Moving Rehabilitation Forward™
System Models: 2782 - Serial numbers 1000 and above 2759 - Serial numbers 1000 and above
ISO 13485 CERTIFIED
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
TABLE OF CONTENTS FOREWORD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
6.6 Control Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27
1- SAFETY PRECAUTIONS . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
6.7 Keymats. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27
1.1 Percautionary Definitions . . . . . . . . . . . . . . . . . . . . . . 2
6.8 Applicator Sound Head. . . . . . . . . . . . . . . . . . . . . . . .28
1.2 Percautionary Instructions . . . . . . . . . . . . . . . . . . . . . . 2
6.9 Applicator Probe Spring Holder . . . . . . . . . . . . . . . .29
2- THEORY OF OPERATION . . . . . . . . . . . . . . . . . . . . . . . . . . 4
6.10 Applicator PC Board. . . . . . . . . . . . . . . . . . . . . . . . . .30
2.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
6.11 Applicator Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31
2.2 Power Supply. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
7- ULTRASOUND APPLICATOR CALIBRATION . . . . . . . .32
2.3 Control Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
8- PARTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33-39
2.4 Ultrasound Board and Applicator . . . . . . . . . . . . . . . 4
8.1 Top Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33
2.5 User Interface and Accessories. . . . . . . . . . . . . . . . . . 4
8.2 Ultrasound Assembly. . . . . . . . . . . . . . . . . . . . . . . . . .34
3- NOMENCLATURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
8.3 Control Board Assembly . . . . . . . . . . . . . . . . . . . . . . .35
3.1 Intelect Transport/Vectra Genisys Ultrasound Therapy Systems . . . . . . . . . . . . . . . . . 5
8.4 Base Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36
3.2 Hardware and Software Symbol Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
9- SCHEMATICS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40-44
4- SPECIFICATIONS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-9 4.1 Intelect Transport/Vectra Genisys Ultrasound Therapy Systems . . . . . . . . . . . . . . . . . . 8 4.2 Intelect Transport/Vectra Genisys Ultrasound Therapy Systems Specifications . . . 9 5- TROUBLESHOOTING . . . . . . . . . . . . . . . . . . . . . . . . . 10-20 5.1 Intelect Transport/Vectra Genisys Ultrasound Therapy Systems Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10 5.2 Intelect Transport/Vectra Genisys Ultrasound Therapy Systems Diagnostics. . . . . .12 5.3 Electrical Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13 5.4 Leakage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13 5.5 Visual Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14 5.6 Unit Startup and Fan Testing . . . . . . . . . . . . . . . . . . .14 5.7 Ultrasound Board Test . . . . . . . . . . . . . . . . . . . . . . . . .15 5.8 Power Supply Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . .16 5.9 Ultrasound Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17 5.10 Ultrasound Applicator Ouptut Test . . . . . . . . . . . .18 5.11 Ultrasound Duty Cycle Test . . . . . . . . . . . . . . . . . . .20 6- REMOVAL & REPLACEMENT . . . . . . . . . . . . . . . . . . 21-31 6.1 Separating Top From Bottom . . . . . . . . . . . . . . . . . .21 6.2 Fan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22 6.3 Power Supply. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23 6.4 Ultrasound Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25 6.5 LCD Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26
8.5 Final Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38 10- WARRANTY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .45
FOREWORD
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
Read, understand, and follow the safety precautions and all other information contained in this manual. This manual contains the necessary safety and field service information for those field service technicians, certified by Chattanooga Group, to perform field service on Intelect Transport/Vectra Genisys Ultrasound Therapy Systems. At the time of publication, the information contained herein was current and up-to-date. However, due to continual technological improvements and increased clinical knowledge in the field of ultrasound therapy, as well as Chattanooga Groups policy of continual improvement, Chattanooga Group reserves the right to make periodic changes and improvements to their equipment and documentation without any obligation on the part of Chattanooga Group. As significant changes occur to Intelect Transport/Vectra Genisys Ultrasound Therapy Systems, service bulletins will be made available on our website (chattgroup.com) in lieu of reprinted manuals. Technicians repairing Intelect Transport/Vectra Genisys Ultrasound Therapy Systems agree to assume all risk and liability associated with this process. This system is to be used only under the supervision of a licensed practitioner.
©2008 Encore Medical, L.P. and its affiliates, Austin, Texas, USA. Any use of editorial, pictorial, or layout composition of this publication without express written consent from Chattanooga Group of Encore Medical, L.P. is strictly prohibited. This publication was written, illustrated, and prepared for print by Chattanooga Group of Encore Medical, L.P.
1
1 SAFETY PRECAUTIONS
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems 1.2 PRECAUTIONARY INSTRUCTIONS Read, understand, and follow all safety precautions found in this manual. Below are general safety precautions that must be read and understood before attempting any service techniques on these systems.
1.1 PRECAUTIONARY SYMBOL DEFINITIONS
The precautionary instructions found in this manual are indicated by specific symbols. Understand these symbols and their definitions before operating or servicing this equipment. The definitions of these symbols are as follows:
A. CAUTION
• Read, understand and practice the precautionary and operating instructions. Know the limitations and hazards associated with using any ultrasound device. Observe the precautionary and operational decals placed on the unit. • Do not operate the unit in an environment of short-wave diathermy use. • The Ultrasound modality should be routinely checked to determine that all controls function normally using the information provided within this manual. • Use of controls or adjustments or performance of procedures other than those specified herein may result in hazardous exposure to ultrasonic energy. • Do Not use sharp objects such as a pencil point or ballpoint pen to operate the buttons on the control panel as damage may result. • Operate, transport and store this unit in temperatures between 59 °F and 104 °F (15 °C and 40 °C), with Relative Humidity ranging from 30%-60%. • Inappropriate handling of, and subjecting the ultrasound applicator to physical abuse, may adversely affect its characteristics. • Inspect Applicator for cracks, which may allow the ingress of conductive fluid before each use. • Inspect all cables, and associated connectors before each use.
Text with a “CAUTION” indicator will explain possible safety infractions that have the potential to cause minor to moderate injury or damage to equipment.
B. WARNING
Text with a “WARNING” indicator will explain possible safety infractions that will potentially cause serious injury and equipment damage.
C. DANGER
Text with a “DANGER” indicator will explain possible safety infractions that are imminently hazardous situations that would result in death or serious injury. D. EXPLOSION HAZARD
Do not use this equipment in the presence of flammable anesthetics. This symbol is also prominently displayed on the serial number plate of the unit. E. DANGEROUS VOLTAGE
Text with a “Dangerous Voltage” indicator serves to inform the Service Technician of possible hazards resulting in the electrical charge retained by faulty power supplies. F. CORROSIVE HAZARD (NiMH Battery)
Text with a “Corrosive Hazard” indicator will explain possible safety infractions if the chemical components of this product are exposed to air, skin or other materials. NOTE: Throughout this manual “NOTE” may be found. The Notes are helpful information to aid in the particular area or function being described.
2
1 SAFETY PRECAUTIONS
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
1.2 SAFETY PRECAUTIONS CONTINUED
• Do not apply the Ultrasound Applicator to the patient during the Head Warming period. Applicator must remain in Applicator Hook during the Head Warming period. • Use only degassed water in power meter for testing Ultrasound Applicators. Use of other types of water will cause false test results. see page 12 for degassed water recipes. • Use of other brands or types of tools, equipment, fixtures, materials, and supplies other than those specifically listed on page 12 will give bad test and calibration results. • If proper equipment is not available or cannot be obtained, send the Ultrasound Applicators to the factory for calibration. • Do not aerate water when filling power meter. • Unplug the unit from the power source and/or remove Battery before attempting any removal or replacement procedures to prevent electrical shock. • Dielectric Withstand Test must always be performed after a Power Supply has been replaced. Field technicians unequipped to preform the Dielectric Withstand Test should contact the manufacturer for Power Supply replacement.
• Handle Power Supply shield with caution. Penetration of the shield may result in damage to the unit and/or personal injury. • The Power Supply will retain high voltage in the Power Supply Capacitor C4. Discharge Capacitor using a multimeter prior to any service of the Power Supply • Applicator must be calibrated after removal and replacement of any part, using an Ohmic UPM DT 10 or UPM DT 100. Failure to calibrate a repaired Applicator, using the proper equipment, may result in serious injury or harm to the patient.
3
2 THEORY OF OPERATION
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
2.1 OVERVIEW The Intelect Transport/Vectra Genisys Ultrasound Therapy Systems are comprised of several PC board assemblies housed within a common enclosure. These assemblies each support a distinct function in the product. The basic elements are User Interface, Control Board, Ultrasound Board, Ultrasound Applicator, and Power Supply. 2.2 POWER SUPPLY A universal input Power Supply provides all parts of the system with 24 VDC. The supply is connected to the mains at all times when the cord is attached. The power switch on the top of the unit, switches the 24 VDC output of the Power Supply to the different components in the system. The 24 VDC supply is regulated locally at each PC Board as required. 2.3 CONTROL BOARD The Control Board serves just as its name implies. It controls the operation of the Ultrasound Board, User Interface and optional accessories. The Control Board communicates to the Ultrasound Board. The Control Board drives the Display. The Control Board reads the menu buttons. The Control Board also reads the amplitude and the Contrast Control on the system. Sound output is generated by the Control Board and routed to an internal speaker. 2.4 ULTRASOUND BOARD AND APPLICATOR The Ultrasound Board generates the 1 or 3.3 MHz output to drive the ultrasound Sound Head. The Ultrasound Board is accessed much like an I/O port by the Control Board. It can provide current and voltage information about the ultrasound output of the board. The calibration data for the Sound Head is passed through the Ultrasound Board from the Applicator up to the Control Board. By storing the calibration data in the Applicator there is no calibration necessary for the Ultrasound Board. Calibrated Chattanooga Group Ultrasound Applicators, designed for the Intelect Transport/Vectra Genisys Ultrasound Therapy Systems, can be connected and operated on any Intelect Transport/Vectra Genisys Therapy System to provide accurate coupling and output. 2.5 USER INTERFACE AND ACCESSORIES The LCD display panel provides the operator visible feedback in the way of menu choices. Pressing the menu buttons makes selections from the menus. The Control Board interprets these user inputs and responds accordingly. Audible feedback is given as well for events such as key presses and end of treatment. The Optional Battery mounts in the system bottom and supplies 24 VDC to the power supply. The Power Supply distributes the power required to the system as needed. The charging circuit for the battery is incorporated on the Control Board. An icon on the display reflects the available amount of charge in the battery. When the system is connected to an approved power outlet and a Battery is installed in the system, the system will charge the Battery to full capacity. The incorporated charger will not overcharge the Battery.
