ab65390 HDL and LDL/VLDL Cholesterol Assay kit (Colorimetric/ Fluorometric) Instructions for use: For sensitive and accurate measurement of HDL and LDL/VLDL cholesterol in serum samples.
This product is for research use only and is not intended for diagnostic use.
Version 7 Last Updated 27 January 2016
Table of Contents INTRODUCTION
1
1.
BACKGROUND
1
2.
ASSAY SUMMARY
2
GENERAL INFORMATION
3
3.
PRECAUTIONS
3
4.
STORAGE AND STABILITY
3
5.
LIMITATIONS
4
6.
MATERIALS SUPPLIED
4
7.
MATERIALS REQUIRED, NOT SUPPLIED
5
8.
TECHNICAL HINTS
6
ASSAY PREPARATION
7
9.
REAGENT PREPARATION
10.
STANDARD PREPARATION
11.
SAMPLE PREPARATION
7 8 10
ASSAY PROCEDURE
12
12.
12
ASSAY PROCEDURE
DATA ANALYSIS
15
13.
CALCULATIONS
15
14.
TYPICAL DATA
17
RESOURCES
19
15.
QUICK ASSAY PROCEDURE
19
16.
TROUBLESHOOTING
20
17.
INTERFERENCES
22
18.
FAQS
22
19.
NOTES
24
INTRODUCTION INTRODUCTION
1. BACKGROUND HDL and LDL/VLDL Cholesterol Assay Kit (Colorimetric/Fluorometric) (ab65390) provides a simple quantification method of HDL and LDL/VLDL after a convenient separation of the different lipoprotein fractions [HDL from LDL and VLDL (very low-density lipoprotein)] in serum samples. In this assay, cholesterol oxidase specifically recognizes free cholesterol and produces a component that will react with a probe to generate color (OD 570 nm) and fluorescence (Ex/Em = 538/587 nm). Because cholesterol esterase is able to hydrolyze cholesteryl ester into free cholesterol, both cholesterol ester and free cholesterol can be detected separately, depending on the presence or absence of cholesterol esterase in the reactions. Regulation of HDL (high-density-lipoprotein)-cholesterol and LDL (lowdensity-lipoprotein)-cholesterol plays a central role in various disease developments. It is well known that low levels of HDL and high level of LDL are associated with an increased risk of cardiovascular events.
ab65390 HDL and LDL/VLDL Cholesterol Assay Kit
1
INTRODUCTION 2. ASSAY SUMMARY
Separate HDL and LDL/VLDL
Prepare standard curve
Prepare samples
Prepare and add reaction mix
Measure absorbance (OD570 nm) or fluorescence (Ex/Em = 538/587 nm)
ab65390 HDL and LDL/VLDL Cholesterol Assay Kit
2
GENERAL INFORMATION GENERAL INFORMATION
3. PRECAUTIONS Please read these instructions carefully prior to beginning the assay.
All kit components have been formulated and quality control tested to function successfully as a kit.
We understand that, occasionally, experimental protocols might need to be modified to meet unique experimental circumstances. However, we cannot guarantee the performance of the product outside the conditions detailed in this protocol booklet.
Reagents should be treated as possible mutagens and should be handle with care and disposed of properly. Please review the Safety Datasheet (SDS) provided with the product for information on the specific components.
Observe good laboratory practices. Gloves, lab coat, and protective eyewear should always be worn. Never pipet by mouth. Do not eat, drink or smoke in the laboratory areas.
All biological materials should be treated as potentially hazardous and handled as such. They should be disposed of in accordance with established safety procedures.
4. STORAGE AND STABILITY Store kit at -20°C in the dark immediately upon receipt. Kit has a storage time of 1 year from receipt, providing components have not been reconstituted. Refer to list of materials supplied for storage conditions of individual components. Observe the storage conditions for individual prepared components in the Materials Supplied Section. Aliquot components in working volumes before storing at the recommended temperature. Reconstituted components are stable for 2 months.
ab65390 HDL and LDL/VLDL Cholesterol Assay Kit
3
GENERAL INFORMATION 5. LIMITATIONS
Assay kit intended for research use only. Not for use in diagnostic procedures.
