imc C-SERIES: CS-4108-N, CL-4124-N

CS-4108-N, CL-4124-N Technical Data Sheet 17.07.2013 imc C-SERIES: CS-4108-N, CL-4124-N Compact measurement systems with isolated inputs Data Sheet ...
Author: Alicia Welch
4 downloads 0 Views 1MB Size
CS-4108-N, CL-4124-N Technical Data Sheet

17.07.2013

imc C-SERIES: CS-4108-N, CL-4124-N Compact measurement systems with isolated inputs Data Sheet Version 1.0

device type: CS, 8 analog channels

device type: CL, 24 analog channels

The devices of the imc C-SERIES CS-4108-N and CL-4124-N are 8- and 24-channel all-purpose measurement systems, respectively, with sampling rates of up to 50 kHz per channel. They are specially designed for measurement tasks in environments with unclear voltage fields such as test rigs or largescale machinery. The input channels are electrically isolated, differential and equipped with per-channel signal conditioning including filters. imc C-SERIE - complete, compact and portable measurement devices The devices of the imc C-SERIES are equipped with a defined standard configuration and available in two different housing types: "CS" in a Alu-Profile housing and "CL" in a dark flat plastic housing with carrying leafs. Highlights Stand-alone startup and power-failure control logic Real-time signal processing and test control with imc Online FAMOS (as standard equipment) intelligent power supply with UPS function and saving of data when power-failure integrated CAN-Interface Counter inputs (measurement of angle, events, velocity etc.) digital inputs and outputs analog outputs (DAC) Onboard flash storage (CF card) or network-harddrive (NAS etc.) complex triggering system PC independent possible equipment with internal WiFi (WLAN) adaptor supports platform independent remote access via standard interner browser (optionally integrated imc REMOTE Webserver) adapted to synchronization with other imc measurement devices via: isolated Sync-Signal (DCF-77, IRIG-B) network based via NTP GPS Measurement channel extension via direct connection of measurement modules belonging to imc's CANSAS series In conjunction with the operating software imc STUDIO and im c DEVICES the devices are immediately ready to take measurements, and all of their functions are operable.

Errors and changes excepted For more information, contact: Berlin: +49 - 30 - 46 70 90 - 0

1

CS-4108-N, CL-4124-N Technical Data Sheet

17.07.2013

Overview of the available devices Order Code

article number

CS-4108-N

1400073

CS-4108-N-ET

1410037

CL-4124-N

1400074

CL-4124-N-ET

1410038

housing

analog channels

CS alu profile

8

extra extended temp.- range

CL portable housing

24

extended temp.- range

Mechanical drawings with dimensions Housing type: CS (95x111x185 mm)

Housing type: CL (270x85x300 mm)

Errors and changes excepted For more information, contact: Berlin: +49 - 30 - 46 70 90 - 0

2

CS-4108-N, CL-4124-N Technical Data Sheet

17.07.2013

Required software version Supported by imc STUDIO 4.0R1 and imc DEVICES 2.8R3, or higher, included with all functions Supported by imc STUDIO 3.0R4 and imc DEVICES 2.7 R3, with slightly limited functions (in particular without Web-Server). Experiments (configurations) that are configured with older devices or predecessor devices and also with software versions downwards to imc DEVICES 2.6 are completely transferable (assignable) imc Plug & Measure (TEDS) This device supports TEDS and can be used together with imc SENSORS. Included accessories 230/110 V power adapter (optionally with country-specific power cable) Installation-DVD of the imc operating software with manual (German and English) "Getting Started with imc C-SERIES" (printed) Manufacturer's Calibration Certificate 1x Ethernet network cable crossed and 1x uncrossed 1x LEMO.1B plug with CS devices and with CL devices 1x LEMO.0B Terminal connections 2x or 6x ACC/DSUB(M)-T4

1x ACC/DSUB(M)-DI4-8 1x ACC/DSUB(M)-DO8 1x ACC/DSUB(M)-ENC4 1x ACC/DSUB(M)-DAC4

DSUB-15 plug with screw terminals for 4-channel measurement of voltages as well as temperatures with PT100 and thermocouples with integrated cold junction compensation (CJC). DSUB-15 plug with screw terminals for 8 digital inputs DSUB-15 plug with screw terminals for 8 digitale Ausgänge DSUB-15 plug with screw terminals for 4 counter inputs DSUB-15 plug with screw terminals for 4 analog outputs

Optional accessories Terminal connections ACC/DSUB(M)-U4 ACC/DSUB(M)-I4 ACC/DSUB-ICP4

DSUB-15 plug with screw terminals for 4-channel voltage measurement DSUB-15 plug with screw terminals for 4-channel current measurement of up to 50 mA (50 shunt, scaling factor: 0.02A/V) DSUB-15 plug with screw terminals for conditioning of 4 IEPE/ICP inputs

Errors and changes excepted For more information, contact: Berlin: +49 - 30 - 46 70 90 - 0

