Lecture 1
The Principles of Microscopy •
BMS 524 - “Introduction to Confocal Microscopy and Image Analysis” Purdue University Department of Basic Medical Sciences, School of Veterinary Medicine
J.Paul Robinson, Ph.D. • Professor of Immunopharmacology • Director, Purdue University Cytometry Laboratories These slides are intended for use in a lecture series. Copies of the graphics are distributed and students encouraged to take their notes on these graphics. The intent is to have the student NOT try to reproduce the figures, but to LISTEN and UNDERSTAND the material. All material copyright J.Paul Robinson unless stated. Textbook for this lecture series in Jim Pawley’s “Handbook of Confocal Microscopy” Plenum Press which has been used extensively for material and ideas to support the class. UPDATED December 27, 1998 J.Paul Robinson - Purdue University Cytometry Laboratories
Slide 1 t:/PowerPoint/confoc/524lect1.ppt
Evaluation • End of term quiz - 100% grade
J.Paul Robinson - Purdue University Cytometry Laboratories
Slide 2 t:/PowerPoint/confoc/524lect1.ppt
Introduction to the Course • • • • • • • •
Microscopy Fluorescence Basic Optics Confocal Microscopes Basic Image Analysis 3D image analysis Live Cell Studies Advanced Applications
J.Paul Robinson - Purdue University Cytometry Laboratories
Slide 3 t:/PowerPoint/confoc/524lect1.ppt
Introduction to Lecture 1 • • • • •
Early Microscopes Modern Microscopes Magnification Nature of Light Optical Designs
J.Paul Robinson - Purdue University Cytometry Laboratories
Slide 4 t:/PowerPoint/confoc/524lect1.ppt
Microscopes • • • •
Upright Inverted Köhler Illumination Fluorescence Illumination "Microscope" was first coined by members of the first "Academia dei Lincei" a scientific society which included Galileo
J.Paul Robinson - Purdue University Cytometry Laboratories
Slide 5 t:/PowerPoint/confoc/524lect1.ppt
Earliest Microscopes •
1590 - Hans & Zacharias Janssen of Middleburg, Holland manufactured the first compound microscopes • 1660 - Marcello Malpighi circa 1660, was one of the first great microscopists, considered the father embryology and early histology - observed capillaries in 1660 • 1665 - Robert Hooke (1635-1703)- book Micrographia, published in 1665, devised the compound microscope most famous microscopical observation was his study of thin slices of cork. He wrote:
“. . . I could exceedingly plainly perceive it to be all perforated and porous. . . these pores, or cells, . . . were indeed the first microscopical pores I ever saw, and perhaps, that were ever seen, for I had not met with any Writer or Person, that had made any mention of them before this.”
J.Paul Robinson - Purdue University Cytometry Laboratories
Slide 6 t:/PowerPoint/confoc/524lect1.ppt
Earliest Microscopes •1673 - Antioni van Leeuwenhoek (1632-1723) Delft, Holland, worked as a draper (a fabric merchant); he is also known to have worked as a surveyor, a wine assayer, and as a minor city official. •Leeuwenhoek is incorrectly called "the inventor of the microscope" •Created a “simple” microscope that could magnify to about 275x, and published drawings of microorganisms in 1683 •Could reach magnifications of over 200x with simple ground lenses - however compound microscopes were mostly of poor quality and could only magnify up to 20-30 times. Hooke claimed they were too difficult to use - his eyesight was poor. •Discovered bacteria, free-living and parasitic microscopic protists, sperm cells, blood cells, microscopic nematodes •In 1673, Leeuwenhoek began writing letters to the Royal Society of London - published in Philosophical Transactions of the Royal Society •In 1680 he was elected a full member of the Royal Society, joining Robert Hooke, Henry Oldenburg, Robert Boyle, Christopher Wren
J.Paul Robinson - Purdue University Cytometry Laboratories
Slide 7 t:/PowerPoint/confoc/524lect1.ppt
Secondary Microscopes •
George Adams Sr. made many microscopes from about 1740-1772 but he was predominantly just a good manufacturer not inventor (in fact it is thought he was more than a copier!)
•
Simple microscopes could attain around 2 micron resolution, while the best compound microscopes were limited to around 5 microns because of chromatic aberration
•
In the 1730s a barrister names Chester More Hall observed that flint glass (newly made glass) dispersed colors much more than “crown glass” (older glass). He designed a system that used a concave lens next to a convex lens which could realign all the colors. This was the first achromatic lens. George Bass was the lens-maker that actually made the lenses, but he did not divulge the secret until over 20 years later to John Dolland who copied the idea in 1759 and patented the achromatic lens.
