DS-NET is a measurement and control system designed for many demanding applications, especially in the fields of ƒ ƒComponent Testing ƒ ƒEngine Testing ƒ ƒProcess Performance Testing ƒ ƒStructural Monitoring

The DS-NET system is rugged and scalable from e.g. a two channel control unit to a large synchronized measurement grid with thousands of channels. It is as flexible as being a stand alone data logger, a channel expansion of Dewesoft instruments, an Ethernet based distributed measurement system or a full-featured independent data acquisition instrument. The completely modular architecture ensures always a perfect fit of the system configuration for the application at hand. A wide range of DS-NET modules is available to support almost any type of input and output signals. These multi-function modules can be combined in countless ways and provide top notch data recording and process control. The system is designed for practical industrial appliance and thus is comprised of all metal housings and robust electronics offering galvanic isolation. Popular connector options enable convenient sensor connection and in combination with the easy-touse software ensure time saving system setup. Considering all these facts, DS-NET will serve you many years and is a safe investment.

DEWESoft™ measurement innovation

KEY FEATURES ƒ ƒMedium speed data acquisition up to 10 kS/s/ch ƒ ƒApplication areas ƒ ƒ D istributed data acquisition, Ethernet based ƒ ƒ S tand alone data logging ƒ ƒ Complete instrument running local DEWESoft™ software ƒ ƒ Customized LabVIEW™ based solution ƒ ƒ C hannel expansion for Dewesoft instruments ƒ ƒCompletely modular and thus very flexible in configuration ƒ ƒScalable from two to several thousand channels ƒ ƒPortable and 19” rack-mount lines ƒ ƒREAL-TIME performance ƒ ƒRedundant data storage ƒ ƒOperating temperature -20° C to +60° C

45

DS-NET DS-NET

Application Areas Ethernet data acquisition system DS-NET is a very flexible and compact Ethernet based data acquisition system. There is a portable line as well as a 19” rack-mount line. Both lines offer very precise galvanically isolated signal conditioning and enhanced features and reliability. Usually the DS-NET system is connected to a host computer running DEWESoft™ online data acquisition software. Up to 160 kS/s can be received from a single DS-NET system and then be processed, visualized and stored on the host computer. But DS-NET also offers real-time performance! Since Microsoft Windows ® is no real-time operating system it can not guarantee certain reaction times. DS-NET runs its own internal real-time operating system and can handle output and alarm functions directly inside the instrument. Thus accurately defined response times are guaranteed - completely independent of any PC.

Fixed latency time Alarm handling inside module

MEASUREMENT vALUE

500 µsec dIGITAL OUT ALARM

Stand alone data logging Every DS-NET system is ready to be used as a rugged stand-alone data logger - without any additional costs! The logging process is configured by a single click in DEWESoft™ software. Measurement data and calculated values can be stored to a USB memory-stick: up to a limit of 32 GB. For triggered storing an aggregate sampling rate of max. 160 kS/s and up to 2 million samples per trigger event are the limit. For continuous storing an aggregate sampling rate of max. 20 kS/s is the limit. Data is stored into files of max. 2 million samples each without any gap between the files. USB sticks can be hot-swapped during measurement without losing any data thanks to the internal buffer memory. Data analysis can be done offline in DEWESoft™ software.

Redundancy in data acquisition The combination of the data logging feature and DEWESoft™ online recording software gives you redundancy in data acquisition for maximum reliability. Both, a USB stick and a measurement PC (via Ethernet), are connected to the DS-NET system in parallel. Data is logged to the USB stick while you are using DEWESoft™ to process, analyze and store the very same data at the same time! As a result, even if your Ethernet connection should break during a measurement, your data is safe, since it is logged to the USB stick.

