Muse Cell Analyzer. User s Guide. For Research Use Only; not for use in diagnostic procedures

Muse™ Cell Analyzer User’s Guide For Research Use Only; not for use in diagnostic procedures. Copyright © 2012 Millipore Corporation, Inc. All righ...
139 downloads 2 Views 3MB Size
Muse™ Cell Analyzer User’s Guide For Research Use Only; not for use in diagnostic procedures.

Copyright

© 2012 Millipore Corporation, Inc. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, stored in retrieval systems, or translated into any form, or by any means: electronic, mechanical, magnetic, optical, or otherwise, without the prior written permission of Millipore Corporation, 25801 Industrial Blvd, Hayward, CA 94545, United States of America.

Disclaimer

Millipore reserves the right to change its products and services at any time to incorporate the latest technological developments. This guide is subject to change without notice. Millipore has not validated the analysis of Muse™ System data using third-party programs and cannot warrant that the results using these programs will be correct. Millipore does not provide support for any third-party programs.

Limitations

The Muse™ System is for research use only; not for use in diagnostic or therapeutic procedures. The results of the assays are dependent upon the proper use of the reagents and instrument. Please refer to the appropriate reagent kit user’s guide for specific instructions and limitations.

Trademarks

The M logo is a trademark of Merck KGaA, Darmstadt, Germany. Millipore, Muse, and guava are registered trademarks or trademarks of Millipore Corporation. Microsoft and Excel are registered trademarks of Microsoft Corporation in the United States or other countries.

Patents

U.S. patents 5,798,222; 6,816,257; 6,710,871 and foreign equivalents owned by Millipore Corporation. Patent App No. 61482946 owned by Millipore Corporation. The Muse™ Count & Viability Reagent is the subject of U.S. Patent 6,403,378 owned by Millipore Corporation.

Contents Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 About this Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Conventions Used in This Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . Help . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . General Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Biological Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Electrical Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Laser Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Precaution Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Limitations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2 2 2 3 3 3 3 3 4 4

Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Muse™ System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Setting Up Your System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Unpacking the Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5 6 6 6

System Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Muse™ Instrument Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Sample Loader . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Cleaning Solution and Waste Bottles . . . . . . . . . . . . . . . . . . . . . 11 Fluid System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 Software Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Navigating Through the Software . . . . . . . . . . . . . . . . . . . . . . . . 12 Main Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Entering and Selecting Information . . . . . . . . . . . . . . . . . . . . . . . 15 Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 Instrument Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 Assay Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

iii

Getting Started . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27 Turning On the Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Turning Off the Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Logging Onto the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Logging On as the First User . . . . . . . . . . . . . . . . . . . . . . . . . . . . Logging On Once Users are Added . . . . . . . . . . . . . . . . . . . . . . . . Selecting an Assay to Run . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

27 28 29 29 32 36

System Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 Running a System Check Procedure . . . . . . . . . . . . . . . . . . . . . . . . . System Check Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Printing Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Viewing Past System Check Results . . . . . . . . . . . . . . . . . . . . . . . System Check Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Clearing the System Check Log . . . . . . . . . . . . . . . . . . . . . . . . . . Exporting the Check Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Exporting the Service Check File . . . . . . . . . . . . . . . . . . . . . . . . . Viewing a Trend Graph . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Viewing the System Check Event Log . . . . . . . . . . . . . . . . . . . . .

37 42 43 44 45 45 46 46 47 48

Cleaning and Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . 51 Cleaning Protocols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Cleaning the Outside of the Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . Rinsing the Capillary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Running Quick Clean . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Backflushing the Capillary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Performing a Complete System Clean . . . . . . . . . . . . . . . . . . . . . . . . Printing the Cleaning Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Cleaning Log Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Clearing the Cleaning Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Exporting the Cleaning Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Viewing and Exporting the Cleaning Event Log . . . . . . . . . . . . . Resetting Fluid Levels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Filling the Cleaning Solution Bottle . . . . . . . . . . . . . . . . . . . . . . . . . Emptying the Waste Bottle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iv

Muse™ Cell Analyzer User’s Guide

51 52 53 53 54 55 58 59 60 60 61 63 64 65

Cleaning the Capillary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66 Replacing the Flow Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67 Preparing the Unit for Depot Service . . . . . . . . . . . . . . . . . . . . . . . . 69

Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71 Setup Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Users & Access Levels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Adding a New User . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Removing a User . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Editing User Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Recommended Quick Clean . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Printing Layout Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Editing Favorites . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Managing Instrument Data Storage . . . . . . . . . . . . . . . . . . . . . . . . . Exporting by User . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Exporting by Date . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Importing All User Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Exporting All User Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Software Update . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Service Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

71 73 74 75 76 76 77 78 80 81 82 83 84 85 86

Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87 Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87

Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91 Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97

v

vi

Muse™ Cell Analyzer User’s Guide

Preface The Muse™ System is the first compact, easy-to-use, desktop cell analysis system that can perform a wide range of cellular assays.

Assay and Reagent Overview For information on performing the assay protocols, refer to the specific kit user’s guide, which can be found at www.Millipore.com/muse. For order information, see page 95. For Research Use Only; not for use in diagnostic procedures. •

Muse System Check Kit Verifies the performance of the system by assessing counting accuracy and fluorescence detection using a standardized fluorescent bead reagent. The kit contains a bead reagent and diluent.



Muse™ Count & Viability Kit Used to determine viability and total cell count. Accurate assessments can be made with a wide variety of cell lines, even those with unusual culture conditions or a tendency to aggregate.



Muse™ Count & Viability Reagent (200X) Enables accurate counting and discrimination of viable and nonviable cells for a variety of difficult cell sample.



Muse™ Annexin V & Dead Cell Kit Used to assess early and late apoptosis. The assay relies on the translocation of phosphatidyl serine (PS) to the outer surface of the cell membrane, an event often associated with the onset of apoptosis.



Muse™ Cell Cycle Kit Identifies and measures the number of cells within the various phases (G0/G1, S, and G2/M) of the cell cycle.



Muse Cell Dispersal Reagent Enzymatic reagent that gently disaggregates clumped cells in suspension, improving the accuracy and precision of cell counts. The formulation of CDR has been optimized for use with Muse™ Count & Viability Reagent.

Preface

1

About this Guide The Muse™ Cell Analyzer User’s Guide provides detailed information on operating and maintaining the Muse™ System. This guide provides instructions for setting up the system, managing users and data files, running a System Check procedure, and cleaning the system. It does not include instructions for using the software modules to acquire samples and analyze data. For information on preparing samples and running them on the instrument, refer to the specific kit user’s guide located at www.millipore.com/muse.

Conventions Used in This Guide ■ NOTE: Points out additional information that may be helpful. ▲ WARNING: Alerts you to situations that could result in bodily harm, instrument damage, failure in a procedure, or incorrect results.

Help 1

Read through the section of the guide specific to the operation you are performing. Refer to the table of contents and index to locate information. A glossary is included to assist you with any unfamiliar terms.

2

See the troubleshooting section for a list of problems and suggested solutions.

3

Refer to the technical support contact information below:

4

2



For ordering information or technical support, call toll-free in the USA and Canada: Phone: +1 (800) MILLIPORE (645-5476) Fax: +1 (800) 645-4539



Outside the US, visit www.millipore.com/offices for up-to-date worldwide contact information.

For information on performing an assay, refer to the specific kit user’s guide, which can be found at www.millipore.com/muse.

Muse™ Cell Analyzer User’s Guide

Safety The Muse™ System is equipped with safety features for your protection. Use the system only as directed in this user’s guide. Do not perform instrument maintenance or service except as specifically stated. Read the following safety information before using the system.

General Safety ▲ WARNING: If this instrument is not used in the manner indicated by the instructions in this guide, the safety features of the instrument may be impaired. Follow these guidelines: •

Use only the tubes specified. The use of tubes other than those specified may result in damage to the instrument.

Biological Safety ▲ WARNING: All biological specimens and materials that come into contact with them can transmit potentially fatal disease. To prevent exposure to biohazardous agents, follow these guidelines: •

Handle all biological specimens and materials as if capable of transmitting infection. Dispose of waste using proper precautions and in accordance with local regulation. Never pipette by mouth. Wear suitable protective clothing, eyewear, and gloves.



After emptying waste bottle, add bleach to the first (fill) line. It takes approximately 10 mL of bleach to reach the fill line. Dispose of waste in accordance with federal, state, and local regulations.

Electrical Safety ▲ WARNING: Power off the instrument using the Power Options tool on the main menu and disconnect the power cord before replacing fuses.

Laser Safety The Muse™ System contains a Class IIIb laser operating at 532 nm in CW mode. Light shields within the instrument enclose the path of laser radiation. Additionally, the instrument enclosure provides secondary protection from any laser radiation.

Preface

3

▲ WARNING: To avoid exposure to laser radiation or electric shock, follow these guidelines: •

Do not open the instrument or attempt to perform any internal maintenance. There are no user serviceable parts.



Turn off the power to the system before removing the flow cell.

Precaution Labels The following labels are affixed to the Muse™ System.

Muse™ Cell Analyzer EMD Millipore Corporation 25801 Industrial Blvd. Hayward, CA 94545 (510) 576-1400

Class 1 Laser Product

Complies with 21 CFR Chapter 1 Subchapter J. Assembled in USA with Components of Foreign Origin The M logo is a registered trademark of Merck, KGaA, Darmstadt, Germany. Muse is a trademark of Merck, KGaA, Darmstadt, Germany.

