Gradient Considerations in LC Operation in Gradients of Solvent Strength in LC
Layout
Gradient Application link
•Introduction % Organic
Final Hold
•LC System Considerations
Gradient Time
•Gradient Parameters Affecting Resolution
Inject Hold-up time
Re-equilibration Total Cycle Time 1998 Waters Corporation
Why do we use gradients?... Gradient Separations: Example: Peptides
...because... Properties of analytes
Why do we use gradients?
Retention (k) of the solutes has a steep dependence on the % organic in the mobile phase Wide range of differing hydrophobicities of the analytes
1
Gradient Considerations in LC Why do we use gradients?
Analyte Retention as a Function of Gradient Slope
The retention (k) of peptides have a steep dependence on the % organic in the mobile phase - The steeper the slope of the line the shallower the gradient must be to achieve maximum resolution 50 40
Peptide
Protein
30 Small Molecule
Slope=B
3.2
k (retention) for each analyte changes independently as the gradient slope changes.
3
Thus, the resolution between ln k peaks changes.
20 ln (k)10
Slope=B
3.1
peak 8 peak 9 peak 10 peak 11
2.9 2.8
0 -10
2.7
-20
2.6
-30 Neue
Volume Fraction Acetonitrile
% Organic
•LC System Considerations
0.67
0.71
System Considerations When Working with Gradients
Layout
•Introduction
0.62
% Acetonitrile
Carmody
Final Hold Gradient Time Inject
•Gradient Parameters Affecting Resolution
Hold-up time
Re-equilibration Total Cycle Time
2
Gradient Considerations in LC Different Configurations Create Different Time Delays
Volumes of Pumping Systems High Pressure Mixer: Multi-Pump Smaller System Volume = Dwell volume Pump 1
Mixer
Injector
Column
Detector
Pump 2
Low Pressure Gradient: Single Pump Larger System Volume = Dwell volume Solvent delivery
B
Mixing Valve Low – More Pipe Volume – More Time Delay
Mixing Valve High – Less Pipe Volume – Less Time Delay
A
C D
Injector
Delay Volume is Significant in Narrow BoreLow Flow Applications Theoretical Gradient No column , 1.00 mL/min 3.9x150 mm, 1.00 mL/min 2.1x150 mm, 0.29 mL/min
System Volume Column Volume A
D
C
Solvent delivery
Injector
Column
Detector
Proportioning Valve
Delay Volume and Extra-Column Band Broadening Volume
B
Column
%B Detector
Proportioning Valve
Extra Column Volume
3
Delay Volume 0
5
Re-equilibration Time 10
15 Minutes
20
25
30
Gradient Considerations in LC Application
Linear Gradient Separation 1 mL/min, 200 and 50 µL/min
System Considerations When Working with Gradients
0.12
1 mL/min 0.10
% Organic
0.08 AU 0.06
200 µ L/min
0.04
50 µ L/min 0.02 0.00
Final Hold Gradient Time Inject Adjust Hold-up time
Re-equilibration Total Cycle Time
10.0
20.0
30.0 Minutes
40.0
50.0
60.0
Delay Volume is Significant in Narrow BoreLow Flow Applications
Re-Equilibration Time For good system/column equilibration
Theoretical Gradient No column , 1.00 mL/min 3.9x150mm, 1.00 mL/min 2.1x150mm, 0.29 mL/min
%B
Delay Volume 0
4
5
where: tr is the re-equilibration time in minutes, VT is the total system volume, Vc is the column volume in mL F is the flowrate in mL/min. column volume = 0.7(π πr2L)
Re-equilibration Time 10
15 Minutes
20
25
tr = (3VT + 5Vc)/F
For example, Alliance system’s volume = 650 µL
30
Gradient Considerations in LC Basic Resolution Equation
Layout
-Factors Influencing Resolution for an Isocratic Separation t w
~
Selectivity
Resolution Equations -Factors Influencing Resolution for an Isocratic Separation Rs =
t
~
w
α-1 α
}
} Efficiency
Selectivity
k k +1
}
N 4
Retention
-Factors Influencing Resolution for a Gradient Separation Rs =
t w
~
ln α Selectivity
1 Bct0 + 1
}
Efficiency
}
}
N 4
Retention
-Factors are similar, however...