46

http://www.dewesoft.com

DS-NET DS-NET Industrial monitoring – data cloud Multiple DS-NET systems are connected to an Ethernet LAN so that the measurement data can be processed, analyzed and stored by any computer within the network. For highest data quality all the measurement nodes are synchronized to each other either by hardware for best performance (time skew typ. ±2 µs) or software (time skew typ. 1 ms). Usually LabVIEW™ software applications are created for each particular installation. Clients also can read the online values via Modbus TCP/IP and the data can be stored to your SQL database, even to a remote server over the Internet! Of course DEWESoft™ software can be used too.

MASTER

SLAVE

SLAVE

Hardware SYNC ±2 µs

Software SYNC typ. 1 ms

Sync cable up to 400 m

Switch

LabView™

DEWESoft™ X

Modbus TCP/IP

SQL Database

Application software

Complete instrument running local DEWESoft™ software For maximum portability there is a compact CPU module available which adds a fullfeatured PC to your DS-NET system and turns it into a complete instrument. The CPU module enables you to run DEWESoft™ software locally. Measurement data is stored directly onto the internal flash disk and the local PC can be remote accessed via WLAN! Attach a keyboard and a display to the instrument for full control and to display the measurement data. Analog IN Digital / Counter In CAN BUS

Remote Control (iPad via WLAN)

WLAN GPS WLAN

UPS Power supply

USB

CAN

DIGITAL

SENSORS

ANALOG

CAR-UPS

GPS

12 V

DS-NET; Gate and Modules

Dewesoft Display MOB-DISP-10

DEWESoft™ measurement innovation

47

DS-NET DS-NET

System Architecture Any DS-NET system starts with one DS-GATE as the base and then up to 16 DS-NET-modules can be chosen to complete one system. An optional DS-NET CPU module can be added to include a local computer. The ingenious mechanic and electronic design of the portable system enable configurations of any number of modules without wasting any space. This way a configuration of e.g. one GATE with 4 modules is exactly 5 slots wide – plus handles on both sides.

Option: DS-NET-CPU2

Option: DS-NET-CPU2i

DS-GATE

DS-GATE The DS-GATE is the central controller for each DS-NET system. It powers and controls the DS-NET-modules and provides an Ethernet TCP/IP interface to a host computer. A 12 MB data buffer guarantees gap-free data transmission. The DS-GATE has Sync In/Out interfaces to daisy-chain multiple systems and offers a serial interface for GPS or IRIG-B devices. One USB interface is included for direct data logging to a USB memory stick. Advanced calculations, alarm handling, boolean combinations, etc. can be defined and run within the DS-GATE independent from any PC or Windows operating system. The max. total sampling rate supported by one DS-GATE is 160 kS/s.

Option: DS-NET-CPU The DS-NET-CPU adds a full-featured PC to your DS-NET system and turns it into a complete instrument. The CPU module enables you to run DEWESoft™ software locally. The module includes an Intel ® Atom 1.6 GHz CPU with 1 GB RAM and a 64 GB flash disk. It connects to DS-GATE via Ethernet and offers further interfaces like 4x USB, WLAN and DVI. Including software option: DEWESoft NET (see “Distributed acquisition”).

48

CPU

Intel® Atom Z530 1.6GHz

RAM

2GB

Hard-disk

64 GB FLASH hard disk MLC

Operating System

Windows XP or Windows 7

Display & Graphics

HDMI output up to 1920 x 1200 HDMI to DVI adapter included DVI to VGA converter included

Networking

1x 1000 BaseT Ethernet

802.11g WLAN

WiFi antenna included

USB

4 USB 2.0 High Speed ports

Operating Temperature

0 to 60° C

Power Consumption

max. 8 W at full CPU load

This module is not available for the rack line.

http://www.dewesoft.com

DS-NET DS-NET

DS-NET Modules DS-NET modules are available for almost any type of input and output signals. These modules are intelligent and contain all the setup data that you can easily define within our award winning DEWESoft™ software, for maximum reliability and deterministic control. Basic calculations like minima, maxima, data reduction and filtering are done directly inside each module. All DS-NET-modules are hot swappable with auto-loading configurations and addressing, making system scalability and serviceability extremely efficient and effective – at system start up or several years down the road.