4
3 NOMENCLATURE
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
components of the Intelect Transport/Vectra Genisys Ultrasound Therapy Systems. Know the components and their functions before performing any operation on or service to the Intelect Transport/Vectra Genisys Ultrasound Therapy Systems.
3.1 INTELECT TRANSPORT/VECTRA GENISYS ULTRASOUND THERAPY SYSTEMS The nomenclature graphics below, Figure 3.1, indicates the general locations of the exterior
Power On/Off
LCD Intensity/ Contrast Dial
LCD
Sound Head
Clinical Library Time
Duty Cycle/ Up Arrow
Head Warming/ Back
Power Cord Plynth
Intensity LED Indicator (Output Power)
Intensity Selection Watts/Watts per cm² / Parameter Display / Enter
Applicator
Applicator Holder Stop Pause Frequency/ Down Arrow
Accessory Panel
Start
Power Cord Connection
Battery Compartment Serial Decal
FIGURE 3.1
5
Ultrasound Applicator Connection
3 NOMENCLATURE
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
3.2 HARDWARE AND SOFTWARE SYMBOL DEFINITIONS The symbols below are found on the system as well as within the software. These symbols are defined for the purpose of recognition and functionality when operating or performing service on the Intelect Transport/Vectra Genisys
Ultrasound Therapy Systems. Know the symbols and their definitions before performing any operation of or service to the Intelect Transport/Vectra Genisys Ultrasound Therapy Systems.
Intelect Transport/Vectra Genisys Ultrasound Therapy Systems Hardware Symbols
Power On/Off The Power On/Off button controls the flow of electricity to the unit. Illuminates steady when system is On. Flashes when power is available.
LCD The LCD (Liquid Crystal Display) allows the user to view and monitor the information displayed before, during, and after therapy.
Clinical Library Select this button to access the following functions:
• Retrieve User Protocol • Restore Factory Settings • Restore Factory Protocols • Languages • View Unit Information Time Press the Up or Down arrow buttons to set total treatment time of therapy.
Back Use this button to return to the previous window.
STOP Select this button to stop a treatment session.
Down Arrow When the window displays a list of options, press the Down Arrow button to scroll down the list.
Pause Press this button to pause and restart therapy sessions.
Sound Head The aluminum face of the applicator that contacts the patient’s skin. It covers a transducer mechanism that converts electrical energy to mechanical energy in the form of a vibrating crystal.
LED Indicator (Output Power) When illuminated, this green light signifies that ultrasound energy is being emitted from the Sound Head.
Applicator The hand held assembly used to deliver ultrasonic energy. The Applicator includes the Sound Head, Transducer, and related electronics.
6
3 NOMENCLATURE
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
3.2 HARDWARE AND SOFTWARE SYMBOL DEFINITIONS CONTINUED
Ultrasound Applicator Connection This port serves as the connection point between the unit and the Ultrasound Applicator.
START Select Start to begin a treatment session.
Accept and Return Select to accept menu selection and to return to the home screen INTENSITY Use the up or down arrow on the Intensity button to adjust Ultrasound Intensity to the prescribed level. Duty Cycle Press to select and display the Duty Cycle of the Ultrasound as prescribed. Options available are: 10%, 20%, 50,% or Continuous (100%).
Up Arrow When the window displays a list of options, press the Up Arrow button to scroll up the list.
Battery Indicator When displayed on the LCD, this symbol indicates the Battery pack option is present on the unit. This symbol also displays the charge status of the Battery.
LCD Intensity/Contrast Dial If the intensity of the LCD display diminishes, turn the dial until the Display contrast is optimal.
Charge Indicator This symbol displays when the unit is connected to Mains Power and the Battery Pack is charging.
Frequency Press to change Frequency to 3 MHz or 1 MHz.
Intensity Display Press to display “Watts” or “Watts per cm²“ for the output of the ultrasound.
Head Warming Press to turn Off or On the Head Warming feature of the system.
Increase Found in two locations on the system, press to increase session time and/or ultrasound intensity respectively.
Decrease Found in two locations on the system, press to decrease session time and/or ultrasound intensity respectively. NOTE: During Battery operation, if the unit is left on, but is not active for more than five minutes, it will power off to conserve Battery power. To restore power, press the Power On/Off button.
7
4 SPECIFICATIONS
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
4.1 INTELECT TRANSPORT/VECTRA GENISYS ULTRASOUND THERAPY SYSTEMS
Figure 4.1 below provides physical details of the Intelect Transport/Vectra Genisys Ultrasound Therapy Systems. This section also provides ultrasound specifications to aid in troubleshooting.
Refer to this section when performing troubleshooting, replacement, and repair of the Intelect Transport/Vectra Genisys Ultrasound Therapy Systems.
A. Intelect Transport/Vectra Genisys Ultrasound Therapy Systems Physical Specifications
DEPTH
HEIGHT
WIDTH
FIGURE 4.1
Dimensions Width.. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. 28.8 cm (11.3") Height . .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. . 16.3 cm (6.4") Depth.. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. 32.8 cm (12.9") Weight Standard Weight (with base) .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .2.3 kg (5.07 lb) Battery Pack. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .0.85 kg (1.87 lb ) Power Input. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. . 100 - 240 VAC, 50/60 Hz 75 VA Output. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. . +24 V, 1.0 A Electrical Class .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. CLASS I Electrical Type Ultrasound TYPE B.. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .
Operating Environment Temperature .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. ..Between 59° F and 104° F (15° C and 40° C) Relative Humidity .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. 30%-60% Atmospheric Pressure .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. . 950-1,050 h Pa Complies with:
9700675
Listed by Intertek Testing Services NA Inc. Conforms to UL Standard 60601-1 Certified to CAN/CSA Standard C22.2 No. 601.1-M90 w/A2 Compliant to IEC/EN 60601-1, IEC 60601-1-2, 60601-2-5
Battery Type . .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. Nickel Metal Hydride (NiMH) (1.2 V x 20 size AA)
8
4 SPECIFICATIONS
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
4.2 INTELECT TRANSPORT/VECTRA GENISYS ULTRASOUND THERAPY SYSTEMS ULTRASOUND SPECIFICATIONS This section provides the necessary ultrasound specifications to aid in troubleshooting. Refer to these specifications as necessary when troubleshooting the ultrasound PC Board and Applicators. A. Ultrasound Frequency . . . . . . . . 1 MHz, ± 5%; 3.3 MHz, ± 5% Duty Cycles . . . 10%, 20%, 50%, and Continuous Pulse Frequency. . . . . . . . . . . . . . . . . . . . . . . . .100 Hz Pulse Duration . .1 mSec, ± 20%; 2 mSec, ± 20% 5 mSec, ± 20% Output Power 10 cm2 Crystal . . . . . . . . . . . . 0-20 Watts at 1 MHz, 0-10 Watts at 3.3 MHz 5 cm2 Crystal . . . . . . . . 0-10 Watts, 1 and 3.3 MHz 2 cm2 Crystal . . . . . . . . . .0-4 Watts, 1 and 3.3 MHz 1 cm2 Crystal . . . . . . . . . . .0-2 Watts, 3.3 MHz Only Amplitude . 0 - 2.5 w/cm2 in Continuous mode, 0-3 w/cm2 in Duty Cycle modes Output accuracy . . . . . . . . . .± 20% above 10% of maximum Temporal Peak to Average Ratios: 2:1, ± 20%, at 50% Duty Cycle 5:1, ± 20%, at 20% Duty Cycle 9:1, ± 20%, at 10% Duty Cycle Beam Non uniformity Ratio. . . .. . . . . . . . . . 5.0 : 1 maximum Beam Type . . . . . . . . . . . . . . . . . . . . . . . . . Collimating Effective Radiating Areas 10 cm2 Crystal - 6.8 cm2 - 10.0 cm2 5 cm2 Crystal - 3.5 cm2 - 5.0 cm2 2 cm2 Crystal - 1.4 cm2 - 2.0 cm2 1 cm2 Crystal - 0.7 cm2 - 1.0 cm2 Treatment Time . . . . . . . . . . . . . . . . . . 1-30 Minutes
B. Head Warming Feature Specifications The Head Warming feature of an Intelect Transport/Vectra Genisys Ultrasound Therapy System utilizes Ultrasound output resulting in warming of the Applicator to increase patient comfort. With Head Warming enabled, ultrasound is emitted without pressing the Start button. The Applicator LED will not illuminate during the Head Warming period. Display will indicate "Warming". Output . . . . 0 - 50% Cycling of maximum power Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.3 MHz Sound Head Temperature . . . . . . 85 °F - 110 °F (29.4 °C - 43.3 °C)
Do not apply the Ultrasound Applicator to the patient during the Head Warming period. Applicator must remain in applicator hook during the Head Warming period.