•
In 1827 Giovanni Battista Amici, built high quality microscopes and introduced the first matched achromatic microscope in 1827. He had previously (1813 designed “reflecting microscopes” using curved mirrors rather than lenses. He recognized the importance of coverslip thickness and developed the concept of “water immersion”
J.Paul Robinson - Purdue University Cytometry Laboratories
Slide 8 t:/PowerPoint/confoc/524lect1.ppt
Lister, Abbe, Zeiss & Schott •
•
In 1830, by Joseph Jackson Lister (father of Lord Joseph Lister) solved the problem of Spherical Aberration - caused by light passing through different parts of the same lens. He solved it mathematically and published this in the Philosophical Transactions in 1830 Ernst Abbe together with Carl Zeiss published a paper in 1877 defining the physical laws that determined resolving distance of an objective. Known as Abbe’s Law “minimum resolving distance (d) is related to the wavelength of light (lambda) divided by the Numeric Aperture, which is proportional to the angle of the light cone (theta) formed by a point on the object, to the objective”.
•
•
Abbe and Zeiss developed oil immersion systems by making oils that matched the refractive index of glass. Thus they were able to make the a Numeric Aperture (N.A.) to the maximum of 1.4 allowing light microscopes to resolve two points distanced only 0.2 microns apart (the theoretical maximum resolution of visible light microscopes). Leitz was also making microscope at this time. Dr Otto Schott formulated glass lenses that color-corrected objectives and produced the first “apochromatic” objectives in 1886.
J.Paul Robinson - Purdue University Cytometry Laboratories
Slide 9 t:/PowerPoint/confoc/524lect1.ppt
Modern Microscopes • Early 20th Century Professor Köhler developed the method of illumination still called “Köhler Illumination” • Köhler recognized that using shorter wavelength light (UV) could improve resolution
J.Paul Robinson - Purdue University Cytometry Laboratories
Slide 10 t:/PowerPoint/confoc/524lect1.ppt
Köhler • Köhler illumination creates an evenly illuminated field of view while illuminating the specimen with a very wide cone of light • Two conjugate image planes are formed – one contains an image of the specimen and the other the filament from the light
J.Paul Robinson - Purdue University Cytometry Laboratories
Slide 11 t:/PowerPoint/confoc/524lect1.ppt
Köhler Illumination condenser
Field iris
Specimen
eyepiece Field stop retina
Conjugate planes for image-forming rays
Field iris
Specimen
Field stop
Conjugate planes for illuminating rays
J.Paul Robinson - Purdue University Cytometry Laboratories
Slide 12 t:/PowerPoint/confoc/524lect1.ppt
Some Definitions • Absorption – When light passes through an object the intensity is reduced depending upon the color absorbed. Thus the selective absorption of white light produces colored light.
• Refraction – Direction change of a ray of light passing from one transparent medium to another with different optical density. A ray from less to more dense medium is bent perpendicular to the surface, with greater deviation for shorter wavelengths
• Diffraction – Light rays bend around edges - new wavefronts are generated at sharp edges - the smaller the aperture the lower the definition
• Dispersion – Separation of light into its constituent wavelengths when entering a transparent medium - the change of refractive index with wavelength, such as the spectrum produced by a prism or a rainbow
J.Paul Robinson - Purdue University Cytometry Laboratories
Slide 13 t:/PowerPoint/confoc/524lect1.ppt
Refraction Short wavelengths are “bent” more than long wavelengths
dispersion
Light is “bent” and the resultant colors separate (dispersion). Red is least refracted, violet most refracted. J.Paul Robinson - Purdue University Cytometry Laboratories
Slide 14 t:/PowerPoint/confoc/524lect1.ppt
Refraction
He sees the fish here… .
