L-40. X-ray High Voltage Generator GENERAL

PD501-002B X-ray High Voltage Generator UD150B-40/V-40/L-40 GENERAL The Shimadzu’s UD150B-40/L-40/V-40 is a high-performance X-ray high voltage gen...
Author: Valentine Lewis
12 downloads 2 Views 1MB Size
PD501-002B

X-ray High Voltage Generator

UD150B-40/V-40/L-40

GENERAL The Shimadzu’s UD150B-40/L-40/V-40 is a high-performance X-ray high voltage generator. It has greatly improved operability and safety with the use of computers and provides a low-ripple output with high X-ray quantum efficiency through a high-frequency inverter system.

The UD150B-40, L-40 and V-40 are suitable for the following applications; (A) General radiography (B) Bucky radiography (C) Auto changer radiography (D) Tomography (E) Fluoroscopy

1

UD150B-40/V-40/L-40 FEATURES (1) Inverter system The high frequency inverter system adopted for high voltage generation, generates low ripple tube voltage with high X-ray quantum efficiency. Moreover, by cutting off X-rays regardless of power phase, it is possible to perform high-accuracy phototimer control in the ultra-short time.

(7) Lights up hand switch The hand switch lights up to indicate Ready Up and X-ray exposure status.

(8) Illumination panel The operation panel indicates the status of the device (e.g., ready to take images or image capture in progress) using color and sound.

(2) Microcomputer To enhance the ergonomic design of the UD150B -40/L40/V-40, a microcomputer is utilized for the improvement of operability and safety.

(9) 2-factors control system In addition to the ordinary setting system of KV, mA, and sec., technique conditions can be also set with 2-factors of kV and mAs. In this case, mA and sec. are automatically set to the maximum allowable value and the shortest value.

(3) Detailed setting of exposure factors Exposure time and mAs can be set more in detail in every 12.5% than the conventional.

(10) Compact size (4) Preset memory

Compact 2-unit construction. The unique spacesaving design of the 2 unit consists of a compact control panel that can be desk or wall mounted and one control cabinet, requiring a very small installation space.

The preset memory allows the most commonly used techniques to be programmed. By pressing the program selector key, automatic selection can be made for radiographic factor, tube focus, etc.

(5) Color LCD Touch Panel

(11) Two standard fluoroscopic modes

A color LCD panel with high visual clarity is used as the screen for setting X-ray conditions. The panel, which uses different colors for extra clarity, allows simple APR selection.

Two standard fluoroscopic modes are provided. Each mode is customised to meet the requirements of clinical application.

(12) Automatic fluoroscopic diagnosis (6) Hybrid Dials

Tube voltage is automatically controlled even if a radiograpic area or subject thickness is changed during fluoroscopy. The image brightness stabilizer (IBS) always ensures a fluoroscopic image with optimum brightness.

X-ray conditions can be changed easily using hybrid dials. Large changes can be made using jog/shuttle and small changes can be made precisely using the up/down buttons. Using both types of adjustment method allows conditions to be set quickly.

CONFIGURATION STANDARD CONSTITUTION (a)

Control console ·················· 1 unit

(b)

Control cabinet ··················· 1 unit

(c)

High voltage generator ······· Incorporated in the control cabinet

(d)

Connection cable················ 1 set Cable length; ∗ Power source-Control cabinet: 10 m (USA and CANADA: 16m) ∗ Control console-Control cabinet: 14 m

2

OPTIONAL ITEMS ∗ ∗ ∗ ∗ ∗ ∗ ∗

SPT-C42, photo-timer control Direct photo-timer pickup Starter Third tube selector Communication unit Voltage regulator Fluoroscopy unit

UD150B-40/V-40/L-40 SPECIFICATIONS (1)

Standard model and nominal maximum power:

(7)

Model

UD150B-40

UD150V-40

UD150L-40

Nominal maximum Power

80 kW

65 kW

50 kW

(2)