7200120001

1000-5880

Rev A

Limitations •

The Muse™ System is for research use only; not for use in diagnostic or therapeutic procedures.



The results of the assays are dependent upon the proper use of reagents and instrument. Refer to the appropriate kit user’s guide located at www.millipore.com/muse for specific instructions and limitations.

4

Muse™ Cell Analyzer User’s Guide

Introduction Muse™ System The Muse™ System consists of a compact, portable, and easy-to-use cell analyzer, software, and optimized reagents. The Muse™ Cell Analyzer uses patent-pending, miniaturized fluorescent detection and micro-capillary technology to deliver quantitative cell analysis of both suspension and adherent cells 2 to 60 μm in diameter. The software includes a dedicated module for each assay, as well as a tools for verifying instrument performance and cleaning the instrument’s fluid system. Convenient mix-and-read assays are optimized for accuracy and convenience. The available assays include: •

Count & Viability



Annexin V & Dead Cell



Cell Cycle

Refer to the assay user’s guide for detailed instructions on preparing samples and running an assay using the assay-specific software module.

Muse™ Cell Analyzer

Introduction

5

Setting Up Your System Your Muse™ System is easy to install. Simply plug it in and start operating.

Components Your Muse™ System shipped with the following components: •

Muse™ Cell Analyzer and power cord



cleaning and waste bottles and corresponding color-coded tubing



flow cell



USB flash drive with software recovery files and user’s guide

Unpacking the Instrument Follow the instructions in the Quick Start Guide to unpack and set up your instrument. Here are the steps outlined briefly. 1

Remove the instrument from the shipping box. You may wish to save the box and foam insert in the event the instrument needs to be moved to a different location.

2

Place the unit on a stable surface. Although it is compact and portable, it contains precisely aligned optical components that are sensitive to jarring movements.

3

Connect the color-coded fluid tubing to the back of the instrument.

4

Add 10 mL of bleach to the waste bottle and fill the cleaning solution bottle to the indicator line with Guava ICF. Place the fluid bottles in their respective receptacles. Match the icon on the fluid bottle with the icon in the receptacle.

5

Connect the fluid tubing from the back of the instrument to the top of the corresponding bottle.

6

Insert the flow cell. See “Replacing the Flow Cell” on page 67 from step step 4 on for details.

7

Plug the instrument into a grounded three-prong AC outlet. When you plug in the instrument for the first time, it will automatically turn on.

8

Log on as an Administrator. See “Logging On as the First User” on page 29.

6

Muse™ Cell Analyzer User’s Guide

9

Perform a Complete System Clean to prime the fluid system. See “Performing a Complete System Clean” on page 55.

10 Perform a System Check procedure. See “Running a System Check Procedure” on page 37.

Introduction

7

8

Muse™ Cell Analyzer User’s Guide

System Overview Muse™ Instrument Overview The Muse™ Cell Analyzer is a compact instrument for fluorescence-based, three parameter cell analysis. It is designed for easy and intuitive operation and minimal maintenance. The integrated touch screen provides an easyto-use software interface for operating the system. To turn on the instrument, press the power button located below the touch screen on the front of the unit. To turn off the instrument, use the Power Options feature on the main menu displayed on the touch screen. Five USB ports—one below the touch screen and four on the back of the unit—allow you to connect a Muse™-compatible printer, USB flash drive, and/or keyboard and mouse. For a list of approved printers, see “Approved Printers” on page 98. waste bottle cleaning solution bottle flow cell access door

touch screen

tube holder sample loader arm power button

System Overview

9

Sample Loader The sample loader holds an individual sample tube. The loader arm can be lifted easily with your finger and released with the touch of a button using the touch screen. The following tube is supported: •

1.5-mL microcentrifuge tube with conical tip and screw cap (if snapcap tubes are used, cut off the cap)

The system automatically detects tubes with caps and will not load the tube if you install a tube with a cap. Loading and Unloading a Tube To load a tube, place the tube in the tube holder and lift up on the loader arm. As you lift the arm, the tube automatically slides back to the loaded position and is seated under the capillary.

During a run, the loader arm drops automatically when sample acquisition is complete. To manually unload a tube, use the eject button on the touch screen. Additionally, a lever located in the flow cell access door allows you to drop the loader arm to remove a sample tube in the rare event of a power outage. Open the flow cell access door and push back on the switch shown in the following illustration.

loader arm release lever Push back to drop loader arm.

10

Muse™ Cell Analyzer User’s Guide

Cleaning Solution and Waste Bottles The cleaning solution bottle can be filled with Guava Instrument Cleaning Fluid (ICF) for system cleaning. For information on filling the bottle, refer to “Filling the Cleaning Solution Bottle” on page 64. The waste bottle captures the sample fluid after it exits the fluid system. Add 10 mL of bleach to the waste bottle after you empty it and before installing it in the unit. For information on emptying the waste bottle, refer to “Emptying the Waste Bottle” on page 65.

Fluid System Sample uptake occurs through the capillary part of the flow cell assembly. Sampling is regulated by a variable-speed fluid pump. The pump does not require sheath fluid or supplementary fluids for sample acquisition.

capillary

Because the system’s sampling precision depends on the integrity of the fluid pathway, it is important to maintain a clean system. Do not allow samples to remain in the capillary for extended periods of time as they may eventually clog the system. Perform frequent cleaning cycles to prevent the build-up of cellular debris that may restrict sample flow. If a clog does occur, you can clear it by using the backflush feature, which reverses the flow of fluid and flushes it out of the capillary at high speed. See “Backflushing the Capillary” on page 54 for detailed instructions on using the backflush feature. At the start of each day perform a Complete System Clean to prime the fluid system. See “Performing a Complete System Clean” on page 55. Always leave a tube of deionized water on the instrument when not in use.

System Overview

11

Software Overview Use the software to acquire samples, view results, and control the instrument. In addition, each software module allows you to adjust the markers and/or gates to fine-tune the analysis. Instructions for running samples using the individual assay modules can be found in the reagent kit user’s guide for each individual assay kit.

Navigating Through the Software A navigation bar at the top of the screen indicates where you are in a given procedure. The current step is highlighted gray. These buttons vary depending on the procedure you are performing (for example, an assay, Complete System Clean, or System Check procedure). If you access one of these procedures and wish to exit, simply press Home to return to the main menu. If the screen you’re on does not display a navigation bar, select the Back, Close, or Cancel button from the current screen to access a screen with a navigation bar. For simplicity, the navigation bar has been cropped out of most of the screens appearing in this guide. Navigation bar

A status bar at the bottom of the screen indicates the user currently logged on, instrument status, and the volume remaining in the cleaning bottle and capacity remaining in the waste bottle.

Example shows waste bottle has 23% capacity remaining (77% full) and cleaning bottle has 20% of the solution remaining (80% has been used).

12

Muse™ Cell Analyzer User’s Guide

Main Menu The main menu allows you to choose an assay or an Essential Tool. Navigation bar Title bar with Setup and tube eject buttons

Select an assay.

Select a tool.

Below the navigation bar, is the title bar containing the Setup button and eject button . The eject button releases the sample loader arm, allowing you to manually unload a tube. Setup, which is accessible from the main menu only, allows you to set specific system features such as managing users and data, and setting specific software options. See “Setup” on page 71 for details on the Setup menu. Title bar

The lower part of the screen is divided into two main sections—Favorites (assays) and Essential Tools. Favorites allows you to select an assay to run. Essential Tools allows you to check the system performance, clean the system, and power off or log off of the system.

System Overview

13

Favorite Assay Buttons The main menu allows you to select from the following assays: •

Count & Viability for performing cell counting and viability assays



Annexin V & Dead Cell for performing apoptosis assays



Cell Cycle for performing cell cycle assays

Essential Tools The following tools are available: •

System Check to check the system’s counting and fluorescence performance prior to running samples. The System Check Beads are standard particles used with this tool. For more information on running System Check, see page 37.



Complete System Clean to clean the fluid system. For information on this cleaning procedure, see “Performing a Complete System Clean” on page 55.



Power Options to turn off the system or log off as the current user

14

Muse™ Cell Analyzer User’s Guide

Entering and Selecting Information The touch screen provides various ways of entering and selecting information, most of which are intuitive. If you connect an external mouse, you can use the cursor to make selections. •

Touch buttons, such as the assay buttons on the main menu.



Drag sliders to increase or decrease values, or drag handles to adjust gates and markers.



Use the keypad to enter information into text or numeric fields. Touch the text field to automatically open the keypad. Type the information and select Done or touch outside the keypad to close it.

System Overview

15

Files For each assay, the software saves two files: an FCS 3.0 data file and a CSV spreadsheet results file, each containing data for all samples in the run. Additionally, you can optionally choose to save a separate file of the instrument and analysis settings. FCS 3.0 Data Files A single FCS file includes the results and sample information for all samples acquired within a run, as well as an event log At the completion of the run (after you click Finish to save and close the data set), you can enter a new file name or use the default file name that appears. The default file name is the user’s initials, followed by the date and time (for example, PT_15Feb2012_115618.FCS). The file name assigned to the FCS file is also assigned to the spreadsheet results file. An extension is automatically appended to the file name you enter. The first three characters represent the assay type, followed by FCS. Assay

File name

Count & Viability

filename.VIA.FCS

Annexin V & Dead Cell

filename.NEX.FCS

Cell Cycle

filename.CCY.FCS

Spreadsheet File A spreadsheet file containing the data for all samples within a data set is automatically saved along with the FCS file in the same directory. The spreadsheet file contains a summary of the statistical results, sample information, and instrument settings for each sample run within a data set. It can be opened and analyzed using a spreadsheet program such as Microsoft Excel. The same file name assigned to the FCS data file is also used for the spreadsheet file, except it has the extension .CSV appended to the file name.

