RSNA 2002 – Role of Imaging Physicist/Scientist in Quality Control – Course #132
Overview
Physicists’ Quality Control for MR Equipment
• MRI Standards • MRI Accreditation Program • MRI Quality Control Program – Definitions – Physicist’s Responsibilities – Technologist’s Daily & Weekly QC Tasks – Other Tests for the Physicist
Geoffrey D. Clarke, Ph.D. University of Texas Health Science Center at San Antonio Radiological Sciences Division
ACR Standard for Diagnostic Medical Physics Performance Monitoring of MRI Equipment 1999 (Res. 19)
Standards & Accreditation • ACR Standard for Performing and Interpreting MRI (Res. 16-2000)
Qualifications Qualified Medical Physicist ••certified certified in diagnostic radiological physics ••continuing continuing education in MRI physics Qualified MR Scientist ••graduate graduate degree in physical science ••3 3 years documented clinical MR experience ••continuing continuing education in MRI physics
• ACR Standard for Diagnostic Medical Physics Performance Monitoring of MRI Equipment (Res. 19-1999) • ACR Magnetic Resonance Imaging Accreditation Program
ACR Standard for Diagnostic Medical Physics Performance Monitoring of MRI Equipment 1999 (Res. 19)
ACR Standard for Diagnostic Medical Physics Performance Monitoring of MRI Equipment 1999 (Res. 19)
Responsibilities Acceptance Testing Annual Performance Evaluation Establish Continuous Quality Control Program • set action criteria • review QC program records during annual survey Provide written survey reports on timely basis
Continuous QC Program Performed by Technologist: •RF calibration for the head coil •Magnetic field gradient calibration •Image SNR & uniformity for the head coil •Processor sensitometry •Physical and mechanical inspection •Hard copy fidelity
1
12/1/2002
Physicists MRI QC
ACR Standard for Diagnostic Medical Physics Performance Monitoring of MRI 1999 (Res. 19) Equipment Physicist’s Annual Equipment Survey: All of the routine QC tests plus: – Phase stability – Magnetic field homogeneity – Calibration of all radiofrequency coils – Image signal-to-noise ratio & uniformity for all coils – Inter-slice RF interference – Artifact evaluation
ACR MRI Accreditation Program Features • Evaluates effectiveness of quality control measures • Will collect findings to further the development of quality control information • Qualified Medical Physicist should be responsible for overseeing the equipment quality control program
ACR MRI Accreditation Program
ACR MRI Accreditation Program
MRI Survey Agreement
Quality Control Section
• Official request for ACR Accreditation • Site agrees to provide all documentation, including but not limited to quality control logs, films, records, or any necessary information requested by the survey team • Agree to use the ACR MRI phantom
• Refers to tests put forth in the ACR Standard for MRI • All tests to be carried out in accordance with written procedures and methods • Preventative maintenance:
ACR Magnetic Resonance Imaging Quality Control Manual 2001
ACR Definition of Quality Assurance
G.D. Clarke (UTHSCSA)
– Documented by qualified service engineer – Repairs documented and records maintained by the MR site
All of the management practices instituted to ensure that: 1. every imaging procedure is necessary and appropriate to the clinical problem at hand, 2. the images generated contain information critical to the solution of that problem, 3. the recorded information is correctly interpreted and made available in a timely fashion to the patient's physician, and 4. the examination results in the lowest possible risk, cost, and inconvenience to the patient.
2 of 8
12/1/2002
Physicists MRI QC
Quality Assurance Policy Manual
ACR Definition of Quality Control
• responsibilities and procedures for QC testing • records of the most recent QC tests • a description of the orientation program procedures for use and maintenance of equipment; • MRI techniques to be used • precautions to protect the patient • proper maintenance of records, including records of testing, equipment service and QA meetings • procedures for cleaning and disinfection
Part of quality assurance: a series of distinct technical procedures that ensure the production of high-quality diagnostic images.
Radiologist’s Responsibilities
Responsibilities of the Qualified Medical Physicist or MRI Scientist
1. Acceptance testing 2. Establishment of baseline performance 3. Detection of changes in equipment performance 4. Verification that equipment performance faults have been corrected.
Eleven specific responsibilities including: • • • • •
To ensure that an effective quality control program exists for all MRI To select the technologist to be the primary quality control technologist To ensure that appropriate test equipment and materials are available to perform the technologist's QC tests. To arrange staffing and scheduling so that quality control tests can be carried out. To select a qualified medical physicist or MRI scientist To ensure that records are properly maintained and updated in the MRI QC procedures manual.