5
Rs = Resolution t = Retention time w = Peak width N = Plate Count α = Selectivity Factor k = Retention Factor
w1
w2
Analyte Retention as a Function of Gradient Slope
B = Slope of ln (k) with solvent composition (an analyte dependent property) c = Gradient Slope
t0 = Time of elution for an unretained peak
Retention
t {
•Gradient Parameters Affecting Resolution
Rs = Resolution Efficiency N = Plate Count t = Retention Time w = Average peak width α = Selectivity Factor k = Retention Factor
}
•LC System Considerations
k k +1
α-1 α
}
}
N 4
{
Rs =
•Introduction
3.2
k (retention) for each analyte changes independently as the gradient slope changes. Thus, the resolution between ln k peaks changes.
Slope=B
3.1 3
peak 8 peak 9 peak 10 peak 11
2.9 2.8 2.7 2.6
Carmody
0.62
0.67
% Acetonitrile
0.71
Gradient Considerations in LC What Factors Influence Gradient RP-HPLC Separations...
Part I - Factors influencing efficiency and retention Gradient Slope; c
Factors Influencing Resolution in Gradient RP-HPLC Separations... Part I - Factors influencing efficiency and retention
Column Length; L and N Flow Rate; F
Gradient Slope; c (%B/min.) - increase in organic concentration per unit time
Part II - Factors influencing selectivity Concentration and Type of Modifier
Rs =
t
~
w
Part III - Factors influencing reproducibility
Efficiency
ln α Selectivity
1 Bct0 + 1
}
}
N 4
}
Temperature Chemistry and Pore Size of the Packing Material
Retention
Column HPLC system
Principle of Gradient Separations What Factors Influence Gradient Slope? % tg
Two ways to change the slope change the percent organic (∆ ∆ %) of the mobile phase across a specified gradient run time.
Rs (Resolution)
c = %B/minute =
change the gradient run time (tg) while keeping the ∆% organic of the mobile phase constant.
0 0.01 Neue
6
approaching isocratic conditions
0.1
1
General slope of the gradient
10
Gradient Considerations in LC Resolution as a Function of Gradient Slope
What Factors Influence Gradient RP-HPLC Separations...
-Slope of the gradient = 0.66%/min
Part I - Factors influencing efficiency and retention Gradient Slope; c (%B/min.) - varied by changing the % organic across a specified gradient run time. All other variables are kept constant.
Conditions
10
AU
-Column: Symmetry300™, C18, 5 m, 3.9 x 150 mm
0.400
Rs =
t
~
w
N 4
1
ln a
9
Bct0 + 1
Selectivity
8
Retention
tg = gradient run time
1 % tg
B.
- Injection: 20µ µL
0.200
}
}
} Efficiency
- Sample: Tryptic digests of bovine cytochrome c
0.300
0.100
- Mobile Phase: Solvent A: 0.1% TFA in water Solvent B: 0.1% TFA in acetonitrile
0.000
- Gradient: 0-45 min., 0-30%B
0.00
.t
0+1
20.00
11
- Flow rate: 0.75 mL/min.
40.00
Minutes
- Temperature: 35 C - Detection: 214 nm
Alden
1998 Waters Corporation
Resolution as a Function of Gradient Slope Analyte Retention as a Function of Gradient Slope Conditions
0.66%/min 0.400 0.300
-Column: Symmetry300™, C18, 5 m, 3.9 x 150 mm - Sample: Tryptic digests of bovine cytochrome c - Injection: 20 L - Mobile Phase: Solvent A: 0.1% TFA in water Solvent B: 0.1% TFA in acetonitrile - Detection: 214 nm - Flow rate: 0.75 mL/min. - Temperature: 35 C
Rs (btw. 8 and 9) = 2.0 Rs (btw. 9 and 10) = 5.4 Rs (btw. 10 and 11) = 6.9
10
9
0.200 AU
11
8
0.100 0.000 30.00
35.00
40.00 Minutes
0.71%/min
0.62%/min 10
9
0.200 AU 0.100
8
11
0.600
Rs (btw. 8 and 9) = 1.7 Rs (btw. 9 and 10) = 4.7 Rs (btw. 10 and 11) = 6.8
0.500 0.400 0.300 AU 0.200
9 8
30.00
35.00
Alden
Minutes
40.00
0.000 30.00
Slope=B
3.1 3
peak 8 peak 9 peak 10 peak 11
2.9 2.8 2.7
11
2.6
0.100
0.000
7
Thus, the resolution between ln k peaks changes.