16

Muse™ Cell Analyzer User’s Guide

Event Log Each time you run an assay, the system saves a log containing a list of all events that occurred during the assay. This information is contained within the FCS data file. To view this list of events, select Options from the Results screen for the assay, then select Event Log. ■ NOTE: There is also an event log for the System Check procedure and the Complete System Clean procedure. For information on those event logs, refer to the System Check and Cleaning chapters, respectively. You can filter the list to view statuses, actions, errors, and/or warnings. Select the appropriate check box(es) to display the types of events you wish to view.

Example of Count & Viability event log

If errors or warnings occur during a run, a message appears in red below the plots, indicating that errors/warnings have been logged. Errors System Check: •

System Check failed due to avg MFI for [channel] – actual [x], Target value = [y] System Check failed due to percent CV for [channel] –value = [x] System Overview

17

Warnings System Check: •

Loader arm is down.



Concentration failed for replicate [1/2/3]



Avg conc failed



%CV failed

All assays: •

Sample Loader is down. Please raise the Loader arm.



Adjust Settings timed out. Please re-enter Adjust Settings if necessary to complete the instrument set-up. (not applicable for System Check)



The run timed out before enough events were acquired.

Annexin V and Count & Viability •

More than 500 particles/μL. Sample is too concentrated. Please dilute or accuracy may be compromised.



Less than 10 particles/μL. Sample is too dilute. Accuracy may be compromised.

Cell Cycle: •

More than 1200 particles/μL. Sample is too concentrated. Please dilute or accuracy may be compromised.



Less than 25 particles/μL. Sample is too dilute. Accuracy may be compromised.

18

Muse™ Cell Analyzer User’s Guide

Folder and File Structure When you export data to a USB drive using any of the export features, a folder structure is created on the USB drive. The following diagram outlines this folder structure. Muse User Name Assay Name Data set files (.fcs, .csv) Spreadsheet files (.csv) Event log (.log) Instrument settings files (.gst) Assay reports (.pdf) System Clean Cleaning log (.log) Cleaning log (.pdf) System Check Service Check file (.zip) Replicate1.CHK.fcs Replicate2.CHK.fcs Replicate3.CHK.fcs MuseCheckLog.csv System Check log (.csv) Event log (.log) Trend Graph report (.pdf) System Check report (.pdf) Public

Files/folders generated by Muse software Files generated by users through Export Files generated by users through “Print to PDF” function.

The Public folder is created automatically and allows administrators to share data. If an administrator wants other users to have access to certain data and/or instrument settings files, they can export the data to the Public folder. While operator-level users can access this data, they cannot modify the files or save files to the Public folder. ■ NOTE: The System Check and System Clean folders are created only if a user exports System Check and System Clean files.

System Overview

19

File Storage The Muse™ System has an internal hard drive for saving data files. To see the available storage, select Setup from the main menu, then select Manage Instrument Storage. The amount of available storage is displayed, as well as the amount of storage being used. For more information on managing instrument storage, see “Managing Instrument Data Storage” on page 80.

Available and used space

Appending Files You can open an existing data set and append data to this file. 1

20

At the start of the assay select View Results from the assay screen. The Retrieve Data Set screen appears.

Muse™ Cell Analyzer User’s Guide

2

Select the location where the file is located. Keep in mind that you will not have access to the files from other users unless the files were exported to the Public folder.

3

Use the Filter field to sort by the file name or any part of the name. Or, use the calendar to search by date. Touch the calendar to open it and select the date.

4

Select the file and then select Retrieve.

Instrument Settings The system allows you to save instrument settings files. These files contain instrument settings, as well as gates and marker settings. The extension .GST is appended to the file name. You can retrieve these files from the MUSE directory, the Public folder, or USB drive when you start an assay. If you are an operator-level user, you can only access instrument settings from your own data folders in the Muse directory and USB drive. Assay

Instrument Settings File Name

Count & Viability

filename.VIA.GST

Annexin V & Dead Cell

filename.NEX.GST

Cell Cycle

filename.CCY.GST

You can recall this file later to: •

download the instrument settings to the instrument for acquisition



apply the gates and markers to data during acquisition

System Overview

21

Retrieving Instrument Settings 1

From the Settings screen at the start of a run, select Retrieve Settings.

2

Select the location where the settings file is stored.

3

Select the file and then select Retrieve. You can filter the list by date. Select the calendar and choose the date when the settings were saved. Or, if you know the name of the file, enter any part of it in the Filter text field.

22

Muse™ Cell Analyzer User’s Guide

The settings are downloaded to the instrument. The system prompts you to run the adjust settings step. •

Select NO if you wish to skip the adjust settings step and use the settings you just retrieved to run the assay. This will advance you to the Sample Info screen in preparation to run the first sample.



Select YES if you wish to perform the adjust settings step using the settings you just retrieved. This will help to ensure that the retrieved settings are correct.

Saving Instrument Settings 1

Select Options from the Results screen after the sample is acquired.

2

Select Save Current Settings.

3

Select the location where you want to save the file. The default directory is the MUSE directory on the unit’s hard drive. If you want to save the file to a USB drive, select the USB drive from the Save In field. A new directory will be created on the USB drive (MUSE\user name\assay name\instrument settings file).

4

Enter a file name or use the default file name.

5

Select Save.

System Overview

23

Assay Overview Detailed steps for each assay are covered in the reagent kit user’s guide which can be found at www.millipore.com/muse. Some additional information on various tasks that can be performed from the assay screens is listed in the following section.

Example of Count & Viability results screen

24

Muse™ Cell Analyzer User’s Guide

Data Set Options Use the Options button on the Results screen to access features that allow you to rename the data set, export data, save instrument settings, and view the event log. 1

Select Options in the title bar of the results screen. The Data Set Options screen appears.



To rename a data set, select Rename Data Set, edit the file name, and select Apply. You can also rename the data set at the completion of the run when you select Finish.



To export the data set, select Export Data Set, select the USB drive from the Export Location field, enter a file name or leave the default. Select Export. You may want to export the data set so that you can share the data with other users.



To export the data to a spreadsheet file, select Export to Spreadsheet, select the USB drive from the Export Location field, enter a file name or leave the default. Select Export.



To save the current instrument settings, see “Saving Instrument Settings” on page 23.



To view or export the event log, see “Event Log” on page 17.

System Overview

25

About Dialog The About dialog provides the software version number and copyright information, as well as technical support information. To access the About dialog touch the title bar of the main menu or any assay screen. Touch the title bar to open the About dialog.

26

Muse™ Cell Analyzer User’s Guide

Getting Started To prepare the system for use perform the following steps each day that you use the instrument. •

Empty the waste bottle and add bleach. See “Emptying the Waste Bottle” on page 65.



Fill the cleaning solution bottle with ICF. See “Filling the Cleaning Solution Bottle” on page 64.



Reset the fluid levels. See “Resetting Fluid Levels” on page 63.



Perform a Complete System Clean. See “Performing a Complete System Clean” on page 55.

Turning On the Unit 1

Ensure the instrument is plugged in.

■ NOTE: If you are plugging in the instrument for the first time, it will automatically turn on. After a brief startup sequence, the logon screen appears. Refer to “Logging On as the First User” on page 29. 2

Press the power button located on the front of the unit below the touch screen. A screen appears allowing you to select your user name from a list. Refer to “Logging On Once Users are Added” on page 32. Getting Started

27

3

Fill the cleaning solution bottle to the fill line with Guava ICF. Empty the waste solution bottle, then add bleach to the first fill line (approximately 10 mL). See “Filling the Cleaning Solution Bottle” on page 64 and “Emptying the Waste Bottle” on page 65 for details.

4

Whenever you fill the cleaning solution bottle and empty the waste bottle, you will need to reset the fluid levels. Resetting the fluid levels sets the fluid indicators in the status bar to 100% and 100%. See “Resetting Fluid Levels” on page 63 for information.

5

Run a Complete System Clean at the start of each day to prime the fluid system. See “Performing a Complete System Clean” on page 55.

Turning Off the Unit 1

Run the Complete System Clean procedure at the end of the day before shutting down the unit. See “Performing a Complete System Clean” on page 55 for details.

2

Leave the tube of DI water on the sample loader.

▲ WARNING: Do not leave a tube of Guava ICF, bleach, or any other cleaning agent loaded on the instrument overnight or for an extended period of time. Prolonged exposure to strong oxidizing agents will damage the flow cell. Always leave a fresh tube of DI water on the system when shutting it down. Change the tube of water regularly to ensure it is clean and free of particles. 3

28

Select Power Options from the main menu under Essential Tools.

Muse™ Cell Analyzer User’s Guide

4

Select Power Off to turn off the system.

Logging Onto the System The system allows two levels of users—administrator and operator. For information on the different user levels and access privileges, see “Users & Access Levels” on page 73.