Do not mix or substitute reagents or materials from other kit lots or vendors. Kits are QC tested as a set of components and performance cannot be guaranteed if utilized separately or substituted.
6. MATERIALS SUPPLIED
Cholesterol Assay Buffer
25 mL
Storage Condition (Before Preparation) -20°C
2X LDL/VLDL Precipitation Buffer Cholesterol Probe (in DMSO, anhydrous) Enzyme Mix (Lyophilized)
10 mL
-20°C
-20°C
200 µL
-20°C
-20°C
1 vial
-20°C
-20°C
1 vial
-20°C
-20°C
100 µL
-20°C
-20°C
Item
Cholesterol Esterase (Lyophilized) Cholesterol Standard (2 µg/µL)
Amount
ab65390 HDL and LDL/VLDL Cholesterol Assay Kit
Storage Condition (After Preparation) -20°C
4
GENERAL INFORMATION 7. MATERIALS REQUIRED, NOT SUPPLIED These materials are not included in the kit, but will be required to successfully perform this assay:
Microplate reader capable of measuring absorbance at OD 570 nm or fluorescence at Ex/Em = 535/587 nm
MilliQ water or other type of double distilled water (ddH2O)
Pipettes and pipette tips, including multi-channel pipette
Assorted glassware for the preparation of reagents and buffer solutions
Tubes for the preparation of reagents and buffer solutions
96 well plate with clear flat bottom (for colorimetric assay) / 96 well plate with clear flat bottom, preferably black (for fluorometric assay)
ab65390 HDL and LDL/VLDL Cholesterol Assay Kit
5
GENERAL INFORMATION 8. TECHNICAL HINTS
This kit is sold based on number of tests. A ‘test’ simply refers to a single assay well. The number of wells that contain sample, control or standard will vary by product. Review the protocol completely to confirm this kit meets your requirements. Please contact our Technical Support staff with any questions.
Selected components in this kit are supplied in surplus amount to account for additional dilutions, evaporation, or instrumentation settings where higher volumes are required. They should be disposed of in accordance with established safety procedures.
Avoid foaming components.
Avoid cross contamination of samples or reagents by changing tips between sample, standard and reagent additions.
Ensure plates are properly sealed or covered during incubation steps.
Ensure all reagents and solutions are at the appropriate temperature before starting the assay.
Samples which generate values that are greater than the most concentrated standard should be further diluted in the appropriate sample dilution buffer.
Make sure you have the right type of plate for your detection method of choice.
Make sure all necessary equipment is switched on and set at the appropriate temperature.
or
bubbles
when
ab65390 HDL and LDL/VLDL Cholesterol Assay Kit
mixing
or
reconstituting
6
ASSAY PREPARATION ASSAY PREPARATION
9. REAGENT PREPARATION 9.1.
Briefly centrifuge small vials at low speed prior to opening Cholesterol Assay Buffer: Ready to use as supplied. Equilibrate to room temperature before use. Store at -20°C.
9.2.
Precipitation Buffer: Ready to use as supplied. Equilibrate to room temperature before use. Store at -20°C.
9.3.
Cholesterol Standard: Ready to use as supplied. Equilibrate to room temperature before use. Aliquot standard so that you have enough volume to perform the desired number of assays. Store at -20°C.
9.4.
Cholesterol Probe : Ready to use as supplied. Warm by placing in a 37°C bath for 1 – 5 minutes to thaw the DMSO solution before use. NOTE: DMSO tends to be solid when stored at -20°C, even when let at room temperature, so it needs to melt for few minutes at 37°C. Aliquot probe so that you have enough volume to perform the desired number of assays. Store at -20°C protected from light and moisture. Use within two months.
