CLINICIAN’S CORNER

CLINICAL REVIEW

Implantable Cardioverter-Defibrillators Expanding Indications and Technologies Zachary Goldberger, MD Rachel Lampert, MD

S

UDDEN CARDIAC DEATH (SCD) causes approximately 450 000 deaths annually in the United States and nearly 50% of all cardiovascular mortality worldwide.1-3 Ventricular tachycardia degenerating into ventricular fibrillation causes two thirds of SCD.4-6 Because there are no warning symptoms to identify persons who might potentially experience SCD,4 successful therapy has focused on identifying high-risk patients and treating them with implantable cardioverter-defibrillators (ICDs), which continuously monitor the heart rhythm and deliver a shock or other appropriate electrical therapy on detection of a potentially fatal sustained ventricular arrhythmia.7 This review will highlight the evolving indications for the ICD and describe technological advances in defibrillator therapy, emphasizing the major clinical trials demonstrating the efficacy of the device in primary and secondary prophylaxis of SCD.

EVIDENCE ACQUISITION We conducted a literature search using the Pubmed and MEDLINE databases from January 1996 to July 2005, using the Medical Subject Heading implantable defibrillator, limited to randomized controlled trials (RCTs) reported in English. Abstracts were reviewed to identify those that included an ICD and a non-ICD group and that had primary outcomes of allCME available online at www.jama.com

Context Sudden cardiac death (SCD) is a major challenge facing contemporary cardiology. For an increasing number of patients, the current standard of care for the treatment and prevention of SCD is the implantable cardioverter-defibrillator (ICD). Since its introduction, there have been numerous advances in ICD technology, and indications for its use have expanded greatly in the past year. Objective To highlight the evolving indications for and the numerous advances in ICD technology, with emphasis on primary and secondary prophylaxis of SCD. Evidence Acquisition Electronic literature search of the Pubmed and MEDLINE databases from January 1996 to July 2005, using the Medical Subject Heading implantable defibrillator. Abstracts and titles were reviewed to identify English-language randomized controlled trials that included an ICD group and a non-ICD group and that had end points of all-cause mortality, cardiac death, and/or arrhythmic mortality as the main outcome. A further MEDLINE search was conducted to identify randomized controlled trials of cardiac resynchronization therapy (CRT) with a CRT and a nonCRT group (including both mortality and other end points). Other studies were included that clarify aspects of device function and other relevant issues. A total of 22 trials were identified. Evidence Synthesis ICD implantation improves survival in patients with a history of life-threatening ventricular arrhythmia. More recent evidence shows that ICD implantation also improves survival as primary prophylaxis against SCD in patients at high risk for ventricular arrhythmias, including those with left ventricular ejection fraction (LVEF) of 35% or less and New York Heart Association class II or III heart failure and those with a history of myocardial infarction and LVEF of 30% or less. Cardiac resynchronization improves symptoms, quality of life, and survival for patients with advanced heart failure and intraventricular conduction delays and ventricular dyssynchrony. Conclusions ICDs have been shown to improve survival as both primary and secondary prophylaxis in an expanding population of patients. Ongoing ICD research may continue to delineate groups with survival benefit from ICDs, and the use and indications of these devices in clinical practice will continue to expand. www.jama.com

JAMA. 2006;295:809-818

cause mortality, cardiac death, and/or arrhythmic mortality. We performed a similar search to identify RCTs of cardiac resynchronization therapy (CRT) including a CRT and a nonCRT group (including both mortality and other end points). A total of 22 trials were identified. Scientific statements and guidelines from official societies (eg, the American College of Cardiology [ACC], American Heart Association [AHA], and the Heart Rhythm Society) were also reviewed.

©2006 American Medical Association. All rights reserved.

We included other studies that clarify aspects of device function and other pertinent issues. Author Affiliations: Department of Medicine, University of Washington Medical Center, Seattle (Dr Goldberger) and Department of Medicine, Yale University School of Medicine, New Haven, Conn (Dr Lampert). Corresponding Author: Rachel Lampert, MD, Section of Cardiology, Yale University School of Medicine, 333 Cedar St, FMP 3, New Haven, CT 06520 ([email protected]). Clinical Review Section Editor: Michael S. Lauer, MD. We encourage authors to submit papers for consideration as a Clinical Review. Please contact Michael S. Lauer, MD, at [email protected].

(Reprinted) JAMA, February 15, 2006—Vol 295, No. 7 809

Downloaded From: http://jama.jamanetwork.com/pdfaccess.ashx?url=/data/journals/jama/5012/ on 01/26/2017

IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS

EVIDENCE SYNTHESIS Expanding Indications

The introduction of the ICD into clinical practice has been an evolving process. While secondary prophylaxis for patients surviving a life-threatening ventricular arrhythmia has long been the standard of care, more recent trials also demonstrate the survival benefit of the ICD in expanding groups of patients at high risk (ie, primary prophylaxis). Secondary Prevention Trials

Several trials have investigated the role of the ICD in secondary prevention of SCD (TABLE 1).8-12 The largest include the Antiarrhythmics vs Implantable Defibrillators (AVID) trial,8 the Cardiac Arrest Study Hamburg (CASH),9 and the Canadian Implantable Defibrillator Study (CIDS).10 Each trial randomly assigned patients to receive ICD vs pharmacological therapy, and in each, the ICD reduced total mortality, although only the AVID results reached statistical significance. AVID was also the largest and bestdesigned trial, including nearly exclu-

sive use of transvenous defibrillators and comparing the ICD against the best available antiarrhythmic drug therapy (amiodarone and sotalol), unlike CASH and CIDS. Further, a meta-analysis of the 3 trials confirmed that ICD therapy resulted in significant relative reductions in total mortality (27%) and arrhythmic mortality (51%).13 Use of the ICD improved survival regardless of underlying structural heart disease, ␤-blockade, surgical revascularization, or presenting arrhythmia (ventricular tachycardia/fibrillation). Patients with the lowest left ventricular ejection fraction (LVEF) and those with advanced congestive heart failure (CHF) benefited most.13 Interestingly, the AVID registry, enrolling patients not qualifying for randomization, showed that some groups previously considered at lower risk for SCD— ie, patients with hemodynamically stable ventricular tachycardia or arrhythmias attributed to “reversible causes”— actually had significantly worse survival than patients randomly assigned to receive ICD, suggesting that these groups may also benefit from ICDs.14

Primary Prevention Trials

The dismal survival rate after cardiac arrest15-17 provides strong impetus to identify high-risk patients who might benefit from an ICD before a first lifethreatening arrhythmia. Historically, patients with left ventricular dysfunction after myocardial infarction (MI) and a history of nonsustained ventricular tachycardia underwent electrophysiologic testing to identify those believed to be at highest risk, ie, those with with inducible, nonsuppressible, ventricular tachyarrhythmias. 18,19 This population was the target for the first primary prevention trials (TABLE 2). The Multicenter Automatic Defibrillator Trial (MADIT)20 compared ICDs with conventional therapy (mainly amiodarone, LVEF ⱕ35%), and the Multicenter Unsustained Tachycardia Trial (MUSTT)22 compared electrophysiologically guided therapy (ICDs or drug therapy) with no guided therapy (LVEF ⱕ40%). MADIT was prematurely aborted after enrolling only 196 patients when preliminary analysis revealed a dra-

Table 1. Secondary Prevention of Sudden Cardiac Death Study AVID,8 1997

Randomization Antiarrhythmic medications (97% amiodarone, 3% sotalol) vs ICD

CASH,9 2000

Antiarrhythmic medications propafenone (withdrawn early), metoprolol, or amiodarone vs ICD Amiodarone vs ICD

CIDS,10 2000

DEBUT,11 2003 ␤-Blocker vs ICD

MAVERIC,12 2004

Electrophysiologically guided therapy (antiarrhythmic, revascularization, or ICD) vs amiodarone

Mean Follow-up, mo

No.