Logging On as the First User The first time you log onto the system, you will be prompted to select Administrator or Guest. If you select Administrator you will have administrator privileges. If you select Guest you will have operator privileges. 1

To add new users and customize the system, select Administrator. ■ NOTE: Select Guest if you do not have a user account but want to use the instrument. Files will be saved with GST initials. ■ IMPORTANT NOTE: Once you log on as an administrator, it is a good idea to add yourself as a user with administrator-level access. When you add a new administrator-level user, the default Administrator user (shown below) will be overwritten. For this reason, we do not recommended running assays and acquiring data

Getting Started

29

sets as a default Administrator, as the data will no longer be accessible.

Default Administrator and Guest users

2

The main menu appears. Touch the date and time located in the lowerright corner of the screen.

Set date and time.

30

Muse™ Cell Analyzer User’s Guide

3

Use the arrow buttons above and below each field to select the day, month, year, hours, and minutes.

4

Select Apply.

5

See “Adding a New User” on page 74 for information on adding new users.

Getting Started

31

Logging On Once Users are Added When you start the system, a screen appears allowing you to select your user name from a list. 1

32

Select your name from the list to display the main menu. If your name does not appear, contact your system administrator.

Muse™ Cell Analyzer User’s Guide

2

Enter your PIN and press OK. If security is not an issue and you choose not to enter a PIN each time you log in, you can select Remember User PIN. The next time you log in, simply select OK to accept the PIN.

The main menu appears after your PIN is correctly entered. Forgetting Your User PIN If you forget your PIN, your administrator can reset it by assigning you a new PIN. If you are an administrator and you forget your PIN, another administrator can reset it. If you are the sole administrator and you forget your pin, you will no longer have administrator access to the system and therefore cannot perform administrator functions. For this reason, it is a good idea to save your PIN in a secure place. If you are the sole administrator and you forget your PIN, you can export your user information to a USB drive and send it to Millipore. We will send a PIN back to you, which you can then import.

Getting Started

33

1

If you forgot your PIN, select Done or touch outside the keypad to close it, then select Forgot User PIN.

2

Insert a USB drive.

34

Muse™ Cell Analyzer User’s Guide

3

Select Export User Info to USB thumb drive. ■ NOTE: Be aware that once you export your PIN you will no longer be able to log onto the system even if you should remember your PIN.

4

Select the drive you want to export to, then select Save.

5

Contact Tech Support. You will be asked to e-mail the zip file from your thumb drive to Tech Support.

6

Tech Support will e-mail you another file which you will save to a USB drive.

7

Insert the USB with the file from Tech Support. Begin to log on—select your user name and Forgot User PIN. Then select Use PIN from Zip in USB thumb drive.

8

Follow the steps to reset the PIN.

Getting Started

35

Selecting an Assay to Run The main menu displays the following assays: •

Count & Viability



Annexin V & Dead Cell



Cell Cycle

For information on performing a System Check, see “Running a System Check Procedure” on page 37. For information on running an assay, refer to the kit user’s guide, which can be found at www.millipore.com/muse.

36

Muse™ Cell Analyzer User’s Guide

System Check The System Check procedure is used to verify the performance of your Muse™ System by assessing counting accuracy and fluorescence detection. The Muse™ System Check Kit contains Muse™ System Check Beads and Muse™ System Check Diluent. Perform the System Check procedure daily before running any of the assays to ensure that the instrument provides reliable, accurate results.

Running a System Check Procedure Run a System Check at the start of each day that you use the instrument to ensure that it is performing properly. Three replicates of the System Check Bead sample are acquired. The results are averaged to determine if they are within the expected range. Run a Complete System Clean at the start of each day and before performing the System Check procedure. A cleaning cycle will prime the fluid system and remove bubbles that may have formed in the tubing. See “Performing a Complete System Clean” on page 55 for details. 1

Prepare a 1:20 dilution of System Check Beads. Refer to the Muse™ System Check Kit User’s Guide for information.

2

Select System Check under Essential Tools at the main menu to display the System Check screen.

3

A message appears prompting you to check the fluid levels in the cleaning and waste bottles. Check the fluids, then click Close. Always remember to reset the fluid levels when you fill the cleaning solution bottle and empty the waste bottle. For information on resetting the fluid levels, see “Resetting Fluid Levels” on page 63.

System Check

37

4

Select Run System Check.

Run System Check.

View the result of the last System Check.

5

The first time you run the procedure, enter the bead lot number, expiration date, and check code. •

Enter the Bead Lot # and press Done on the keypad.



Touch the calendar icon in the Exp. Date field to select the expiration date. Touch outside the calendar to close it.



Touch the Check Code field and enter the code.

All values are required and can be found on the information card that comes with the bead kit. Once you enter this information, it will remain in the software. Each time you run the procedure, check the information to ensure it is accurate. Update the values when a new lot number of System Check Beads is used, if necessary.

38

Muse™ Cell Analyzer User’s Guide

6

Select Next.

Touch a field to display a keypad allowing you to enter the values.

7

Mix the tube of prepared beads and load it on the system.

8

Select Run.

The system performs a prime, then acquires the first replicate. The progress bar and fly wheel provide indicators as to the status of acquisition. The progress bar is divided into three sections—one for each replicate. If the fly wheel is turning but the progress bar in not System Check

39

advancing, the fluid system may be clogged or the beads may have settled to the bottom of the tube. If the beads settled, select Abort, unload the tube and mix. Then reload and select Run System Check again.

Progress bar gives an estimate of the acquisition status for each replicate. Fly wheel lets you know the instrument is acquiring sample.

9

Remove the tube and vortex it to resuspend the beads.

10 Load the tube and select Run. The system acquires the second replicate.

40

Muse™ Cell Analyzer User’s Guide

11 Repeat steps 9 and 10 to acquire the third replicate. Upon completion, the system displays a PASS/FAIL result and the Particles/mL value for each replicate, as well as the average.

View past System Check results. Bead lot, date, and time

PASS/FAIL result

Status field

The bead lot number, date, and time of the run appears just above the Bead Info and printer tabs. All runs in the System Check log are displayed in this format. The Status field at the bottom of the screen indicates that the System Check was completed.

System Check

41

System Check Results The result appears as PASS or FAIL. The bead count (Particles/mL) for each replicate and the average are displayed, as well as the %CV. If any result for Particles/mL falls outside ±10% of the expected value, the result is outside the acceptable range and appears in red. If the procedure fails, touch the Help button (?) to display troubleshooting information. ■ NOTE: The System Check procedure can also fail if the red and yellow mean fluorescence intensities (MFIs) fall outside the expected results. If all Particles/mL values are green, but the result is FAIL, contact Millipore Technical Support.

Touch Help button to get troubleshooting information.

42

Muse™ Cell Analyzer User’s Guide

Printing Results To print the results, select the print tab to the right of the Bead Info tab on the Results screen. You can choose to print the current results, the log list (see “Viewing Past System Check Results” on page 44), or a trend graph. 1

Select the printer from the drop-down list of printers. The list displays all available printers. If no printers are detected, the default printer is PDF to USB, which allows you to export a PDF.

2

Select the Print Current Check button to print the current results. You can also print the entire log list (past System Check results) or the trend graph. If you select to print the trend graph, the number of runs included in the printout is based on the number of runs selected at the Trend Graph screen (see “Viewing a Trend Graph” on page 47).

3

(Optional) Select Print Preview if you want to view the results as they will print.

4

Select Print. If no printers are available or if you wish to export to a PDF, select PDF to USB, then select Print. The Export to PDF screen appears. Select the USB port, enter a file name for the report file, or leave the default name. Select Export.

Log List button Print tab

Print results, log of all results, or trend graph.

System Check

43

Viewing Past System Check Results To see the results from previous runs, select the Log List button below the Clean button on the Results screen. You can also use the arrows to the right of the Log List button to quickly scroll through and display previous System Check results in the Bead Info tab. Each entry in the System Check Log represents one System Check run and includes the user who ran the test, the bead lot number, the date and time the test was run, and whether or not the test passed (Y or N). A hyphen indicates that the run was aborted. Use the arrow buttons at the bottom of the screen to scroll through the pages of results. Select a run from the list to display the selected results in the Bead Info tab.

For information on clearing and exporting the System Check Log, or viewing a trend graph or the event log, see “System Check Options” starting on page 45.

44

Muse™ Cell Analyzer User’s Guide

System Check Options Select Options at the top of the System Check Results screen to export or clear the System Check results log, export the Service Check file, or view the trend graph or event log.

Clearing the System Check Log The System Check Log is a list of all past System Check results. After exporting the System Check Log, you may wish to delete all entries in the log. Once the data is cleared, you will no longer be able to view past results or generate trend graphs. Only administrator-level users can clear the System Check Log. 1

Select Options from the System Check Results screen.

2

Select Clear Check Log.

3

Select Clear to confirm.

System Check

45

Exporting the Check Log The System Check Log is a list of all past System Check results. Only administrator-level users can export the System Check Log. 1

Select Options from the System Check Results screen.

2

Select Export Check Log.

3

Select the USB drive from the Export Location field.

4

Enter a name for the log file, or leave the default name.

5

Select Export.

Exporting the Service Check File The Service Check file is a zipped file containing the detailed results from the most recent System Check run. This file is used by service personnel to troubleshoot your system if your System Check results continue to fail. Use the Export Service Check File feature to export the file so that you can send it to Millipore. 1

Select Options from the System Check Results screen.

2

Select Export Service Check File.

3

Select the USB drive from the Export Location field.

4

Enter a name for the log file, or leave the default name.

5

Select Export.