Ratings: Model

UD150B-40

UD150V-40

UD150L-40

Power Source

380-480 V, 3 phase 50/60 Hz

380-480 V, 3 phase 50/60 Hz

380-480 V, 3 phase 200-240 V, 1 phase 50/60 Hz

Short-time rating *1

150 kV 125 kV 100 kV 80 kV

500 mA 630 mA 800 mA 1000 mA

150 kV 125 kV 100 kV 80 kV

400 mA 500 mA 650 mA 800 mA

150 kV 125 kV 100 kV 80 kV

320 mA 400 mA 500 mA 630 mA

Long-time rating 75 kV 125 kV *1

20 mA 12 mA

125kV

9 mA

125 kV

9 mA

*1. Limited according to the X-ray tube type.

(3)

(4)

UD150V-40

UD150L-40

1000, 800, 630, 500, 400, 320, 250, 200, 160, 125, 100, 80, 63, 50, 40, 32, 25, 20, 16, 12 and 10 mA

800, 630, 500, 400, 320, 250, 200, 160, 125, 100, 80, 63,50, 40, 32, 25, 20, 16, 12 and 10 mA

630, 500, 400, 320, 250,200, 160, 125, 100, 80, 63, 50, 40, 32, 25, 20, 16, 12 and 10 mA

Of the 21 positions above, it is possible to use any 6 positions allowed by the X-ray tube for one focus.

Of the 20 positions above, it is possible to use any 6 positions allowed by the X-ray tube for one focus.

Of the 19 positions above, it is possible to use any 6 positions allowed by the X-ray tube for one focus.

y Display: The set value is displayed digitally. y Setting method: shuttle and UP/DOWN switch Setting and display of exposure time (in the case of 3-control system) y Setting range: UD150B-40

UD150V-40

UD150L-40

1.0, 1.1, 1.2, 1.4, 1.6, 1.8, 2.0, 2.2, 2.5, 2.8, 3.2, 3.6, 4.0, 4.5, 5.0, 5.6, 6.3, 7.1, 8.0, 9.0, 10, 11, 12, 14, 16, 18, 20, 22, 25, 28, 32, 36, 40, 45, 50, 56, 63, 71, 80 ms, 90 ms, 0.10, 0.11, 0.12, 0.14, 0.16, 0.18, 0.20, 0.22, 0.25, 0.28, 0.32, 0.40, 0.45,

Technique factor setting system

It is impossible to set the mAs value to less 0.5 or over 800.

Setting and display of radiographic tube voltage y Setting range: 40 to 150 kV in increment of 1 kV. y Display: The set value is displayed digitally. y Setting method: shuttle and UP/DOWN switch.

(6)

UD150B-40

The exposure is made and broken regardless of the supply voltage phase.

y 2-control system (kV, mAs) A system that the shortest exposure time and maximum tube current are automatically set for sec and mA according to the allowable load of the X-ray tube when kV and mAs are set as the technique factors. y 3-control system (kV, mA, sec.) A system to set kV, mA, and sec as technique factors. (5)

y Setting range:

(8)

Tube voltage and X-ray exposure time control system The DC voltage rectified and smoothed from 3-phase or single-phase power source is applied to the inverter circuit. Tube voltage is controlled by the feedback system, and at the same time, applied to the X-ray tube, making and breaking it.