Population

1016

Survived VT/VF/cardiac arrest; VT with syncope; VT with LVEF ⱕ40%

18

288

Survived VT/VF/cardiac arrest

57

659

Survived VT/VF/cardiac arrest; VT with syncope; VT with LVEF ⱕ35% and cycle length ⱕ400 ms

35

86

Survived VT/VF/cardiac arrest; no structural abnormalities

36

214

Survived VT/VF/cardiac arrest

60‡

Main Findings Reduction in total mortality with ICD therapy (HR, 0.66; 95% CI, 0.51-0.85; P⬍.02) NNT = 9 at 3 y Reduction in total mortality with ICD therapy (HR, 0.82; 95% CI, 0.60-1.11; P = .08)* Reduction in death from any cause with ICD therapy (P = .14) Reduction in risk of death from arrhythmia with ICD therapy (P = .09) HR, 0.85; 95% CI, 0.67-1.10* Total of 7 deaths, all of which occurred in the ␤-blocker group (3 deaths during 2-y follow-up, P = .07; 4 deaths during 3-y follow-up, P = .02)† Lower mortality with ICD therapy vs non-ICD therapy (HR, 0.54; 95% CI, 0.39-0.97; P = .04)§ No advantage to electrophysiologic testing

Abbreviations: AVID, Antiarrhythmics vs Implantable Defibrillators; CASH, Cardiac Arrest Study Hamburg; CI, confidence interval; CIDS, Canadian Implantable Defibrillator Study; DEBUT, Defibrillator vs ␤-Blockers for Unexplained Death in Thailand; HR, hazard ratio; ICD, implantable cardioverter-defibrillator; LVEF, left ventricular ejection fraction; MAVERIC, Midlands Trial of Empirical Amiodarone vs Electrophysiologic-Guided Intervention and Cardioverter Implant in Ventricular Arrhythmias; NNT, number needed to treat; VF, ventricular fibrillation; VT, ventricular tachycardia. *NNT not calculated because of nonsignificant results. †HR and CI not reported. ‡Median follow-up. §Data not available for calculation of NNT.

810

JAMA, February 15, 2006—Vol 295, No. 7 (Reprinted)

©2006 American Medical Association. All rights reserved.

Downloaded From: http://jama.jamanetwork.com/pdfaccess.ashx?url=/data/journals/jama/5012/ on 01/26/2017

IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS

matic benefit of ICD therapy in reducing overall mortality by 54% (P=.009). MADIT had no control group, raising the question of whether the trial proved benefit of ICD or detriment of amiodarone. In addition, use of ␤-blockers was higher in the ICD group. However, MUSTT, while not designed to evaluate the ICD, supported the MADIT findings. The original hypothesis of MUSTT was that electrophysiologically guided therapy, either pharmacological or device-based, could reduce arrhythmic and total mortality in highrisk patients who had arrhythmias induced at electrophysiologic study. In patients randomized to receive electrophysiologically guided therapy, antiarrhythmic drugs were tested first, and, at the physician’s discretion, nonresponders received ICDs. MUSTT did show a decrease in arrhythmic death/ cardiac arrest with electrophysiologically guided therapy, the investigators’ primary hypothesis. However, subgroup analysis revealed that the ben-

P=.005), it was significantly higher than that for the inducible, ICD-treated patients (24%). These data implied that noninducible patients with left ventricular dysfunction and nonsustained ventricular tachycardia may also benefit from a prophylactic ICD, and that electrophysiologic study may be an inadequate stratifier of risk. The second MADIT trial (MADIT II)23 directly addressed the value of prophylactic ICD implantation in patients with coronary artery disease and LVEF of 30% or less, without electrophysiologic risk stratification. Patients receiving the ICD showed a 31% relative reduction in mortality at any interval (P=.02). Of note, among 593 patients in the ICD group who underwent peri-implant electrophysiologic testing (not an entry criterion), inducibility did not predict later ventricular arrhythmia, supporting a low sensitivity for electrophysiologic testing. One concern following MADIT II was the higher rate of CHF in the ICD

efits were due entirely to the ICD: at 5 years, there were absolute reductions in total mortality of 31% when compared with those receiving pharmacological therapy and of 24% when compared with those receiving no therapy (24% mortality in the ICD group, 55% in the pharmacological therapy group, and 48% in the group receiving no therapy). In MUSTT, few patients received amiodarone, and ICD use was not randomized. However, taken together, MUSTT and MADIT clearly demonstrate the benefit of the ICD in the population of patients with coronary artery disease, low LVEF, and inducible ventricular arrhythmia. The MUSTT registry 26 also followed up patients who had clinical criteria for the trial but no inducible arrhythmias. Surprisingly, while 5-year mortality for these patients was statistically lower than for inducible patients randomly assigned to receive no therapy (44% and 48%, respectively;

Table 2. Primary Prevention of Sudden Cardiac Death in Ischemic Cardiomyopathy Mean Follow-up, mo

Study MADIT,20 1996

Randomization Antiarrhythmic therapy (74% amiodarone) vs ICD

No.

Population

196

27

Reduction in total mortality with ICD therapy (HR, 0.46; 95% CI, 0.26-0.92; P = .009) NNT = 4 at 5 y

CABG-Patch,21 1997

CABG surgery plus ICD vs CABG surgery plus conventional therapy Electrophysiologically guided therapy (antiarrhythmic or ICD) vs conventional therapy Conventional therapy vs ICD

900

Prior MI; LVEF ⱕ35%; asymptomatic NSVT; NYHA class I-III; inducible VT refractory to intravenous procainamide on electrophysiologic study Patients scheduled for CABG; LVEF ⱕ35%; positive SAECG result

32

No reduction in total mortality with ICD therapy (HR, 1.07; 95% CI, 0.81-1.42; P = .64)†

Prior MI; LVEF ⱕ40%; CAD; NSVT; inducible VT on electrophysiologic study

39*

Prior MI; LVEF ⱕ30%

20

674

Recent MI (within 4-40 d), LVEF ⱕ35%; impaired cardiac autonomic modulation (heart rate variability)

39

2521

NYHA class II/III CHF (ischemic and nonischemic); LVEF ⱕ35%

45.5*

Reduction in total mortality with ICD therapy (HR, 0.69; 95% CI, 0.32-0.63; P⬍.001) NNT = 3 at 5 y Reduction in total mortality with ICD therapy (HR, 0.69; 95% CI, 0.51-0.93; P = .02) NNT = 18 over mean 20 mo No reduction in death from any cause with ICD therapy (P = .66) Risk of arrhythmic death lower with ICD therapy (P = .009) HR, 1.08; 95% CI, 0.76-1.55† Overall: reduction in mortality with ICD therapy (P = .007) Ischemic heart disease: reduction in mortality with ICD therapy (P = .05) HR, 0.77; 97.5% CI, 0.62-0.96 NNT = 14 at 5 y

MUSTT,22 1999

MADIT II,23 2002

DINAMIT,24 2004 Conventional therapy vs ICD

SCD-HeFT,25 2005

Conventional therapy vs amiodarone vs ICD

704

1232

Main Findings

Abbreviations: CABG, coronary artery bypass graft; CAD, coronary artery disease; CHF, congestive heart failure; CI, confidence interval; DINAMIT, Defibrillator in Acute Myocardial Infarction Trial; HR, hazard ratio; ICD, implantable cardioverter-defibrillator; LVEF, left ventricular ejection fraction; MADIT, Multicenter Automatic Defibrillator Trial; MI, myocardial infarction; MUSTT, Multicenter Unsustained Tachycardia Trial; NSVT, nonsustained ventricular tachycardia; NYHA, New York Heart Association; NNT, number needed to treat; SAECG, signal-averaged electrocardiogram; SCD-HeFT, Sudden Cardiac Death in Heart Failure Trial; VT, ventricular tachycardia. *Median follow-up. †NNT not appropriate, as ICD implantation resulted in nonsignificantly higher mortality.

©2006 American Medical Association. All rights reserved.