10
Rs (btw. 8 and 9) = 2.3 Rs (btw. 9 and 10) = 6.0 Rs (btw. 10 and 11) = 6.8
0.300
3.2
k (retention) for each analyte changes independently as the gradient slope changes.
35.00 Minutes
40.00
Carmody
0.62
0.67
% Acetonitrile
0.71
Gradient Considerations in LC Resolution as a Function of Gradient Duration
What Factors Influence Gradient RP-HPLC Separations...
78 Peaks 15 Minutes
0.1200 0.1000 0.0800 AU 0.0600 0.0400 0.0200 0.0000 0.00
5.00
10.00 Minutes
-Column: Symmetry300™, C18, 5 m, 4.6 x 150 mm - Sample: Tryptic digests of bovine serum albumin - Injection: 20 L - Mobile Phase: Solvent A: 0.1% TFA in water Solvent B: 0.1% TFA in acetonitrile - Gradient: 0-45 min., 0-30%B - Detection: 214 nm - Flow rate: 0.75 mL/min. - Temperature: 35 C
15.00
98 Peaks 25 Minutes
0.0800 0.0600
Part I - Factors influencing efficiency and retention
Conditions
AU 0.0400 0.0200
Gradient Slope; c (%B/min.) - varied by changing the gradient run
0.0000 0.00
time. All other variables are kept constant.
10.00
20.00
Minutes
114 Peaks 50 Minutes
0.0600 0.0400 AU 0.0200 0.0000
Rs =
t
~
w
0.00
N 4
ln α
}
}
Selectivity
B.
% tg
. t0 + 1
0.0400 0.0300 AU
20.00
40.00
Minutes
-Longer RT; Shallower the slope; Increases Rs
137 Peaks 75 Minutes
0.0500
0.0200 0.0100
-Rs ~ 1/c where c = gradient slope. All other variables are kept constant
0.0000
}
Efficiency
1
0.00
Retention
40.00 Minutes
20.00
60.00
162 Peaks 150 Minutes
0.0300 0.0200 AU 0.0100 0.0000 0.00
Gradient Modifications: Initial ACN% in Peptide Separations 0-28% in 75 min., (0.37%/min)
0.2000
0.1500
9
1
4
3 0.0500
5
Gradient SLOPE is more important if peaks elute during gradient (not in initial condition)
7 6
2
0.0000
r.t. = 85 min.
0.0500-0.0500
0.00
20.00
40.00 Minutes Minutes
6-28% in 60 min., (0.37%/min)
0.2000
60.00
80.00 80.00
11
9 10
0.1500
8 0.1000
1 3
AU 0.0500
4
5
7 6
2
0.0000
r.t. = 70 min.
0.0500 0.0500--
0.00
Alden
8
20.00
40.00 Minutes
Minutes
100.00
150.00
Alden
1998 Waters Corporation
Summary of Part I - Gradient Slope
11 10
8 0.1000 AU
50.00
60.00
80.00
-By changing the initial mobile phase conditions, but keeping the gradient slope the same, the run time can effectively be shortened without a loss in resolution.
• Gradient Slope is one of the most powerful operational parameter you have at your disposal • Resolution increases as gradient slope decreases. • Change in the initial percent organic can decrease the run time, maintain the resolution of your separation and preserve your elution pattern.
Gradient Considerations in LC What Factors Influence Gradient RP-HPLC Separations...
Part I - Factors influencing efficiency and retention Gradient Slope; c Column Length; L and N Flow Rate; F
Part II - Factors influencing selectivity
Part I - Factors influencing efficiency and retention
Rs =
t w
~
N 4
ln α
} Selectivity
Approach 2: do not scale the gradient volume in proportion to the column volume (such as keep the gradient run time constant while changing the column length).
9
0
+1
1 B.
% tg
. ε . πr2 .L/F + 1 t
Column Volume to Gradient Volume Relationship (Approach 1)
-Gradient volume scaled to column volume 5 minute gradient = 1 c.v./min Total volume = 10 column volumes
50 mm column
Approach 1: scale gradient volume in proportion to the column volume such as change the gradient run time with the column length).