9
5 TROUBLESHOOTING
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
5.1 INTELECT TRANSPORT/VECTRA GENISYS ULTRASOUND THERAPY SYSTEMS ERROR MESSAGES a “Bad Board” only. A. The following information is provided as an aid No component level troubleshooting in defining the Software Error Messages of the information is or will be provided by Intelect Transport/Vectra Genisys Ultrasound Chattanooga Group for field troubleshooting of Therapy Systems. Once a particular error board components. message is defined, the information will also list B. Once a particular PC board has been probable causes and possible remedies. Once determined as bad, refer to the appropriate the problem area is determined, subsequent removal and replacement section for the tests for verification will be necessary to board affected and follow the instructions for determine a “Bad Board”. replacement of the board. All Troubleshooting and tests will be to validate ERROR CODE
ERROR TYPE
DEFINITION
PROBABLE CAUSES
POSSIBLE REMEDY
USER CORRECTABLE WARNING MESSAGES Icon
Sound Head over temp
Sound Head has overheated to a temperature that could cause damage to the crystal
Icon
No Applicator plugged into system
Applicator not plugged into system connector Applicator connector not completely seated in system connector Applicator Cable disconnected from Applicator PCB
1. Turn Off Head Warming feature 2. Allow Sound Head to cool before continuing therapy 3. Connect a known good Applicator to the system and try again. If system works properly, re-calibrate Applicator or replace with new Applicator. 4. Replace Applicator PC Board Plug Applicator into system Make certain Applicator connector is completely seated in system connector Connect Applicator Cable to Applicator PCB Replace with new Applicator Cable
Applicator Cable damaged
Replace Applicator PC Board
Applicator not calibrated
Calibrate Applicator Replace Ultrasound Board
Icon
100
WARN_US_ UNPLUGGED
101
200
201 202 203
WARN_US_ APPL_ BECAME_ UNPLUGGED ERR_EEPROM
ERR_APPL_ SAVE_CAL_ DATA ERR_APPL_NOT_ CAL'D_OK ERR_APPL_ SAVE_TIMEOUT
Battery plugged into system but not recognized by the unit Technician attempted to calibrate an Ultrasound Applicator but none is plugged into system Applicator came unplugged from system while administering therapy Error accessing the internal EEPROM used to store system configuration settings and protocols Error attempting to save data to applicator EEPROM during Applicator calibration General error attempting to calibrate an Applicator Error attempting to save data to Applicator EEPROM during Applicator calibration
Bad Ultrasound Board Bad Battery Control Board damaged Applicator not plugged into system connector Applicator connector not completely seated in system connector Applicator Cable disconnected from Applicator PCB Applicator Cable damaged Bad Control Board
Replace the Battery Replace the Control Board 1. Plug Applicator into system 2. Make certain Applicator connector is completely seated in system connector 3. Connect Applicator Cable to Applicator PCB 4. Replace with new Applicator Cable 5. Replace Applicator PC Board Replace Control Board
Bad Applicator
1. Replace with new Applicator Cable
Bad Applicator Cable
2. Replace Applicator PC Board
Applicator PCB Connector Bad
3. Replace with known good Applicator
Bad Ultrasound Board
4. Replace Ultrasound Board
10
5 TROUBLESHOOTING
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
5.1 INTELECT TRANSPORT/VECTRA GENISYS ULTRASOUND THERAPY SYSTEMS ERROR MESSAGES CONTINUED ERROR CODE
ERROR TYPE
DEFINITION
PROBABLE CAUSES
POSSIBLE REMEDY
USER CORRECTABLE WARNING MESSAGES Some type of critical Ultrasound Board error has occurred No Ultrasound Board is detected in unit
300
CRIT_ERR_US_BOARD
301
CRIT_ERR_US_BOARD NOT_DETECTED
302
CRIT_ERR_US_RESET
303
CRIT_ERR_US_BOARD_ Error reading from Ultrasound Board READ
304
CRIT_ERR_US_BOARD_ Error writing to Ultrasound Board WRITE CRIT_ERR_US_BOARD_ Error calibrating Applicator CALIBRATION_ERROR
305
Ultrasound Board to Control Board Connecting Header loose or disconnected Bad Ultrasound Board Bad Control Board
1. Reseat Connecting Header into Control Board 2. Replace Ultrasound Board 3. Replace Control Board
Ultrasound Board has reset for some unknown reason
NONE
No indication of problem and no ultrasound output from the Applicator
Probe Spring Holder cracked or broken
Replace Probe Spring Holder
NONE
Interrupted ultrasound output from Applicator
Bad Applicator Cable
Replace Applicator Cable
11
5 TROUBLESHOOTING
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
5.2 INTELECT TRANSPORT/VECTRA GENISYS ULTRASOUND THERAPY SYSTEMS DIAGNOSTICS A. General 1. The following information is intended to aid in troubleshooting the major components of the Intelect Transport/Vectra Genisys Ultrasound Therapy Systems to “Board Level” only. These tests are FACTORY standard testing procedures and methods used at the factory before shipment of any Intelect Transport/Vectra Genisys Ultrasound Therapy Systems. 2. Due to the complex nature of the technology utilized by Chattanooga Group, the recommended troubleshooting techniques are to determine “Bad Board” and board replacement only. No board component level troubleshooting is recommended nor will information or parts be supplied by Chattanooga Group. Any board component level troubleshooting performed will be at sole risk and liability of the Service Technician performing such troubleshooting techniques. B. Special Tools, Fixtures & Materials Required 1. Certain tests require the use of special tools and/or fixtures. These will be listed at the particular test where they are required. Testing with any other special tool or fixture other than those stated could give erroneous readings or test results. Always perform the tests exactly as stated to ensure accurate results. 2. Scope and other standard test equipment settings will be listed for each test performed to aid in performing the test to FACTORY standards and ensure proper readings. 3. The troubleshooting and repair of the Intelect Transport/Vectra Genisys Ultrasound Therapy Systems should be performed only by authorized technicians trained and certified by Chattanooga Group. C. Equipment Required • Digital Multimeter • Milliohm Meter • Applicators • Dielectric Withstand (Hi-Pot) and ground resistance tester • Orange Stick
• • • • • •
NOTE: Adjust Dielectric Withstand tester to indicate fault with 120k Ohm Load across the output when at specified test voltage. # 1 Phillips Screwdriver # 2 Phillips Screwdriver Insulated Needle Nose Pliers 10k Resistor Ohmic Instruments UPM DT 10 or UPM DT 100 Ultrasound Power Meter Degassed Water (< 5 ppm) for Ultrasound Power Meter Recipe(s) for Degassed Water 1) Boil distilled water for 30 minutes. Place water in a canning jar, and cover. Allow to cool to room temperature of approximately 70 °F (21 °C). May be refrigerated to aid cooling time. or 2) Bring distilled water to a boil. Place the container under vacuum for 5 to 10 minutes. Use a Dissolved Oxygen Test Kit to test oxygen level of degassed water. NOTE: Canning jars are ideal storage and transport containers for degassed water. In order to minimize aeration of degassed water during transport fill to a positive meniscus and slide the lid over the surface. Seal tightly. When pouring degassed water into and out of containers pour slowly down the side of the container to minimize aeration. Do not use tap water or distilled water in the ultrasound power meter. Use only degassed water in order to obtain correct test results. The chart below illustrates the oxygen content of degassed, tap, and distilled Water. WATER TYPE Degassed (per recipe 1 or 2)
ppm of Oxygen Less than 5 ppm
Tap Water
Up to 35 ppm
Distilled Water
Up to 20 ppm
D. Full Functional Tests Perform the tests found in this section to verify full functionality of new Intelect Transport/ Vectra Genisys Ultrasound Therapy Systems and accessories.
12
5 TROUBLESHOOTING
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
5.3 ELECTRICAL SAFETY The Intelect Transport/Vectra Genisys Ultrasound Therapy Systems has been tested to IEC/EN 60601-1, Standard for Safety for Medical Equipment. NOTE: This device complies with current leakage, ground continuity, and dielectric withstand (Hi-Pot) limits as prescribed by IEC/EN/UL 60601-1 and CSA/CAN 601.1 Medical Electrical, Part 1: General Requirements for Safety.
9700675
Listed by Intertek Testing Services NA Inc. Conforms to UL Standard 60601-1 Certified to CAN/CSA Standard C22.2 No. 601.1-M90 w/A2 Compliant to IEC/EN 60601-1, IEC 60601-1-2, 60601-2-5
Facility, local and national limits and test methods may vary. Power Requirements
Models: 2782 and 2759 . . . . . . Input : 100-240VAC (Serial numbers 1000 and above) 75 VA, 50/60 Hz 5.4 LEAKAGE Conduct all necessary leakage tests as required per NFPA 99 (National Fire Protection Association) “Health Care Facility” standards.
Unit failing Dielectric Withstand or Leakage tests could indicate serious internal problems. Do not place unit back into service! Send unit to factory for repair. Do not attempt to repair.
13
5 TROUBLESHOOTING
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
5.5 VISUAL INSPECTION Visually inspect the Intelect Transport/Vectra Genisys Ultrasound Therapy Systems. A visual inspection can, to an experienced technician, indicate possible abuse of the unit and internal problems. 5.6 UNIT STARTUP AND FAN TESTING A. Test 1. Place unit face up on work surface. 2. Connect Power Cord to unit and plug into proper power receptacle. 3. Connect an Applicator to the unit. 4. Turn system on. Adjust the Intensity up and press Start. The fan will begin to run, after the audible beep tones. 5. Place hand at the back of system, near Contrast Control, to verify Fan is blowing out. Refer to Figure 5.1. B. Test Results 1. Unit will not start, unit failed test. a) Possible bad Main Power Switch. b) Possible bad Power Supply. c) Possible bad power outlet or Mains Power Cord. d) Possible bad Ultrasound Board. 2. Treatment will not start up. a) Applicator not connected. b) Possible bad Ultrasound Board. c) Possible bad Control Board. 3. Screen does not display, unit failed test. a) Contrast Control needs adjusting. b) Possible bad Display. c) Possible bad Control Board. d) Possible bad Power Supply. e) Visually check power LED. If flashing, the Power Supply is good. Replace the Control Board. f ) If not flashing perform the Ultrasound Board test to check the Ultrasound Board. Refer to section 5.7 page 15. If the Power Supply functions replace the Ultrasound Board.
4. Fan not blowing outward, unit failed test a) Fan blowing inward. Fan wired wrong. Rewire or replace Fan. 5. Fan not blowing. NOTE: The Fan only functions when Ultrasound is being emitted from the Sound Head. a) Possible bad Fan. b) Possible bad Power Supply. c) Possible bad Control Board.
VERIFY FAN IS BLOWING OUT
FIGURE 5.1
14
5 TROUBLESHOOTING
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
5.7 ULTRASOUND BOARD TEST NOTE: This test is designed to eliminate or verify a problem with the Ultrasound Board. A. Test 1. Separate top from bottom. Refer to 6.1, part C. 2. Disconnect the Power Supply Harness from the Control Board. Lay the unit open. Refer to Figure 5.2. 3. Remove the Ultrasound Board. Refer to 6.4, steps 2 and 3. NOTE: The Ultrasound/Control Board Header may stay attached to either the Ultrasound Board or the Control Board. It may be necessary to remove the Header for installation. 4. Place the top of the unit back on the bottom of the unit. While holding the top and bottom together, with the Ultrasound Board missing, plug the unit back into the Mains Power. 5. Press the power button to verify that the unit turns on. NOTE: The Power On/Off button will flash when plugged into Mains Power if the Ultrasound Board is not functioning properly. B. Test Results 1. Unit will power up, unit passed test. a) Replace the Ultrasound Board. 2. Screen does not display, unit failed test. a) Contrast Control needs adjusting. b) Possible bad Display. c) Possible bad Control Board. d) Possible bad Power Supply.
ULTRASOUND BOARD
FIGURE 5.2
15
5 TROUBLESHOOTING
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
5.8 POWER SUPPLY TEST A. Test 1. Separate top from bottom. Refer to 6.1, part C. 2. Disconnect the Power Supply Harness from the Control Board. Lay the unit open. 3. Using a multimeter, hold the red wire of the multimeter to the prong on the connector for the red wire of the Power Supply Harness. Hold the black wire of the multimeter to the prong on the connector for the black wire of the Power Supply Harness. Refer to Figure 5.3. B. Test Results 1. Power Supply reads approximately 24 VDC, Power Supply is good. 2. Power Supply does not reads approximately 24 VDC, Power Supply failed test. a) Replace Power Supply.