46

Muse™ Cell Analyzer User’s Guide

Viewing a Trend Graph You can view a trend graph of the particles/mL data from the last 10, 30, or 60 System Check runs. A data point appears for each of the three replicate values. 1

Select Options from the System Check Results screen.

2

Select View Trend Graph.

3

Choose the number of runs (10, 30, or 60) that you wish to plot.

M

A legend in the lower-left corner lists the information found on the graph. A description of each item in the legend appears in the following table. Legend Item

Description

Replicates Outside Range

Data point appears as a red triangle (value falls outside the high or low 10% limit lines)

Replicates Within Range

Data point appears as a green triangle

Replicate Median

A black line connects the median values from each triangle System Check

47

4

Legend Item

Description

Limit Low

A purple line appears 10% below the expected particle count determined by the Check Code

Target

A green line at the expected particle count determined by the Check Code

Limit High

A pink line appears 10% above the expected particle count determined by the Check Code

(Optional) You cannot export the trend graph to a CSV file, but you can print to a pdf. •

Select the number of runs (last 10, 30, or 60) you wish to include from the Trend Graph screen, then close the screen. Close the System Check Log Options screen.



Select the printer tab from the System Check Results screen.



Select Print Trend Graph (see page 43), then select PDF to USB from the Printer list, and finally select Print.

Viewing the System Check Event Log The system saves a list of all events that occurred during the System Check run to a log file. 1

Select Options from the System Check Results screen.

2

Select View Event Log. The event log contains a list of all events that occurred during the System Check run. Every step that the operator performed and every step that the instrument performed, independent of the operator, are logged, as well as warnings and errors that occurred during the run.

48

Muse™ Cell Analyzer User’s Guide

3

You can filter the list to view statuses, actions, errors, and/or warnings. Select the appropriate check box(es) to display the types of events you wish to view.

Filter the event log list by selecting the types of events you wish to view.

If errors or warnings occur during a run, a message appears in red on the results screen, indicating that errors/warnings have been logged. For a list of System Check errors and warnings, see page 17. Exporting the Event Log You can export the System Check event log to a CSV file. If you are currently viewing the event log, proceed to step 3. The entire event log is exported regardless of whether you filtered the list. 1

Select Options from the System Check Results screen.

2

Select View Event Log.

3

Select Export from the Data Set Event Log screen.

System Check

49

4

Select the USB drive from the Export Location field.

5

Enter a name for the log file, or leave the default name.

6

Select Export.

50

Muse™ Cell Analyzer User’s Guide

Cleaning and Maintenance The system’s sampling precision depends on the integrity of the fluid pathway; therefore, it is important to maintain a clean system. Do not allow samples to remain loaded on the instrument for extended periods of time, as they may eventually clog the fluid system. Perform frequent cleaning cycles to prevent the build-up of cellular debris that may restrict sample flow.

Cleaning Protocols Several cleaning options are available depending on the type of cleaning you wish to do. The following protocols are available by selecting Clean from any assay screen, as well as the System Check screen (shown below).

Select Clean from System Check or any assay screen to access the cleaning options.



Capillary Rinse allows you to rinse the outer surface of the capillary. This can be helpful if you want to reduce carry-over between samples, or when cell counts are critical.



Quick Clean is a short cleaning cycle that allows you to clean the system during and after an assay, or as often as you like throughout the day.



Backflush reverses fluid out of the capillary. Use this feature if you suspect a clog. Always load a tube of 20% bleach when you perform a backflush.



Complete System Clean allows you to thoroughly clean the instrument at the end of the day or between assays, if needed.

Cleaning and Maintenance

51



Reset Fluid Levels sets the fluid level status for the waste bottle and cleaning solution bottle back to 100% and 100%, after you have emptied the waste bottle and filled the cleaning solution bottle.

Cleaning options

Cleaning the Outside of the Unit Clean the outside of the instrument by wiping it down with a soft cloth moistened with a 70% isopropyl alcohol or a 10% bleach solution in DI water. Follow with a cloth moistened with water. To avoid getting excess liquid on the instrument, do not spray these solutions on the instrument. To clean the touch screen, wipe it with 70% isopropyl alcohol. ▲ WARNING: To avoid exposure to laser radiation or electric shock, do not open the unit or attempt to perform any internal maintenance. There are no user-serviceable parts.

52

Muse™ Cell Analyzer User’s Guide

Rinsing the Capillary Capillary Rinse allows you to rinse the outside of the capillary by simply loading and unloading a tube of water. Use this option to avoid carryover. 1

Select Clean from the System Check screen or any assay screen.

2

Select Capillary Rinse from the list of cleaning protocols.

3

Load a full tube of DI water on the unit.

4

Select Eject.

5

Discard the tube of water.

Running Quick Clean Quick Clean is a short cleaning cycle that cleans the fluid system during and after an assay, or as often as you like throughout the day. If you are running lots of samples with high background, run Quick Clean according to the frequency recommended for the assay—a Recommended Quick Clean message will appear for every N number of samples. 1

Select Clean from the System Check screen or any assay screen.

2

Select Quick Clean from the list of cleaning protocols.

3

The system prompts you for ICF, however, you can load a tube containing any one of the following solutions: •

DI water to quickly flush out the system.



Guava ICF to clean the system (follow with a second Quick Clean using water to rinse) Cleaning and Maintenance

53



10% bleach solution in Guava ICF (1 part bleach in 9 parts Guava ICF) [follow with a second Quick Clean using water to rinse]

4

Select Clean to run the solution.

5

If you used water you are done with the Quick Clean procedure. If you used Guava ICF, either straight or with bleach, you will need to run water to rinse the fluid system. Select Quick Clean from the cleaning protocol screen. Load a full tube of DI water, then select Clean.

You may continue running samples, or leave the tube of water loaded on the unit until you are ready to use the system again.

Backflushing the Capillary The Backflush feature reverses the flow of fluid out of the capillary. Use this feature when you suspect that the fluid pathway is clogged. 1

If you suspect a clog during acquisition, select Clean from the System Check screen or any assay screen.

2

Select Backflush from the list of cleaning protocols.

3

Load a tube containing 100 μL of 20% bleach.

54

Muse™ Cell Analyzer User’s Guide

4

Select Clean.

5

When the backflush is complete, throw out the tube of bleach you used in step 3, as it may contain debris from the backflush.

6

Select Quick Clean to rinse the residual bleach from the capillary. Load a full tube of DI water on the unit and select Clean.

7

Replace the sample and continue with the assay or procedure you were running.

Performing a Complete System Clean Run the Complete System Clean feature to clean the unit at the end of each day, or between assays if a thorough cleaning is needed. You can also run this cleaning cycle to prime the fluid system or if you suspect there is air in the fluid lines. Always ensure the cleaning solution bottle is filled with ICF whenever you run the Complete System Clean. 1

Select Complete System Clean under Essential Tools at the main menu. This feature is also available by selecting Clean from the System Check screen or any assay screen, then selecting Complete System Clean from the list of cleaning protocols.

Cleaning and Maintenance

55

2

Select Run Complete Clean.

Displays the last time a complete system clean was performed.

3

56

Load a full tube of Guava ICF on the unit and select Run.

Muse™ Cell Analyzer User’s Guide

4

When the ICF cycle is complete, the system prompts you for a tube of DI water. Load a full tube of DI water on the unit and select Continue.

When the system cleaning procedure is complete, the Cleaning Log appears. Each entry in the log represents one System Cleaning procedure. The list shows the user who performed the cleaning, the date and time the cleaning was performed, and whether the cleaning was completed or aborted.

Scroll through cleaning log entries.

Cleaning and Maintenance

57

5

If you are shutting down the system, leave the tube of DI water used for cleaning loaded on the unit and select Finish to return to the main menu.

▲ WARNING: Do not leave a tube of Guava ICF, bleach, or any other cleaning agent loaded on the instrument overnight or for an extended period of time. Prolonged exposure to strong oxidizing agents will damage the flow cell. Always leave a tube of DI water on the system after cleaning and when shutting down. Change the tube of water regularly to ensure it is clean and free of particles. 6

(Optional) To run the cleaning procedure again, select Next Cleaning.

Printing the Cleaning Log You can print a list of all Complete System Clean procedures performed since the last time the cleaning log was cleared. The list contains the date and time that each Complete System Clean was run, as well as the operator who performed the cleaning procedure and whether the cleaning was completed or aborted. 1

Select the printer tab from the Cleaning Log screen. If you did not just complete the system cleaning procedure, select Complete System Clean under Essential Tools at the main menu, then select View Log/Results from the Complete System Clean screen.

2

(Optional) Select Print Preview to see the list as it will print.

3

If necessary, select the printer from the Printer drop-down list. If a printer is detected, it will appear as the default. If no printers are available, the default is PDF to USB, allowing you to save the results to a PDF file.

4

Select Print to print the list. If no printers are available or if you wish to export to a PDF, select PDF to USB, then select Print. The Export to PDF screen appears. Select the

58

Muse™ Cell Analyzer User’s Guide

USB port, enter a file name for the report file, or leave the default name. Select Export.

Printer tab

Select to print the log. Or, select Export to save to a PDF.

Cleaning Log Options You can view the cleaning log, and clear or export the entries in the log list. While only administrator-level users can clear the log, any user can export the log. 1

Select Options from the Cleaning Log screen. If you did not just complete the system cleaning procedure, select Complete System Clean under Essential Tools at the main menu, then

Cleaning and Maintenance

59

select View Log/Results from the Complete System Clean screen, then Options.