(Reprinted) JAMA, February 15, 2006—Vol 295, No. 7 811

Downloaded From: http://jama.jamanetwork.com/pdfaccess.ashx?url=/data/journals/jama/5012/ on 01/26/2017

IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS

group.23,27 The later Dual Chamber and VVI Implantable Defibrillator (DAVID) trial,28 which randomly assigned patients to receive either dual- or singlechamber ICDs, showed higher incidence of CHF with right ventricular pacing. The mechanism by which right ventricular pacing effectively causes iatrogenic left bundle-branch block and subsequent contractile dyssynchrony likely explains the higher incidence of CHF in ICD-treated patients in MADIT II. This concern was not appreciated at that time. Based on these and other data showing deleterious effects of right ventricular pacing,29,30 devices are now programmed to minimize such pacing.31,32 The recently published Sudden Cardiac Death in Heart Failure Trial (SCDHeFT)25 enrolled patients with either ischemic or nonischemic cardiomyopathy, New York Heart Association (NYHA) class II or III heart failure, and LVEF less than or equal to 35%. The results confirmed the benefit of ICD in ischemic patients as found in MADIT II, as well as the findings of a previous smaller study of patients with nonischemic cardiomyopathy (TABLE 3),24,25,33-35 the Prophylactic Defibrillator Implantation in Patients with Nonischemic Di-

lated Cardiomyopathy (DEFINITE) trial.35 In SCD-HeFT, ICD-treated patients lived longer than those treated with amiodarone (which had no benefit) or conventional medical therapy. While the concern has been raised that ICD benefit in MADIT II may have been skewed by the short follow-up,36 SCDHeFT showed ICD benefit extending to 5 years, independent of heart failure etiology (ischemic vs nonischemic). Subgroup analysis revealed benefit in patients with NYHA class II, but not class III, heart failure. These studies suggest that most patients with LVEF less than or equal to 35% should receive an ICD (specifically those with NYHA class II or III heart failure, or class I with history of MI and LVEF ⱕ30%). However, whether LVEF should now be the primary factor determining ICD eligibility remains controversial.37 In MUSTT, LVEF had poor specificity in predicting SCD38; in other studies, combinations of factors were more predictive.39,40 Neither MADIT II nor SCDHeFT evaluated ICD benefit in patients known to be noninducible. This may explain why the absolute reductions in allcause mortality for MADIT II and SCDHeFT (6% and 7%, respectively) are

much smaller than for MADIT or MUSTT (23% and 31%, respectively), which included electrophysiologic criteria. There are correspondingly larger numbers needed to treat for SCDHeFT/MADIT II (15-17) than for MUSTT/MADIT (3-4). Thus, the benefit of ICD therapy may be greater in patients with SCD risk beyond low LVEF (although these differences in absolute risk reduction may also be due to better medical therapy in the control groups in the later trials41). Editorialists have questioned the validity of ICD implantation in all post-MI patients with reduced LVEF,37,42,43 suggesting that further risk stratification is still needed.37 The potential role for noninvasive riskstratifiers such as T-wave alternans, shown to have good predictive value,44,45 remains undetermined. Two studies showed no benefit of ICD for primary prophylaxis in specific populations: the Coronary Artery Bypass Graft Patch (CABG-Patch) trial21 and the Defibrillator in Acute Myocardial Infarction (DINAMIT) trial (Table 2). 24 CABG-Patch randomly assigned patients with an abnormal signal-averaged electrocardiogram and scheduled for coronary

Table 3. Primary Prevention of Sudden Cardiac Death in Nonischemic Dilated Cardiomyopathy Study CAT,33 2002

Randomization Conventional therapy vs ICD

No. 104

AMIOVIRT,34 2003

Amiodarone vs ICD

103

DEFINITE,35 2004

Conventional therapy vs ICD

458

SCD-HeFT,25 2005

Conventional therapy vs amiodarone vs ICD

2521

Population NYHA class II/III, NIDCM; LVEF ⱕ30%; asymptomatic NSVT NYHA class I-III, NIDCM; LVEF ⱕ35%; asymptomatic NSVT NIDCM; LVEF ⬍36%; NSVT or PVCs

NYHA class II/III CHF (ischemic and nonischemic); LVEF ⱕ35%

Mean Follow-up, mo

Main Findings

66

No reduction in total mortality with ICD therapy (P = .55)*

36

No reduction in total mortality with ICD therapy (P = .80)*

29

Reduction in total mortality with ICD therapy (P = .08) Reduction in death from arrhythmia with ICD therapy (P = .006) HR, 0.65; 95% CI, 0.40-1.06† Overall: reduction in mortality with ICD therapy (P = .007) Nonischemic heart disease: reduction in mortality with ICD therapy (P = .06) HR, 0.77; 97.5% CI, 0.62-0.96 NNT = 14 at 5 y

45.5‡

Abbreviations: AMIOVIRT, Amiodarone vs Implantable Defibrillator in Patients with Nonischemic Cardiomyopathy and Asymptomatic Nonsustained Ventricular Tachycardia; CAT, Cardiomyopathy Trial; CHF, congestive heart failure; CI, confidence interval; DEFINITE, Prophylactic Defibrillator Implantation in Patients with Nonischemic Dilated Cardiomyopathy; HR, hazard ratio; ICD, implantable cardioverter-defibrillator; LVEF, left ventricular ejection fraction; NIDCM, nonischemic dilated cardiomyopathy; NNT, number needed to treat; NSVT, nonsustained ventricular tachycardia; NYHA, New York Heart Association; PVC, premature ventricular contraction; SCD-HeFT, Sudden Cardiac Death in Heart Failure Trial. *HR and CI not reported. †NNT not calculated because of nonsignificant results. ‡Median follow-up.

812

JAMA, February 15, 2006—Vol 295, No. 7 (Reprinted)

©2006 American Medical Association. All rights reserved.

Downloaded From: http://jama.jamanetwork.com/pdfaccess.ashx?url=/data/journals/jama/5012/ on 01/26/2017

IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS

artery bypass graft surgery to receive treatment with either an ICD implanted during surgery or no ICD. The ICD showed no benefit, likely due to the lower-risk profile of the patient population. First, while preoperative ejection fractions were low, they may have improved with surgery. Also, the signal-averaged electrocardiogram may lack specificity.18,19 Revascularization itself may have protected against arrhythmia, although in AVID, the ICD offered similar survival rates independent of revascularization.8 A recent substudy of the large Valsartan in Acute Myocardial Infarction Trial (VALIANT)46 demonstrated that SCD risk is highest in the first 30 days after MI (complicated by reduced LVEF and/or CHF), suggesting that early ICD implantation might save lives. However, the DINAMIT study, which randomly assigned such patients to receive an ICD or conventional medical therapy, showed no ICD benefit early after MI, possibly due to the low event rate in this small study. It is also possible that the VALIANT patients with SCD were sicker than the survivors, with competing risks. DINAMIT supports this theory, as patients who received shocks were also more likely to die. It is unknown whether further risk stratification, or noninvasive measures such a wearable external vest or an automatic external defibrillator, might be more beneficial and costeffective early after MI.47 Overall, these trials strongly support ICD implantation for secondary prevention in patients with prior lifethreatening arrhythmia, and as primary prophylaxis for many patients with a low LVEF and with coronary artery disease, CHF, or both. The ACC/ AHA/North American Society for Pacing and Electrophysiology (now the Heart Rhythm Society) last revised their guidelines for ICD implantation in 2002,48 and, as such, these guidelines do not yet reflect the most recent primary prophylaxis trials. However, the 2005 ACC/AHA Guidelines for the Diagnosis and Management of Chronic Heart Failure in the Adult49 recom-

Box. Centers for Medicare & Medicaid Services (CMS) Coverage Requirements for Implantable Cardioverter-Defibrillator Implantation* Patients with IDCM, documented prior MI, NYHA class II and III heart failure, and measured LVEF ⬍35% Patients with NIDCM ⬎9 months, NYHA class II and III heart failure, and measured LVEF ⬍35% Patients who meet all current CMS coverage requirements for a CRT device and have NYHA class IV heart failure Patients with NIDCM ⬎3 mo, NYHA class II or III heart failure, and measured LVEF ⬍35%† Abbreviations: CABG, coronary artery bypass graft; CRT, cardiac resynchronization therapy; IDCM, ischemic dilated cardiomyopathy; LVEF, left ventricular ejection fraction; MI, myocardial infarction; NIDCM, nonischemic dilated cardiomyopathy; NYHA, New York Heart Association; PTCA, percutaneous transluminal coronary angioplasty. *For all groups, patients must not have (1) cardiogenic shock or symptomatic hypotension while in a stable baseline rhythm; (2) had CABG surgery or PTCA within the past 3 months; (3) had an acute MI within the past 40 days; (4) clinical symptoms or findings that would make them a candidate for coronary revascularization; (5) irreversible brain damage from preexisting cerebral disease; or (6) any disease, other than cardiac disease, associated with a likelihood of survival less than 1 year. From Centers for Medicare & Medicaid Services.50 †The CMS has determined that the evidence, though less compelling at this time, is adequate to conclude that an ICD is reasonable and necessary for this subset of patients.