Ethernet interface

for configuration and data output Protocols: TCP/IP, UDP, PING, ASCII, Modbus TCP/IP Services: DHCP, FTP-Server, FTP-Client, e-mail-send-client (SMTP)

High data rate over Ethernet 160 kS/s total sampling speed per system example: 16 channels with 10 kS/s per channel 80 channels with 2 kS/s per channel Synchronization and clock

Master Slave principle, IRIG standard, DCF77, AFNOR etc. GPS time and position data NMEA0183, SNTP over Ethernet (all channels synchronized)

12 MB int. data buffer memory

Data buffer for block transfer of measurement data, different logger possibilities, expandable by USB device

PAC functionality

data logger, alarm handling, mathematics, numeric, boolean combinations, functions generator

Module connection

up to 16 DS NET modules via 2 UARTS, Baud rate up to 24 MBaud each UART

Galvanic isolation

all channels and power supply 500 V DC

DEWESoft™ measurement innovation

49

DS-NET DS-NET ACC2

CFB2

BR4

BR8

V8

V8-200

V4

V4-HV

TH4

TH8

DIO8

AO4

2

2

4

8

8

8

4

4

4

8

8

4

Data Rates [Hz]

10k

10k

10k

10k

10k

10k

10k

10k

10k

100

10k

10k

Isolation Voltage [V]

500

500

500

500

500

500

1.2k

1.2k

1.2k

500

500

500

Frequency Signal





Pulse Width





Counter





Time









Number of Channels

Input Types ■















±60V

±10V

±10V

±200V

±10V

±1kV

±80mV

±80mV

Current (0..25 mA)







Resistance





Potentiometer





Pt100, Pt1000





Thermocouple









Full, half and quarter bridges



Voltage max. Range



Inductive full and half bridges



LVDT



IEPE/ICP Sensors

Status















Output Signal Voltage (±10 V)





Current (4..20mA)



Frequency Signal





Pulse Width













Status











Connectors Screw





BNC DSUB

■ V8-B

BR4-D



■ V4-B

■ ■



Thermocouple

TH8-C

Spring Terminal



Special Modules CAN 2 SENSOR power supply

max. 2 modules (4 CAN channels) connected to one DS-NET CPU max. 1 module per system

Note: some measurements may require optional adapters and may not be available dependent on the connectors of the module

50

http://www.dewesoft.com

DS-NET DS-NET DS-NET-ACC2

Multiple Input Module

DS-NET-CFB2

Carrier Frequency and AC/DC Bridge Module

2 galvanically isolated universal analog input channels

voltage: ±60 V, ±10 V, ±1 V, ±100 mV current: 0..25 mA potentiometer, resistance: 100 kΩ, 4 kΩ, 400 Ω Pt100 & Pt1000: -200..850 °C thermocouple types: B, E, J, K, L, T, U, N, R, S bridge: ±2.5 mV/V, ±50 mV/V, ±500 mV/V (@ 2.5 V excitation) IEPE sensors: ±10 V; constant current 4 mA

Resolution

24 bit

Sampling rate

10 kHz per channel (thermocouple 8 Hz)

2 digital I/O channels

input: state, tare, memory reset output: state alarm, threshold voltage: max. 30 V

Signal processing

linearisation, digital filter, average, scaling, min/max storage, RMS, arithmetic, alarm

TEDS

class 1 and class 2, according IEEE 1541.4

Galvanic isolation

of I/O-signals (each channel), power supply and interface isolation voltage 500 V

2 isolated analog input channels

Strain gauge and inductive measuring bridges (full, half, quarter), LVDT, RVDT

DC and carrier frequency (CF) principle

DC excitation, 600 Hz CF excitation, 4.8 kHz CF excitation for bridges

2 analog output Resolution

DS-NET-BR4

Multiple Input Module

DS-NET-BR8

Bridge Input Module

Voltage ±10 V, 10 kHz 24 bit

2 digital I/O channels

input: state, tare, memory reset output: state, alarm, threshold

Signal processing

linearisation, digital filter, average, scaling, min/max storage, RMS, arithmetic, alarm