2

Refer to the following sections for information on the task you wish to perform.

Clearing the Cleaning Log After you export the Cleaning Log, you can clear the log. Only administrator-level users can clear the Cleaning Log. 1

Select Options from the Cleaning Log screen.

2

Select Clear Cleaning Log from the Cleaning Log Options screen. A message appears informing you that the action cannot be undone.

3

Select Clear to confirm and clear the log.

Exporting the Cleaning Log You can export the cleaning log to a CSV file. The file contains a list of all Complete System Clean procedures performed since the last time the cleaning log was cleared. The list includes the dates and times that each Complete System Clean was run, as well as the operators who performed the cleaning procedure and whether the cleaning was completed or aborted. 1

Select Options from the Cleaning Log screen.

2

Select Export Cleaning Log from the Cleaning Log Options screen.

60

Muse™ Cell Analyzer User’s Guide

3

Select the USB drive from the Export Location field.

4

Enter a file name or use the default file name.

5

Select Export.

Viewing and Exporting the Cleaning Event Log The system keeps track of and saves a list of all events that occurred during the Complete System Clean procedure to a log file. 1

Select Options from the Cleaning Log screen.

1

Select Export Event Log from the Cleaning Log Options screen. The event log contains a list of all events that occurred during the Complete System Clean. Every step that the operator performed and every step that the instrument performed during the procedure are logged.

Cleaning and Maintenance

61

2

You can filter the list to view statuses, actions, errors, and/or warnings. Select the appropriate check box(es) to display the types of events you wish to view.

Filter the event log list by selecting the types of events you wish to view.

3

62

(Optional) If you wish to export the cleaning event log, select Export from the Event Log screen, select the USB drive from the Export Location field, enter a file name or use the default name, and select Export.

Muse™ Cell Analyzer User’s Guide

Resetting Fluid Levels Use the Reset Fluid Levels option to reset the status indicator for the waste bottle and cleaning solution bottle back to 100% and 100% after you have emptied the waste bottle and filled the cleaning solution bottle. It is important to reset the fluid levels each time you fill and empty the bottles so that the instrument can accurately determine the amount of fluid in each bottle. capacity remaining

fluid remaining

When the waste bottle has 1% capacity remaining or the cleaning solution bottle has 1% fluid remaining, a message appears prompting you to fill the cleaning solution bottle and empty the waste bottle. 1

Select Clean from the System Check screen or any assay screen.

2

Select Reset Fluid Levels from the list of cleaning protocols. You can also select Complete System Clean under Essential Tools at the main menu, then select View Log/Results, then Options.

3

Select Next. The levels are now set to 100% for both the cleaning and waste bottles.

Cleaning and Maintenance

63

Filling the Cleaning Solution Bottle Fill the cleaning solution bottle with Guava ICF at the start of each day before you run the Complete System Clean procedure, and as needed. Do not allow the bottle to empty. This will pull air into the fluid system and require that you prime the system with water. The status bar located at the bottom of the screen indicates the amount of cleaning solution left in the bottle. When the remaining volume reaches 20%, the value appears in red, indicating it is time to fill the bottle with ICF. When the volume reaches 1%, a message appears indicating that you need to immediately fill the cleaning solution bottle and empty the waste bottle before proceeding.

Example shows cleaning solution bottle has 20% solution remaining (80% has been used).

1

Uncrew the tubing from the top of the cleaning solution bottle.

2

Press down on the cleaning solution bottle to release it from the unit. The bottle will pop up slightly allowing you to remove it. Push down on the bottle to remove it.

3

Unscrew the cap.

4

Fill the bottle to the fill line with Guava ICF.

5

Replace the cap and reinstall the bottle in the instrument. Align the shower icon on the bottle with the notch/line on the instrument. Press down on the bottle to engage it.

6

Reconnect the fluid line to the top of the bottle.

64

Muse™ Cell Analyzer User’s Guide

Emptying the Waste Bottle Empty the waste bottle at the start of each day before you run the Complete System Clean procedure, and as needed. The status bar located at the bottom of the screen, indicates the amount of space left in the waste bottle. When the remaining space reaches 20%, the value appears in red, indicating it is time to empty the waste bottle. When the remaining space reaches 1%, a message appears indicating that you need to immediately empty the waste bottle and fill the cleaning solution bottle before proceeding.

Example shows waste bottle has 20% capacity remaining (80% full).

▲ WARNING: Handle all biological specimens and materials they come in contact with as if capable of transmitting infection. Dispose of these materials using proper precautions in accordance with federal, state, and local regulations. 1

Unscrew the tubing from the top of the waste bottle.

2

Press down on the waste bottle to release it from the unit. The bottle will pop up slightly allowing you to remove it. Push down on the bottle to remove it.

3

Carefully unscrew the cap.

4

Empty the contents according to your local and state biohazardous waste disposal guidelines.

5

Rinse the bottle with water.

6

Add approximately 10 mL of bleach to the empty waste bottle.

Cleaning and Maintenance

65

7

Replace the cap and reinstall the bottle in the instrument. Align the trash can icon on the bottle with the notch/line on the instrument. Press down on the bottle to engage it. Reconnect the fluid line to the top of the bottle.

Cleaning the Capillary Use the flow cell cleaning tool (syringe) to clean the flow cell. 1

Use the eject button on the main menu or any assay screen to lower the sample loader arm.

2

Open the flow cell access door at the top of the instrument.

3

Use your fingers to unscrew the tubing from the instrument.

4

Fill the syringe with DI water or Guava ICF, then screw the syringe onto the tubing.

5

Apply gentle, steady pressure to the plunger and watch as the fluid flows from the tip of the capillary. •

66

Make sure the fluid stream is straight. If it is not straight, the tip of the capillary may be chipped or there may be a partial clog in the flow cell.

Muse™ Cell Analyzer User’s Guide



Check the capillary to ensure there are no leaks along the length of the tube or liquid pooling at the top of the flow cell.

6

Unscrew the syringe tool from the tubing.

7

Reconnect the tubing to the instrument. Make sure the tubing is screwed on tightly and there are no constricting kinks or twists.

8

Close the access door.

9

Prime the system by running a Quick Clean procedure. Perform a System Check procedure.

Replacing the Flow Cell You can replace the flow cell if it becomes damaged or clogged so severely that backflushing and cleaning the system do not fix the problem. When removing and replacing the flow cell, handle it with care. The capillary tube is fragile; avoid touching it unnecessarily. Do not force the flow cell into the receptacle. ▲ WARNING: To avoid exposure to laser radiation, turn off the power to the Muse™ System before attempting to remove the flow cell. 1

Use the eject button on the main menu or any assay menu to lower the sample loader arm.

2

Open the flow cell access door at the top of the instrument.

Cleaning and Maintenance

67

3

Use your fingers to unscrew the tubing from the instrument.

4

Release the clamp that holds the flow cell assembly in place. Press down on the left edge of the clamp to release the forked side on the right.

5

Holding onto the tubing as closely to the green connector as possible, gently pull up on the flow cell assembly to remove it from the receptacle. Keep the assembly completely vertical until the capillary is clear of the instrument.

6

Discard the flow cell and replace it with a new flow cell assembly.

7

Install the new flow cell. The flow cell fits only one way into the receptacle. Align the cut-out on one corner of the flow cell with the cut-out in the receptacle. Avoid bumping the capillary tube against the instrument or sides of the receptacle as you install it.

8

Close the clamp to secure the flow cell.

68

Muse™ Cell Analyzer User’s Guide

9

Reconnect the tubing to the instrument. Make sure the tubing is screwed on tightly and there are no constricting kinks or twists.

10 Close the access door. 11 To ensure that the flow cell was correctly installed, run Quick Clean to prime the system. Perform a System Check procedure.

Preparing the Unit for Depot Service Contact EMD Millipore Technical Service for the Decontamination Form and instructions. The form must be returned via e-mail before you can return the instrument. A return authorization (RMA) number will be issued to you. Write this number on the outside of the shipping box. If you did not save your original shipping box, EMD Millipore can send one to you. 1

Perform a Complete System Clean, then power off the system.

2

Empty and rinse out the waste and cleaning solution bottles and place them in their original box.

3

Wipe the outside of the instrument with 70% isopropyl alcohol or a 10% bleach solution in DI water, followed by water.

4

Open the flow cell access door and disconnect the fluid tubing from the instrument.

5

Release the clamp securing the flow cell and carefully remove the flow cell.

6

Use an empty syringe to flush the flow cell with air. Place the flow cell in its original box.

7

Lift the loader arm in the up position.

8

Place the shipping insert (with the foam attached) on a flat surface, then place the instrument in the foam cut-out. The tall block of foam should be on the right, as it will be positioned under the loader arm when sides are folder up.

9

Fold the sides of the insert up, making sure the foam block fits under the loader arm.

10 Place the power supply and power cord behind the instrument. 11 Use both handles on the insert to lift the instrument and lower it into the shipping box. Cleaning and Maintenance

69

12 Place the fluidics box (with flow cell, fluid bottles, and tubing) in the space in front of the instrument. 13 Ship the instrument to: EMD Millipore ATTN: Muse RMA (insert the RMA number issued to you) 25801 Industrial Blvd Hayward, CA 94545

70

Muse™ Cell Analyzer User’s Guide

Setup Setup Options Setup contains the tools necessary to set various system features, such as adding new users and setting their access levels, selecting how often users should clean during an assay, and choosing printer and file storage options. 1

Select Setup from the main menu.

