TECHNICAL SPECIFICATIONS

IQnfc+ Near Field Communication Test System TM

© 2016 LitePoint, A Teradyne Company. All rights reserved.

Introduction Near field communication (NFC) is a set of short-range wireless technologies that communicates through magnetic field induction between electronic devices in close proximity. As NFC offers safe and simple communications between electronic devices, many NFC applications, such as mobile payment and short peer-to-peer transactions, are fast gaining in popularity. It is important to ensure NFC-enabled products meet all measures of quality and provide a consistent and exceptional user experience. IQnfc+ is a wireless test solution for NFC-enabled devices that focuses on characterizing the NFC physical layer and ensuring product quality. IQnfc+ supports measurements of NFC Forum analog test cases, as well as EMVCo PICC/PCD L1 test cases.

Simple and Flexible Tests for NFC devices IQnfc+ consists of two hardware components: the test instrument and test coils. The test instrument contains the main NFC test functional blocks, including analyzer and NFC transceiver. It communicates with an external computer for data transfer and signal processing. The test coils serve as the communication interface with the Device Under Test (DUT). We support all 8 test coils (Poller/PCD and Listener/PICC) that are defined in the NFC Forum and EMVCo.

Sense

Matching circuit

Level TX

USB MON1 MON2 MONH

Listener/PICC Test Coil VNA (Reserved for future)

Matching circuit

Poller/PCD Test Coil

Sense TX

TX MOD TX RF

IQnfc+ Test Instrument

Test Fixture

PC: Tester control and data processing Orange Path: signal output Green Path: signal input

Figure 1. Example test configuration

Near Field Communication Test System 2

IQnfc+ Test Instrument Specifications Analyzer Parameter

Ports

Value

MON1, MON2

2 Vpp, +/-1 V Max.

MONH

30 Vpp, +/-15 V Max.

MON1, MON2

+/-5% (50 mVpp to 2 Vpp, +/- 25 mV to +/-1 V)

MONH

+/-5% (700 mVpp to 30 Vpp, +/- 350 mV to +/-15 V)

Frequency range

MON1, MON2, MONH

12 to 20 MHz

Frequency accuracy

MON1, MON2, MONH

< +/-8 ppm (< +/-100 Hz @ 13.56 MHz)

Quantization

MON1, MON2, MONH

14 bits

Spurious (non-harmonic)

MON1, MON2, MONH

< -60 dBc

Input return loss

MON1, MON2

> 15 dB

MON1, MON2

50 ohms

MONH

1 Mohm || 15 pF

Ports

Value

TXRF

0.2 to 20 Vpp (re: 50 ohms)

TXMOD

Pulse Amplitude: 0 to 5.5 V (re: 50 ohms) Pulse Offset: -2 to +3 V (re: 50 ohms)

Output power P1 dB

TXRF

> 30 dBm

Output voltage accuracy

TXRF, TXMOD

+/- 10% with regard to set level

TXRF

0.25 dB

TXMOD

1 mV

Output return loss

TXRF

> 23.1 dB

Output impedance

TXRF, TXMOD

50 ohms

Frequency range

TXRF

13.56 +/- 0.01 MHz

Frequency accuracy

TXRF, TXMOD

< +/-8 ppm (< +/-100 Hz @ 13.56 MHz)

Frequency resolution

TXRF

10Hz

Spurious

TXRF

< -60 dBc (13.56 +/-2 MHz)

Max input voltage

Input voltage accuracy

Input impedance

Transmitter Parameter

Output voltage range

Output voltage resolution

Near Field Communication Test System 3

NFC Measurement Specifications Parameter

Ports

Value

Standard

ISO14443A/B, ISO18092, EMVco, Felica, NFC A/B/F/P2P

Operation mode

Target and Initiator emulation

Data rate

Support various NFC data rates analysis

106, 212, 424 and 848 kbps

Field strength

Measure carrier field strength (DUT as initiator)

Relative measurement, DUT and Position dependent. Resonator & Q selection dependent.

Frequency accuracy

Measure DUT frequency error (DUT as initiator)

< ±100 Hz

Target frame delay time

Measure the DUT response time (DUT as target)

Tester contribution < ±20 ns

Modulation depth/index/ timing profile

Measure waveform shape (DUT as initiator and target)

This is a relative measurement: dependent on DUT, position, resonator and Q selection.

Overshoot

Overshoot following the rising and falling edges

< ± 0.5 %

Rise and Fall time

Measure the rise and fall time as defined in the standards

< ± 2/fc seconds

Rx sensitivity

Min detectable field strength (DUT as initiator and target)

This is a relative measurement: dependent on DUT, position, resonator and Q selection.

Payload data

Provide results of received payload data (DUT as initiator and target)

CRC error

Provide results of CRC error (DUT as initiator and target)

Near Field Communication Test System 4

IQnfc+ Test Instrument Port Description Front Panel

I/O

Function

Type

MON1

Listener/PICC signal input

BNC female

MON2

Poller/PCD signal input

BNC female

MONH

Listener/PICC level input (DC+RF)

BNC female

VNA

Resonant sweep output

BNC female

TX MOD

Listener/PICC signal output

BNC female

TXRF

Poller/PCD signal output

BNC female

TEST HEAD

Digital control and power for control of test head

Proprietary

Status Indicator

Blinking Green: application loading Green: No fault detected, ready to use Orange: software fault Red: power on without established USB connection Off: Power off

LED

I/O

Function

Type

Reference input

13.56 MHz clock reference

BNC female

Marker out / trigger in 1

TTL compatible

BNC female

Marker out / trigger in 2

TTL compatible

BNC female

USB

USB 2.0 compatible connection to external controller

USB Type B

AC in

AC power input

100 to 240 VAC (automatically switched) 50 to 60 Hz Includes hard power switch

Rear Panel

Near Field Communication Test System 5

General and Environmental Dimensions

14.782” L X 16.707” W X 2.093” H

Weight

7.9 pounds

Power requirements

90-260 VAC, 47-63 Hz;