.t
Retention
Column HPLC system
2 Approaches:
% tg
B.
}
Efficiency
1
Part III - Factors influencing reproducibility
The Number of Column Volumes per Minute Impacts Resolution
Gradient Slope; c – Column Length; L and N
}
Concentration and Type of Modifier Temperature Chemistry and Pore Size of the Packing Material
What Factors Influence Gradient RP-HPLC Separations...
250 mm column
Column volume = 0.5 mL
Column volume = 2.5 mL
5 minute gradient @ 1 mL/min
25 minute gradient @ 1 mL/min
gradient volume = tg x f.r. = 5 Total volume = g.v./c.v. = 10 column vols.
gradient volume = tg x f.r. = 25 Total volume = g.v./c.v. = 10 column vols.
Gradient Considerations in LC Variation in Column Lengths at Equal Ratio of Gradient Volumes to Column Volumes
What Factors Influence Gradient RP-HPLC Separations...
Column Volume = 0.83 mL Total Volume = 4.5 c.v.
Part I - Factors influencing efficiency and retention
5 minutes 4.6 X 50 mm
0.0600
Gradient Slope; c Column Length; L and N
AU0.0400 0.0200 4.00
5.00
6.00 Minutes
Rs =
t w
~
N
0.1500
1 B.
% tg
. ε . πr2 .L/F + 1 t
Selectivity
AU 0.0500
10.00
Retention
L (column length) is varied. Gradient volume is scaled in proportion to the column volume.
0.1500
Terms are constant
15 minutes 4.6 X 150 mm,
0.1000
}
Efficiency
}
}
4
ln α
Column Volume = 2.5 mL Total Volume = 4.5 c.v.
Conditions - Column: Symmetry300™, C18, 5 m - Sample: Tryptic digests of bovine serum albumin - Injection: 20 L - Mobile Phase: Solvent A: 0.1% TFA in water Solvent B: 0.1% TFA in acetonitrile - 0 - 30 %B in the time shown. - Flow rate: 0.75 mL/min. - Detection: 214 nm - Temperature: 35 C
12.00
14.00 Minutes
- Elution pattern stays
16.00
Column Volume = 4.2 mL Total Volume = 4.5 c.v.
the same. 25 minutes 4.6 X 250 mm,
0.1000
- Resolution inc. as the # of plates inc.
AU 0.0500
- Run time inc. as column 20.00
length inc.
25.00
Minutes Alden
Column Volume to Gradient Volume Relationship (Approach 2) -Gradient volume not scaled to column volume 5 minute gradient = 1 c.v./min Total volume = 10 column volumes 50 mm column 250 mm column
5 minute gradient @ 1 mL/min
Column volume = 2.5 mL 5 minute gradient @ 1 mL/min
gradient volume = tg x f.r. = 5 Total volume = g.v./c.v. = 10 column vols.
gradient volume = tg x f.r. = 5 Total volume = g.v./c.v. = 2 column vols.
Column volume = 0.5 mL
1998 Waters Corporation
What Factors Influence Gradient RP-HPLC Separations...
Part I - Factors influencing efficiency and retention Gradient Slope; c Column Length; L and N
Rs =
t w
~
10
Selectivity
1 B.
% tg
. ε . πr2 .L/F + 1 t
}
Efficiency
ln α
}
}
N 4
Retention
Gradient Considerations in LC Column Length Effects on Resolution at a Constant Gradient Duration Column Volume = 0.83 mL Total Volume =40.7 c.v..
Symmetry300™ 4.6 X 50 mm,C18, 5 µm 10
13
5,6
0.2500 0.2000
2
0.1500 AU0.1000
1
8
4 3
0.0500
12
9 11
7
0.0000 0.0500 0.0500-20.00
0.00
40.00
Minutes
Column Volume = 2.5 mL Total Volume =13.5 c.v.
Symmetry300™ 4.6 X 150 mm C18, 5µm 5,6
0.500
13
10
0.300 AU 0.200
4
2
8
3
1
0.100
11
7
-Will observe elution pattern changes.