• Write Purchase Specifications • Perform Acceptance Testing – Baseline Measurements
• Determine Action Limits • Set up Daily/Weekly QC Tests • MRI equipment performance review
Technologist Responsibilities
MRI QC Technologist’s Responsibilities
• Designated QC Technologist(s) • QC Notebook
• Daily magnetic resonance image quality control procedures • Weekly quality control of hard copy and soft copy Images • Routine visual inspection of equipment
G.D. Clarke (UTHSCSA)
– QC policies and procedures – data forms where QC procedure results are recorded – notes on QC problems and corrective actions
• Review QC Data with QA Committee • Only Use Alternative Phantoms & Procedures when documented by physicist • Follow Established Action Limits
3 of 8
12/1/2002
Physicists MRI QC
Documenting the Quality Control Program
Technologist’s Tests • • • • • • • • •
Center frequency Table positioning Setup & Scanning Geometric accuracy High contrast resolution Low contrast detectability Artifact analysis Film quality control Visual Checklist
Daily Daily Daily Daily Daily Daily Daily Weekly Weekly
• Data form for daily equipment quality control – ACR MRI QC manual, pg. 64
• MRI Facility quality control visual checklist – ACR MRI QC manual, pg. 65
• Laser film printer control chart – ACR MRI QC manual, pg. 66
*ACR MRI QC Manual, Table 1 (p.25)
Magnetic Field Homogeneity
Annual Survey Tests • • • •
Ideal Homogeneity Good Homogeneity Poor Homogeneity
Magnetic Field Homogeneity Slice Position Accuracy Slice Thickness Accuracy Radio Frequency Coil Checks
FWHM FWHM
– Volume coils FWHM
1. Signal-to-noise ratio 2. Percent integral uniformity 3. Percent signal ghosting
ωoo
Denotes Denotes aa totally uniform uniform magnetic magnetic field. All signal signal is is at at resonant resonant frequency, ωo. o.
– Surface Coils SNR Tests
• •
Inter-Slice RF Interference Soft Copy Displays (monitors)
Magnetic Field Homogeneity
ωoo
Fourier Fourier transform transform of signal produces produces a Lorentzian Lorentzian peak in well-shimmed magnet magnet
Magnet Magnet field field homogeneity homogeneity can can be be characterized characterized using FWHM of resonance of peak peak
Magnetic Field Homogeneity
Phase images from GRE sequences with 10ms difference in TE’s
Phase and Unwrapped Phase Images
G.D. Clarke (UTHSCSA)
ωoo
4 of 8
• Overall, the phase mapping technique provides the best mechanism for evaluating field homogeneity. • Phase-maps in several planes can be obtained to determine the spherical harmonic coefficients and allows a means of “shimming” the magnet. • For some system service personnel may have to provide use of phase-mapping acquisition and analysis tools. • Filmed copy of vendor’s final homogeneity map and shim coefficients is useful for documentation and establishing a baseline.
12/1/2002
Physicists MRI QC
Slice Position Accuracy
Slice Position Accuracy
• Uses Crossed-Wedges as Reference for Positioning and Slice Spacing Accuracy • MRAP pass criterion: magnitude of bar length difference ≤ 5 mm. • The actual displacement is ½ of the measured difference • ACR Accreditation criteria are very weak, physicist may want to hold manufacturer to a higher standard
Slice Position
SLICE #1
Slice Spacing
SLICE #11
Crossed wedges should be of equal length if position and spacing are accurate
MRI Slice Thickness
Slice Position Accuracy
• Signal ramps have a slope of 10:1
• Causes of poor performance:
• Signal from ramp is 10 x slice thickness
• Operator error • Table positioning shift • Miscalibrated gradients • High Bo inhomogeneities
• Two ramps are used to compensate for inplane rotation of the phantom • Phantom does not compensate for tilting backwards or swaying left-right
Slice Thickness
Slice Thickness Measurement
4. Lower display to one/half the average. 5. Measure lengths of top and bottom ramps and calculate slice thickness.
1. Use slice 1 of ACR T1. 2. Magnify by 2 to 4. Adjust window/level to see signal ramps. (Set window at minimum.) 3. Use rectangular ROI to measure mean of middle of each signal ramp. Take average.