mend that “consideration of ICD implantation is recommended in patients with LVEF less than 30% and mild to moderate symptoms of heart failure and in whom survival with good functional capacity is otherwise anticipated to extend beyond 1 year.” In addition, the Centers for Medicare & Medicaid Services has expanded coverage for ICDs primarily based on the results of MADIT II and SCD-HeFT (BOX).50 The ICD is also used to prevent SCD in other patients at high risk, such as those with cardiac ion-channel abnormalities or structural heart diseases creating an arrhythmogenic substrate. Although prospective randomized trials in these relatively rare conditions are not likely to be pursued,51 case series show efficacy of the ICD in patients with long-QT syndromes,52 Brugada syndrome,53 hypertrophic cardiomyopathy,54,55 and arrhythmogenic right ventricular dysplasia.56 Current guidelines support its use in selected patients with these disorders.48 An evaluation of the cost-effectiveness of the ICD using a Markov model57 revealed that prophylactic ICD implan-

©2006 American Medical Association. All rights reserved.

tation was both more effective and more expensive than control therapy. Consistent with the lower numbers needed to treat, the populations showing greatest cost-effectiveness of the ICD were those in MADIT I and MUSTT, with approximately $25 000 per life-year added. The ICD was less cost-effective in the MADIT II and SCD-HeFT populations, with $40 000 to $50 000 per lifeyear added, respectively. (Other data from SCD-HeFT showed a costeffectiveness of $33 000.57) Costs per quality-adjusted life-year ranged from $34 000 to $70 000. Not surprisingly, lowering the cost of the device or increasing longevity would improve costeffectiveness.57 These costs are well within the range considered acceptable to society.58 Expanding Technologies

Implantable cardioverter-defibrillators were first developed by Dr Michel Mirowski, whose invention was prompted by the sudden death of a colleague. Ironically, at the time of its development, Mirowski emphasized that the ICD was “not a definitive treatment for recurrent malignant arrhyth-

(Reprinted) JAMA, February 15, 2006—Vol 295, No. 7 813

Downloaded From: http://jama.jamanetwork.com/pdfaccess.ashx?url=/data/journals/jama/5012/ on 01/26/2017

IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS

mia and should not be regarded as a substitute for antiarrhythmic medications or other modes of therapy.”7 The 3 main functions of the ICD are detection of arrhythmia; delivery of appropriate electrical therapy; and storage of diagnostic information, including electrograms and details of treated episodes. In addition, ICDs provide antibradycardia pacing, either single- or dual-chamber. The device consists of 2 components, the pulse generator and the lead electrode system. Current ICDs are only slightly larger than a pacemaker (25-45 cm3) and are similarly implanted in a subcutaneous pectoral pocket.59 Leads are inserted transvenously through the subclavian, axillary, or cephalic vein into the right ventricular apex60 (and the right atrial appendage for dual-chamber pacing). The modern ICD combines highenergy defibrillation with 2 other electrical therapies, low-energy cardioversion and antitachycardia pacing, to terminate ventricular arrhythmias. Antitachycardia pacing terminates ventricular tachycardia without delivering a painful shock by pacing at a rate faster than the intrinsic tachycardia, entering and interrupting the re-entrant circuit.61,62 Tiered therapy allows programming of these electrical options to treat tachycardias with rates within defined zones of detection. 63 For example, a slower arrhythmia might be treated by antitachycardia pacing followed by low-energy shock and then high-energy shock, if needed. Ventricular fibrillation falls in a faster zone, prompting high-energy defibrillation. While antitachycardia pacing was initially used only to treat slower, stable ventricular tachycardia, the recent Pacing Reduces Shocks for Fast VT II (PainFREE Rx II) trial64 found that empirically programmed antitachycardia pacing delivered during device charging (a function available in some devices) could treat faster ventricular tachycardias, reducing shocks by 70% with no adverse outcomes. Dual-Chamber ICDs. While early ICDs could pace the ventricle, the deleterious effects of single-chamber pac814

ing (ie, pacemaker syndrome, development of atrial tachyarrhythmias65-69), together with the problems in detection and treatment of ventricular arrhythmias caused by implanting 2 devices, led to the incorporation of dual-chamber pacing capabilities.70-73 Dual-chamber ICDs offer the hemodynamic benefits of dual-chamber pacing in patients with bradyarrhythmias, with recent devices incorporating algorithms that provide atrial-based pacing while minimizing right ventricular pacing.31,32,74 In most, although not all, studies, the dual-chamber device offers improved discrimination between ventricular and supraventricular arrhythmias, decreasing inappropriate shocks due to rapid supraventricular rhythms or physiologic sinus tachycardia. This is achieved using specific algorithms,75-77 such as those that analyze the relative timing of atrial and ventricular electrograms, or by inhibition of ICD discharge when the sensed atrial rate is faster than the ventricular rate, as is the case in atrial fibrillation.76 Further, for patients receiving ICDs, universal dualchamber device implantation may be the least-expensive strategy.78 Cardiac Resynchronization Therapy. Abnormal chamber mechanics— that is, intraventricular conduction delays resulting in dyssynchronous contraction—contribute significantly to CHF.79,80 Left bundle-branch block, whether intrinsic or due to right ventricular pacing, impairs effective contraction by causing septal contraction before activation of the lateral wall of the left ventricle, creating dyssynchrony.81 This abnormality causes a reduced cardiac output due to shortened diastolic filling time, a result of the lengthening of isovolumetric contraction and relaxation times. In addition, delayed activation of the left ventricular lateral wall may lead to a contraction after aortic valve closure, reducing ventricular filling during diastole.82 These factors represent the likely mechanism for the increased mortality with dual-chamber pacing seen in the DAVID trial, as well as the in-

JAMA, February 15, 2006—Vol 295, No. 7 (Reprinted)

creased CHF seen in patients receiving the ICD in MADIT II.23,28 Cardiac resynchronization therapy, also known as biventricular pacing, can improve CHF in patients with left bundle-branch block through placement of a third lead through the coronary sinus into a venous branch along the free wall of the left ventricle (or epicardially). Pacing the left ventricle promotes earlier contraction, restoring physiologic synchrony of the ventricles.80 CRT improves not only cardiac output and ejection fraction83,84 but symptoms as well, including improvements in NYHA class, 6-minute walking distance, and quality of life, along with lower hospitalization rates (TABLE 4).56,85-90 Further, CRT improves survival.91,92 The recent randomized Cardiac Resynchronization–Heart Failure (CAREHF) trial92 evaluated the effects of CRT alone (pacing only) on morbidity and mortality in patients with NYHA class III or IV heart failure due to left ventricular systolic dysfunction and cardiac dyssynchrony. CRT significantly decreased mortality, with 82 deaths compared with 120 in the group receiving medical therapy (P⬍.002), as well as improved quality of life. However, 32% of the deaths in the CRT group were sudden, suggesting that a combination of CRT and defibrillation capabilities might improve survival. The recently published Comparison of Medical Therapy, Pacing, and D e f i b r i l l a t i o n i n H e a rt F a i l u re (COMPANION) trial 9 1 randomly assigned patients with CHF (NYHA class III or IV with LVEF ⬍35%) and conduction delays (QRS intervals ⬎120 ms) to receive conventional therapy alone, CRT alone, or CRT incorporated in an ICD. While patients in the CRT-alone group had a 19% risk reduction in the primary end point of death or hospitalization (P=.01), patients in the CRT plus ICD group had a significant improvement in total mortality, a relative reduction of 36%. A limitation of COMPANION was its lack of power to directly compare CRT with defibrillation vs CRT without defibrillation.

©2006 American Medical Association. All rights reserved.