Setting and display of radiographic tube current (for 3-control system)

Setting and display of radiographic tube current time product (in the case of 2-control system) y Setting range: UD150B-40

UD150V-40

UD150L-40

0.5, 0.6, 0.7, 0.8, 0.9, 1.0, 1.1, 1.2, 1.4, 1.6, 1.8, 2.0, 2.2, 2.5, 2.8, 3.2, 3.6, 4.0, 4.5, 5.0, 5.6, 6.3, 7.1, 8.0, 9.0, 10, 11, 12, 14, 16, 18, 20, 22, 25, 28, 32, 36, 40, 45, 50, 56, 63, 71, 80, 90, 100, 110, 125, 140, 160, 180, 200, 220, 250, 280, 320, 360, 400, 450, 500, 560, 630, 710, 800 mAs Setting is available in 64 positions shown above. The value is limited to 500 mAs when using photo-timer.

y Display: The set value is displayed digitally. y Setting method: shuttle and UP/DOWN switch

0.50, 0.56, 0.63, 0.71, 0.80, 0.90, 1.0, 1.1, 1.2, 1.4, 1.6, 1.8, 2.0, 2.2, 2.5, 2.8, 3.2, 3.6, 4.0, 4.5, 5.0, 6.3, 7.1, 8.0, 9.0, 10 sec

y Display: The set value is displayed digitally. y Setting method: shuttle and UP/DOWN switch. (9)

Setting and display of fluoroscopic tube voltage y Setting range: 50~125kV in increment of 1kV. y Display: The set value is displayed digitally.

(10) Setting and display of fluoroscopic tube current y Setting range: Fluoroscopic tube current is automatically set according to fluoroscopic tube voltage. Presetting is already made so that the fluoroscopic tube current value rises as the fluoroscopic tube voltage rises. y Display: Digital display of actual value during radiation of X-ray (11) Resetting and display of accumulated fluoroscopic time y Accumulated fluoroscopic time reset: Press the “ ” key. The time is reset to “0.0 min”. y Accumulated fluoroscopic time display: Display range … 0.0~99 min (12) Setting of fluoroscopic timer Press the “ ” key. “10 min” is set. The “ “ key flickers with buzzer sounding during the lapse time from 4 minutes and 30 seconds to 10 minutes. Fluoroscopy is shut out after lapse of 10 minutes. 3

UD150B-40/V-40/L-40 SPECIFICATIONS (cont.) (13) Selectable techniques (A) General radiography (B) Bucky radiography (C) Auto changer radiography (D) Tomography (E) Fluoroscopy ∗1 The technique selector doubles as the X-ray tube selector. (14) Selection of X-ray tube: Use the radiography technique key to select the X-ray tube. Up to two tubes can be connected. (15) Selection of focus: The “large” or “small” focus can be selected. (16) Other selector keys: y The exposure modes “MANUAL” or “PHOTOTIMER” are selectable as the “AEC”. y The technique factor setting systems “2-CONTROL” or “3-CONTROL” are selectable as the “mA・sec” key. y The “Radiography program screen” or “Radiography step screen” can be registered as the “HOME” key. y The “System set-up screen” is displayed to the color LCD as the “SET UP” key. y The “Radiography history screen” is displayed to the color LCD as the “LOG” key. (17) Switches POWER ON/OFF switch and READY/X-RAY hand switch. (18) Indicator “Radiography Ready”, “Exposure”, “Stand-by”, “Caution”, “Fault” are LED-indicated. (19) Preset program: The 7 tab keys for examination area selection (ANATOMICAL TAB KEY) and 15 keys for technique selection (PROTOCOL KEY) enable 105 programs of exposure parameter in total. Furthermore each PROTOCOL KEY can memorize up to 7 different exposure parameters, those are automatically renewed by every examination, a max. of 400 programs can be stored in memory. Technique factors (kV, mAs or kV, mA, sec.), radiographic mode (manual, photo-timer), technique factor setting system (2-control, 3-control), focus, photo-pickup field selection and film density selection are memorized.

4

(20) Direct photo-timer exposure (option) A photo-timer exposure is available in combination with Shimadzu’s photo-pickup detector by adding the pertinent control circuit. y Manual operation or photo-timer can be selected. y SPEED (sensitivity) selector “SPEED. H”, “SPEED. M” and “SPEED. L” are selectable at installation of the system. y Film density adjustment Adjustment in 11 stages is possible. y Photo-pickup field selection Four kinds of photo-pickup fields are selectable. (21) Display of error In the occurrence of error on each function, errorcodes are displayed on the color LCD touch panel.