FIGURE 5.3
16
5 TROUBLESHOOTING
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
5.9 ULTRASOUND TESTS A. Equipment Required for 5.9 and 5.10 • Degassed Water. Refer to page 12 for degassed water recipes • Ohmic Instruments UPM DT 10 or UPM DT 100 Ultrasound Power Meter • Known good Ultrasound Applicator NOTE: Use any Ultrasound Applicator supplied with a Chattanooga Group unit for this test. B. Ultrasound Applicator Identification Test Procedures 1. Without Ultrasound Applicator installed, turn unit on. 2. Look at the Display center. The Applicator not detected icon should appear. Refer to Figure 5.4. 3. Connect the Applicator into Applicator Receptacle. Refer to Figure 5.5. Watch Applicator LED while connecting to system. The LED should flash green five times. NOTE: The flashing green light indicates that the Applicator is programmed and calibrated. 4. Look at the display center. The Applicator detected icon should appear. Refer to Figure 5.6. C. Ultrasound Applicator Identification Test Results 1. Unit operates as described in steps 2 and 4. a) Unit passed test. 2. “NO CAL” icon displays center of system Display. a) Applicator not calibrated. b) Possible bad Applicator Cable. Re-test with known good Applicator. c) Possible bad Ultrasound Board. 3. Not detected icon displayed after ten seconds of Applicator being plugged in. a) Possible bad Applicator or Applicator Cable. Re-test with known good Applicator. b) Possible bad Ultrasound Board.
FIGURE 5.4
FIGURE 5.5
FIGURE 5.6
17
5 TROUBLESHOOTING
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
5.10 ULTRASOUND APPLICATOR OUTPUT TEST NOTE: The position of the Applicator over the stainless steel cone is critical to the test results. The Applicator must be level and centered.
Perform this test using all available Ultrasound Applicators used with the system being tested. NOTE: See page 17 for a list of equipment required for Output Tests. A. Ultrasound Applicator Output Test Procedures 1. Set up Ohmic Instruments UPM DT 10 or UPM DT 100 ultrasound power meter per operator’s instructions and fill test reservoir with degassed water. 2 Place an Applicator into the power meter holder. Make certain the Applicator’s Sound Head is completely submerged in the degassed water and centered directly over the stainless steel cone. Refer to Figure 5.7. 3. Zero or Tare the power meter. 4. Turn the unit on. 5. Press the Intensity Display button until “Watts” appears in the intensity area of the Display. 6. Increase Intensity to maximum. 7. Press “Start”. 8. Press the Frequency button until "1 MHz” is displayed in the frequency area of Display. 9. Compare power meter readings to Figure 5.8 for each of the power settings listed in the chart. 10. Press the Frequency button until “3.3 MHz” is displayed in the frequency area of Display. Repeat test and compare readings to Figure 5.8. B. Ultrasound Applicator Output Test Results 1. Output ranges fall within the specified ranges as listed in Figure 5.8. Unit passed test. 2. Readings fall outside specified ranges of Figure 5.8. a) Possible bad degassed water in power meter. b) Possible use of power meter other than Ohmic Instruments UPM DT 10 or UPM DT 100 ultrasound power meter. c) Possible bad or out of calibration Applicator. d) Check Ultrasound Board. e) Check Control Board.
FIGURE 5.7
APPLICATOR OUTPUT SPECIFICATIONS APPLICATOR SIZE
POWER SETTING WATTS
OUTPUT RANGE
1 cm2
1* 2*
0.8 - 1.2 1.6 - 2.4
2 cm2
1 2 4
0.8 - 1.2 1.6 - 2.4 3.2 - 4.8
5 cm2
1 2 5 10
0.8 - 1.2 1.6 - 2.4 4.0 - 6.0 8.0 - 12.0
10 cm2
1 5 10 15** 20**
0.8 - 1.2 4.0 - 6.0 8.0 - 12.0 12.0 - 18.0 16.0 - 24.0
* 3.3 MHz Only **1.0 MHz Only FIGURE 5.8
18
5 TROUBLESHOOTING
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
5.10 ULTRASOUND APPLICATOR OUTPUT TEST (CONTINUED) C. Timer Control/End of Treatment Test NOTE: Prior to testing, verify that Head Warming is off. Head Warming emitts ultrasound and the intensity will not reset to zero if Head Warming is on. 1. Turn on the unit. 2 Plug in an Applicator. 3. The Intensity Display button can read “Watts” or "Watts per cm²" in the intensity area of the Display. 4. Increase Intensity to 0.1 MHz 5. Set the time to 1 minute. 6. Press “Start”. 7. Verify that the green light on the Applicator goes out when the timer reaches zero and that the intensity resets to zero on the display.
19
5 TROUBLESHOOTING
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
5.11 ULTRASOUND DUTY CYCLE TEST Perform this test using all Ultrasound Applicators used with the system being tested. A. Ultrasound Duty Cycle Test Procedures 1. Set up Ohmic Instruments UPM DT 10 or UPM DT 100 ultrasound power meter per operator’s instructions and fill test reservoir with degassed water.
e) Check Ultrasound Board. f) Check Control Board.
DUTY CYCLE SPECIFICATIONS
Use only degassed water in power meter for testing
ultrasound applicators. Use of other types of water will cause false test results. See page 12 for degassed water recipes. Do not aerate water when filling power meter.
OUTPUT FOR 10%, 20% AND 50% DUTY CYCLES
DUTY CYCLE
OUTPUT RANGE
1 cm² at 2.7 Watts
10%
0.2 0.3
20%
0.4 0.7
50%
1.1 1.6
10%
0.2 0.3
20%
0.9 1.3
50%
2.2 3.2
10%
1.0 1.4
20%
1.9 2.9
50%
4.8 7.2
10 cm² at 20 Watts Operating at 1 MHz
10%
1.6 2.4
10 cm² at 10 Watts Operating at 3.3 MHz
2 cm² at 5.4 Watts
2. Place an Applicator into the power meter holder. Make certain the Applicator’s Sound Head is submerged in the degassed water and centered directly over the stainless steel cone. 3. “Zero” or Tare the power meter. 4. Turn the unit on. 5. Press the Intensity Display button until “Watts” appears in the Intensity area of the Display. 6. Increase Intensity to maximum. 7. Press “Start”. 8. Press the Duty Cycle button until “10%” is displayed in the duty cycle area of the Display. 9. Compare power meter readings to Figure 5.9 for each of the following duty cycle settings listed in the chart. 10. Press the Frequency button until “3.3 MHz” is displayed in the frequency area of display. Repeat test and compare readings to Figure 5.9. B. Ultrasound Duty Cycle Test Results 1. Duty Cycles fall within the specified ranges as listed in Figure 5.9. a) Unit passed test 2. Readings fall outside specified ranges of Figure 5.9. a) Possible bad degassed water in power meter. b) Applicator not level and centered over cone in power meter. c) Possible use of power meter other than Ohmic Instruments UPM DT 10 or UPM DT 100 ultrasound power meter.
5 cm² at 12 Watts
20%
3.2 4.8
50%
8.0 12.0
10%
0.8 1.2
20%
1.6 2.4
50%
4.0 6.0
FIGURE 5.9
d) Possible bad or out of calibration Applicator.
20
6 REMOVAL & REPLACEMENT
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
6.1 SEPARATING TOP FROM BOTTOM
REMOVE MAINS POWER CORD
REMOVE PLYNTH
Unplug the unit from the power source and/or remove battery before attempting any removal or replacement procedures to prevent electrical shock. A. Part Numbers Intelect TranSport Top .............................. 27399 Vectra Genysis Top........................................ 27244 Base ......................................................................... 27252 B. Tools & Equipment Required • #1 Phillips Screwdriver C. Separating Top from Bottom 1. Remove the Plynth from the bottom of the system. Refer to Figure 6.1. 2. Remove the Mains Power cord from the system. Refer to Figure 6.1. 3. Remove the rear Fan Grill from the system to gain access to the two rear mounting screws. Refer to Figure 6.2. 4. Remove the two front feet from the system to gain access to the two front screws. Refer to Figure 6.3. 5. Remove the four mounting screws from the bottom housing using the #1 Phillips screwdriver. Refer to Figure 6.3. 6. Carefully separate the upper and lower housings of the system. 7. Disconnect the Power Supply and Fan Harnesses from the Control Board of the system. Refer to Figure 6.3. 8. Reassemble the unit by reversing the steps above. NOTE: Do not overtighten the screws. Overtightening will damage the threads of the brass inserts.
FIGURE 6.1
REMOVE FAN GRILL
REMOVE FRONT FEET
FIGURE 6.2
REMOVE FOUR SCREWS
DISCONNECT POWER SUPPLY AND FAN HARNESSES
FIGURE 6.3
21
6 REMOVAL & REPLACEMENT
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
6.2 FAN
REMOVE SCREWS
FAN BAFFLE
Unplug the unit from the power source and/or remove battery before attempting any removal or replacement procedures to prevent electrical shock. A. Part Numbers ....................................... 27158 B. Tools & Equipment Required • #1 Phillips Screwdriver C. Fan 1. Separate top from bottom. Refer to 6.1, part C. 2. With #1 Phillips screwdriver, remove the 2 screws securing the Fan. Refer to Figure 6.4. 3. Remove Fan from the unit. 4. Remove Fan Baffle from Fan. Refer to Figure 6.4. 5. Install the replacement by reversing the steps above. NOTE: Refer to Figure 6.5 to route the Fan Harness when installing. Make certain that the Fan Baffle sits onto the system housing baffle tabs. Refer to Figure 6.6. Make sure the Fan mounting screws are tightly secured. Refer to Figure 6.7.
FIGURE 6.4
ROUTE HARNESS AS SHOWN
FIGURE 6.5 BAFFLE TABS
FIGURE 6.6 SECURELY TIGHTEN MOUNTING SCREWS
FIGURE 6.7
22
6 REMOVAL & REPLACEMENT
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
6.3 POWER SUPPLY ROUTE HARNESS AS SHOWN
Unplug the unit from the power source and/or remove battery before attempting any removal or replacement procedures to prevent electrical shock. POWER SUPPLY
The Power Supply will retain high voltage in the Power Supply Capacitor C4. discharge Capacitor using a multimeter prior to any service of the Power Supply
CAPACITOR C4
FIGURE 6.8
A. Part Numbers ...........................................27265 B. Tools & Equipment Required • #1 Phillips Screwdriver • Digital Multimeter • 10k Resistor C. Removing the Power Supply 1. Separate top from bottom. Refer to 6.1, part C. 2. Discharge the capacitor by wrapping a 10k resistor around the probes of a multimeter. Touch the leads of the mulitmeter to the prongs on the capacitor to discharge. Figure 6.8 Insert. 3. Watch the multimeter to verify that the voltage across the capacitor discharges close to zero volts DC. 4. With #1 Phillips screwdriver, remove the 2 screws securing the Power Supply. Refer to Figure 6.9. 5. Lift Power Supply and disconnect the Power Supply Harnesses from the Mains disconnect. When removing the Harnesses from the Mains disconnect, note the locations of the wire plug ins. Refer to Figure 6.10.