3

CS-4108-N, CL-4124-N Technical Data Sheet

17.07.2013

General technical specs for all devices of imc C-SERIES Parameter Housing type Ingress Protection

Value

Remarks

Alu profile

CS

plastic portable housing

CL

IP20

Terminal connection Terminal connection (DSUB-15)

ACC/DSUB(M)-T4

analog inputs

ACC/DSUB(M)-U4 ACC/DSUB(M)-I4 ACC/DSUB-ICP4

Terminal connection (DSUB-15)

ACC/DSUB(M)-DI4-8

8 digital inputs

DI, DO, INC, DAC

ACC/DSUB(M)-DO8

8 digital outputs

ACC/DSUB(M)-ENC4

4 counter inputs

ACC/DSUB(M)-DAC4

4 analog outputs

Further terminal connections

RJ45 CF-Card slot 2x DSUB-9

Ethernet (10/100 MBit), PC/network removable storage two CAN-nodes

DSUB-9

external Display (CS)

DSUB-9

external GPS module

BNC

synchronization

LEMO FGG.1B.302.CLAD62Z

supply (CS)

LEMO FGG.0B.302.CLAD62Z

supply (CL)

Weight without table-top power adapter

approx. 2 kg

CS

approx. 3.5 kg

CL

Dimensions (WxHxD) in mm

95 x 111 x 185

CS

270 x 85 x 300

CL

Power supply Parameter

Value

Remarks

DC input supply voltage

10 V to 32 V DC

Isolation of supply input

not-isolated

CS

isolated

CL

Power adapter Auto start upon power up Automatic shutdown with data saving upon power fail UPS UPS buffer time constant

110 V / 230 V AC configurable

automatic start of measurement

yes battery: lead-gel 1 sec (with CS) 30 sec. (with CL)

Internal battery voltage

external adapter, included in delivery

uninterruptable power supply maximum duration of a continuous outage before triggering device shutdown

4V

CS

24 V

CL

Errors and changes excepted For more information, contact: Berlin: +49 - 30 - 46 70 90 - 0

4

CS-4108-N, CL-4124-N Technical Data Sheet

17.07.2013

Power supply Parameter

Value

Remarks

Effective buffer capacity

typ. 23°C, battery fully charged

Minimum charging time for 1 min. buffer duration Charging time for empty battery

3,5 Wh

CS

5,1 Wh

CL

19 min

for empty battery, 23°C CS

21 min.

CL

6h

device activated

Charging capacity

automatic charge control 1.1 W

CS

1.5 W

CL

Operating conditions Operating environment (standard)

dry, non corrosive environment within specified temperature range

Operating temperature (standard)

-10°C to +55°C no condensation

Operating temperature(extended)

-20°C to +85°C with condensation

Operating altitude

up to 2000 m

Relative humidity

80 % for less than 31°C, for more than 31°C linear declining to 50%, according DIN EN61010-1

Data acquisition and hardware options Max. aggregate sampling rate

400 kS/s

Sampling rate channel wise configurable in steps of 1-, 2-, 5 Number of simultaneously applicable sampling rates (in one configuration)

2

Monitor channels (doubled channels with independent sampling and trigger configurations) Multi-triggered data acquisition: multitrigger and multi-shot Independent trigger machines (start/stop, arbitrary channel assignments)

48

Extensive intelligent trigger functions Direct onboard data reduction: arithmetic mean, min, max Extensive real-time calculation and control functions

Synchronization

External GPS signal receiver

included in standard deliveries (via imc Online FAMOS) DCF 77, IRIG-B (auto detect) NTP GPS O

Internal WiFi (WLAN) adaptor

O IEEE 802.11g (1 Antenna) max. 54 MBit/s

Errors and changes excepted For more information, contact: Berlin: +49 - 30 - 46 70 90 - 0

5

CS-4108-N, CL-4124-N Technical Data Sheet

17.07.2013

Data storage internal removable storage

CF-Card (covered CF slot)

internal hard drive

O (with CL)

Any memory depth with pre- and post triggering Circular buffering

Synchronization and time base Time base per device without external synchronization Parameter

Value typ.

Accuracy internal time base RTC Drift

±20 ppm

Ageing

min. / max.

Remarks

±50 ppm

balanced (default), at 25°C

±50 ppm

-40°C to +85°C operating temperature

±10 ppm

@ 25°C; 10 years

Time base per device with external synchronization signal Parameter Supported formats

GPS

DCF77

NMEA / PPS*

IRIG-B

NTP

B002

Version 4 (downwards compatible)

B000, B001, B003** Precision

±1 µs

Jitter (max.)