UD150B-40/V-40/L-40 POWER REQUIREMENTS (a)

System: 3-phase alternating current y Frequency: 50 or 60 Hz Model

UD150B-40

UD150V-40

UD150L-40

Standard voltage and allowable voltage range at no load and power source impedance

380 V ± 10%, 0.10 Ω or less 400 V ± 10%, 0.11 Ω or less 415 V ± 10%, 0.12 Ω or less 440 V ± 10%, 0.13 Ω or less 480 V ± 10%, 0.16 Ω or less

380 V ± 10%, 0.10 Ω or less 400 V ± 10%, 0.11 Ω or less 415 V ± 10%, 0.12 Ω or less 440 V ± 10%, 0.13 Ω or less 480 V ± 10%, 0.16 Ω or less

380 V ± 10%, 0.16 Ω or less 400 V ± 10%, 0.17 Ω or less 415 V ± 10%, 0.19 Ω or less 440 V ± 10%, 0.21 Ω or less 480 V ± 10%, 0.21 Ω or less

Switch at Type hand Capacity

(b)

y Relations between the length and sectional area of lead-in conductor with respect to the transformer capacity: Length

75 A or less

75 A or less

10 m or less

20 m

30 m

40 m

50 m

60 m

70 m

80 m

90 m

100 m

UD150B-40 75 kVA (3-phase)

5.5

8

14

22

22

22

38

38

38

38

UD150V-40 75 kVA (3-phase)

5.5

8

14

22

22

22

38

38

38

38

UD150L-40 50 kVA (3-phase)

5.5

5.5

5.5

5.5

8

8

8

14

14

14

Transformer Capacity

Molded case circuit breaker 50 A or less

3-phase alternating current (Optional auto transformer is necessary.) y Frequency: 50 or 60 Hz Model

UD150B-40

Standard voltage and allowable voltage range at no load Power source impedance

Switch at hand

(c)

Wire Size (Nominal Sectional Area) mm2

UD150V-40

y Relation between the length and sectional area of lead-in conductor with respect to the transformer capacity (in case of conduit):

UD150L-40

Length

200 V ± 10% 220 V ± 10% 240 V ± 10% 0.054 Ω or less

0.054 Ω or less

Type

Knife switch and fuse or molded case circuit breaker

Capacity

100 A or less

100 A or less

10 m or less

20 m

30 m

40 m

50 m

60 m

70 m

80 m

90 m

100 m

UD150B-40 75 kVA (3-phase)

14

22

38

38

60

60

60

100

100

100

UD150V-40 75 kVA (3-phase)

14

22

38

38

60

60

60

100

100

100

UD150L-40 50 kVA (3-phase)

8

14

14

22

38

38

38

60

60

60

Transformer Capacity

0.087 Ω or less

Wire Size (Nominal Sectional Area) mm2

100 A or less

Single-phase alternating current (UD150L-40 only) y Frequency: 50 or 60 Hz y Standard voltage and allowable voltage range at no load: 200 V +10, -5% 220 V ± 10% 240 V ± 10% y Power source impedance: 0.051 Ω

y Relations between the length and sectional area of lead-in conductor with respect to the transformer capacity (in case of conduit): Wire Size (Nominal Sectional Area) mm2

Length Transformer Capacity

50 kVA (single-phase)

10 m or less

20 m

30 m

40 m

50 m

60 m

70 m

80 m

90 m

100 m

14

22

38

60

60

100

100

100

100

100

y Line switch Type: Knife switch or circuit breaker Capacity: 100 A or more y Rating of fuse or circuit breaker: 60 A or less

AMBIENT CONDITION When the ambient conditions of each room do not satisfy the following conditions, it is suggested to install an exclusive air conditioner. The calorie generated by the apparatus in each room is as follows.