0.000
Column Length Effects on Resolution at a Constant Gradient Duration (cont'd) Conditions 130 Peaks Symmetry300™ 4.6 X 50 mm, C18, 5 µm 0.02000 0.01500 0.01000 AU 0.00500 0.00000 0.00500 0.00500-0.00
20.00
40.00 Minutes
131 Peaks µm Symmetry300™ 4.6 X 150 mm, C18, 5µ
0.1200 0.1000 0.0800 AU0.0600 0.0400 0.0200 0.0000 0.00
20.00
Minutes
40.00
40.00
20.00
0.00
- Sample: Tryptic digests of bovine cytochrome c - Injection: 20 L - Mobile Phase: Solvent A: 0.1% TFA in water Solvent B: 0.1% TFA in acetonitrile - Gradient: 0-45 min., 030%B - Flow rate: 0.75 mL/min. - Detection: 214 nm - Temperature: 35 C
12
9
0.400
Conditions
Minutes
0.800
Column Volume = 4.2 mL Total Volume = 8.0 c.v.
Symmetry300™ 4.6 X 250 mm, C18, 5µ µm 9 10
0.600
56
AU0.400
4
2 1
0.200
8
12
13
11
-Resolution changes -Run time remains the same.
7
3
0
Alden
0.000 0.00
20.00
Minutes
0.00
20.00
Minutes
-50 mm column has a similar resolving power as 250 mm column if the gradient duration remains the same.
40.00 1998 Waters Corporation
40.00
Summary Part I - Column Length
What Factors Influence Gradient RP-HPLC Separations...
If the gradient volume is scaled proportionally to the column volume
Part I - Factors influencing efficiency and retention
Elution pattern does not change Resolution increases with column length.
If the gradient volume is not scaled in proportion to the column volume Elution pattern and resolution changes 50 mm column exhibits similar resolving power to a 250 mm column.
11
134 Peaks µm Symmetry300™ 4.6 X 250 mm, C18, 5µ
0.1200 0.1000 0.0800 AU0.0600 0.0400 0.0200 0.0000 0.0
- Sample: Tryptic digests of bovine serum albumin - Injection: 20 L (7 L for 4.6 X 50 mm) - Mobile Phase: Solvent A: 0.1% TFA in water Solvent B: 0.1% TFA in acetonitrile - Gradient: 0-45 min., 0-30%B - Temperature: 35 C - Flow rate: 0.75 mL/min. - Detection: 214 nm
Gradient Slope; c Column Length; L and N Flow Rate; F Part II - Factors influencing selectivity Concentration and Type of Modifier Temperature Chemistry and Pore Size of the Packing Material
Part III - Factors influencing reproducibility Column HPLC system
Alden
Gradient Considerations in LC What Factors Influence Gradient RP-HPLC Separations...
Resolution as a Function of Flow Rate at a Constant Gradient Duration
Part I - Factors influencing efficiency and retention Gradient Slope; c Column Length; L and N Flow Rate; F
12
Rs (Resolution)
10
F (flow rate) is varied. All other variables are kept constant
Rs =
t w
~
N 4
1
ln α
Efficiency Selectivity
% tg
. εt . πr2 .L/F + 1
}
}
}
B.
Retention
8
6
1 mL/min
4
2
0
0.33
3.3
Flow Rate (mL/min) Neue 1998 Waters Corporation
Flow Rate Effects on Resolution at a Constant Gradient Duration
Conditions
75 Peaks AU
0.0500 0.0400 0.0300 0.0200 0.0100 0.0000 0.00
20.00
Minutes
40.00
0.3 mL/minute
100 Peaks AU
0.0400 0.0300 0.0200 0.0100 0.0000
0.5 mL/minute 0.00
AU
0.02500 0.02000 0.01500 0.01000 0.00500 0.00000 0.00500 0.00500-0.00
20.00
40.00
135 Peaks
Minutes
20.00
Minutes
40.00
1.0 mL/minute
130 Peaks 0.02000
- Column: Symmetry300™, C18, 5µ µm, 4.6x50mm - Sample: Tryptic digests of bovine serum albumin - Injection: 20µ µL - Mobile Phase: Solvent A: 0.1% TFA in water Solvent B: 0.1% TFA in acetonitrile - Gradient: 0-45 min., 0-30%B - Temperature: 35 C - Detection: 214 nm
-Best resolution occurred at a flow rate of 1.0 mL/min. for this peptide under these conditions.