±8 µs 5 V TTL level

Voltage level

TTL (PPS*)

5 ms is necessary

---

** using BCD information only

Synchronization with DCF77 for several devices (Master/Slave) Parameter

Value typ.

max. cable length

min. / max. 200 m

max. number of devices

20

Common mode

max. 50 V

Voltage level DCF input/output

Remarks BNC cable RG58 Slaves, plus 1 Master SYNC-signal is already internally isolated, for reliable operation even with different ground voltage level (ground loops)

5V "SYNC" connection

BNC

Errors and changes excepted For more information, contact: Berlin: +49 - 30 - 46 70 90 - 0

6

CS-4108-N, CL-4124-N Technical Data Sheet

17.07.2013

Cx-41xx analog inputs Technical Specs Parameter

Value

Analog inputs Measurement modes

Remarks

8

CS

24

CL

voltage thermocouple, RTD (Pt100) current current fed sensors

Terminal connection (DSUB-15)

ACC/DSUB(M)-T4

analog inputs

ACC/DSUB(M)-U4

with thermo plug (ACC/DSUB(M)-T4) with shunt plug (ACC/DSUB(M)-I4) with plug (ACC/DSUB-ICP4)

ACC/DSUB(M)-I4 ACC/DSUB-ICP4

Sampling rate, Bandwidth, Filter, TEDS Parameter

Value

Remarks

Sampling rate

100 kHz

per channel

Bandwidth

0 Hz to 11 kHz

-3 dB

0 Hz to 8 kHz

-0.2 dB

Resolution

16 Bit

internal processing 24 Bit

Filter (digital) cut-off frequency, characteristic, order

2 Hz to 5 kHz

TEDS - Transducer Electronic DataSheets

conform IEEE 1451.4 Class II MMI

Butterworth, Bessel low pass filter: 8th order high pass filter: 4th order band pass: LP 4th and HP 4th order Anti-aliasing filter: Cauer 8.order with fcutoff = 0.4 fa ACC/DSUB(M)-TEDS-xx

General Parameter

Value typ.

min. / max.

Isolation nominal rating

60 V

test voltage

300 V (10 sec.)

Overvoltage protection

60 V ESD 2 kV

Input coupling configuration

transient protection: automotive load dump ISO 7637, Test impulse 6 DC differential, isolated

Errors and changes excepted For more information, contact: Berlin: +49 - 30 - 46 70 90 - 0

Remarks channel to case (earth) and channel-to-channel not with ICP plug differential input voltage (continuous) human body model test pulse 6 with max. -250 V Ri=30 , td=300 µs, tr 5 V on ranges < 5 V or device powered-down

Remarks with IEPE (ICP)-extension plug

+5 V >0.26 A 1.0

5% >0.2 A 1 G , < 40 pF

channel-to-ground (housing)

> 1 G , < 10 pF

channel-to-channel

> 165 dB (50 Hz) > 92 dB (50 Hz)

range range

Channel isolation (crosstalk)

Errors and changes excepted For more information, contact: Berlin: +49 - 30 - 46 70 90 - 0

2V 5V

Rsource = 0

Rsource

100

8

CS-4108-N, CL-4124-N Technical Data Sheet

17.07.2013

Temperature measurement - thermocouples Parameter

Value typ.

Measurement range

min. / max.

R, S, B, J, T, E, K, L, N

Resolution

Remarks according IEC 584

0.063 K (1/16 K)

Measurement uncertainty

< 0,6 K

type K, range -150 °C to 1200 °C type T, range -150 °C to 400 °C type N, range 380 °C to 1200 °C

< 1.0 K

type K, range -200 °C to -150 °C type T, range -200 °C to -150 °C

< 1.5 K

Temperature drift

0.02 K/K Ta

Ta= |Ta -25 °C| ambient temperature Ta

Uncertainty of cold junction compensation Temperature drift

type N, range -200 °C to 380 °C

< 0.15 K

with ACC/DSUB-T4 Tj = |Tj -25°C| cold junction temperature Tj

0.001 K/K Tj

Temperature measurement – PT100 Parameter Measurement range

Value typ.

min. / max.

Remarks

-200 °C to +850 °C -200 °C to +250 °C

Resolution

0.063 K (1/16 K)

Measurement uncertainty < 0.2 K < 0.05 % Temperature drift Sensor feed (PT100)

0.01 K/K Ta 250 µA

Errors and changes excepted For more information, contact: Berlin: +49 - 30 - 46 70 90 - 0

4-wire connection -200 °C to +850 °C plus of reading Ta=|Ta -25 °C|; ambient temp. Ta non-isolated

9

CS-4108-N, CL-4124-N Technical Data Sheet

17.07.2013

Technical Specs: Features (for all devices of imc C-SERIES) Digital Inputs Parameter

Value

Channels Configuration options

Remarks

8

common ground reference for each 4-channel group, isolated from the other input group

TTL or 24 V input voltage range

configurable at the DSUB globally for 8 Bits: jumper from LCOM to LEVEL:activates TTLmode LEVEL unconnected: activates 24 V-mode

Sampling rate

10 kHz

per channel

Isolation strength

±150 V

tested ±200 V isolated to system ground, supply and untereinander

Input configuration

differential

Input current

max. 500 µA

Switching threshold

isolated mutually and from supply

1.5 V (±200 mV)

5 V level

8 V (±300 mV)

24 V level

Switching time