G.D. Clarke (UTHSCSA)
Slice Thickness = 0.2 ×
5 of 8
(top × bottom) (top + bottom)
12/1/2002
Physicists MRI QC
Slice Thickness • ACR-MRAP: slice thickness measured should be + 0.7 mm of prescribed value –
MRI Equipment Performance Evaluation
Site & Equipment Data
Site: _____________________________ MRAP Number: ____________________
Date: ________ Serial Number: ___________
Equipment: MRI System Manuafacturer: _________________ Model : ________ Processor Manufacturer:
_________________ Model: _________
PACS Manufacturer:
_________________ Model: _________
ACR MRAP Phantom Number used: _________
1. Magnetic Field Homogeneity
Bo Homogeneity
+ 14% error on 5mm slice, may be too generous
Method Used (check one): Spectral Peak ___ Phase Difference ___ Other (describe) __________________________ Measured Homogeneity:
Diameter of Spherical Volume (cm)
• Corrective actions:
Homogeneity (ppm)
________ ________ ________
– Check Axial Site Series Images – Replace cables & connectors, look for other sources of distorted RF pulse shape in RF electronics – Try switching RF coils – Check gradient calibration
Slice Position Accuracy Slice Thickness Accuracy
_______ _______ _______
2. Slice Position Accuracy From Slice Positionss #1 and #11 of the ACR Phantom: -
Wedge (mm)
=-
=+
Slice Location #1
________
Slice Location #11
________
=+
=
3. Slice Thickness Accuracy From Slice Position #1 of the ACR Phantom: Slice Thickness
Top
______
Bottom
______
(fwhm in mm)
Calculated slice Thickness (mm) ______
Duplicate these forms so they will be available for repeated use.
Uniformity Pattern
Bird-Cage Head Coil
Birdcage Coil High Field
RF coils produce uniformity patterns characteristic of their design.
Volume RF Coil Measurements
Solenoid Coil Low Field
Image Intensity Uniformity • Performance criteria: PIU ≥ 90% (high − low) percent integral uniformity = 100 × 1 − (high + low)
• Measurement Considerations: ACR Phantom Slice #7
G.D. Clarke (UTHSCSA)
6 of 8
• Display may not show signal values • Display may not allow user to set signal display level • There may not be a well-defined high/low intensity level
12/1/2002
Physicists MRI QC
Phased- Array Coils
Image Intensity Uniformity
•The signal is viewed from more “angles” •Using four channels does not produce √4 * S/N
• Causes of failure: • Poor phantom centering in head coil (usually AP) • Ghosting • Motion or vibration • Mechanical failure in head coil
Abdomen-pelvis phased-array
Torso Phased Array
Cervical-Cranial Phased Array
Surface RF Coil Measurements
Volume Coil Data % Image Uniformity
Max Signal Min Signal
Signal-to-Noise
Mean Signal SD of Background Signal
Percent Signal Ghosting
Ghost Signal Mean Signal Background Signal
Surface Coil Data Maximum Signal-toNoise
Maximum signal SD of Background Signal
4. RF Coil Performance Evaluation A. VOLUME RF COIL RF Coil Description: __________________________ Date: ____________
Slice Cross-Talk Measurements
Phantom Description: ___________________________________________
Volume Coil
Pulse Sequence: degrees
Type: ____
FOV: _____ cm2
Data Recorded
TR: _____
TE: ______
flip angle _____
Matrix: ___________ BW: _________kHz ; NSA ___ Slice thickness ______mm; spacing _______ mm TX attenuation (or gain) __________
1. Position 5mm slices on the uniform volume
Data Collected: Mean Signal
Maximum Signal
Minimum Signal
Background Signal
Calculated Values: Uniformity SNR Ghosting
Noise Standard Deviation
Ghost Signal
2. Repeat measurements decreasing the slice gap :
Calculated Values: Signal-to-Noise Ratio
Percent Image Uniformity
Percent Signal Ghosting
B. RF SURFACE COIL RF Coil Description: __________________________ Date: ____________ Phantom Description: ___________________________________________ Pulse Sequence:
Surface Coil Signal-to-Noise
Slice thickness ______mm; spacing _______ mm
Maximum Signal
Noise Standard Deviation
Image uniformity distribution OK?
________
Image ghosting OK?