Galvanic isolation

of I/O-signals (each channel), power supply and interface isolation voltage 500 V

4 galvanically isolated universal analog input channels

voltage: ±10 V, ±1 V, ±100 mV current: 0..25 mA, potentiometer, resistance: 100 kΩ, 4 kΩ, 400 Ω Pt100 &Pt1000: -200..850 °C thermocouple types (not for BR4‑D): B, E, J, K, L, T, U, N, R, S bridge: ±2.5 mV/V, ±50 mV/V, ±500 mV/V (@ 2.5 V excitation)

Resolution

24 bit

Sampling rate

10 kHz per channel (thermocouple 8 Hz)

Signal processing

linearisation, digital filter, average, scaling, min/max storage, RMS, arithmetic, alarm

Galvanic isolation

of I/O-signals (each channel), power supply and interface isolation voltage 500 V

Option

D-SUB connectors model: DS-NET BR4-D Lemo 10 pin (see page 48)

8 strain gage input channels

+/-1 mV/V, +/-5 mV/V quarter bridge +/-2 mV/V, +/-10 mV/V half-, full bridge excitation: 2V, 4V selectable selectable shunt: 100 kOhm full, half and quarter bridge 3-, 4-, 5-, 6-wire

Resolution

24 bit A/D with AAF filters 1 kHz

Sampling rate

10 kHz per channel

Signal processing

linearisation, digital filter, average, scaling, min/max storage, RMS, arithmetic, alarm

Galvanic isolation

of I/O-signals (each channel), power supply and interface isolation voltage 500 V

Option

adapter for screw terminal available

DEWESoft™ measurement innovation

NE

W

51

DS-NET DS-NET DS-NET-V8

Voltage Module

DS-NET-V8-200 Voltage Module

DS-NET-V4

High Isolation Voltage Module

DS-NET-V4-HV

High Voltage Module

52

8 galvanically isolated input channels

differential voltage ±10 V , current via shunt 25 mA ( V8-SHUNT - not for V8‑B), common mode voltage: 100 V permanent

Resolution

24 bit

Sampling rate

10 kHz

2 digital I/O channels (not for V8‑B)

input: state, tare, reset output: state alarm max. 30 V

Signal processing

linearisation, digital filter, average, scaling, min/max storage, RMS, arithmetic, alarm

Galvanic isolation

of I/O-signals (each channel), power supply and interface isolation voltage 500 V

Option

BNC connectors model: DS-NET V8-B

8 galvanically isolated input channels

isolated differential input voltage ±200 V

Resolution

24 bit

Sampling rate

10 kHz

Signal processing

linearisation, digital filter, average, scaling, min/max storage, RMS, arithmetic, alarm

Galvanic isolation

of I/O-signals (each channel), power supply and interface isolation voltage 500 V

4 galvanically isolated input channels

Voltages at high potential, ranges 100 mV, 1 V, 10 V current via an external shunt

Resolution

24 bit

Sampling rate

10 kHz per channel

Signal processing

linearisation, digital filter, average, scaling, min/max storage, RMS, arithmetic, alarm

Galvanic isolation

1200 V short-term 5 kVpk

Option

BNC connectors model: V4‑B

4 galvanically isolated input channels

Voltages, range 40 V, 120 V, 400 V, 1000 V

Resolution

24 bit

Sampling rate

10 kHz per channel

Signal processing

linearisation, digital filter, average, scaling, min/max storage, RMS, arithmetic, alarm

Galvanic isolation

1200 V short-term 5 kVpk

Option

HV BNC connector cable

http://www.dewesoft.com

DS-NET DS-NET DS-NET-TH4

High Isolation Thermocouple Module

DS-NET-TH-8

Thermocouple Module

DS-NET-DIO8

Digital Input/Output Module

4 galvanically isolated input channels

for non-isolated thermocouples at high potential

Cold junction compensation

internal

Dynamic linearisation

Optimum positioning of interpolation points in selected range, types B, E, J, K, L, T, U, N, R, S programmable