2

Select the option you wish to perform. ■ NOTE: Depending on your user access level, some options may not be available to you. Refer to “Users & Access Levels” on page 73 for information on the options available for each user access level. Eject USB drive and adjust volume— features available to all users.

Setup options

Setup

71

3

Select Eject USB in the title bar to safely eject a USB thumb drive. The USB drives currently connected to the instrument are listed. Select the eject button to the right of the drive you want to eject.

eject button

4

Select the volume control in the title bar to adjust the audio feedback on the instrument. Use the center slider or the arrows on either end of the slider to increase or decrease the volume. Select Mute if you wish to turn off the volume.

Tap arrows. Or drag slider.

5

72

Refer to the following sections for more information on the Setup task you wish to perform.

Muse™ Cell Analyzer User’s Guide

Users & Access Levels The Muse™ System supports two levels of access control—Administrator and Operator. The access level is assigned by the laboratory administrator when a new user is added to the system. Administrators have privileges to all options within the Setup menu. Operators have the following privileges only: •

eject a USB drive



adjust the volume



edit their own user name and/or PIN



select the default printer



edit favorites (assay list that appears on the main menu)



export their own data by user or date (however, cannot delete after exporting)



export a Service Check file



export the cleaning log

Tasks that operator-level users cannot perform are: •

change the cleaning options



import or export all user data



edit printing layout options



update the software



export the System Check log



export data (or save instrument settings) to the Public folder

Setup

73

To Access the Users & Access Levels Screen 1

Select Setup from the main menu.

2

Select Users & Access Levels from the Setup menu. The Users & Access Levels screen appears. Administrators can add and remove users. Any user can edit their name and PIN.

3

Refer to the following sections for more information on adding, removing, and editing user information.

Adding a New User Only administrators can add new users to the system. When you add the first new Administrator user, the default Administrator (ADM) will be deleted. 1

Select Setup from the main menu.

2

Select Users & Access Levels from the Setup menu.

3

Select Add New User. ■ NOTE: All fields are required.

4

74

Enter the user name using the keypad.

Muse™ Cell Analyzer User’s Guide

5

Enter up to 3 characters in the Initials field to identify the user. Select Done. These initials will appear in the status bar when the user is logged on.

6

Enter up to 4 numeric characters for the password. Select Done. Confirm the password by reentering it. Select Done.

7

Select an access level—Administrator or Operator.

8

Select Add.

Removing a User Only administrators can remove a user from the system. The default Administrator will be overwritten by the first Administrator-level user added to the system. The default Guest user cannot be removed. ■ NOTE: Removing a user does not remove the user’s FCS data from the system. 1

Select Setup from the main menu.

2

Select Users & Access Levels from the Setup menu.

3

Select the user name from the list of users, then select Remove User.

4

Select Yes at the confirmation screen to remove the user.

Setup

75

Editing User Information Any user can edit their user name and/or password. Only administrators can change a user’s access level. 1

Select Setup from the main menu.

2

Select Users & Access Levels from the Setup menu.

3

Select the user name, then select Edit User.

4

Modify the information and select Save.

Recommended Quick Clean The Recommended Quick Clean feature allows your administrator to select how often a Quick Clean cycle should be performed during a run. When a user is running an assay a message appears prompting the user to perform a Quick Clean. Only administrators can modify the recommended Quick Cleans. 1

Select Setup from the main menu.

2

Select Recommended Cleaning Options from the Setup menu.

3

Select edit to change the current setting for a given assay. The n represents the number of samples that can be run before the user will be prompted to perform a Quick Clean.

76

Muse™ Cell Analyzer User’s Guide

4

Use the + and – buttons to increase or decrease the number of samples between Quick Cleans. You can also select to allows users to skip the recommended cleaning for this assay. In this case, when the cleaning message appears during the assay, the user can select Cancel to skip the cleaning.

5

Select Apply to apply the changes.

Printing Layout Options Any user can select the default printer. Only administrators can edit the layout options. 1

Select Setup from the main menu.

2

Select Printing Layout Options from the Setup menu.

3

Select the printer you wish to use as the default.

4

Select the layout options you wish to use. •

Include Signature Field adds a signature field to the printout.



Include Plots includes plots on the printouts.

Setup

77

■ NOTE: Only administrators can edit the layout options.

5

Click Back to save changes.

Editing Favorites You can add assays and select the order in which they appear on the main menu using the Edit Favorites option. All user levels can add and organize assays in the Favorites list. 1

Select Setup from the main menu.

2

Select Edit Favorites from the Setup menu.

3

Select a slot, either with a current assay (to remove the assay) or an empty slot (to add an assay). ■ NOTE: If you want to change the order of the assays, first remove the assays then add them in the order you want them to appear.

78

Muse™ Cell Analyzer User’s Guide

4

Select the assay that you want to add (or replace) for the selected slot. Or, select Remove Item to remove the assay from the selected slot.

Setup

79

The Edit Favorites screen now appears with the selected assay in the slot (or the assay removed from the slot). 5

Select Save.

Managing Instrument Data Storage The Manage Instrument Storage options allow you to import data to and export data from the system. Use the export features to back up or share data. For a description of the folders created on the USB drive when you export data, see “Folder and File Structure” on page 19. 1

Select Setup from the main menu.

2

Select Manage Instrument Storage from the Setup menu.

3

Select the option you wish to perform. Only administrators can import and export all user data. Operators can export their own data (by user and date).

Displays the free disc space and how much space is being used.

Only administrators can perform these functions.

4

80

Refer to the following sections for more information.

Muse™ Cell Analyzer User’s Guide

Exporting by User You can export data by the user. Operator-level users can only export their own user data. 1

Select Setup from the main menu.

2

Select Manage Instrument Storage from the Setup menu.

3

Select Export by User to export data for a particular user.

4

Select the USB drive from the Export Location field.

5

Select the user whose data you wish to export from the Users list. Use the arrow buttons in the bottom-right corner to scroll through the pages of users.

6

If you wish to delete the data once it has been successfully exported, select Delete files from instrument after successful export. ■ NOTE: Only administrators can delete data after exporting.

7

Select Export. If you selected to delete the data after exporting, a dialog appears requesting confirmation. Select Yes. Otherwise, a message appears indicating the export status. Select Close.

Setup

81

Exporting by Date You can export data from the last 30, 60, or 90 days or older. Operator-level users can only export their own user data. 1

Select Setup from the main menu.

2

Select Manage Instrument Storage from the Setup menu.

3

Select Export by Date to export data for a particular time frame.

4

Select the USB drive from the Export Location field.

5

Select the files to export—30, 60, or 90 days or older.

6

If you wish to delete the data once it has been successfully exported, select Delete files from instrument after successful export. ■ NOTE: Only administrators can delete data after exporting.

7

82

Select Export. If you selected to delete the data after exporting, a dialog appears requesting confirmation. Select Yes. Otherwise, a message appears indicating the export status. Select Close.

Muse™ Cell Analyzer User’s Guide

Importing All User Data If you exported all user data, for example, if you sent your system out for depot service, you may wish to import the data back to the system. Only administrators can import all user data. ▲ WARNING: Importing all user data will overwrite any data that you currently have on your system. 1

Select Setup from the main menu.

2

Select Manage Instrument Storage from the Setup menu.

3

Select Import All User Data.

4

Select the USB drive where the data is stored. If you wish to also import the data in the Public folder, select Include Public Folder.

5

Select Import.

6

The system automatically searches the drive for the user data. A message appears indicating that any data currently on the system will

Setup

83

be overwritten. Select YES to proceed, or Cancel to cancel the importing.

7

Select Close to the message that data was successfully imported.

Exporting All User Data Use the Export All User Data feature to back up all data files. Only administrators can export all user data. 1

Select Setup from the main menu.

2

Select Manage Instrument Storage from the Setup menu.

3

Select Export All User Data.

4

(Optional) If you wish to delete the data from the system after it’s exported, select Delete files from the instrument after successful export.

5

The system displays the drive space needed for the export and displays only the USB drives with the available space. Select the USB drive and then select Export.

84

Muse™ Cell Analyzer User’s Guide

■ NOTE: This could take several minutes depending on how much data you are exporting.

6

Select Close to the message that data was successfully exported.

Software Update Use Software Update to load the latest version of Muse™ software on your instrument. 1

Download the software update from www.millipore.com onto a USB thumb drive.

2

Install the USB drive into an available port on the instrument.

3

Select Setup from the main menu.

4

Select Software Update from the Setup menu. The system automatically scans the available USB drives for a software update file. If an update is found, a confirmation message appears.

Setup

85

5

Select YES to install the software update.

Example

The new software will install and the system will automatically reboot in the new version. A message appears that the software successfully installed. The Select User screen appears.

Service Tools Service Tools allows a qualified service technician to connect a laptop and run a service application for troubleshooting purposes.

86

Muse™ Cell Analyzer User’s Guide

Troubleshooting This section lists possible problems you might encounter or messages you might see during operation. If you have a problem that you cannot resolve by using the troubleshooting section, call technical support, toll-free in the USA and Canada at 1 (800) MILLIPORE (645-5476) or visit www.millipore.com/techservice. For assay specific troubleshooting, refer to the assay-specific kit user's guide at www.millipore.com/muse.

Problems Problem

Recommended Solution

During startup, computer freezes on particular screens.