Downloaded From: http://jama.jamanetwork.com/pdfaccess.ashx?url=/data/journals/jama/5012/ on 01/26/2017

IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS

Further, in COMPANION, device benefit began immediately, while in CAREHF, no treatment benefit was seen until 1 year, likely because the reverse remodeling with CRT is not immediate. While the issue of CRT alone vs CRT plus ICD remains controversial, these data, in combination with those from SCD-HeFT, showing minimal benefit of a single-lead device in NYHA class III CHF, suggests that the combination of CRT plus ICD may provide the most benefit for patients with severe CHF. In response to this growing body of evidence, the recently revised ACC/AHA CHF Guidelines for

the Diagnosis and Management of Chronic Heart Failure in the Adult49 state that “there is strong evidence to support the use of CRT to improve symptoms, exercise capacity, quality of life, LVEF, and survival and to decrease hospitalizations in patients with persistently symptomatic heart failure undergoing optimal medical therapy who have cardiac dyssynchrony.” In addition to QRS duration, methods such as tissue Doppler imaging may further delineate patients with dyssynchrony most likely to benefit from CRT. 8 2 Three early case reports describe ventricular tachycardia result-

ing from CRT,93-95 attributed to epicardial left ventricular pacing. However, a recent analysis of 2 of the large CRT trials (totaling 880 CRT-treated patients) showed no increase in arrhythmia frequency with CRT.96 The ongoing Multicenter Automatic Defibrillator Implantation Trial with Cardiac Resynchronization Therapy trial (MADIT-CRT) will determine whether, in patients receiving a prophylactic ICD who have a widened QRS interval but less than NYHA class III heart failure, CRT in combination with the ICD will slow the progression of CHF to class III or IV.97

Table 4. Cardiac Resynchronization Therapy Study MUSTIC,85 2001

Randomization Conventional therapy vs CRT

No. 67

MIRACLE,86 2002*

Conventional therapy vs CRT

453

PATH-CHF,87 2002

Crossover: biventricular vs optimal univentricular stimulation followed by opposite treatment Conventional therapy vs CRT/ICD

CONTAK-CD,88 2003*

42

490

MIRACLE-ICD,89 ICD vs CRT/ICD 2003*

369

PATH-CHF II,90 2003

101

COMPANION,91 2004*

Conventional therapy vs CRT, stratified by QRS interval into long-QRS (⬎150 ms) and short-QRS (120 to 150 ms) groups Conventional therapy vs CRT vs CRT/ICD

CARE-HF,92 2005

Conventional therapy vs CRT

1520

813

Mean Population Follow-up, mo Main Findings NYHA class III; QRS interval 3 CRT resulted in improvement in 6-min walking ⬎150 ms; LVEF ⬍35%; distance (P⬍.001) sinus rhythm NYHA class III/IV; QRS 6 CRT/ICD resulted in improved quality-of-life interval ⱖ130 ms; LVEF scores (P = .001), improved NYHA class ⱕ35%; LVEDD ⱖ55 (P⬍.001), and 6-min walking distance mm; 6-min walking (P = .005) distance ⱕ450 m NYHA class III/IV; QRS 12 No significant difference in peak oxygen or 6-min interval ⱖ120 ms walking distance between biventricular and optimized univentricular pacing (left ventricular in 36 patients) NYHA class II-IV; QRS 6 CRT improved peak oxygen consumption interval ⱖ120 ms; (P = .03) and 6-min walking distance (P = .04) LVEF ⱕ35% Changes in NYHA class (P = .10) and quality-of-life scores (P = .40) were not statistically significant NYHA class III/IV; QRS 6 CRT/ICD improved quality-of-life scores (P = .02), interval ⱖ130 ms; NYHA class (P = .007), peak oxygen LVEF ⱕ35% consumption (P = .04), and exercise function on a treadmill (P⬍.001) NYHA class II-IV; QRS 3 Long-QRS group improved peak oxygen interval ⬎120 ms; consumption (P⬍.001), 6-min walking LVEF ⱕ30% distance (P = .02), and quality-of-life scores (P = .004) Short-QRS group did not improve in any end point with active pacing NYHA class III/IV; LVEF 17 Reduction in total mortality with CRT alone ⬍35%; QRS interval (P = .06) ⱖ120 ms; Reduction in mortality with CRT/ICD (P = .003) hospitalization for CHF HR, 0.64; 95% CI, 0.48-0.86 within 12 mo NNT = 11 over median 15-16 mo NYHA class III/IV; LVEF 29 Reduction in all-cause mortality with CRT vs ⱕ35%; LVEDD ⱕ30 conventional therapy (P⬍.002) mm; QRS interval ⱖ120 CRT reduced the interventricular mechanical ms; if QRS interval delay, end-systolic volume index, and area of 120-149 ms, additional the mitral regurgitant jet; increased LVEF; and criteria for dyssynchrony improved symptoms and quality-of-life scores (P⬍.01) HR, 0.64; 95% CI, 0.48-0.85 NNT = 12 at 2 y

Abbreviations: CARE-HF, Cardiac Resynchronization—Heart Failure; CHF, congestive heart failure; CI, confidence interval; COMPANION, Comparison of Medical Therapy, Pacing, and Defibrillation in Heart Failure; CRT, cardiac resynchronization therapy; HR, hazard ratio; ICD, implantable cardioverter-defibrillator; LVEDD, left ventricular end diastolic dimension; LVEF, left ventricular ejection fraction; MIRACLE, Multicenter InSync Randomized Clinical Evaluation; MUSTIC, Multisite Stimulation in Cardiomyopathy; NNT, number needed to treat; NYHA, New York Heart Association; PATH-CHF, Pacing Therapies for Congestive Heart Failure. *CRT/ICD indicates CRT incorporated into an ICD.

©2006 American Medical Association. All rights reserved.

(Reprinted) JAMA, February 15, 2006—Vol 295, No. 7 815

Downloaded From: http://jama.jamanetwork.com/pdfaccess.ashx?url=/data/journals/jama/5012/ on 01/26/2017

IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS Atrial Defibrillators

Because pharmacological treatment for atrial fibrillation is frequently ineffective98-100 and carries potentially life-threatening proarrhythmic effects,101-104 attention has focused on nonpharmacological therapies, including the atrial ICD, which recognizes and delivers a shock for atrial fibrillation. The overall efficacy in terminating atrial fibrillation is 76% to 90%, with no ventricular proarrhythmia105 and good discrimination between atrial and ventricular tachyarrhythmias.106 Rapid atrial pacing can terminate atrial tachycardias (often the precursor to atrial fibrillation) painlessly,107 but efficacy is low.106 Because atrial shocks are painful, devices programmed to be triggered by the patient may therefore be underused.108 Only a very small percentage of all ICDs implanted (2%) are atrial defibrillators (2003 industry data courtesy of Medtronic; C. Gennaro, BSME, MBA, written communication, March 2004). Device Malfunction

With expanding device technology has come an increase in malfunctions resulting in advisories and recalls, and with increasing indications for the ICD, the number of patients affected by advisories also will increase.109 Death due to device malfunction is rare110,111 but emotionally devastating when it occurs, as recently illustrated by the death of a college student with hypertrophic cardiomyopathy whose ICD failed to deliver therapy due to a short-circuiting device.54,109,111 Since February 2005, there have been advisories involving all 3 leading device manufacturers (Guidant, Medtronic, and St Jude).112 Patients and physicians faced with an advisory must weigh the risks of malfunction, which vary with the nature of the specific advisory and the patient’s underlying arrhythmia, with the risks of replacement. The Heart Rhythm Society convened a conference in September 2005 that brought together patients, physicians, industry representatives, and the US Food and Drug Administration to address device performance issues113 and 816

is currently working on guidelines to improve surveillance, analysis, and communication of device performance. CONCLUSIONS In summary, ICDs improve survival in an expanding population of patients. In addition to providing secondary prophylaxis in survivors of life-threatening ventricular arrhythmia, ICD implantation is now first-line therapy for primary prophylaxis in most patients with marked left ventricular dysfunction, regardless of etiology. Recent technological advances include resynchronization therapy and the incorporation of atrial defibrillation in some devices. Ongoing research and refined methods of risk stratification may continue to delineate groups with survival benefit from ICDs. Author Contributions: Drs Goldberger and Lampert had full access to all of the data in the study and take responsibility for the integrity of the data and the accuracy of the data analysis. Study concept and design; analysis and interpretation of data; critical revision of the manuscript for important intellectual content: Goldberger, Lampert. Acquisition of data; drafting of the manuscript; statistical analysis: Goldberger. Study supervision: Lampert. Financial Disclosures: Dr Lampert has received research funding and honoraria from Medtronic and has participated in research sponsored by Medtronic, Guidant, and St Jude. REFERENCES 1. Myerburg RJ, Kessler KM, Castellanos A. Sudden cardiac death: epidemiology, transient risk, and intervention assessment. Ann Intern Med. 1993;119:11871197. 2. Kannel WB, Schatzkin A. Sudden death: lessons from subsets in population studies. J Am Coll Cardiol. 1985; 5(6 suppl):141B-149B. 3. Josephson M, Wellens HJ. Implantable defibrillators and sudden cardiac death. Circulation. 2004;109: 2685-2691. 4. Luu M, Stevenson WG, Stevenson LW, Baron K, Walden J. Diverse mechanisms of unexpected cardiac arrest in advanced heart failure. Circulation. 1989; 80:1675-1680. 5. Huikuri HV, Castellanos A, Myerburg RJ. Sudden death due to cardiac arrhythmias. N Engl J Med. 2001; 345:1473-1482. 6. Bayes de Luna A, Coumel P, Leclercq JF. Ambulatory sudden cardiac death: mechanisms of production of fatal arrhythmia on the basis of data from 157 cases. Am Heart J. 1989;117:151-159. 7. Mirowski M, Reid PR, Mower MM, et al. Termination of malignant ventricular arrhythmias with an implanted automatic defibrillator in human beings. N Engl J Med. 1980;303:322-324. 8. Antiarrhythmics versus Implantable Defibrillators (AVID) Investigators. A comparison of antiarrhythmicdrug therapy with implantable defibrillators in patients resuscitated from near-fatal ventricular arrhythmias. N Engl J Med. 1997;337:1576-1583. 9. Kuck KH, Cappato R, Siebels J, Ruppel R. Random-