Resolution

24 bit

Sampling rate

10 kHz per channel

Signal processing

digital filter, average, scaling, min/max storage, arithmetic, alarm

Galvanic isolation

1200 V short-term 5 kVpk

8 galvanically isolated input channels

thermocouples and voltages in the range of ±80 mV, common mode voltage: 100 V permanent

Cold junction compensation

DS-NET TH8-C: internal DS-NET TH8: TH8-CJC connectors available (option)

Dynamic linearisation

Optimum positioning of interpolation points in selected range, types B, E, J, K, L, T, U, N, R, S programmable

Resolution Sampling rate

24 bit

Signal processing

digital filter, average, scaling, min/max storage, arithmetic, alarm

Galvanic isolation

of I/O-signals (each channel), power supply and interface isolation voltage 500 V

Option

DS-NET TH8-C: with integrated CJC

8 digital inputs and 8 digital outputs

configurable as counter, frequency, PWM and time inputs, frequency or PWM output, state in or output

State in- and outputs

process- and host controlled, programmable threshold

Frequency inand outputs

frequency measurement up to 1 MHz (Chronos method), frequency output up to 10 kHz

Counter

forward/backward counter, quadrature counter with reference zero recognition (reset/enable), up to 1 MHz

PWM in- and outputs

measurement of duty cycle and frequency, output with variable frequency and/or duty cycle

100 Hz per channel (~8 Hz with activated mains rejection)

Time measurement

DS-NET-AO4

Analog Output Module

Galvanic isolation

of I/O-signals (group/group), power supply and interface isolation voltage 500 Veff

4 galvanically isolated analog outputs

voltage ±10 V, current 4..20 mA selectable

DAC resolution 16 bit

10 kHz sample rate

4 digital input and 4 digital output channels

configurable as 2 counter, 2 frequency, or 2 PWM inputs, 2 frequency or PWM output, state in- or output, max. 30 V

Frequency inand outputs

frequency measurement up to 1 MHz (Chronos method), frequency output up to 10 kHz

Counter

Forward/backward counter, quadrature counter with reference zero recognition (reset/enable), up to 1 MHz

PWM in- and output

measurement of duty cycle and frequency, output with variable frequency and/or duty cycle

Time measurement Outputs freely scalable Galvanic isolation

DEWESoft™ measurement innovation

of I/O-signals (each channel), power supply and interface isolation voltage 500 V

53

DS-NET DS-NET DS-NET-CAN2

DS-NET-SUPPLY / - BNC

CAN Bus Input Module

Sensor Power Supply Module

2 high speed CAN interface channels, up to 1MBit/s Isolated CAN input 500V OBDII, J1939, CAN output Supports CAN 2.0b standard Functions: send, receive, listen (silent), buffer max. 2 modules connected to one DS-NET CPU

4 galvanically isolated DC sensor supply voltages Voltages combinations

+5 V +12 V +15 V +24 V

Supplied Power

5 W per output voltage

Galvanic isolation

each voltage is galvanically isolated with 1.5 kV

the voltages can be can be con­n ected to get any possible voltage combination (e.g. 17 V, 20 V, ...)

Only one SUPPLY module can be used per DS NET system.

DS-NET WiFi WiFi Module

The DS-NET WiFi module can connect your DS-NET system to any wireless network. It can even be configured to create a WLAN network, that other computers can connect to via DHCP! Network Standard

802.11b/g/n

Data Rate

Up to 150Mbps

Interface

1x 10/100 Base-T Ethernet port

Antenna

1x RSMA connector (incl. 2DBi antenna)

Operating Modes

Access Point, Access Point Client, WISP Client, Router

DHCP

Support for client and server

Security

64/128-bit WEP, WPA, WPA2, WPS

DS-CAR-UPS-1

UPS CAR Power Supply

9 to 36 V DC Vehicle power supply

Small UPS for in cars usage 12 V car battery input (cigarette lighter), internal buffer battery 58 Wh / 4A, 12 V DC internal charging and temperature control electronics 1x 2 pin Lemo for DEWE-43A, DS-NET, 101-DISP

