Magna-FX and Mini Magna-FX Cannulated Screw Fixation System. Surgical Technique

Magna-FX® and Mini Magna-FX Cannulated Screw Fixation System Surgical Technique 2 Magna-FX® and Mini Magna-FX Cannulated Screw Fixation Surgical Te...
Author: Dana James
122 downloads 2 Views 786KB Size
Magna-FX® and Mini Magna-FX Cannulated Screw Fixation System Surgical Technique

2

Magna-FX® and Mini Magna-FX Cannulated Screw Fixation Surgical Technique

Magna-FX and Mini Magna-FX Surgical Technique Table of Contents Indications for Use for Magna-FX and Mini Magna-FX.................3 Surgical Technique for Magna-FX and Mini Magna-FX Guide Pin Placement.............................................................. 3 Measuring.............................................................................. 3 Drilling................................................................................... 3 Tapping.................................................................................. 3 Screw Insertion...................................................................... 3 Surgical Technique for Femoral Neck Fractures Guide Pin Placement.............................................................. 4 Measuring.............................................................................. 4 Drilling................................................................................... 4 Tapping.................................................................................. 4 Screw Insertion...................................................................... 4 Surgical Technique for Other Metaphyseal indications Guide Pin Placement.............................................................. 5 Measuring.............................................................................. 5 Drilling................................................................................... 5 Tapping.................................................................................. 5 Screw Insertion...................................................................... 5 Instruments and Implants Magna-FX Instruments and Implants...................................... 6 Mini Magna-FX Instruments and Implants............................... 7

Magna-FX® and Mini Magna-FX Cannulated Screw Fixation Surgical Technique

Indications for Use The Magna-FX Cannulated Bone Screws are indicated for fractures of: • The intracapsular hip, femoral condyles, tibial condyles, ankle, acetabulum, pelvis, and other areas where accurate screw placement is required. • The screw can also be applied to slipped capital femoral epiphysis if image intensification fluoroscopy is used to ascertain the proper position of the screw. The Mini Magna-FX Cannulated Bone Screws are indicated for: • Fractures of the medial malleolus, distal tibia, distal radius, calcaneus, talus, patella, and pelvis. • It can also be used in the reduction of small bone fragments, reduction and fixation with cerclage wire, the fixation of bone prominences (i.e., lateral end of clavicle, coracoid process of scapula).

The Magna-FX and Mini Magna-FX Cannulated Screw systems share a common approach to fracture fixation. Each system has it’s respective purpose, but is applied in the same fashion.

1. Guide Pin Placement

3. Drilling Using the cannulated reamer, drill to a depth of no more than 5 mm shorter than the depth gauge reading per the surgeon’s preference of screw purchase. (See Table 1)

Following open or closed reduction of a fracture site, use a guide pin under fluoroscopy to ensure proper screw placement. A 1.7mm, 2.5mm, or 3.2mm guide pin should be used with respect to bone size and density in order to achieve maximum fixation(See Table 1). For larger fractures, multiple guide pins can be used.

4. Tapping (Optional)

If multiple guide pins are required, position the parallel guide pin on the bone over the guide pin. This will ensure accurate parallel guide pin placement.

5. Screw Insertion

2. Measuring Place the depth gauge over the guide pin and read the actual depth of the pin in the bone. The surgeon may elect to use a screw 5mm less than the depth gauge reading per his/her fracture reduction techniques for increased bone purchase.

Table 1: Instrument Sizing Guide Screw Diameter

7.0mm

5.0mm

4.0mm

Guide Pin

3.2mm

2.5mm

1.7mm

Tap

7.0mm

5.0mm

4.0mm

Cannulated Drill/Reamer

5.0mm

3.5mm

2.7mm

The self-cutting threads of Magna-FX and Mini MagnaFX screws allow tapping to be optional. Place the cannulated tap over the guide pin and tap the proximal cortex (See Table 1). In patients with hard bone, it may be necessary to tap the entire reamed length.

Drilling manually, or using a power drill, with a cannulated screw driver bit, insert the proper length Magna-FX or Mini MagnaFX Screw over the guide pin. A washer may be used to distribute stress over the cortex of the bone. When inserting screws, if the cortex is thin or damaged, washers must be used to prevent the screws from pulling through the cortex. Using the manual cannulated driver, finish seating the screw and check fracture impaction with x-ray. Threads must not extend across the fracture site. Remove guide pin. Repeat steps 2-5 for additional screws, and close.