JAMA, February 15, 2006—Vol 295, No. 7 (Reprinted)

ized comparison of antiarrhythmic drug therapy with implantable defibrillators in patients resuscitated from cardiac arrest: the Cardiac Arrest Study Hamburg (CASH). Circulation. 2000;102:748-754. 10. Connolly SJ, Gent M, Roberts RS, et al. Canadian Implantable Defibrillator Study (CIDS): a randomized trial of the implantable cardioverter defibrillator against amiodarone. Circulation. 2000;101: 1297-1302. 11. Nademanee K, Veerakul G, Mower M, et al. Defibrillator versus beta-Blockers for Unexplained death in Thailand (DEBUT): a randomized clinical trial. Circulation. 2003;107:2221-2226. 12. Lau EW, Griffith MJ, Pathmanathan RK, et al. The Midlands Trial of Empirical Amiodarone versus Electrophysiology-guided Interventions and Implantable Cardioverter-defibrillators (MAVERIC): a multicentre prospective randomised clinical trial on the secondary prevention of sudden cardiac death. Europace. 2004;6:257-266. 13. Connolly SJ, Hallstrom AP, Cappato R, et al. Metaanalysis of the implantable cardioverter defibrillator secondary prevention trials: AVID, CASH and CIDS studies. Eur Heart J. 2000;21:2071-2078. 14. Anderson JL, Hallstrom AP, Epstein AE, et al; AVID Investigators. Design and results of the Antiarrhythmics vs Implantable Defibrillators (AVID) registry. Circulation. 1999;99:1692-1699. 15. Bigger JT. Expanding indications for implantable cardiac defibrillators. N Engl J Med. 2002;346:931-933. 16. de Vreede-Swagemakers JJ, Gorgels AP, DuboisArbouw WI, et al. Out-of-hospital cardiac arrest in the 1990’s: a population-based study in the Maastricht area on incidence, characteristics and survival. J Am Coll Cardiol. 1997;30:1500-1505. 17. Weaver WD, Hill D, Fahrenbruch CE, et al. Use of the automatic external defibrillator in the management of out-of-hospital cardiac arrest. N Engl J Med. 1988;319:661-666. 18. Bhandari AK, Widerhorn J, Sager PT, et al. Prognostic significance of programmed ventricular stimulation in patients surviving complicated acute myocardial infarction: a prospective study. Am Heart J. 1992;124:87-96. 19. Wilber DJ, Olshansky B, Moran JF, Scanlon PJ. Electrophysiological testing and nonsustained ventricular tachycardia: use and limitations in patients with coronary artery disease and impaired ventricular function. Circulation. 1990;82:350-358. 20. Moss AJ, Hall WJ, Cannom DS, et al; Multicenter Automatic Defibrillator Implantation Trial Investigators. Improved survival with an implanted defibrillator in patients with coronary disease at high risk for ventricular arrhythmia. N Engl J Med. 1996;335: 1933-1940. 21. Bigger JT Jr; Coronary Artery Bypass Graft (CABG)Patch Trial Investigators. Prophylactic use of implanted cardiac defibrillators in patients at high risk for ventricular arrhythmias after coronary-artery bypass graft surgery. N Engl J Med. 1997;337:1569-1575. 22. Buxton AE, Lee KL, Fisher JD, Josephson ME, Prystowsky EN, Hafley G; Multicenter Unsustained Tachycardia Trial Investigators. A randomized study of the prevention of sudden death in patients with coronary artery disease. N Engl J Med. 1999;341:18821890. 23. Moss AJ, Zareba W, Hall WJ, et al. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med. 2002;346:877-883. 24. Hohnloser SH, Kuck KH, Dorian P, et al. Prophylactic use of an implantable cardioverter-defibrillator after acute myocardial infarction. N Engl J Med. 2004; 351:2481-2488. 25. Bardy GH, Lee KL, Mark DB, et al. Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure. N Engl J Med. 2005;352:225237.

©2006 American Medical Association. All rights reserved.

Downloaded From: http://jama.jamanetwork.com/pdfaccess.ashx?url=/data/journals/jama/5012/ on 01/26/2017

IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS 26. Buxton AE, Lee KL, DiCarlo L, et al; Multicenter Unsustained Tachycardia Trial Investigators. Electrophysiologic testing to identify patients with coronary artery disease who are at risk for sudden death. N Engl J Med. 2000;342:1937-1945. 27. Hurst TM, Hinrichs M, Breidenbach C, Katz N, Waldecker B. Detection of myocardial injury during transvenous implantation of automatic cardioverter-defibrillators. J Am Coll Cardiol. 1999;34: 402-408. 28. Wilkoff BL, Cook JR, Epstein AE, et al; Dual Chamber and VVI Implantable Defibrillator Trial Investigators. Dual-chamber pacing or ventricular backup pacing in patients with an implantable defibrillator: the Dual Chamber and VVI Implantable Defibrillator (DAVID) trial. JAMA. 2002;288:3115-3123. 29. Wonisch M, Lercher P, Scherr D, et al. Influence of permanent right ventricular pacing on cardiorespiratory exercise parameters in chronic heart failure patients with implanted cardioverter defibrillators. Chest. 2005;127:787-793. 30. Sweeney MO, Hellkamp AS, Ellenbogan KA, et al. Adverse effect of ventricular pacing on heart failure and atrial fibrillation among patients with normal baseline QRS duration in a clinical trial of pacemaker therapy for sinus node dysfunction. Circulation. 2003; 107:2932-2937. 31. Sweeney MO, Ellenbogan KA, Betzold R, et al. Multicenter, prospective, randomized trial of a new atrial-based managed ventricular pacing mode (MVP) in dual chamber ICDs [abstract]. Circulation. 2004;110: 444. 32. Sweeney MO, Shea JB, Fox V, et al. Randomized pilot study of a new atrial-based minimal ventricular pacing mode in dual chamber implantable cardioverter-defibrillators. Heart Rhythm. 2004;1:160167. 33. Bansch D, Antz M, Boczor S, et al. Primary prevention of sudden cardiac death in idiopathic dilated cardiomyopathy: the Cardiomyopathy Trial (CAT). Circulation. 2002;105:1453-1458. 34. Strickberger SA, Hummel JD, Bartlett TG, et al; AMIOVIRT Investigators. Amiodarone versus implantable cardioverter-defibrillator: randomized trial in patients with nonischemic dilated cardiomyopathy and asymptomatic nonsustained ventricular tachycardia—AMIOVIRT. J Am Coll Cardiol. 2003;41: 1707-1712. 35. Kadish A, Dyer A, Daubert JP, et al. Prophylactic defibrillator implantation in patients with nonischemic dilated cardiomyopathy. N Engl J Med. 2004;350: 2151-2158. 36. Coats AJ. MADIT II, the Multi-center Automatic Defibrillator Implantation Trial II stopped early for mortality reduction, has ICD therapy earned its evidencebased credentials? Int J Cardiol. 2002;82:1-5. 37. Buxton AE. Should everyone with an ejection fraction less than or equal to 30% receive an implantable cardioverter-defibrillator? not everyone with an ejection fraction ⬍ or=30% should receive an implantable cardioverter-defibrillator. Circulation. 2005;111: 2537-2549. 38. Buxton AE, Lee KL, Hafley GE, et al. Relation of ejection fraction and inducible ventricular tachycardia to mode of death in patients with coronary artery disease: an analysis of patients enrolled in the Multicenter Unsustained Tachycardia Trial. Circulation. 2002;106:2466-2472. 39. La Rovere MT, Pinna GD, Hohnloser SH, et al. Baroreflex sensitivity and heart rate variability in the identification of patients at risk for life-threatening arrhythmias: implications for clinical trials. Circulation. 2001;103:2072-2077. 40. Bailey JJ, Berson AS, Handelsman H, Hodges M. Utility of current risk stratification tests for predicting major arrhythmic events after myocardial infarction. J Am Coll Cardiol. 2001;38:1902-1911. 41. Kadish A. Prophylactic defibrillator implantation—