REMOVE SCREWS
FIGURE 6.9
GREEN WITH YELLOW STRIPE
Dielectric Withstand Test must always be preformed after a power supply has been replaced. Field technician unequipped to perform the Dielectric Withstand Test should contact the manufacturer for Power Supply replacement.
BLUE
FIGURE 6.10
23
RED
6 REMOVAL & REPLACEMENT
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
6.3 POWER SUPPLY (CONTINUED) 6. Remove Power Supply from system. 7. Carefully remove the copper tape securing the shield to the Power Supply. Remove from the inside of the Power Supply and retain the Shield for installation of the replacement. NOTE: The black side of the shield is non conductive and should always face the Power Supply. 8. Reverse steps 4 - 7 to install the replacement. Refer to Figure 6.8 to correctly route Power Supply Harness. Make certain the Power Supply sits onto alignment pins as shown in Figure 6.11.
POWER SUPPLY ALIGNMENT PINS
FIGURE 6.11
Handle Power Supply Shield with caution. Penetration of the Shield may result in damage to the unit and/or personal injury.
24
6 REMOVAL & REPLACEMENT
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
6.4 ULTRASOUND BOARD A. Part Number .......................................................... 27269
SCREW
SCREW
Unplug the unit from the power source and/or remove battery before attempting any removal or replacement procedures to prevent electrical shock.
SCREW
B. Tools and Equipment Required • #1 Phillips Screwdriver C. Ultrasound Board 1. Separate top from bottom. Refer to 6.1, part C. 2. With #1 Phillips screwdriver remove the 4 screws securing the Ultrasound Board to the unit. Refer to Figure 6.12. 3. Carefully remove Ultrasound Board from the Control Board/Ultrasound Board Header. Refer to Figure 6.13. 4. Remove Control Board/Ultrasound Board Header. Be careful not to bend or damage any of the pins. Refer to Figure 6.13. NOTE: The Ultrasound/Control board Header may stay attached to either the Ultrasound Board or the Control Board. It may be necessary to remove the Header for installation. 5. Install the Ultrasound Board by reversing the steps above. 6. Verify the ultrasound output with a known good Applicator. Refer to Section 5.10. NOTE: Install the Header with the long pins inserted into Ultrasound Board. Make certain Header is completely seated against Ultrasound Board. Refer to Figure 6.14. Refer to Figure 6.15 to align the Header with the Control Board. Carefully press the Header and board into place until the Header is completely seated into Control Board
SCREW
FIGURE 6.12 REMOVE CONNECTOR FROM CONTROL BOARD
REMOVE US BOARD FROM CONNECTOR
FIGURE 6.13
INSTALL CONNECTOR TO US BOARD AS SHOWN
FIGURE 6.14 ALIGN CONNECTOR PINS WITH CONTROL BOARD CONNECTOR
FIGURE 6.15
25
6 REMOVAL & REPLACEMENT
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
6.5 LCD DISPLAY 2 SCREWS EACH END RELEASE RETAINING BRACKET TABS
Unplug the unit from the power source and/or remove battery before attempting any removal or replacement procedures to prevent electrical shock. A. Part Number .......................................................... 27264 B. Tools and Equipment Required • #1 Phillips Screwdriver • Orange stick C. Ultrasound Board 1. Separate top from bottom. Refer to 6.1, part C. 2. Remove Ultrasound Board. Refer to 6.4, part C. 3. Using a # 1 Phillips screwdriver remove the 4 LCD retaining screws. Refer to Figure 6.16. 4. With an orange stick or flat edged piece of plastic gently push in and up on each end of the LCD retaining brackets to free the locking tabs from the Control Board. Refer to Figure 6.16. 5. Set retaining brackets aside for reassembly. 6. Carefully remove the LCD from the Control Board being careful not to damage the LCD header pins. Refer to Figure 6.17 . 7. Install the LCD by reversing steps 1 through 6.
FIGURE 6.16
REMOVE LCD FROM CONTROL BOARD
FIGURE 6.17
NOTE: Make certain to align the LCD header pins with the top row on the Control Board LCD connector. Refer to Figure 6.17.
26
6 REMOVAL & REPLACEMENT
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
6.6 CONTROL BOARD
RETAINING SCREWS
Unplug the unit from the power source and/or remove battery before attempting any removal or replacement procedures to prevent electrical shock. A. Part Number .......................................................... 27268 B. Tools and Equipment Required • #1 Phillips Screwdriver • Orange Stick C. Control Board 1. Separate top from bottom. Refer to 6.1, part C. 2. Remove Ultrasound Board. Refer to 6.4, part C. 3. Remove LCD from Control Board. Refer to 6.5, part C. 4. Remove the 7 retaining screws securing the Control Board to the housing. Refer to Figure 6.18. 5. Using an orange stick or flat edged piece of plastic, carefully push locking tabs while lifting the Control Board to release Control Board from housing. Refer to Figure 6.19. 6. Remove the LCD Contrast Knob and retain for installation on the new Control Board. Refer to Figure 6.20. 7. Install the replacement Control Board by reversing steps 1 - 6. NOTE: Do not overtighten the screws. Overtightening will damage the threads of the brass inserts.
FIGURE 6.18
PUSH TABS IN AND PULL UP ON BOARD SIMULTANEOUSLY
FIGURE 6.19
6.7 KEYMATS A. Part Number .......................................................... 27368 B. Tools and Equipment Required • #1 Phillips Screwdriver • Flat Blade Screwdriver C. Keymat NOTE: Follow 6.6 steps 1 through 5, for Control Board removal. When the Control Board is removed lift the Keymats out and replace. Keymats set includes the On/Off button. CONTRAST KNOB
FIGURE 6.20
27
6 REMOVAL & REPLACEMENT
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
6.8 APPLICATOR SOUND HEAD A. Part Numbers (Applicators) Intelect Applicators 1 cm² .............................................................................. 27381 2 cm² .............................................................................. 27382 5 cm² .............................................................................. 27383 10 cm² ........................................................................... 27384 Genisys Applicators 1 cm² .............................................................................. 27333 2 cm² .............................................................................. 27334 5 cm² .............................................................................. 27335 10 cm² ........................................................................... 27336 Sound Heads 1 cm² .............................................................................. 27126 2 cm² .............................................................................. 27127 5 cm² .............................................................................. 27128 10 cm² ........................................................................... 27129 B. Tools and Equipment Required • None C. Sound Head Removal and Replacement NOTE: The Transducer is contained within the Sound Head of the Applicator. 1. Unscrew the Sound Head from the body of the Applicator. Refer to Figure 6.21. 2. After removing the Sound Head inspect the plastic Probe Spring Holder in the top of the Applicator Assembly for cracks or discoloration. Refer to Figure 6.22. NOTE: If the Probe Spring Holder is dark or contains crack it should be replaced. Refer to Section 6.9, part C. 3. Reverse steps 1 and 2 to install a replacement Sound Head. 4. Recalibrate the Applicator following the calibration procedure in Section 7, part B.
FIGURE 6.21
PROBE SPRING HOLDER
FIGURE 6.22
Applicator must be calibrated after removal and replacement of any part, using an Ohmic UPM DT 10 or UPM DT 100. Failure to calibrate a repaired Applicator, using the proper equipment, may result in serious injury to the patient.
28
6 REMOVAL & REPLACEMENT
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
6.9 APPLICATOR PROBE SPRING HOLDER A. Part Number .......................................................... 27040 B. Tools and Equipment Required • #1 Phillips Screwdriver C. Probe Spring Holder Removal and Replacement 1. Remove the Sound Head per Section 6.8 part C. 2. Remove the brass Probe Spring and inspect for damage. Retain the Probe Spring for installation if undamaged. Refer to Figure 6.25. 3. Using a # 1 Phillips screwdriver remove the screw securing the Probe Spring Holder to the Applicator housing. Refer to Figure 6.25. 4. Reverse steps 1 through 3 for installation of the replacement spring. 5. Recalibrate the Applicator. NOTE: Do not over tighten the screw holding the Probe Spring Holder. A cracked or damaged Probe Spring Holder will affect the operation of the Applicator.
BRASS PROBE SPRING
PROBE SPRING HOLDER SCREW
FIGURE 6.25
Applicator must be calibrated after removal and replacement of any part, using an Ohmic UPM DT 10 or UPM DT 100. Failure to calibrate a repaired Applicator, using the proper equipment, may result in serious injury to the patient.
29
6 REMOVAL & REPLACEMENT
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
6.10 APPLICATOR PC BOARD A. Part Numbers 1 cm² .............................................................................. 27600 2 cm² .............................................................................. 27071 5 cm² .............................................................................. 27073 10 cm² ........................................................................... 27075 B. Tools and Equipment Required • #1 Phillips Screwdriver C. PC Board Removal and Replacement 1. Remove the Sound Head per Section 6.8, part C. 2. Remove the Probe Spring and Probe Spring Holder per Section 6.9, part C. 3. Remove the Applicator Cable per Section 6.11, part C. 4. Pull the PC Board from the Applicator housing. Refer to Figure 6.24. NOTE: If necessary needle nose pliers can be used to grasp the edge of the PC Board for removal. 6. Reverse steps 1 through 4 for installation of the replacement PC Board. 7. Recalibrate the Applicator. NOTE: When replacing the PC Board it should be oriented with the Cable Connector toward the inside of the Applicator Handle. Push the PC Board as far into the Applicator Handle as possible, with the Cable end fitting into the two PC Board slots. Refer to Figure 6.25. The Sound Head end of the PC Board fits against the screw standoff for the Probe Spring Holder. Refer to Figure 6.26.
FIGURE 6.24
GROOVES FOR PC BOARD
FIGURE 6.25
PROBE SPRING HOLDER STANDOFF
Applicator must be calibrated after removal and replacement of any part, using an Ohmic UPM DT 10 or UPM DT 100. Failure to calibrate a repaired Applicator, using the proper equipment, may result in serious injury to the patient.