Cycle power by pressing the power button located below the touch screen.

Message: Instrument could not be detected.

Cycle power. If message appears after rebooting, contact Technical Support.

Instrument settings could not be retrieved

Ensure settings are located in the correct user folder. If settings are to shared among user, ensure settings files are saved to the Public folder.

Forgot user PIN

See “Forgetting Your User PIN” on page 33, or contact Tech Support.

Continued System Check failures after Complete System Clean.

Ensure the cleaning solution bottle is full of ICF and all connections are finger tight.

Loader arm is stuck in the up position. Eject button on screen does not lower it.

Use the loader arm release lever in the flow cell hatch. See “Loading and Unloading a Tube” on page 10.

Loader arm is not dropping/ loading.

Ensure the loader arm release lever in the flow cell access door is in the proper position (see page 10).

Troubleshooting

87

Problem

Recommended Solution

System Check failures: one ore more Particles/mL results fall outside the acceptance range (appears in red).

1.Ensure the correct Bead Lot # and Check Code are entered. Refer to the System Check Beads vial label and information card for values. 2.Run Quick Clean, then rerun the System Check procedure. If results still fail, run the Complete System Clean procedure. 3.Prepare a fresh bead sample and rerun the System Check procedure. 4.Rerun the Complete System Clean procedure. Ensure that the cleaning solution vial has sufficient fluid and that there are no kinks in the tubing from the flow cell or the cleaning solution vial.

Data sets saved by the default Administrator user are no longer accessible after a new Administrator user is added to the system.

The default Administrator is overwritten when a new Administrator user is added to the system. Add another administrator-level user with the name Administrator.

Sample acquisition is taking longer than expected or progress bar stops during acquisition.

Ensure that the System Check procedure was run and passed. If the progress bar stops during acquisition, the fluid system may be clogged. Run Quick Clean.

Instrument clogging; too many cells

Run Quick Clean to clean out the capillary. Resume running sample.

Lower Cell Concentration warning during acquisition

The sample concentration may be too low. The assay instructions are optimized to give you cell concentration between 100 and 500 cells/μL for accurate results. Repeat sample preparation with a lower dilution factor to allow for adequate cell numbers.

High Cell Concentration warning during acquisition

If the concentration of the stained cell sample for acquisition is high (>500 cells/μL), the accuracy of the data will most likely be compromised. Repeat sample preparation with a higher dilution factor to allow for adequate cell numbers.

88

Muse™ Cell Analyzer User’s Guide

Problem

Recommended Solution

Background staining and/or non-specific staining of cells

The cells may be damaged, as dead cells tend to aggregate and non-specifically bind fluorescent reagent. Avoid damaging the cells when handling and processing them in culture.

Low level of staining

Although the assay procedure was optimized for multiple cell types, every cell line behaves differently. A low signal may indicate that the cell concentration may be too high for the amount of reagent use. Restain cells at a lower concentration. Ensure proper controls are used.

Variability in day-to-day experiments

• If the results are inconsistent, check that the samples were well mixed prior to acquisition. Cells may quickly settle in your samples and your results will be inaccurate unless the cells are mixed just prior to acquisition. • Monitor experimental cell cultures to ensure that cell viability and cell numbers being analyzed are consistent. Any drop in cell numbers or viability can influence experimental results. • If there appears to be day-to-day variation of the staining pattern, ensure the Muse™ Cell Analyzer is working properly. Run the Muse™ System Check procedure to verify proper instrument function and accuracy. • Always monitor threshold settings, especially if using different cell types, to ensure cell events are not excluded.

Troubleshooting

89

90

Muse™ Cell Analyzer User’s Guide

Glossary

acquisition

The electronic and software function of collecting various types of information from a cell sample.

analysis

The software function of numerically and graphically manipulating data to identify and separate cell populations for the purpose of calculating relevant statistical information.

coefficient of variation (%CV)

The ratio of the standard deviation to the mean, expressed as a percent. It is calculated using the formula:

%CV = SD x

x

100

data set

A series of samples included within one file for a selected assay. An FCS file and a spreadsheet file are saved for each data set.

detector

A device used to measure light intensity. The fluorescence detectors are photodiodes (yellow and red). The FSC detector that measures cell size is also a photodiode. Both output a current that is proportional to the intensity of incident light.

dot plot

A graphical representation of two-parameter data. Each axis of the plot displays values for one parameter. A dot represents the values for a cell or particle.

FCS file

Flow Cytometry Standard file. A data file containing the results for an individual sample as well as all acquisition information at the time of data collection. FCS files are defined by the Data Files Standards Committee of the Society for Analytical Cytology. Cytometry. 1990;11:323– 332.

Glossary

91

flow cell

An optical assembly within the Muse™ Cell Analyzer. The flow cell consists of a metal shuttle holding a glass capillary with a tiny chamber where the laser beam illuminates the sample stream and cellular measurements occur.

fluorescence

The phenomenon of light emission that occurs when a fluorochrome’s excited electrons drop to a lower energy level.

fluorochrome

A fluorescent dye used as a detection reagent in cell analysis applications. A molecule capable of absorbing light at a certain wavelength, then emitting light at a longer wavelength (fluorescence) as it releases energy.

gate

A graphical boundary that defines a subset of data. Gates may be set on a single-parameter histogram or a twoparameter dot plot.

histogram

A graphical representation of single-parameter data. The horizontal axis of the graph represents the increasing signal intensity of the parameter and the vertical axis represents the number of events (cells).

laser

Light amplification by stimulated emission of radiation. A light source that is highly directional, monochromatic, coherent, and bright. The emitted light is in one or more narrow spectral bands, and is concentrated in an intense, narrow beam.

marker

A boundary or set of boundaries used to segregate data into subsets for statistical analysis. Set a marker on a histogram to obtain statistics on a certain region. Set quadrant markers on a dot plot to obtain statistics on data within four quadrants.

mean fluorescence

The average of the fluorescence intensities of each event acquired within a given set of events.

92

Muse™ Cell Analyzer User’s Guide

median

The axis value for the event that falls in the middle of the distribution.

parameter

A specific cell property that is measured as the cell passes in front of the laser beam. Each parameter is the output from a photomultiplier (which measures fluorescence) or a photodiode (which measures forward scatter).

population

A group of cells that express similar values within one or more parameters. For example, cells that are positive for a particular antibody appear in the same location within a histogram or dot plot.

red detector

A device used for measuring light intensity. The red channel on the Muse™ Cell Analyzer can measure fluorescent light from dyes that emit signals similar to PE-Cy5.

threshold

The minimum level of discrimination to electronically eliminate unwanted signal. A threshold setting allows you to specify events you wish to acquire based on signal intensity of the event. Anything below the threshold in not acquired.

yellow detector

A device used for measuring light intensity. The yellow channel on the Muse™ Cell Analyzer can measure fluorescent light from dyes that emit signals similar to PE.

Glossary

93

94

Muse™ Cell Analyzer User’s Guide

Ordering Information For ordering information contact the nearest Millipore office by calling 1 800 645-5476 or visiting us on our website at www.millipore.com/offices. Millipore and its distribution network will provide Muse® products to all sectors of life science research in certain countries outside North America and Europe. Part

Catalog Number

Muse™ Cell Analyzer

0500-3115

cleaning solution bottle and corresponding fluid tubing

0110-7865

waste bottle and corresponding fluid tubing

0110-7860

flow cell assembly

0500-3120

flow cell cleaning tool

6000-2820

1.5-mL tubes, Eppendorf (quantity 500)

1000-0785

Muse™ Resource flash drive (software and user’s guide)

0110-7880

Muse™ Cell Analyzer User’s Guide

0110-7895

Reagents for Muse™ System

Catalog Number

Muse™ System Check Kit (100 tests)

MCH100101

Muse™ Count & Viability Kit (100 tests)

MCH100102

Muse™ Count & Viability Kit (600 tests)

MCH600103

Muse™ Count & Viability Kit (200X)

MCH100104

Muse™ Annexin V & Dead Cell Kit (100 tests)

MCH100105

Muse™ Cell Cycle Kit (100 tests)

MCH100106

Muse™ Cell Dispersal Reagent (100 tests)

MCH100107 Ordering Information

95

96

Muse™ Cell Analyzer User’s Guide

Specifications Muse™ System Physical Characteristics instrument weight: instrument size height: width: depth: connectors:

13.1 lb (5.94 kg) 8.69 in (22.07 cm) 8.12 in (20.62 cm) 11.11 in (28.22 cm) power – Kycon KPJX-PM back panel for motherboard – Intel D525MW five USB connectors—one on front and four on back of instrument

Operating Environment temperature: external power supply (input voltage range): main unit (input voltage): fuse rating:

16–35°C (60–95°F) 100–240 VAC, 50/60 Hz 80 W 15VDC, 5A auto-resettable

Optics laser: forward scatter detector: fluorescence detectors:

532-nm green laser photodiode photodiodes (YLW 576/28, RED 680/30)

Signal Processing parameter dynamic range: pulse processing: time:

4.0 decade digital signal processing every particle time stamped

Specifications

97

Fluidics flow cell dimension: pump: sample flow rate: cleaning / waste bottles: waste generation: dead volume: sample concentration: sample requirement:

rectangular capillary (1.5 mm x 0.8 mm) with 100-μm round bore positive displacement 7 μL/min to 36 μL/min 50-mL plastic (co-polyester) bottles with screw tops