toward an evidence-based approach. N Engl J Med. 2005;352:285-286. 42. Buxton AE. The clinical use of implantable cardioverter defibrillators: where are we now? where should we go? Ann Intern Med. 2003;138:512514. 43. Reynolds MR, Josephson ME. MADIT II (second Multicenter Automated Defibrillator Implantation Trial) debate: risk stratification, costs, and public policy. Circulation. 2003;108:1779-1783. 44. Bloomfield DM, Steinman RC, Namerow PB, et al. Microvolt T-wave alternans distinguishes between patients likely and patients not likely to benefit from implanted cardiac defibrillator therapy: a solution to the Multicenter Automatic Defibrillator Implantation Trial (MADIT) II conundrum. Circulation. 2004;110:18851889. 45. Gold MR, Bloomfield DM, Anderson KP, et al. A comparison of T-wave alternans, signal averaged electrocardiography and programmed ventricular stimulation for arrhythmia risk stratification. J Am Coll Cardiol. 2000;36:2247-2253. 46. Solomon SD, Zelenkofske S, McMurray JJ, et al. Sudden death in patients with myocardial infarction and left ventricular dysfunction, heart failure, or both. N Engl J Med. 2005;352:2581-2588. 47. Buxton AE. Sudden death after myocardial infarction—who needs prophylaxis, and when? N Engl J Med. 2005;352:2638-2640. 48. Gregoratos G, Abrams J, Epstein AE, et al. ACC/ AHA/NASPE 2002 guideline update for implantation of cardiac pacemakers and antiarrhythmia devices: summary article: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (ACC/AHA/NASPE Committee to Update the 1998 Pacemaker Guidelines). J Cardiovasc Electrophysiol. 2002;13:1183-1199. 49. American College of Cardiology. Clinical Statements/Guidelines. Available at: http://www.acc .org/clinical/topic/topic.htm. Accessed December 19, 2005. 50. Centers for Medicare & Medicaid Services Web site. Available at: https://www.cms.hhs.gov. Accessed December 19, 2005. 51. DiMarco JP. Implantable cardioverter-defibrillators. N Engl J Med. 2003;349:1836-1847. 52. Zareba W, Moss AJ, Daubert JP, Hall WJ, Robinson JL, Andrews M. Implantable cardioverter defibrillator in high-risk long QT syndrome patients. J Cardiovasc Electrophysiol. 2003;14:337-341. 53. Watanabe H, Chinushi M, Sugiura H, et al. Unsuccessful internal defibrillation in Brugada syndrome: focus on refractoriness and ventricular fibrillation cycle length. J Cardiovasc Electrophysiol. 2005; 16:262-266. 54. Maron BJ, Estes NA III, Maron MS, Almquist AK, Link MS, Udelson JE. Primary prevention of sudden death as a novel treatment strategy in hypertrophic cardiomyopathy. Circulation. 2003;107:2872-2875. 55. Begley DA, Mohiddin SA, Tripodi D, Winkler JB, Fananapazir L. Efficacy of implantable cardioverter defibrillator therapy for primary and secondary prevention of sudden cardiac death in hypertrophic cardiomyopathy. Pacing Clin Electrophysiol. 2003;26: 1887-1896. 56. Corrado D, Leoni L, Link MS, et al. Implantable cardioverter-defibrillator therapy for prevention of sudden death in patients with arrhythmogenic right ventricular cardiomyopathy/dysplasia. Circulation. 2003; 108:3084-3091. 57. Sanders GD, Hlatky MA, Owens DK. Costeffectiveness of implantable cardioverter-defibrillators. N Engl J Med. 2005;353:1471-1480. 58. Goldman L. Cost-effectiveness in a flat world— can ICDs help the United States get rhythm? N Engl J Med. 2005;353:1513-1515. 59. Kusumoto FM, Goldschlager N. Device therapy for cardiac arrhythmias. JAMA. 2002;287:1848-1852.

©2006 American Medical Association. All rights reserved.

60. Yee R, Klein GJ, Leitch JW, et al. A permanent transvenous lead system for an implantable pacemaker cardioverter-defibrillator: nonthoracotomy approach to implantation. Circulation. 1992;85:196-204. 61. Saksena S, Krol RB, Kaushik RR. Innovations in pulse generators and lead systems: balancing complexity with clinical benefit and long-term results. Am Heart J. 1994;127:1010-1021. 62. Estes NA III, Haugh CJ, Wang PJ, Manolis AS. Antitachycardia pacing and low-energy cardioversion for ventricular tachycardia termination: a clinical perspective. Am Heart J. 1994;127:1038-1046. 63. Gollob MH, Seger JJ. Current status of the implantable cardioverter-defibrillator. Chest. 2001;119: 1210-1221. 64. Wathen MS, DeGroot PJ, Sweeney MO, et al. Prospective randomized multicenter trial of empirical antitachycardia pacing versus shocks for spontaneous rapid ventricular tachycardia in patients with implantable cardioverter-defibrillators: Pacing Fast Ventricular Tachycardia Reduces Shock Therapies (PainFREE Rx II) trial results. Circulation. 2004;110:2591-2596. 65. Zanini R, Facchinetti AI, Gallo G, Cazzamalli L, Bonandi L, Dei Cas L. Morbidity and mortality of patients with sinus node disease: comparative effects of atrial and ventricular pacing. Pacing Clin Electrophysiol. 1990;13:2076-2079. 66. Santini M, Ansalone G, Auriti A, Magris B, Pandozi C, Altamura G. Indications for dual-chamber (DDD) pacing in implantable cardioverterdefibrillator patients. Am J Cardiol. 1996;78: 116-118. 67. Markewitz A, Schad N, Hemmer W, Bernheim C, Ciavolella M, Weinhold C. What is the most appropriate stimulation mode in patients with sinus node dysfunction? Pacing Clin Electrophysiol. 1986;9:11151120. 68. Hesselson AB, Parsonnet V, Bernstein AD, Bonavita GJ. Deleterious effects of long-term singlechamber ventricular pacing in patients with sick sinus syndrome: the hidden benefits of dual-chamber pacing. J Am Coll Cardiol. 1992;19:1542-1549. 69. Feuer JM, Shandling AH, Messenger JC. Influence of cardiac pacing mode on the long-term development of atrial fibrillation. Am J Cardiol. 1989; 64:1376-1379. 70. Blanck Z, Niazi I, Axtell K, et al. Feasibility of concomitant implantation of permanent transvenous pacemaker and defibrillator systems. Am J Cardiol. 1994; 74:1249-1253. 71. Lampert R, McPherson CA, Lewis RJ, Batsford WP, Rosenfeld LE. Inappropriate sensing of atrial stimuli in patients with third-generation defibrillators and DDD pacemakers. Pacing Clin Electrophysiol. 1998;21:12251229. 72. Pfeiffer D, Jung W, Fehske W, et al. Complications of pacemaker-defibrillator devices: diagnosis and management. Am Heart J. 1994;127:1073-1080. 73. Mattke S, Markewitz A, Muller D, et al. The combined transvenous implantation of cardioverter defibrillators and permanent pacemakers. Pacing Clin Electrophysiol. 1997;20:2775-2782. 74. Nielsen JC, Kristensen L, Andersen HR, Mortensen PT, Pedersen OL, Pedersen AK. A randomized comparison of atrial and dual-chamber pacing in 177 consecutive patients with sick sinus syndrome: echocardiographic and clinical outcome. J Am Coll Cardiol. 2003;42:614-623. 75. Stadler RW, Gunderson BD, Gillberg JM. An adaptive interval-based algorithm for withholding ICD therapy during sinus tachycardia. Pacing Clin Electrophysiol. 2003;26:1189-1201. 76. Israel CW, Gronefeld G, Iscolo N, Stoppler C, Hohnloser SH. Discrimination between ventricular and supraventricular tachycardia by dual chamber cardioverter defibrillators: importance of the atrial sensing function. Pacing Clin Electrophysiol. 2001;24: 183-190.