Temperature 5 to 35°C

Humidity 45 to 85%

Caution No dew condensation allowed Generated Calorie

Apparatus used

Model

Control console Control cabinet X-ray high voltage generator

UD150B-40

UD150V-40

UD150L-40

Approx. 50 kcal/h Approx. 680 kcal/h

Approx. 680 kcal/h

Approx. 600 kcal/h

y The above value is at stand-by state. y Generated calorie for 1 kW ∗ 1 kW = 860 kcal/h

5

UD150B-40/V-40/L-40 CONTROL CONSOLE PANEL VIEW

Operation Panel

Home screen display switch

Color TFT touch panel

Radiography history display switch

Hand switch

Register program display switch

Power ON and OFF switch

AEC ON /OFF switch (optional)

Radiography ready switch

Density UP and DOWN switch (optional)

X-ray exposure switch

6

[kV],[mA/mAs],[sec.] display

Radiography ready indicator

[kV],[mA/mAs],[sec.] setting shuttle and switch

X-ray exposure indicator

Focus select switch

Radiography stand-by indicator

[mA]/[mAs] select switch

Caution indicator

X-ray tube selected indicator

Fault indicator

UD150B-40/V-40/L-40 DISPLAY OF THE COLOR TFT TOUCH PANEL



Radiography programs tab key Radiography technique key Photo pickup field select key Body thickness correction key Speed (film sensitivity) key (optional) Heat unit gauge Radiographic programs key Automatic stepping of process item key Radiographic process key

7

UD150B-40/V-40/L-40

IBS Key IBS(Image Brightness Stabilizer)of ON/OFF Press the [IBS] to make the key light up. Fluoroscopic tube voltage down key. Fluoroscopic tube voltage setting. Press key to decrease the displayed value. Fluoroscopic tube voltage up key. Fluoroscopic tube voltage setting. Press key to increase the displayed value. Pulse rate key Pulse rate setting NOTE) Pulsed fluoroscopy can be selected if the DR unit incorporates a pulsed fluoroscopy function. Fluoroscopic mode key. Fluoroscopic mode setting. Preset key to alter the modes. Preset fluoroscopy conditions (fluoroscopy mode) and preset conditions are shown below. "PULSE N" is optional. Fluoroscopic mode

Preset conditions

Remarks

DOSE N

Standard fluoroscopy (continuous fluoroscopy)

-

DOSE L

Fluoroscopy at low X-ray dosage (continuous fluoroscopy)

-

PULSE N

Pulsed fluoroscopy

Digital image processing system Used.

Fluoroscopic Timer key. Setting the Fluoroscopic Timer Hold down the [Fluoroscopic timer] key for three seconds when in other to reset the timer to zero.

8

UD150B-40/V-40/L-40 DIMENSIONS AND WEIGHT Control console 308

345

82

Front view

Side view Unit: mm Weight: 2.5 kg

Control cabinet

Top view

Side view

Front view

Unit: mm Weight: D150BC-40 (for UD150B-40) 250 kg D150VC-40 (for UD150V-40) 250 kg D150LC-40 (for UD150L-40) 240 kg

Remarks y Every value in this Product Data Sheet is a standard value, and it may vary a little from the actual at each site. y The appearances and specifications are subject to change for reasons of improvement without notice.

SHIMADZU CORPORATION. International Marketing Division 3. Kanda-Nishikicho 1-chome, Chiyoda-ku, Tokyo 101-8448, Japan Phone: 81(3)3219-5641

URL http://www.shimadzu.com

Fax: 81(3)3219-5710

Shimadzu Corporation Medical Systems Group has been certified by TUV Rheinland as a manufacturer of medical equipment and systems in compliance with ISO9001: 2000 Quality Management Systems and EN ISO13485: 2003 Medical Equipment Quality Management Systems.

9