FIGURE 6.26
30
6 REMOVAL & REPLACEMENT
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
6.11 APPLICATOR CABLE A. Part Numbers Gray Applicator Cable..................................... 27351 Blue Applicator Cable ..................................... 27106 B. Tools and Equipment Required • #1 Phillips Screwdriver C. PC Board Removal and Replacement 1. Using a counterclockwise motion, turn the end of the Applicator to remove the Applicator Cable. Refer to Figure 6.29. 2. Once the end of the cable is free from the Applicator housing, grasp the wires of the cable and gently pull the applicator connector from the holder on the PC Board. 3. To install position the two black tabs of the replacement Cable away from the inside of the Applicator. Using a small screwdriver push the Connector onto the pins of the Connector Holder. Gently push the sides of the Connector to verify that it is well seated. 4. Once the Cable Connector is well seated, hold the Cable and twist the wires 2 1/2 turns counterclockwise. Refer to Figure 6.30. NOTE: Twisting the Cable wires prior to screwing the Cable onto the Applicator housing will allow the wires to untwist and straighten out when the Cable is screwed into place. 5. Verify the operation of the Applicator.
FIGURE 6.29
FIGURE 6.30
31
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems 7 ULTRASOUND APPLICATOR CALIBRATION
7.1 GENERAL PROCEDURES A. Tools and Equipment Required • All Ultrasound Applicators for the unit being serviced • Ohmic Instruments UPM DT 10 or UPM DT 100 ultrasound power meter, set to "Watts" • Degassed Water. Refer to page 12 for degassed water recipes
Use only degassed water in power meter for calibrating ultrasound applicators. Use of other types of water will cause false readings and bad test results. See page 12 for degassed water recipes. Use of other brands or types of tools, equipment, fixtures, materials, and supplies other than those specifically listed on page 12 will give bad test and calibration results. If proper equipment is not available or cannot be obtained, send the ultrasound applicators to the factory for calibration.
FIGURE 7.1
B. Ultrasound Applicator Calibration Procedures 1. Perform the following on all Ultrasound Applicators annually or after repairing the Applicator. 2. With the system on, press the Clinical Libraries button once. Refer to Figure 7.1. 3. Simultaneously press and hold the Treatment Time and Intensity Decrease buttons for approximately 2 seconds. Refer to Figure 7.2. The calibration procedures screen should display. 4. Press the Frequency button once to highlight "Calibrate Applicator". 5. Press the Intensity Display button once to select. 6. Position the Ultrasound Applicator being calibrated into the Ohmic Instruments UPM UPM DT 10 or DT 100 power meter holder and set meter to "Watts”. Refer to Figure 7.3. 7. Follow the instructions on the LCD Display. 8. When calibration is complete, repeat the process with the remaining Applicators. Follow the test procedures in 5.9 and 5.10 to verify that the Applicator is calibrated.
FIGURE 7.2
SET METER TO WATTS
FIGURE 7.3
32
8 PARTS
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
8.1 TOP ASSEMBLY
1
3
4
2
5 ITEM NUMBER
PART NUMBER
1
27399 27244
2
3 4
DESCRIPTION
TOP MOULDING ASSEMBLY (GREY) TOP MOULDING ASSEMBLY (BLUE) SEE CONTROL CONTROL ASSEMBLY BOARD ASSY DRAWING 27368 KEYMAT (GREY) 27255 KEYMAT (BLUE) N/A PART OF KEYMAT
33
QTY REQ'D
1 1
1
8 PARTS
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
8.2 ULTRASOUND ASSEMBLY
1
3
2
NOTE: INSERT #3 INTO PCB #2 FIRST, THEN INTO #1
4
ITEM NUMBER
PART NUMBER
1
2
SEE TOP ASSY DRAWING 27269
3 4
DESCRIPTION
QTY REQ'D
TOP ASSEMBLY
1
ULTRASOUND BOARD
1
27580
HEADER
1
27142
M3 X 6MM SCREWS PAN HEAD
4
34
8 PARTS
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
8.3 CONTROL BOARD ASSEMBLY
NOTE: INSERT #4 THROUGH TOP ROW OF SOCKET CONNECTOR OF #1
1 3
5 4 2
ITEM NUMBER
PART NUMBER
DESCRIPTION
1
27268
CONTROL BOARD
1
2
27261
LCD BRACKETS
2
3
27012
CONTRAST KNOB
1
4
27264
LCD
1
5
27588
HEADER
1
35
QTY REQ'D
8 PARTS
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
9 4
9
6
10
NOTE TUCK FERRITE UNDER POWER SUPPLY 10 5 11
3
BLUE WIRE
RED WIRE GREEN/YELLOW WIRE
8.4 BASE ASSEMBLY
L
N
DETAIL A PLACE INSULATED SIDE OF SHIELD NEXT TO POWER SUPPLY
TUCK FERRITE UNDER POWER SUPPLY AND INTO CORNER SEE DETAIL A
1
7
2 NOTE:
COPPER TAPE LOCATIONS
THE COPPER TAPE, PART 12, SECURES THE SHIELD, PART # 11, TO THE POWER SUPPLY HOUSING.
12 CUT COPPER TAPE 3/4" LONG (4 PLACES) ITEM NUMBER
PART NUMBER
DESCRIPTION
1
27252
BASE
1
2
27256
APPLICATOR HOLDER
1
36
QTY REQ'D
8 PARTS
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
8.4 BASE ASSEMBLY (CONTINUED) ITEM NUMBER
PART NUMBER
DESCRIPTION
QTY REQ'D
3
27259
CONNECTOR INFILL PANEL
1
4
27367
FAN SEAL
1
5
27265
POWER SUPPLY
1
6
27158
FAN
1
7
27152
UNIVERSAL SNAP-IN INLET (IEC SOCKET)
1
9
27136
M4 X 35MM SCREW PAN HEAD
2
10
27142
M3 X 6MM SCREW PAN HEAD
2
11
27592
SHIELD T- PORT POWER SUPPLY
1
12
28152
COPPER TAPE 1/2"
0.25FT
37
8 PARTS
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
8.5 FINAL ASSEMBLY
1
FAN CONNECTOR
5 9
12 POWER SUPPLY CONNECTOR
10
2 10 10
6
10 6
LEFT
RIGHT
BATTERY CONNECTOR 7 (OPTIONAL)
3
4
11 ITEM NUMBER
PART NUMBER
1
SEE TOP ASSY DRAWING SEE BASE ASSY DRAWING
2
DESCRIPTION
QTY REQ'D
TOP ASSEMBLY
1
BASE ASSEMBLY
1
38
8 PARTS
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
8.5 FINAL ASSEMBLY (CONTINUED) ITEM NUMBER
PART NUMBER
DESCRIPTION
QTY REQ'D
3
27410
BATTER COMPARTMENT COVER ASSEMBLY
1
4
27253
PLYNTH
1
5
27361 27263
1
6 7
27363 27365 27478
REAR VENT (GREY) REAR VENT (BLUE) FOOT LEFT FOOT RIGHT BATTERY PACK
9
27373
FOAM BAFFLE
1
10
27142
M3 X 6MM PAN HEAD SCREW
4
11
27150
FEET 1/2"D X 1/4" H BLK
2
12
27821
TAPE 3M
1 1 1
.17FT
39
LEDYW
D19
R132 1.00K
VIN
22
3
33uF/25V
C87
GND
220K
R133
1
VCC
9
R135 330K
1
Q17 2N7002
R134 4.7K
74HC04
U17D
GND
R136 270
3
8
LEDGN
D21
Ready
74HC04
U17B
C79 18pF
47
2
4
R138 2.2K
GND
R119
GND
GND
C81 0.1uF
FB11 HZ1206
1
R120 10K
1
10
9
19
20
8
11
1
2
MCLR/Vpp
VCC
GND
GND
74HC04
U17E
74HC04
10
2
C89 0.1uF
BZX84C5V1
D15
2
2
2
D16 BAS16
RC0/T1OSO/T1CKI RC1/T1OSI/CCP2 RC2/CCP1 RC3/SCK/SCL RC4/SDI/SDA RC5/SDO RC6/TX/CK RC7/RX/DT
RB0/INT RB1 RB2 RB3/PGM RB4 RB5 RB6/PGC RB7/PGD
RA0/AN0 RA1/AN1 RA2/AN2/Vref-/CVref RA3/AN3/Vref+ RA4/TOCKI/C1OUT
U17A
1
PIC18F252-20/SO
OSC2/CLKOUT
OSC1/CLKIN
Vss
Vdd
Vss
RA5/AN4/SS
GND R139 2.2K
1
21
3 D22 BAS16
11 12 13 14 15 16 17 18
21 22 23 24 25 26 27 28
2 3 4 5 6
VCC
5
R141 33K
13
1
VCC
2
74HC04
U17F
74HC04
U17C
12
6
Duty Cycle Shutdown Active High
R115 270
1
TP13
R142 270
TESTPOINT
22 D23 LEDRD
Duty Cycle
DUTY_CYCLE
/BUS_RST
GND
1
C76 0.1uF
1 GND
2
1
BUS_DI
BUS_DO
BUS_CLK
CAL
I2C_CLOCK
I2C_DATA
/CS_US
DUTY_CYCLE
V_SUPPLY
V_SAMPLE
I_SAMPLE
1_3MHZ
TP15
/BUS_RST
STATUS 2
STATUS 1
PWM
2
1 47
R137
GND
2
2
C30 1000pF
2
1
GND
2
1
D18 LEDGN
D17 LEDRD
HEAD_DATA
C96 470pF
R158 1.00K
C78 1000pF
1
1
C101 33pF
PGD
PGC
/CS_US
GND
HEAD_CLOCK
BUS_DI BUS_DO BUS_CLK
R122 270
R121 270
GND
C97 470pF
R159 1.00K
GND_US
24V_US
C88 1000pF
J3
47
C92 33pF
R147
1 1 3 5 7 9
2
GND
C93 33pF
47
VCC
TX MD1
RX
R148
GND
C77 3.3uF/16V
GND_US
Header 2x5 PUB
2 4 6 8 10
LM4040DIM3-4.1
D14
3
4.096V
VCC
2 1 1 2
7
U16
GND
12
10uF
C91
R124 33K
R123 33K 1
47
C94 33pF
R149
180
R117
VCC
4
3
GND
47
C95 33pF
R150
GND
R126 10K
R125 33K
Precision PWM Reference
V_REF
-
+
J4 39 37 35 33 31 29 27 25 23 21 19 17 15 13 11 9 7 5 3 1
1
GND
U18 LMV7239M5
HEADER 20X2
40 38 36 34 32 30 28 26 24 22 20 18 16 14 12 10 8 6 4 2
GND
1 2
1
GND
VCC
20MHz
Y1
C100 3.3uF/16V
1
GND
GND
D20 BAS16
C80 18pF
GND
11
1
2
R131 2.21K
1
VIN
1
2
HEAD_DETECT
2
11
2
1
2
1 2
2
2
11
23
3
2 2 1
1 2 1
5 2
2
1
2
1
2
1
2
3
2
2
2 1
1 14 7
D13 LEDRD
1 2
1
C83 .01uF/5%
R127 12.1K 21
GND
GND_US
24V_US
GND
GND
Digital Ground
C90 1000pF
VCC
Digital Ground
R128 2.61K 21
3pole Bessel, -3dB @ 1KHz, -51dB@10KHz, -111dB @100KHz
GNDD_US GNDD_US GNDD_US GNDD_US GNDD_US
0.1uF
C82
1 2
2 1 2
2 C84 .068uF/5%
.033uF/5%
C85
R129 20.5K
1
2 1 GND
2
1
1 2
GND
1 2
2
J5 CON2_NL
R130 20.5K
3
1
C86 0.1uF 1
VCC
-
+
5 GND
2
FREQ.
