Not IEC 61000 applicable -4-6

20

Note1: At 80 MHz and 800 MHz, the higher frequency range applies. Note2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people. a

Field strengths from fixed transmitters, such as base stations for radio (cellular/ cordless) telephones and land mobile radio, AM and FM radio broadcast, and TV broadcast cannot be predicted the oretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Electrical Stimulator are used exceeds the applicable RF compliance level above, the Electrical Stimulator should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the Electrical Stimulator. b Over the frequency range 150 kHz to 80MHz, field strengths should be less than 3 V/m. (Table 6) Recommended separation distance between portable and mobile RF communications equipment and the Electrical Stimulator The Electrical Stimulator are intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customers or the users of these Electrical Stimulator can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Electrical Stimulator as recommended below, according to the maximum ouptut power of the communications equipment.

21

Output Power of Transmitter in Watt

Separation distance according to frequency of transmitter in meter

0.01

150 kHz to 80 MHZ d = 1.2 √P 0.12

80 MHz to 800 MHZ d = 1.2 √P 0.12

800 MHz to 2.5GHz d = 1.2 √P 0.23

0.1

0.38

0.38

0.73

1

1.2

1.2

2.3

10

3.8

3.8

7.3

23 100 12 12 For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. Note: At 80MHz and 800MHz, the separation distance for the higher frequency range applies Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.

Note: EMC tests conducted including attached electrode cord of 1 m length

22

NORMALIZED SYMBOLS Applied part of Type BF Disposal in accordance with Directive 2002/96/EC (WEEE)

0197

Complies with the Eurpean Medical Device Directive(93/42/EEC) and amended by directive 2007/47/EC requirements. Notified body TUV Rheinland (0197). The name and the address of the manufacturer The name and the address of the Authorized EC-representative in Europe Refer to Instruction Manual.

DISPOSAL Exhausted battery does not belong in household waste. Dispose of the battery according to the current federal, state and local regulation. As a consumer, you are obligated by law to return spent battery. NOTES: Please dispose of the device in accordance with the directive 2002/96/EC -WEEE (Waste of Electronic Equipment). If you have any queries, please refer to the local authorities responsible of waste disposal.

23

WARRANTY One year warranty is available from purchasing date, excluding failures caused by improper use listed below: 1) Failure resulting from unauthorised disassembly or modification; 2) Failure resulting from dropping during application or transportation; 3) Failure resulting from operation other than that described in the User’s Manual. Present the Warranty Card stamped by the distribution center.

24

Name: Shenzhen Dongdixin Technology Co., LTD. Add: No.3 Building XiliBaimang Xusheng Industrial Estate 518108,Nanshan Shenzhen, China. Tel: 0086-755-27652316 E-mail: [email protected]

Shanghai International Holding Corp. GmbH (Europe) Eiffestraße 80, 20537 Hamburg Germany Tel: 0049-40-2513175 Fax: 0049-40-255726 Manufactured for TensCare Ltd 9 Blenheim Road Epsom Surrey KT19 9BE UK Tel. NO 01372723434 www.tenscare.co.uk Copyright 2013 by Shenzhen Dongdixin Technology Co., LTD. Edition: V1.5