3

4

Magna-FX® and Mini Magna-FX Cannulated Screw Fixation Surgical Technique

Femoral Neck Fractures - Alignment Guide Technique

The Magna-FX Cannulated Screw Fixation System is appropriate for use in Stabilization of femoral neck fractures with multiple screws, as well as for fixation of other metaphyseal fractures. Exposure and reduction are performed appropriate to the site and alignment technique used.

1. Guide Pin Placement* A. Following open or closed reduction of the fracture, use the Variable Angle Drill Guide under fluoroscopy. Insert a 3.2mm, 9” long Guide Pin through the center of the femoral neck and across the fracture site, engaging the subchondral bone.

Through the placement holes, drive C. three to four 3.2mm, 9” long Guide Pins in a triangular pattern until they engage the subchondral bone. A minimum of three screws is required to adequately secure femoral neck fractures. Remove the Parallel Pin Guide and the center pin.

2. Measuring* Place the Depth Gauge over the Guide Pin and read the actual depth of the pin in the bone. The surgeon may elect to use a screw 5mm less than the Depth Gauge reading.

4. Tapping* (Optional) The self-cutting threads of the Magna-FX Screw allow tapping to be optional. Place the Cannulated Tap over the Guide Pin and tap the proximal cortex. In patients with hard bone, it may be necessary to tap the entire reamed length.

5. Screw Insertion* A. Using a power drill with the Cannulated Screwdriver bit, insert the proper length Magna-FX Screw over the Guide Pin. A washer may be used to distribute stress over the cortex of the bone. When inserting screws, if the cortex is thin or damaged, washers must be used to prevent the screws from pulling through the cortex.

B. Position the Parallel Pin Guide on the bone over the Guide Pin.

3. Drilling* (Optional in poor bone) Using the Cannulated Reamer, drill to a depth of 10mm less than the actual depth of the pin.

B. Using the manual Cannulated Driver, finish seating the screw and check fracture impaction with x-ray. Threads must not extend across the fracture site. Remove Guide Pin. Repeat steps 2-5 for additional screws, and close.

Magna-FX® and Mini Magna-FX Cannulated Screw Fixation Surgical Technique

Freehand Technique

The Magna-FX Cannulated Screw Fixation System is also appropriate for freehand insertion.

1. Guide Pin Placement* A. Following fracture reduction under fluoroscopy, insert a 3.2mm, 9” long Guide Pin across the fracture site, either freehand or using the Freehand 3.2mm Pin Guide, engaging the opposite cortex if necessary

B. If additional fixation screws are to be used, reposition and repeat A.

2. Measuring* Place the Depth Gauge over the Guide Pin and read actual depth of the pin in the bone. The surgeon may elect to use a screw 5 mm less than the Depth Gauge reading.

3. Drilling* (Optional in poor bone) Using the Cannulated Reamer, Drill to a depth 10mm less than the actual depth of the pin. (Optional in poor bone)

4. Tapping* (Optional) The self-cutting threads of the Magna-FX Screw allow tapping to be optional. Place the Cannulated Tap over the Guide Pin and tap the proximal cortex. In patients with hard bone, it may be necessary to tap the entire reamed length.

*PRECAUTION: During placement of the guide pins, measuring, reaming, tapping, and screw insertion, check the guide wire position frequently using fluoroscopy to prevent unintended guide wire advancement and/or penetration into the surrounding tissues.

5. Screw Insertion* A. Using a power drill with the Cannulated Screwdriver bit, insert the proper length Magna-FX Screw over the Guide Pin. A washer may be used to distribute stress over the cortex of the bone. When inserting screws, if the cortex is thin or damaged, washers must be used to prevent the screws from pulling through the cortex.

B. Using the manual Cannulated Screw- driver, finish seating the screw and check fracture impaction with x-ray. Remove Guide Pin. Repeat steps 2-5 for additional screws, and close.

5

6

Magna-FX® and Mini Magna-FX Cannulated Screw Fixation Surgical Technique

Magna-FX Ordering Information Item No.

Description

Qty. in Set

Magna-FX Cannulated Screws Thread Dia.: 7.0mm, Thread Length: 16mm

Item No.