But it is really here!! J.Paul Robinson - Purdue University Cytometry Laboratories
Slide 15 t:/PowerPoint/confoc/524lect1.ppt
Control
Absorption
No blue/green light red filter J.Paul Robinson - Purdue University Cytometry Laboratories
Slide 16 t:/PowerPoint/confoc/524lect1.ppt
Light absorption white light
blue light
J.Paul Robinson - Purdue University Cytometry Laboratories
red light
green light
Slide 17 t:/PowerPoint/confoc/524lect1.ppt
Absorption Chart Color Colorin inwhite whitelight light
Color Colorof oflight lightabsorbed absorbed
red
blue
green
blue green
red red
green
yellow
blue
magenta
blue
green
black
red red
green
gray
pink
green
cyan
J.Paul Robinson - Purdue University Cytometry Laboratories
blue blue Slide 18 t:/PowerPoint/confoc/524lect1.ppt
The light spectrum Wavelength ---- Frequency Blue light 488 nm short wavelength high frequency high energy (2 times the red)
Photon as a wave packet of energy
Red light 650 nm long wavelength low frequency low energy J.Paul Robinson - Purdue University Cytometry Laboratories
Slide 19 t:/PowerPoint/confoc/524lect1.ppt
Magnification • An object can be focussed generally no closer than 250 mm from the eye (depending upon how old you are!) • this is considered to be the normal viewing distance for 1x magnification • Young people may be able to focus as close as 125 mm so they can magnify as much as 2x because the image covers a larger part of the retina - that is it is “magnified” at the place where the image is formed J.Paul Robinson - Purdue University Cytometry Laboratories
Slide 20 t:/PowerPoint/confoc/524lect1.ppt
Magnification
1000mm
35 mm slide 24x35 mm
1000 mm M = 35 mm = 28
p
The projected image is 28 times larger than we would see it at 250 mm from our eyes. If we used a 10x magnifier we would have a magnification of 280x, but we would reduce the field of view by a factor of 10x.
J.Paul Robinson - Purdue University Cytometry Laboratories
Slide 21 t:/PowerPoint/confoc/524lect1.ppt
Some Principles • Rule of thumb is is not to exceed 1,000 times the NA of the objective • Modern microscopes magnify both in the objective and the ocular and thus are called “compound microscopes” - Simple microscopes have only a single lens
J.Paul Robinson - Purdue University Cytometry Laboratories
Slide 22 t:/PowerPoint/confoc/524lect1.ppt
Basic Microscopy • Bright field illumination does not reveal differences in brightness between structural details - i.e. no contrast • Structural details emerge via phase differences and by staining of components • The edge effects (diffraction, refraction, reflection) produce contrast and detail
J.Paul Robinson - Purdue University Cytometry Laboratories
Slide 23 t:/PowerPoint/confoc/524lect1.ppt
Microscope Basics • Originally conformed to the German DIN standard • Standard required the following – real image formed at a tube length of 160mm – the parfocal distance set to 45 mm – object to image distance set to 195 mm
Object to Image Distance = 195 mm
Mechanical tube length = 160 mm
Focal length of objective = 45 mm
• Currently we use the ISO standard
J.Paul Robinson - Purdue University Cytometry Laboratories
Slide 24 t:/PowerPoint/confoc/524lect1.ppt
The Conventional Microscope
Mechanical tube length = 160 mm
Object to Image Distance = 195 mm
Focal length of objective = 45 mm Modified from “Pawley “Handbook of Confocal Microscopy”, Plenum Press
J.Paul Robinson - Purdue University Cytometry Laboratories
Slide 25 t:/PowerPoint/confoc/524lect1.ppt
Upright Scope Epiillumination Source
Brightfield Source
J.Paul Robinson - Purdue University Cytometry Laboratories
Slide 26 t:/PowerPoint/confoc/524lect1.ppt
Inverted Microscope Brightfield Source
Epiillumination Source J.Paul Robinson - Purdue University Cytometry Laboratories
Slide 27 t:/PowerPoint/confoc/524lect1.ppt
Conventional Finite Optics with Telan system
Modified from “Pawley “Handbook of Confocal Microscopy”, Plenum Press
Ocular Intermediate Image 195 mm
160 mm
Telan Optics Other optics Objective 45 mm
Sample being imaged J.Paul Robinson - Purdue University Cytometry Laboratories
Slide 28 t:/PowerPoint/confoc/524lect1.ppt
Infinity Optics Ocular Primary Image Plane
Tube Lens Infinite Image Distance
Other optics Other optics Objective
The main advantage of infinity corrected lens systems is the relative insensitivity to additional optics within the tube length. Secondly one can focus by moving the objective and not the specimen (stage)
Modified from “Pawley “Handbook of Confocal Microscopy”, Plenum Press
Sample being imaged J.Paul Robinson - Purdue University Cytometry Laboratories
Slide 29 t:/PowerPoint/confoc/524lect1.ppt
Summary Lecture 1 • • • • •
Simple versus compound microscopes Achromatic aberration Spherical aberration Köhler illumination Refraction, Absorption, dispersion, diffraction • Magnification • Upright and inverted microscopes • Optical Designs - 160 mm and Infinity optics
J.Paul Robinson - Purdue University Cytometry Laboratories
Slide 30 t:/PowerPoint/confoc/524lect1.ppt