9.5.
Cholesterol Esterase: Reconstitute in 220 µL Cholesterol Assay Buffer. Keep on ice during the assay. Aliquot esterase so that you have enough volume to perform the desired number of assays. Store aliquots at -20°C.
9.6.
Enzyme Mix: Reconstitute in 220 µL Cholesterol Assay Buffer. Keep on ice during the assay. Aliquot enzyme mix so that you have enough volume to perform the desired number of assays. Store aliquots at -20°C.
ab65390 HDL and LDL/VLDL Cholesterol Assay Kit
7
ASSAY PREPARATION 10. STANDARD PREPARATION
Always prepare a fresh set of standards for every use.
Discard the working standard dilutions after use as they do not store well.
10.1. For colorimetric assay 10.1.1. Prepare a 250 µL 0.2 µg/µL Cholesterol Standard by diluting 25 µL Cholesterol Standard in 225 µL of Cholesterol Assay Buffer. 10.1.2. Using the 0.2 µg/µL Cholesterol Standard, prepare standard curve dilution as described in the table in a microplate or microcentrifuge tubes:
Standard #
Volume of Cholesterol standard (µL)
Assay Buffer (µL)
1
0
150
Final volume standard in well (µL) 50
2
12
138
50
1
3
24
126
50
2
4
36
114
50
3
5
48
102
50
4
6
60
90
50
5
End Conc Cholesterol in well (µg/well) 0
Each dilution has enough amount of standard to set up duplicate reading (2 x 50 µL).
ab65390 HDL and LDL/VLDL Cholesterol Assay Kit
8
ASSAY PREPARATION 10.2. For fluorometric assay 10.2.1. Prepare a 0.25 µg/µL Cholesterol standard by diluting 5 µL Cholesterol Standard in 35 µL of Cholesterol Assay Buffer. 10.2.2. Prepare a 250 µL 0.025 µg/µL Cholesterol Standard by diluting 25 µL Cholesterol Standard in 225 µL of Cholesterol Assay Buffer. 10.2.3. Using 0.025 µg/µL Cholesterol Standard, prepare standard curve dilution as described in the table in a microplate or microcentrifuge tubes:
Standard #
Volume of Cholesterol standard (µL)
Assay Buffer (µL)
1
0
150
Final volume standard in well (µL) 50
2
12
138
50
0.1
3
24
126
50
0.2
4
36
114
50
0.3
5
48
102
50
0.4
6
60
90
50
0.5
End Conc Cholesterol in well (µg/well) 0
Each dilution has enough amount of standard to set up duplicate reading (2 x 50 µL).
ab65390 HDL and LDL/VLDL Cholesterol Assay Kit
9
ASSAY PREPARATION 11. SAMPLE PREPARATION General Sample Information
We recommend performing several dilutions of your sample to ensure the readings are within the standard value range.
We recommend that you use fresh samples. If you cannot perform the assay at the same time, we suggest that you complete the Sample Preparation step before storing the samples. Alternatively, if that is not possible, we suggest that you snap freeze your samples in liquid nitrogen upon extraction and store them immediately at 80°C. When you are ready to test your samples, thaw them on ice. Be aware however that this might affect the stability of your samples and the readings can be lower than expected.
11.1. Biological fluids (Serum and Plasma): Quantification of TOTAL CHOLESTEROL: 11.1.1. Use samples directly; no preparation step is required. Proceed to Section 12. Separation of HDL and LDL/VLDL: 11.1.2. Mix 100 µL of sample with 100 µL of 2X Precipitation Buffer in microcentrifuge tubes. 11.1.3. Incubate 10 minutes at room temperature. 11.1.4. Centrifuge at 2,000 x g (5,000 rpm on a bench-top microcentrifuge) for 10 minutes. 11.1.5. Transfer the supernatant into new labeled tubes. This is the HDL fraction. 11.1.6. Precipitates contain the LDL/VLDL fraction. To measure the LDL/VLDL fraction, centrifuge the precipitate again as in step 11.1.4. 11.1.7. Remove trace amount of HDL supernatant carefully. 11.1.8. Resuspend the precipitate in 200 µL PBS. This is the LDL/VLDL fraction.