Description

Qty. in Set

00-1146-030-00

30mm

2

00-1146-040-32

40mm

1

00-1146-035-00

35mm

2

00-1146-045-32

45mm

1

00-1146-040-00

40mm

2

00-1146-050-32

50mm

1

00-1146-045-00

45mm

2

00-1146-055-32

55mm

1

00-1146-050-00

50mm

2

00-1146-060-32

60mm

1

00-1146-055-00

55mm

2

00-1146-065-32

65mm

1

00-1146-060-00

60mm

2

00-1146-070-32

70mm

2

00-1146-065-00

65mm

2

00-11460-75-32

75mm

2

00-1146-070-00

70mm

4

00-1146-080-32

80mm

2

85mm

2

00-1146-075-00

75mm

4

00-1146-085-32

00-1146-080-00

80mm

4

00-1146-090-32

90mm

2

00-1146-085-00

85mm

4

00-1146-095-32

95mm

2

00-1146-090-00

90mm

4

00-1146-100-32

100mm

2

00-1146-095-00

95mm

4

00-1146-105-32

105mm

1

00-1146-100-00

100mm

4

00-1146-110-32

110mm

1

00-1146-105-00

105mm

2

00-1146-115-32

115mm

1

00-1146-110-00

110mm

2

00-1146-120-32

120mm

1

00-1146-115-00

115mm

2

00-1146-120-00

120mm

2

00-1146-125-00

125mm

2

Magna-FX Cannulated Screws Thread Dia.: 7.0mm, Thread Length: Full

00-1146-130-00

130mm

2

00-4900-065-51

Large Washer

1

Description

Qty. in Set

Magna-FX Instrumentation

Magna-FX Cannulated Screws Thread Dia.: 7.0mm, Thread Length: 32mm

Optional Implant

Item No.

00-1146-004-00

Percutaneous Guide 3.2

1

00-1146-006-00

Percutaneous Guide 5.0

1

00-1146-008-00

Percutaneous Guide 8.0

1

00-1146-082-00

Screw Case

1

00-1179-042-00

Cann Reamer 3.2mm X 5.0mm Dia

1

00-1179-043-00

Cann Reamer 2.0mm X 5.0mm Dia

1

00-1181-021-00

THD GDE Pin Trocar 3.2mm X 460mm

1

00-1181-022-00

THD GDE Pin Trocar 2.0mm X 460mm

1

00-2237-043-00

5mm Hex Head Screwdriver

1

00-4812-001-00

Self Holding Screw Forceps, STD

1

00-1179-080-00

Pelvic Instrument Case

1

00-1146-001-00

Cannulated Screwdriver (for T-handle)

1

00-1146-002-00

Cannulated Screwdriver (for Jacobs Chuck)

1

00-1146-003-00

T-handle

1

00-1146-007-00

7.0mm Cannulated Tap

1

00-1146-010-00

Adjustable Pin Guide

1

00-1146-030-99

30mm

2

00-1146-035-99

35mm

2

00-1146-040-99

40mm

2

00-1146-045-99

45mm

2

00-1146-050-99

50mm

2

00-1146-055-99

55mm

2

00-1146-060-99

60mm

3

00-1146-012-00

Parallel Pin Guide

1

00-1146-065-99

65mm

3

00-1146-015-00

3.2mm Pin Guide

1

00-1146-070-99

70mm

4

00-1146-017-00

1

00-1146-075-99

75mm

4

Screw Extractor

00-1146-020-00

80mm

4

5.0mm Cannulated Reamer

2

00-1146-080-99 00-1146-085-99

85mm

4

00-1146-081-00

90mm

4

Cannulated Screw/ Instrument Case

1

00-1146-090-99 00-1146-095-99

95mm

4

00-1180-040-00

Depth Gauge

1

00-1146-100-99

100mm

4

00-1181-020-00

3.2 mm Guide Pins (Box of 5)

1

00-1146-105-99

105mm

2

00-0851-000-00

Hip Ruler

1

00-1146-110-99

110mm

2

00-1146-115-99

115mm

2

00-1146-120-99

120mm

2

00-1146-125-99

125mm

2

00-1146-130-99

130mm

2

Magna-FX® and Mini Magna-FX Cannulated Screw Fixation Surgical Technique

Mini Magna-FX Ordering Information Item No.

Description

Qty. in Set

Mini Magna-FX Cannulated Screws Fully Threaded Thread Dia.: 4.0mm

Item No.

Description

Qty. in Set

Mini Magna-FX Cannulated Screws Fully Threaded Thread Dia.: 5.0mm

Item No.