1 2
40
1 2
0
R162
LMV321M
U15 4
PWR
9 SCHEMATICS Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
9.1 ULTRASOUND BOARD (1) - PORTS
GND
I2C_CLOCK I2C_DATA
DUTY_CYCLE
HZ1206
3
2
U2
TPS2819DBV
IN
GND
VDD
VCC
R25 2.2K
C25 0.1uF
GND
OUT
VCC
4
5
1
1
GND
FB2
R26 2.2K
HZ1206
VCC
2
GND
2.2
R9
1 2 3 7 5 6 8 4
C15 0.1uF
1
GND
R10 1.0M
2
SCL SDA ADDR0 ADDR1 XIN XOUT VDD VSS
2
GND
1
U5
CLKN CLKP FBK REF EXTLF LOCK/IPRG VDD VSS
C26 4.7uF/25V
FS6131-01
Q4 IRLR110
C16 4.7uF/25V
C27 0.1uF
12
GND
GND
16 15 13 12 10 9 14 11
GND
GND
2
1
74HC27D
C103
.01uF 50V
OSCILLATOR
GND
EOH
OSC1
13 2 1
1
VCC
OUT
R11 1.0M
Q7 IRLR110
C2 100pF/100V
1
VCC
4
3
620
R27
1
C4 33uF/63V
C3 4.7uF/25V
+40VADJ
2
R12 2.2
39pF
C17
2
R28 620
2
C18 0.1uF
GND
C6 0.1uF/200V
C5 0.1uF/200V
GND
1
GND
GND
4
5
C28
U3 VDD
9 10 11
39pF
OUT
3 4 5
U1B
GND
1
VCC
8
GND
2
74HC27D
U1C
0.1uF
C29
74HC27D
TPS2819DBV
IN
GND
8
1
VCC
6 7
3 2
ENG00787
5
4
T1
6
3
2
1
TESTPOINT
TP14
GND
HZ1206
FB10 VCC
D3 BAS16
VIN
3
NL
C98
R160 22_NL
Q8 2N7002
GND
1 12
9
4
1
TX2-24V
K1
R13 10K
8
1_3MHZ
10
5
3
2
2
L2 2.8uH/2A
L1 11.5uH/2A
1
C10 1000pF/250V 1
C9 2200pF/250V 1 1
VCC
2
2
FB1 HZ1206
8
10
5
3
4.7uF/25V
TX2-24V
K2
C11
1 2 1 12
9
4
GND
1 1
2 R5 0.56
GND
R112 0
2
IRF
US_GND
2
5
8 8 2
5
6
US Ground
TP1 TESTPOINT
GND
1
3
4
6
L3 100uH
C12 0.1uF
SOUNDHEAD CONNECTOR
1
3
4
J1 KMDG-8S-BS
GND
1
1.00K
R18
14.0K
R17
0.1uF
C19
L4 100uH_NL
US_GND
GND
2
GND
2
2
GND
C20 0.1uF_NL
0.1uF
C21
2
3
0.1uF
C22
GND
1
1
-
+
1
1.00K
-5V
LMH6642MA
R21
R22
U4
6
1.5K_NL
VCC
R20 300K_NL
1
R19 1.5K_NL
1
1
2
L10 10uH
L9 10uH
2
GND
C23 0.1uF_NL
C13 10nF
R7 100
2
2
R23 0_NL
GND
0
R151
0
R152
HEAD_DETECT
C14 10nF
R8 100
HEAD_CLOCK VRF
HEAD_DATA
7 7 S1 S2 S3 S1 S2 S3
1 2 2 U1A
R2 47
R1 47
2 1 1 2
1 1
4
1 2
1 1 2 2
1 21 2
1 21 2
GND
1 2
1 2
1
3
1
1
2
1
2
1
2
3 2
1 1 2 2
1 2 1 2
1
2 2 4 1
1 2
1 2
FB9
1
2
VCC
1
2
1 2
3 2
1 2
1 2 2 12 1
1 2
2 12 1 7 4
2 1
C1 100pF/100V
14 7
41 1
9 SCHEMATICS Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
9.2 ULTRASOUND BOARD (2) - AMPS
DUTY_CYCLE
C70 NL
PWR
FEEDBK
I_SAMPLE
IRF
CAL
VRF
GND
47
R64
TP8 TESTPOINT
IOUT- FULL-WAVE RECTIFIER
1
GND
GND
GND
R99 1.0M
1
Q9 2N7002
0- 4.1V
1
1
1
R49 499
1
GND
1
R29 499
2
R50 1.00K
R30 1.00K
R66
R65
4.99K
4.99K
2
2
2
2
2
-
+
-
+
-5V
C55 NL
1
1
12
13
C56 3.3uF/16V_NL 1 2
6
5
2
3
1
GND
R68 1M_NL
1
2
1
2
2
14
4.096V
V_REF
2
LMV824M
U9D
10nF_NL
2
2
GND
GND
C57
R69 100_NL
1.00K
R51
7
1.00K
R31
C32 0.1uF
U7B LMH6644MA
1 1
29.4K
R67
1
1
1
U7A LMH6644MA
C31 0.1uF
2
2
TP9 TESTPOINT
1
1
1
1
2
649
R71
C58 1nF_NL
R53 1.00K
D7 BAT54S
1.00K
R33
D4 BAT54S
R72 1.78K
R52 499
R32 499
2
2
2
1
9
10
13
12
2
2
4
11
1 2
1
GND
GND
R75 511
6.81K
R74
R107 3.32K
47
R73
2
GND
1
2
R76 1.0M
R55 2.00K
2
8
2
1
Q10 2N7002
C43 18pF_NL
U7C LMH6644MA
2
1
14
C33 18pF_NL
U7D LMH6644MA
2.00K
R35
R34 2.00K
R54 2.00K
-
+
-
+
3
2
1
1
2.21K
R77
1
1
GND
GND
C37 0.1uF
U9A LMV824M
1
C44 33nF
C34 33nF
VCC
1
R56 499
R36 499
GND
C45 0.1uF
GND
GND
2
2
1
1
1
1
VCC
TP10 TESTPOINT
GND
2
R58 1.00K
R57 301
R38 2.00K
R37 301
3
2
1
6
5
9
10
MAX4644EUT
GND
V+
IN
U8
R80 4.99K
2
2
2
2
1
1
1
-
+
-
+
NO
C35 22nF
C46 4.7nF
R59 4.02K
U9C LMV824M
NC
COM
C60 10nF
2
2
R39 1.47K
U9B LMV824M
C73 100pF
1
8
4
5
6
1
7
TP2
2
2
GND
1
4.99K
R84
1
TESTPOINT
TP5
2.2K
R108
MMBT3906
560pF
C61
1 2 3 4
2
2
Q13
U13
UCC3808AN-2
COMP FB CS RC
GND
C62 470pF
220K
R109
VDD OUTA OUTB GND
4.7uF/25V
1
C63
1
8 7 6 5
GND
GND
VCC
GND
Q14 MMBT3906
9.09K
R110
2
GND
47
R86
R87 2.2
C64 0.1uF
2
1 R88 1.0M
1
U12 MAX660M
1
D8
D10
8
1 C49
Q15 NDT3055L
R89 0.100 Ohm
V+
3.3uF/16V
2
1
3
2
3
2
2 1 2 1
1 1 2
1 2 1 2
1
FB3 HZ1206
C41 0.1uF
2 R90 200
R60 200
C47 1nF/100V
GND
TP11 TESTPOINT
0-0.300V (0-3A)
R91 200
R61 200
TP3 TESTPOINT 1 1 C66 1nF/100V
Q16 NDT3055L
2
C65
OUT
680uF/35V
VIN
GND
GND
2
TESTPOINT
1
5
C48 0.1uF/50V
3.3uF/16V
C42 0.1uF
VCC
1
GND
-5V SUPPLY
C51
2
GND
1
50mA max.