0.01500
AU
0.01000 0.00500 0.00000 0.00500-0.00500 0.00
12
2.0 mL/minute 20.00
Minutes
40.00
-Elution pattern changes. 1998 Waters Corporation
Alden
Summary of Part I - Flow Rate Maximum resolution is achieved at an optimal flow rate: As flow rate changes N changes As flow rate changes the elution pattern changes.
Gradient Considerations in LC What Factors Influence Gradient RP-HPLC Separations...
Type of Modifiers
Part I - Factors influencing efficiency and retention Gradient Slope; c Column Length; L and N Flow Rate; F
– Solvation
Part II - Factors influencing selectivity – Ionization
Concentration and Type of Modifier Temperature Chemistry and Pore Size of the Packing Material
– Ion-pairing
Part III - Factors influencing reproducibility
– Volatility
Column HPLC system
The Power of Different TFA Concentrations in Your Mobile Phase
Effects of TFA Concentration on Resolution - Typical gradient conditions Conditions - Column: Symmetry300™, C18, 5 µm, 3.9x150mm,
0.1% TFA in solvents A and B
- Injection: 20µ µL
0.200
13 12
9 5,6 2
0.100
1 0.000
4
8 7
3
11
- Mobile Phase: Solvent A: water Solvent B: acetonitrile
- Gradient: 0-45 min., 0-30%B - Flow rate: 0.75 mL/min.
0.00
1
Minutes
1
1
- Temperature: 35 C - Detection: 214 nm
13
12 11
8
7
10 13 9
5,6
2
4
7
3
12
8
11
10
- 0.2% TFA in solvents A and B
0.00 Alden
4
3
- 0.1% TFA in solvents A and B
40.00
20.00
13 9
56
2
0.300
AU
10
- 0.05% TFA in solvents A and B
- Sample: Tryptic digests of bovine cytochrome c
10
(Collection)
13
5,6 2
3
4
7
20.00
9 8
12 11
40.00
Minutes Alden
Gradient Considerations in LC The Power of Different TFA Concentrations in Your Mobile Phase
Alternate Ion Pairing Reagents TFA and HFBA (Heptafluorobutyric Acid)
Conditions 10
0.1% TFA in Solvent A 0.1% TFA in Solvent B
- Column: Symmetry300™, C18, 5µ µm, 3.9x150mm
0.200
9
12
5,6
AU
2
0.100
1
4
8 7
3
11
20.00
- Gradient: 0-45 min., 0-30%B
0.05% TFA in Solvent A 0.1% TFA in Solvent B
0.2500
- Sample: Tryptic Digests of Bovine Cytochrome c
- Mobile Phase: Solvent A: water Solvent B: acetonitrile
40.00
Minutes
0.3000
10
- Flow rate: 0.75 mL/min.
0.2000
12 13 AU 0.1500
1
0.0500
9
56
2
0.1000
TFA
7
3 4
8
Column: Delta-Pak™ C18, 5 µm, 300 , 2.0 x 150 mm
0.02
Eluents: A=water/ 0.1% TFA or HFBA B=acetonitrile/ 0.1% TFA or HFBA
- Injection: 20µ µL
0.000
0.00
11
Gradient: O-60 % B 120 min
0.00
0.40
Temp: 35 C
0.20
0.00 20
20.00
Flow: 0.18 mL/min
HFBA
- Temperature: 35 C - Detection: 214 nm
0.0000 0.00
Sample: Rabbit cytochrome c tryptic digest, 500 pmol
0.40
13
Absorbance (214 nm)
0.300
40.00
40
60
80
Time (min)
Minutes
Alden
What Factors Influence Gradient RP-HPLC Separations...
Temperature Effects on Resolution
Part I - Factors influencing efficiency and retention Gradient Slope; c Column Length; L and N Flow Rate; F
Part II - Factors influencing selectivity Concentration and Type of Modifier Temperature Chemistry and Pore Size of the Packing Material
Part III - Factors influencing reproducibility Column HPLC system
14
Resolution is temperature dependent Temperature is a critical parameter to control in order to achieve reproducible separations.