DS-CAR-UPS-2

UPS CAR Power Supply

DS-NET-BR4 Lemo 10 pin

NE

W

For easy and secure connection now available with Lemo 10 pin

9 to 36 V DC Vehicle power supply

Small UPS for in cars usage 12 V car battery input (cigarette lighter), internal buffer battery 108 Wh / 4A, 12 V DC internal charging and temperature control electronics 1x 2 pin Lemo for DEWE-43A, DS-NET, 101-DISP

54

http://www.dewesoft.com

DS-NET

DS-NET Optional Connector-Adapters CFB2-120

Module: CFB2

1 channel quarter bridge completion adapter 120 Ω

TH8-CJC

Module: TH8

4 channel thermocouple adapter with integrated CJC

BR4-CJC

Module: BR4

2 channel thermocouple adapter with integrated CJC

ACC2-CJC

Module: ACC2

1 channel thermocouple adapter with integrated CJC

ACC2-BNC

Module: ACC2

1 channel screw to BNC adapter: ICP ® input

CFB2-350

Module: CFB2

1 channel quarter bridge completion adapter 120 Ω

ACC2-120

Module: ACC2

1 channel quarter bridge completion adapter 120 Ω

ACC2-350

Module: ACC2

1 channel quarter bridge completion adapter 350 Ω

BR4-120

Module: BR4

2 channel quarter and half bridge completion adapter 120 Ω

BR4-350

Module: BR4

2 channel quarter and half bridge completion adapter 350 Ω

DEWESoft™ measurement innovation

BR4-D-120

Module: BR4-D

1 channel quarter and half bridge completion adapter 120 Ω

BR4-D-350

Module: BR4-D

1 channel quarter and half bridge completion adapter 350 Ω

DSUB-BNC

Module: BR4-D

1 channel DSUB9 to BNC adapter for voltage input

V8-SHUNT

Module: V8

4 channel shunt connector for current measurement (25 mA)

55

DS-NET DS-NET

DS-NET Modular Automotive Data Recorder ƒ ƒAnalog In ƒ ƒDigital / Counter In ƒ ƒCAN BUS ƒ ƒWLAN ƒ ƒGPS ƒ ƒUPS Power supply GPS for absolute time sync and position

WLAN

Remote Control (iPad via WLAN)

USB

CAN

DIGITAL

SENSORS

VGA

3G/UMTS modem

ANALOG

3G/UMTS telephone network

Dewesoft Display MOB-DISP-10

ƒ ƒUp to 128 isolated (>500 V) multi sensor inputs ƒ ƒSample rate up to 10 kS/s per channel @ 24 bit ƒ ƒWiFi (WLAN), LAN, UMTS/GSM, Internet ... modem ƒ ƒGPS for position, synchronization and absolute clock ƒ ƒDatalogging to USB stick ƒ ƒ128 GB flash HDD ƒ ƒRealtime operating system (3 sec fast startup) ƒ ƒ0° C ... +60° C operating temperature (fanless!)

OPTIONS: ƒ ƒCPU: 1.6 GHz processor ƒ ƒUp to 4 CAN ports per system (requires CPU) ƒ ƒVideo webcam ƒ ƒWindows® 7 and DEWESoft™ monitoring/analysis software ƒ ƒDASYLab® and LabVIEW™ - compatible

Fleet tests, collective tests, component tests, R&D

56

http://www.dewesoft.com

DS-NET DS-NET Dewesoft offers the new DS-NET modular data-recorder for automotive applications NEW DESIGN With the new DS-NET modular data recorders, Dewesoft offers a wide range of applications in the automotive field. The new instruments cover applications for all kinds of fleet tests, homologation tests, fatigue tests and component tests. The modular design provides flexible adaption of the instrument to the required job to do.