2 1
ENG00789
T3
1
6 7
5
Analog Ground
GND
8
GND
TP12 TESTPOINT
2.2
R93
40 VOLT ADJUSTABLE SUPPLY
Ground Links
R92 1.0M
C67 0.1uF/50V
3 2
4
GND
C53 0.1uF
FB4 HZ1206
TP4 TESTPOINT
3
1
C54 0.1uF
1
VCC
1
2
7 OSC
1 2 CAP+ GND 3
2 1
4 CAP2 12
2 1
1 FC LV 6
VOUT- FULL-WAVE RECTIFIER
2
1 1
2
3
2
1 2
MBRS1100 MBRS1100
1 1 2
2 1 1 2
V_SAMPLE
1
2
1
2
1 2
1 2
1 2
VIN
20 - 25.5 VDC
RECTIFIER
D24
-5V
C68 0.1uF
4
FB5 FERRITE
R96 2.00K
R95 18.0K
C74 0.1uF
FB12
GND
1
1
1
C75 4.7uF/35V
FB6 FERRITE
GND
2
150
R94
150
R63
R62 150
3
1
C69 0.1uF
L8 47uH
2
2 1
2
1
-
+
GND
VCC
U14 LMV321M
C50 1nF/200V
TP6 TESTPOINT 1 1 4
GND
2
2
2 1
1 2
1 1 2 2 1 2
1 2
3 2 11 4 1 2
1 2
1 1 2
2 1 3 2
2 4 3
1
TESTPOINT
C52 0.1uF/200V
TP7
CON2
J6 1 2
+40VADJ
R161 33K
1
2 1 1 2 4 2 3
1 2 2 1 3 2
2 1
1 1 2 1 2 1
1
1 1
1 2
1 2 1 2 1 2
5 2 1 2
+
-
+
-
42
2
V_SUPPLY
C99 0.1uF
GND_US
24V_US
GND
2
9 SCHEMATICS Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
9.3 ULTRASOUND BOARD (3) - POWER SUPPLY
/BUS_RST
C2 .01uf
VCC
R31 10K
VCC
C52 .1uf
3
2
RST
Q4 2N7002
R6 100K
BUS_DI BUS_DO BUS_CLK
DS1233-10(3)
GND
Vdd
U4
10K
R27
1
/CS_STIM /CS_US /CS_AUX
R38 R39 R40
C4 22pf
MD1 TxD1 RxD1
R41 R42 R43
100 100 100
20MHz
Y1
NMI /STBY MD0
C19 .1uF
RxD0 TxD0 SCK0
VCC
100 100 100
C5 22pf
VCC
43
C16 .1uf
8 7 3 4
9
15 70 71 92
96 95 94 93 46
48 47 31 30 21 20 11 10
91 90 81 80 69 68 58 57
13 14 12
2 5 97 98 99
3
1 100
7 8 6
59 4 37 36
10K
R57
25AA080
Vcc /HOLD /WP Vss
U8
H8S2144AF
VCC2
VSS VSS VSS VSS
P83 P82 P81 P80 AVSS
PA0 PA1 PA2 PA3 PA4 PA5 PA6 PA7
PB0 PB1 PB2 PB3 PB4 PB5 PB6 PB7
RxD0 TxD0 SCK0
XTAL MD1 TxD1 RxD1 SCK1
EXTAL
/RES RES0
NMI /STBY MD0
VCC1 VCC1 AVCC AVref
U1
/CS SCLK SI SO
1 6 5 2
P90 P91 P92 P93 P94 P95 P96 P97
P70 P71 P72 P73 P74 P75 P76 P77
TxD2 RxD2 SCK2
P60 P61 P62 P63 P64 P65 P66 P67
P30 P31 P32 P33 P34 P35 P36 P37
P20 P21 P22 P23 P24 P25 P26 P27
P10 P11 P12 P13 P14 P15 P16 P17
P43 P44 P45 P46 P47
/EEPROM_CS
25 24 23 22 19 18 17 16
38 39 40 41 42 43 44 45
49 50 51
26 27 28 29 32 33 34 35
SCK0 TxD0 RxD0
R56
R66 R67
10K
24V_MON
FAN
BATT_IN BACKLIGHT
82 83 84 85 86 87 88 89 /EEPROM_CS /BATT_EN
C_INH DQ AC_IN SD_REQ SHUTDOWN 24V_EN Vee_EN BEEPER
/CS RS R/W E /RST
67 66 65 64 63 62 61 60
79 78 77 76 75 74 73 72
52 53 54 55 56
VCC
1K 1K
R15 1K
1N6263
D17
1N6263
D16
1N6263
D15
1N6263
D14
R16 1K
22pf
R14 1K
C40
R82 R83 R84 R85 R86
22pf
1 2 3 4 5 6 7 8
C39
/CS RS R/W NMI /STBY MD0 E /RST 10K
RP1
R17 1K
1 2 3 4
22pf
C41
TxD2 RxD2
DIP1 SW DIP-4
22pf
C42
100 100 100 100 100
16 15 14 13 12 11 10 9
R22 1K
22pf
C43
VCC
SW12
SW6
SW2
8 7 6 5
22pf
C44
R23 1K
22pf
C45
SW13
SW10
SW7
SW3
IND_1
LED1
R18 1K
22pf
C46
R24 1K
1000pf
C54
SW14
SW11
SW8
SW4
IND_2
LED2
R19 1K
1000pf
C55
R25 1K
SW5
IND_3
LED3
R20 1K
1000pf
C56 1000pf
C57
IND_4
LED4
R21 1K
VCC
1000pf
C58
7 8 9 10 11 12 13 14
15 4 5 6 16
VSS VSS VSS VSS
CC
VCC VSS +24V VSS +24V VSS
VSS VSS VSS VSS VSS
VSS RxD1 TxD1 MD1 VCC
RxD2 TxD2
R49 10K
C6 .1uf
LCD Display Connector: Samtec HLE-122-02-S-DV-BE-A
DB0 DB1 DB2 DB3 DB4 DB5 DB6 DB7
/CS RS R/W E /RST
DSP1
R50 10K
R51 10K
VCC
LCD 240 * 128
1 3 5 7 9
2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 36 38 40
2 4 6 8 10
HEADER 20X2 Connector: Samtec TLE-120-01-G-DV-A
1 3 5 7 9 11 13 15 17 19 21 23 25 27 29 31 33 35 37 39
JP1
PUB BUS
J1
R74 100K
VCC
Cathode VSS
VEE
VO
VCC Anode
R47
VEE
VCC
33
CC VCC
/BUS_RST
VSS VSS
VSS +24V VSS +24V VSS VCC
BUS_DI BUS_DO BUS_CLK
B7
B6
MCLR
K 1
17
3
2 A
100
R53
CON4
1 2 3 4
100
R52
CN2
1K
R13
18K
R12
Q6 2N7002
22pf
C47
R75 10K
R4 100K
Panasonic: EVUF3AF25B14 R35 10K
22pf
C48
R76 10K
BACKLIGHT
22pf
C49
R77 10K
VCC
/CS_AUX
/CS_US /CS_STIM
FAN
BEEPER
24V_MON
R54 100K
R7 100K
+24V
VCC
VCC
1 2
Q7 ZVN4306GV
FAN
J3
Q5 2N7002
BEEPER
BP1
C22 .01uf
D20 1N6263
R60 3.01K
R59 22K
+24V
9 SCHEMATICS Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
9.4 CONTROL BOARD (1)
Vee_EN
C61 10uf / 35V
C_INH
24V_DC
C24 1uf
10uH
L2
6.8
R81
R62 39K
LT1617
GND
/SHDN
Vin
U7
4 2 5 3
C17 100uf
+
C25 .022uf
NFB
SW
C18 100uf
3
1
D7 1N5819
C35 10uF / 50V
R63 4.7K
5 4
2
2
4
5
470pF
C51
+
8 7
VCC
FB OUT ON/OFF GND
LM2575T5.0(5)
IN
U6
C14 2.2uf
1N6263
1
C13 .01uf
D4
C7 .1uF
C36 4.7uF
R34 10K
C8 330uf / 35V
+24V
C12 .01uf
C37 .01uf
C1 .01uf
15
LM3478
PGND Vss
COMP
Vcc FA/SD
U9
10uH
L3
15
R78
.22uF
C10
R79
L1 330mH
FB
ISEN
DR
C9 330uF
.001uf
C38
D9 1N5819
3
1
6
10uH
L5
VCC
15K
R10
R64 26.1K
50K
VR1 Vee Cal
C59 2.2uf
R48 270K
R65 .033 OHM
Q8 IRFR4105
VeeTP1
680
R46
C50 100pf
D21 B560C-13
C27 100pf
VEE
C11 10uF / 50V
LED6 5V OK
R80 100
R61 560K
D22 1N6263
/BATT_EN
C53 .1uF
C23 .1uF
10K
R33
VCC
L4 10uH
3.9K
C20 100uf
R71
6
5
C30 .1uF
+
VSS_BATT
5V_BATT
C33 10uF / 50V
Q3 FMMT551
150K
U5B LM358D
LED7 Vee- OK
1K
R73
VCC
R55
.1uF
C29
7
C32 10uF / 50V
1K
R72
8
R70 10K
R32 .033 OHM
C26 100uF / 35V
VSS_BATT
CC
C28 .01uf
R30 10K
VCC
R37 2.7K
SHUTDOWN
24V_DC
C3 .01uf
R44 18K
24V_DC
24V_BATT
SD_REQ
2
3
1N4148
C21 1uf
VCC
U5A
LM358D
1
D19 1N6263
ON/OFF
1N4148 D11
SW1
C34 .1uF
D10
8 4
VCC
+
44 4
1K
R29
R3 1M
R58 10K
R26 10K
R2 10K
D1 1N6263
15K
R9
Q2 2N7002
C31 .1uF
R28 10K
D12 1N4148
D13 1N4148
BATT_IN
D18 5.1V
R1 100K
VCC
24V OK
LED5
4.7K
R36
Q1 2N7002
10M
R11
VCC
7.5K
R45
1
2
3
D2 1N4148
24V_DC
AC_IN
VCC
D5 1N6263
IPS521
COM
IN
DG
U2
100K
R5
.1uF
C15
1M
R8
4
5
C_INH DQ
C60 2.2uf
R68 R69
B560C-13
D6
B560C-13
D3
24V_BATT
100 100
+24V
24V_BATT CC VSS_BATT 5V_BATT
1 2 3 4 5 6 CON6
J2
75W
1 2
CN1
9 SCHEMATICS Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
9.5 CONTROL BOARD (2)
10 WARRANTY
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
Chattanooga Group, a division of Encore Medical L. P., ("Company") warrants that the Intelect Transport/Vectra Genisys Ultrasound Therapy Systems ("Product") is free of defects in material and workmanship. This warranty shall remain in effect for two years (24 months) from the date of original consumer purchase. If these Products fail to function during the two years warranty period due to a defect in material or workmanship, at the Company’s option, the Company or selling dealer will repair or replace this Product without charge within a period of thirty (30) days from the date on which the Product is returned to the Company or the dealer. All repairs to the Product must be performed by a service center authorized by the Company. Any modifications or repairs performed by unauthorized centers or groups will void this warranty. The warranty period for the applicators is one year (12 months). This Warranty Does Not Cover: • Replacement parts or labor furnished by anyone other than the Company, the selling dealer, or a certified Company service technician. • Defects or damage caused by labor furnished by someone other than Company, the selling dealer or a certified Company service technician. • Any malfunction or failure in the Product caused by product misuse, including, but not limited to, the failure to provide reasonable and required maintenance or any use that is inconsistent with the Product User's Manual. COMPANY SHALL NOT BE LIABLE IN ANY EVENT FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES. Some locations do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation or exclusion may not apply to you. To obtain service from Company or the selling dealer under this warranty: 1. A written claim must be made within the warranty period to the Company or the selling dealer. Written claims made to the Company should be sent to: Chattanooga Group 4717 Adams Road Hixson, TN 37343 USA 1-800-592-7329 1-423-870-2281 1-423-875-5497 FAX chattgroup.com and 2. The Product must be returned to the Company or the selling dealer by the owner. A Return Authorization (RA) Number must be obtained before returning any product to the Company. This warranty gives you specific legal rights and you may also have other rights which vary from location to location. The Company does not authorize any person or representative to create for it any other obligation or liability in connection with the sale of the Product. Any representative or agreement not contained in the warranty shall be void and of no effect. THE FOREGOING WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY WARRANTY OR MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
45
Moving Rehabilitation Forward™
ISO 13485 CERTIFIED
*27670*
4717 Adams Road P.O. Box 489 Hixson, TN 37343 1-800-592-7329 1 423-870-2281 1 423-875-5497 FAX chattgroup.com
27670C
© 2008 Encore Medical, L.P.