(Reprinted) JAMA, February 15, 2006—Vol 295, No. 7 817

Downloaded From: http://jama.jamanetwork.com/pdfaccess.ashx?url=/data/journals/jama/5012/ on 01/26/2017

IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS 77. Barold HS, Newby KH, Tomassoni G, Kearney M, Brandon J, Natale A. Prospective evaluation of new and old criteria to discriminate between supraventricular and ventricular tachycardia in implantable defibrillators. Pacing Clin Electrophysiol. 1998;21: 1347-1355. 78. Goldberger Z, Elbel B, McPherson CA, Paltiel AD, Lampert R. Cost advantage of dual-chamber versus single-chamber cardioverter-defibrillator implantation. J Am Coll Cardiol. 2005;46:850-857. 79. Rosenqvist M, Isaaz K, Botvinick EH, et al. Relative importance of activation sequence compared to atrioventricular synchrony in left ventricular function. Am J Cardiol. 1991;67:148-156. 80. Leclercq C, Kass DA. Retiming the failing heart: principles and current clinical status of cardiac resynchronization. J Am Coll Cardiol. 2002;39:194201. 81. Likoff MJ, Chandler SL, Kay HR. Clinical determinants of mortality in chronic congestive heart failure secondary to idiopathic dilated or to ischemic cardiomyopathy. Am J Cardiol. 1987;59:634638. 82. Ansalone G, Giannantoni P, Ricci R, et al. Doppler myocardial imaging in patients with heart failure receiving biventricular pacing treatment. Am Heart J. 2001;142:881-896. 83. Cazeau S, Ritter P, Lazarus A, et al. Multisite pacing for end-stage heart failure: early experience. Pacing Clin Electrophysiol. 1996;19:17481757. 84. Blanc JJ, Etienne Y, Gilard M, et al. Evaluation of different ventricular pacing sites in patients with severe heart failure: results of an acute hemodynamic study. Circulation. 1997;96:32733277. 85. Cazeau S, Leclercq C, Lavergne T, et al. Effects of multisite biventricular pacing in patients with heart failure and intraventricular conduction delay. N Engl J Med. 2001;344:873-880. 86. Abraham WT, Fisher WG, Smith AL, et al. Cardiac resynchronization in chronic heart failure. N Engl J Med. 2002;346:1845-1853. 87. Auricchio A, Stellbrink C, Sack S, et al. Longterm clinical effect of hemodynamically optimized cardiac resynchronization therapy in patients with heart failure and ventricular conduction delay. J Am Coll Cardiol. 2002;39:2026-2033. 88. Higgins SL, Hummel JD, Niazi IK, et al. Cardiac resynchronization therapy for the treatment of heart failure in patients with intraventricular conduction delay and malignant ventricular tachyarrhythmias. J Am Coll Cardiol. 2003;42:14541459.

818

89. Young JB, Abraham WT, Smith AL, et al. Combined cardiac resynchronization and implantable cardioversion defibrillation in advanced chronic heart failure: the MIRACLE ICD Trial. JAMA. 2003;289:26852694. 90. Auricchio A, Stellbrink C, Butter C, et al. Clinical efficacy of cardiac resynchronization therapy using left ventricular pacing in heart failure patients stratified by severity of ventricular conduction delay. J Am Coll Cardiol. 2003;42:2109-2116. 91. Bristow MR, Saxon LA, Boehmer J, et al. Cardiacresynchronization therapy with or without an implantable defibrillator in advanced chronic heart failure. N Engl J Med. 2004;350:2140-2150. 92. Cleland JG, Daubert JC, Erdmann E, et al. The effect of cardiac resynchronization on morbidity and mortality in heart failure. N Engl J Med. 2005;352:15391549. 93. Guerra JM, Wu J, Miller JM, Groh WJ. Increase in ventricular tachycardia frequency after biventricular implantable cardioverter defibrillator upgrade. J Cardiovasc Electrophysiol. 2003;14:12451247. 94. Mykytsey A, Maheshwari P, Dhar G, et al. Ventricular tachycardia induced by biventricular pacing in patient with severe ischemic cardiomyopathy. J Cardiovasc Electrophysiol. 2005;16:655-658. 95. Rivero-Ayerza M, Vanderheyden M, Verstreken S, de Zutter M, Geelen P, Brugada P. Polymorphic ventricular tachycardia induced by left ventricular pacing. Circulation. 2004;109:2924-2925. 96. McSwain RL, Schwartz RA, Delurgio DB, Mera FV, Langberg JJ, Leon AR. The impact of cardiac resynchronization therapy on ventricular tachycardia/ fibrillation. J Cardiovasc Electrophysiol. 2005;16:11681171. 97. ClinicalTrials.gov. MADIT-CRT: Automatic Defibrillator Implantation With Cardiac Resynchronization Therapy. Available at: http://www.clinicaltrials .gov/ct/show/NCT00180271. Accessed January 4, 2006. 98. Ganz LI, Antman EM. Antiarrhythmic drug therapy in the management of atrial fibrillation. J Cardiovasc Electrophysiol. 1997;8:1175-1189. 99. Nattel S, Hadjis T, Talajic M. The treatment of atrial fibrillation: an evaluation of drug therapy, electrical modalities and therapeutic considerations. Drugs. 1994; 48:345-371. 100. Sharif MN, Wyse DG. Atrial fibrillation: overview of therapeutic trials. Can J Cardiol. 1998;14:12411254. 101. Coplen SE, Antman EM, Berlin JA, Hewitt P, Chalmers TC. Efficacy and safety of quinidine therapy for maintenance of sinus rhythm after cardioversion:

JAMA, February 15, 2006—Vol 295, No. 7 (Reprinted)

a meta-analysis of randomized control trials. Circulation. 1990;82:1106-1116. 102. Flaker GC, Blackshear JL, McBride R, Kronmal RA, Halperin JL, Hart RG; Stroke Prevention in Atrial Fibrillation Investigators. Antiarrhythmic drug therapy and cardiac mortality in atrial fibrillation. J Am Coll Cardiol. 1992;20:527-532. 103. Middlekauff HR, Stevenson WG, Gornbein JA. Antiarrhythmic prophylaxis vs warfarin anticoagulation to prevent thromboembolic events among patients with atrial fibrillation: a decision analysis. Arch Intern Med. 1995;155:913-920. 104. Prystowsky EN. Proarrhythmia during drug treatment of supraventricular tachycardia: paradoxical risk of sinus rhythm for sudden death. Am J Cardiol. 1996; 78:35-41. 105. Daoud EG, Timmermans C, Fellows C, et al; Metrix Investigators. Initial clinical experience with ambulatory use of an implantable atrial defibrillator for conversion of atrial fibrillation. Circulation. 2000;102: 1407-1413. 106. Schoels W, Swerdlow CD, Jung W, Stein KM, Seidl K, Haffajee CJ. Worldwide clinical experience with a new dual-chamber implantable cardioverter defibrillator system. J Cardiovasc Electrophysiol. 2001;12: 521-528. 107. Friedman PA, Dijkman B, Warman EN, et al. Atrial therapies reduce atrial arrhythmia burden in defibrillator patients. Circulation. 2001;104:1023-1028. 108. Cooper JM, Katcher MS, Orlov MV. Implantable devices for the treatment of atrial fibrillation. N Engl J Med. 2002;346:2062-2068. 109. Maisel WH, Sweeney MO, Stevenson WG, Ellison KE, Epstein LM. Recalls and safety alerts involving pacemakers and implantable cardioverterdefibrillator generators. JAMA. 2001;286:793-799. 110. Hauser RG, Kallinen L. Deaths associated with implantable cardioverter defibrillator failure and deactivation reported in the United States Food and Drug Administration Manufacturer and User Facility Device Experience Database. Heart Rhythm. 2004;1:399405. 111. Hauser RG, Maron BJ. Lessons from the failure and recall of an implantable cardioverter-defibrillator. Circulation. 2005;112:2040-2042. 112. Steinbrook R. The controversy over Guidant’s implantable defibrillators. N Engl J Med. 2005;353:221224. 113. Heart Rhythm Society; US Food and Drug Administration. Proceedings document from the policy conference on pacemaker and ICD performance, September 16, 2005. Available at: http://www.hrsonline .org/advocacyDocs/HRS-device_conference.pdf. Accessibility verified January 4, 2006.

©2006 American Medical Association. All rights reserved.