Description

Qty. in Set

00-1142-040-12

12mm

1

00-1142-050-12

12mm

1

00-1142-040-16

16mm

1

00-1142-050-16

16mm

1

00-1142-040-20

20mm

1

00-1142-050-20

20mm

1

00-1142-040-24

24mm

1

00-1142-050-24

24mm

1

Mini Magna-FX Instrumentation 00-1142-000-00 Mini Magna-FX Cannulated Screw Set (Set includes the following instruments, and sterilization case w/lid)

00-1142-040-28

28mm

2

00-1142-050-28

28mm

2

00-1142-017-00 Countersink

1

00-1142-040-32

32mm

2

00-1142-050-32

32mm

2

1

00-1142-040-36

36mm

2

00-1142-050-36

36mm

2

00-1142-040-40

40mm

2

00-1142-050-40

40mm

2

00-1142-013-00 Parellel Placement Jig 00-1142-001-17 Guide Pin, Partially Threaded, 1.7mm

00-1142-040-44

44mm

1

00-1142-050-44

44mm

1

00-1142-040-48

48mm

1

00-1142-050-48

48mm

1

00-1142-040-55

55mm

1

00-1142-050-55

55mm

1

00-1142-040-65

65mm

1

00-1142-050-65

65mm

1

Mini Magna-FX Cannulated Screws Partially Threaded Thread Dia.: 4.0mm

Mini Magna-FX Cannulated Screws Partially Threaded Thread Dia.: 5.0mm

00-1142-041-24

24mm

1

00-1142-051-24

24mm

1

00-1142-041-26

26mm

1

00-1142-051-26

26mm

1

00-1142-041-28

28mm

1

00-1142-051-28

28mm

1

00-1142-041-30

30mm

1

00-1142-051-30

30mm

1

00-1142-041-32

32mm

1

00-1142-051-32

32mm

1

00-1142-041-34

34mm

1

00-1142-051-34

34mm

1

00-1142-041-36

36mm

2

00-1142-051-36

36mm

1

00-1142-041-38

38mm

2

00-1142-051-38

38mm

1

00-1142-041-40

40mm

2

00-1142-051-40

40mm

1

00-1142-041-42

42mm

2

00-1142-051-42

42mm

1

00-1142-041-44

44mm

2

00-1142-051-44

44mm

1

00-1142-041-46

46mm

2

00-1142-051-46

46mm

1

00-1142-041-48

48mm

1

00-1142-051-48

48mm

1

00-1142-041-50

50mm

1

00-1142-051-50

50mm

1

00-1142-041-55

55mm

1

00-1142-051-55

55mm

1

00-1142-041-60

60mm

1

00-1142-051-60

60mm

1

00-1142-041-65

65mm

1

00-1142-051-65

65mm

1

00-1142-041-70

70mm

1

00-1142-051-70

70mm

1

00-1142-018-00

Washer

6

00-1142-001-25 Guide Pin, Partially Threaded, 2.5mm 00-1142-002-00 Cannulated Screwdriver, 3.5mm 00-1142-011-27 Cannulated Drill, 2.7mm 00-1142-011-35 Cannulated Drill, 3.5mm 00-1142-012-00 Free-Hand Drill Guide 00-1142-014-00 Depth Gauge 00-1142-015-00 Screw Extractor 00-1142-016-27 Insert Drill, Sleeve, 2.7mm 00-1142-016-35 Insert Drill, Sleeve, 3.5mm 00-1142-019-00 Guide Pin Sheath 00-1142-020-40 Cannulated Tap, 4.0mm 00-1142-020-50 Cannulated Tap, 5.0mm 00-1142-080-00 Mini Magna-FX Implant/ Instrument Sterilization Case Set

1

1 Box of 5 1 Box of 5 1 1 1 1 1 1 1 1 1 1 1 1

7

DISCLAIMER: This documentation is intended exclusively for physicians and is not intended for laypersons. Information on the products and procedures contained in this document is of a general nature and does not represent and does not constitute medical advice or recommendations. Because this information does not purport to constitute any diagnostic or therapeutic statement with regard to any individual medical case, each patient must be examined and advised individually, and this document does not replace the need for such examination and/or advise in whole or in part. Please refer to the package inserts for important product information, including, but not limited to, indications, contraindications, warnings, precautions, and adverse effects.

Contact your Zimmer representative or visit us at www.zimmer.com

The CE mark is valid only if it is also printed on the product label.

97-1146-103-00 Rev. 1 12-17-15 ©2015 Zimmer, Inc.