ab65390 HDL and LDL/VLDL Cholesterol Assay Kit
10
ASSAY PREPARATION If the supernatant is cloudy, the sample should be re-centrifuged. If the sample remains cloudy, dilute the sample 1:1 with PBS and repeat the separation procedure from step 11.1.2 For serum and plasma: use 1 – 20 µL of the fractions. Adjust total volume to 50 µL with Cholesterol Assay Buffer. NOTE: We suggest using different volumes of sample to ensure readings are within the Standard Curve range.
ab65390 HDL and LDL/VLDL Cholesterol Assay Kit
11
ASSAY PROCEDURE ASSAY PROCEDURE
12. ASSAY PROCEDURE
Equilibrate all materials and prepared reagents to room temperature prior to use.
We recommend that you assay all standards, controls and samples in duplicate.
Prepare all reagents, working standards and samples as directed in the previous sections.
12.1. Set up Reaction wells: Standard wells = 50 µL Standard dilutions. Sample wells for TOTAL cholesterol = 2 – 50 µL samples (adjust volume to 50 µL/well with Assay Buffer). Sample wells for FREE cholesterol = 2 – 50 µL samples (adjust volume to 50 µL/well with Assay Buffer). 12.2. Cholesterol Reaction Mix (COLORIMETRIC ASSAY): 12.2.1. Prepare 50 µL of Reaction Mix for each reaction. Mix enough reagents for the number of assays (samples and controls) to be performed. Prepare a master of the Reaction Mix to ensure consistency. We recommend the following calculation: X µL component x (Number reactions +1) Component
Total Cholesterol Reaction Mix (µL) 44
Free Cholesterol Reaction Mix (µL)
Cholesterol Probe
2
2
Enzyme Mix
2
2
Cholesterol Esterase*
2
0
Cholesterol Assay Buffer
46
*NOTE: Cholesterol Esterase hydrolyzes cholesteryl ester to free cholesterol. If no esterase is added to the reaction, the assay only detects free cholesterol only. With the addition of Cholesterol Esterase, the assay detects total cholesterol (cholesterol and cholesteryl esters).Cholesterol Esterase must be added to the standard curve wells to convert all the cholesterol in the standard solution. ab65390 HDL and LDL/VLDL Cholesterol Assay Kit
12
ASSAY PROCEDURE 12.2.2. Add 50 µL of Total Cholesterol Reaction Mix to Standard wells. 12.2.3. Add 50 µL of Total Cholesterol Reaction Mix to Total Cholesterol sample wells. 12.2.4. Add 50 µL of Free Cholesterol Reaction Mix to Free Cholesterol sample wells. 12.2.5. Mix and incubate at 37°C for 60 minutes protected from light. 12.2.6. Measure output immediately on a microplate reader at OD 570 nm. 12.3. Cholesterol Reaction Mix (FLUOROMETRIC ASSAY) 12.3.1. Prepare 50 µL of Reaction Mix for each reaction. Mix enough reagents for the number of assays (samples and controls) to be performed. Prepare a master of the Reaction Mix to ensure consistency. We recommend the following calculation: X µL component x (Number reactions +1) Component
Cholesterol Assay Buffer Cholesterol Probe
Total Cholesterol Reaction Mix (µL) 45.6
Free Cholesterol Reaction Mix (µL) 47.6
0.4
0.4
Enzyme Mix
2
2
Cholesterol Esterase*
2
0
NOTE: For the fluorometric assay, use 0.4 μL/well of the Probe to decrease the background readings, therefore increasing detection sensitivity. *NOTE: Cholesterol Esterase hydrolyzes cholesteryl ester to free cholesterol. If no esterase is added to the reaction, the assay only detects free cholesterol only. With the addition of Cholesterol Esterase, the assay detects total cholesterol (cholesterol and cholesteryl esters).Cholesterol Esterase must be added to the standard curve wells to convert all the cholesterol in the standard solution.