________
Maximum Signal-toNoise Ratio
Window width ________
1
2
3
4
11
11
11
11
Slice Gap (mm)
min
0.5
1.0
5
3. Measure the signal-to-noise ratio (SNR) for each of the image sets.
TX attenuation (or gain) __________
HARD COPY IMAGE:
ACR MRI QC QC Manual, Manual, pg. pg. 125 125
Type: ____ TR: _____ TE: ______ FOV: _____ cm2 Matrix: ___________ BW: _________kHz ; NSA ___
-
Series # No. Slices
4. Plot the SNR vs. percentage slice gap Window level _______
Several copies of this page may be required to report on all RF coils.
G.D. Clarke (UTHSCSA)
7 of 8
12/1/2002
Physicists MRI QC
Signal-to-Noise vs. Inter-slice Gap
Soft Copy Displays • Requires precision luminance meter • Four tests
Signal-to-Noise Ratio (percent of Maximum)
100%
SE, TE = 20 FE, TE =8 TFE, TE =4
90%
– Maximum and minimum luminance – Luminance uniformity – Resolution – Spatial Accuracy
80%
70% 0%
25%
50%
75%
100%
Percentage Slice Gap
5. Interslice RF Interference Phantom Description: ___________________________________________ Pulse Sequence:
2
Type: ____ TR: _____ TE: ______ FOV: _____ cm
MRI Equipment Evaluation Summary
Matrix: ___________ BW: _________kHz ; NSA ___
S eries Number
Slice Gap (mm)
Site ___________________ Report Date: __________ System MRAP #_____________ Survey Date: __________ MRI System Manufacturer ___________ Model: __________ Physicist/MRI Scientist: ____________________ Signature: ________________________________
100 %
S ignalto-Noise Ratio M e a sure d S NR
RF Slice Interference
Number of slices______
1 2 3 4
90 %
Equipment Evaluation Tests 80 %
70 %
Summary Sheet
0% 25% 50% 75% 100% Inte r-slice Ga p (pe rce nt of slice thickne ss)
6. Soft Copy Displays Monitor Description: __________________________________________
Soft Copy Displays
-2
Maximum Luminance: ________________________ Cd m . -2
Minimum Luminance: _________________________ Cd m . Luminance Uniformity:
(pg. 131)
Average of values obtained in four corners of screen: ______ Cd m-2. Luminance measured in center of screen:
______ Cd m-2.
Percent difference: ________ % -
Review of Routine QC Program
|(Center – Average Corners)/(Center) x 100% < 30%|
7. Evaluation of Site’s Technologist QC Program 4) Set up and positioning accuracy:
(daily)
5) Center Frequency: (daily) 6) Transmitter Attenuation or Gain: (daily)
_________
1. Magnetic Field Homogeneity: 2. Slice Position Accuracy 3. Slice Thickness Accuracy 4. RF Coils’ Performance a. Volume Coils’ Signal-to-Noise Ratio b. Volume Coils’ Image Uniformity c. Volume Coils’ Ghosting Ratios d. Surface Coils’ Signal-to-Noise Ratio 5. Inter-slice RF Interference 6. Soft copy displays
Pass / Fail _________ _________ _________ _________ _________ _________ _________ _________ _________
Medical Physicist’s or MRI Scientist’s Recommendations for Quality Improvement:
______________________________________________ ______________________________________________ ______________________________________________ ______________________________________________ ______________________________________________ ______________________________________________ ______________________________________________ ______________________________________________ ______________________________________________
_________ _________
7) Geometric Accuracy Measurements: (daily)
_________
8) Spatial Resolution Measurements: (daily)
_________
9) Low Contrast Detectability: (daily)
_________
10) Film Quality Control (weekly)
_________
Visual Checklist: (weekly)
_________
ACR ACR MRI MRI QC Manual, pg. 127
MRI QC Program Summary
Successful MRI QC Program
• MRI Physicist – runs baseline tests of system performance – sets action criteria for routine ACR phantom tests – performs quarterly calibration checks with appropriate phantoms – reviews QC program
1. Technologists run QC scans on a daily basis 2. If exceed action criteria – repeat QC procedure 3. Physicist reviews QC data annually 4. Record data - report problems to service 5. Have service record problems and solutions in a service log
• Technologist – performs daily tests to assess image quality using ACR phantom – Weekly checks of hard copy output
• All measurements made, problems discovered, and actions required to resolve the problems are recorded for review
G.D. Clarke (UTHSCSA)
8 of 8
12/1/2002