Gradient Considerations in LC What Factors Influence Gradient RP-HPLC Separations...
Temperature Effects on Resolution Conditions
10
Rs (5,6) = 0 Rs (8,9) = 2.39
0.300
30 C
- Column: Symmetry300™, C18, 5µ µm, 3.9x150mm
9
0.200
5, 6 AU
Part I - Factors influencing efficiency and retention
11
8
7
0.100
- Sample: Tryptic digests of bovine cytochrome c
0.000 25.00
30.00
35.00
Gradient Slope; c – Column Length; L and N –
40.00
Minutes
- Injection: 20µ µL
0.200
10
Rs (5,6) = 0 Rs (8,9) = 2.07
0.300
35 C
AU
8
7
0.100
Flow Rate; F–
- Mobile Phase: Solvent A: 0.1% TFA in water Solvent B: 0.1% TFA in acetonitrile
9
5, 6
11
0.000
Part II - Factors influencing selectivity
- Gradient: 0-45 min., 0-30%B 25.00
30.00
35.00
40.00
Minutes
Rs (5,6) = 0.84 Rs (8,9) = 1.63
0.300
0.200
- Flow rate: 0.75 mL/min.
10
40 C - Detection: 214 nm
Part III - Factors influencing reproducibility
9
5 6
AU
8
7
0.100
Column– HPLC system –
11
0.000 25.00
Alden
30.00
35.00
Concentration and Type of Modifier – Temperature– Pore Size and Chemistry of the Packing Material
40.00
Minutes
Pore Size Effects on Resolution
Selectivity Differences Between Packings
Conditions
Symmetry , C18, 5µm, 4.6 X 150 mm, 100 10
0.30
4
23
AU 0.10
13 12
9
56
0.20
8
1
11
7
0.00
0.10 0.10-0.00
20.00
40.00 Minutes
Symmetry300™, C18, 5µm, 4.6 X 150 mm, 300
- Sample: Tryptic digests of cytochrome c (bovine) - Injection: 20µ µL - Mobile Phase: Solvent A: 0.1% TFA in water Solvent B: 0.1% TFA in acetonitrile - Gradient: 0-50 min., 030%B - Temperature: 35 C - Flow Rate: 0.75 mL/min. - Detection: 214 nm
10 13
5,6
0.500
12
9
0.400 0.300 AU 0.200
4
2 1
0.100
8
3
7
0.10
5 4
AUFS 0.05
1
2
3
0.00 5.00
10.00
5 4
AUFS 0.05
1 2
Carmody
20.00
Minutes
40.00
Gradient: Sample:
3
Detector: Temperature:
0.00
0.00
Conditions: Columns: Flow Rates: Mobile Phase:
15.00
Symmetry C8, 100A
0.10
-Different pore sizes change selectivity.
Ac-Arg-Gly-X-X-Gly-Gly-Leu-Gly-LysAmide -X-X-: 1: Ala-Gly with free alpha amino group 3: Ala-Gly 2: Gly-Gly 5: Val-Val 4: Val-Gly
SymmetryShield™ RP8, 100A
11
0.000
15
Alberta Peptides on Symmetry Reversed-Phase Columns
5.00
10.00
15.00 Minutes
3.9 mm x 150 mm 0.8 mL/min A. 0.1% TFA aqueous; B. acetonitrile with 0.1% TFA 10% to 40% B in 30 minutes, step to 60% B for 5 min 9 L Alberta Peptides, mix with 5 decapeptides 214 nm 35 C
Gradient Considerations in LC What Factors Influence Gradient RP-HPLC Separations...
Peptide Mapping Validation -Robustness Testing
Part I - Factors influencing efficiency and retention Gradient Slope; c Column Length; L and N
Choice and quality of enzyme Digestion conditions HPLC conditions Equipment
Flow Rate; F
Part II - Factors influencing selectivity Concentration and Type of Modifier Temperature Chemistry and Pore Size of the Packing Material
System Column
Part III - Factors influencing reproducibility Column HPLC system
Separations Module
Effects of Irreproducible Gradient Delivery Traditional HPLC System
-
Gradient Accuracy & Precision
Experience resolution differences and retention time shifts
10%B
0.08 0.07
1% Step Gradient at 1.0 mL/min Accuracy