ab65390 HDL and LDL/VLDL Cholesterol Assay Kit
13
ASSAY PROCEDURE
12.3.2. Add 50 µL of Total Cholesterol Reaction Mix to Standard wells. 12.3.3. Add 50 µL of Total Cholesterol Reaction Mix to Total Cholesterol sample wells. 12.3.4. Add 50 µL of Free Cholesterol Reaction Mix to Free Cholesterol sample wells. 12.3.5. Mix and incubate at 37°C for 60 minutes protected from light. 12.3.6. Measure output immediately on a microplate reader at Ex/Em= 535/587 nm.
ab65390 HDL and LDL/VLDL Cholesterol Assay Kit
14
DATA ANALYSIS DATA ANALYSIS
13. CALCULATIONS
Samples producing signals greater than that of the highest standard should be further diluted in appropriate buffer and reanalyzed, then multiply the concentration found by the appropriate dilution factor.
13.1. Average the duplicate reading for each standard and sample. 13.2. Subtract the mean absorbance value of the blank (Standard #1) from all standard and sample readings. This is the corrected absorbance. 13.3. Plot the corrected absorbance values for each standard as a function of the final concentration of Cholesterol. 13.4. Draw the best smooth curve through these points to construct the standard curve. Most plate reader software or Excel can plot these values and curve fit. Calculate the trendline equation based on your standard curve data (use the equation that provides the most accurate fit). 13.5. Concentration of Cholesterol in the test samples is calculated as: 𝐶ℎ𝑜𝑙𝑒𝑠𝑡𝑒𝑟𝑜𝑙 𝐶𝑜𝑛𝑐𝑒𝑛𝑡𝑟𝑎𝑡𝑖𝑜𝑛 = 𝐴 𝑉 ∗𝐷
()
Where: A = amount of cholesterol in the sample well calculated from the standard curve (µg). V = volume of sample added to the sample reaction well (µL). D = Dilution Factor. For Total cholesterol, D = 1; for HDL and LDL/VLDL fractions, D = 2. Cholesterol Molecular Weight: 386.6 g/mol 1 µg/µL = 100 mg/dL.
ab65390 HDL and LDL/VLDL Cholesterol Assay Kit
15
DATA ANALYSIS
Total Cholesterol (Free Cholesterol + Cholesteryl esters): use Total Cholesterol Reaction Mix. Free Colesterol: use Free Cholesterol Reaction Mix. Cholesteryl esters: Total Cholesterol value – Free Cholesterol value.
ab65390 HDL and LDL/VLDL Cholesterol Assay Kit
16
DATA ANALYSIS 14. TYPICAL DATA TYPICAL STANDARD CURVE – Data provided for demonstration purposes only. A new standard curve must be generated for each assay performed
Figure 1: Typical Cholesterol Standard calibration curve using colorimetric assay.
Figure 2: Measurement of total cholesterol, HDL, LDL/VLDL from serum samples from various species. Total Cholesterol (blue bar), HDL (green bar), and LDL/VLDL (yellow bar) cholesterol. ab65390 HDL and LDL/VLDL Cholesterol Assay Kit
17
DATA ANALYSIS
Figure 3: Wild type (WT) or farnesoid X receptor knock-out (FRX-KO) mice were treated with either vehicle (Veh) or a high fat-containing diet (HFD) for 16 weeks. Serum HDL (top) and serum LDL/VLDL cholesterol (bottom) from 5-7 mice/group was measured using ab65390 following protocol instructions. An asterisk (*) means P