Rugged mechanical design The CNC aluminium construction provides maximal ruggedness and stability at low weight and fanless cooling at operating conditions from -20° to +60° C (CPU: 0° to +60° C) . No moving parts or cables are used in the internal electronics. Storage is performed on internal flash harddisk or external flash memorysticks.

Rugged electronical design All inputs are galvanically isolated with 500..1200 V from channel to channel and to chassis GND. Most of the inputs offer a wide range of sensor connections which can be choosen by software and a synchronous sampling speed of up to 10 kS/s per channel. The aggregate sampling rate per gate is 160 kS/s. Multiple gates offer an aggregate sampling rate of up to 1.28 MS/s. Due to its modularity the system offers up to 128 measurement channels including analog, digital, counters plus CAN, USB, LAN, GPS.

Multiple features DS-NET recorders can act as dynamic datarecorders which automatically start the continous acquisition when powered up or store data only on software-definable trigger conditions. It is easy to use absolute timestamps via GPS for synchronization and the data can be sent during recording via WiFi or G3/UMTS connection to a database or the control center.

Software included Every system comes with the award winning DEWESoft™ 7 measurement and analysis software. But also DASYLab® and LabVIEW™ drivers are available for additional applications.

Sensor connections A wide range of sensors can be connected to the isolated inputs such as: voltages up to 1000 V, currents 0-20 mA, 10..1000 A (clamps), thermocouple B,E,J,K,L,T,U,N,R,S (build in CJC) PT100, PT1000, strain gauge ¼ , ½ , full bridge (DC or carrier frequency excitation), IEPE, ICP sensors, resistance, potentiometers, LVDT, frequency, pulse width, encoders, counter, time, status. For hybrid applications we offer high isolation with 1200 V (5 k V pk) for voltage and thermocouple inputs. WLAN

EXPAND YOUR SYSTEM USB SYNC

8x high dynamic analog 24 bit@200 kS/s per channel 8x super-counter synchron 2x CAN high speed synchron

monitor connector 4 USB 64 GB flash HDD ATOM 1.6 GHz operates DEWESoft™ 7

DEWESoft™ measurement innovation

32 isolated universal high speed thermosensors GPS 1 Hz 2 CAN

12 isolated multi sensor inputs

8 isolated digital IN and 8 digital OUT 2 isolated multi sensor and ICP © inputs

57

DS-NET-CAN / DS-CAN2 DS-NET-CAN / DS-CAN2

DS-CAN-LOGGER ƒ ƒUp to 4 high speed CAN interfaces (isolated) ƒ ƒHardware synchronization ƒ ƒGPS (1Hz) for absolute time and position (option) ƒ ƒWebcam video (option) ƒ ƒLAN, WLAN (WiFi), UMTS/GSM, 3G, internet ….modem ƒ ƒ128 GB flash-disk, monitor connection ƒ ƒWindows® 7 (option) + DEWESoft™ CAN monitoring/analysis software ƒ ƒ0°C…+60°C operating temperature (fanless!)

Fleet-test, fatigue test, automotive monitoring, R&D, ...

GPS 1 Hz for absolute time sync and position

3G/UMTS modem

Webcamera for documentation

up to 4 high speed CAN

USB

VGA

3G/UMTS telephone network

WLAN

Remote Control (iPad via WLAN)

DEWESoft™ Display MOB-DISP-10

CAN instruments

Connect GPS antenna directly to the CAN Logger Add time and position information to the recorded CAN messages for easier post processing

VIDEO information Add a simple webcam to record video information

Isolated CAN BUS interfaces All CAN bus interfaces are optically isolated for use in noisy environments

UMTS/3G or WIFI support Transfer the data online or offline to any WLAN or UMTS/3G network to be stored in a database or online monitored

58

http://www.dewesoft.com

DS-NET-CAN / DS-CAN2 DS-CAN2

DS-NET-CAN / DS-CAN2

DS-NET-CAN2 In-vehicle measurement CAN data recording: ƒ ƒAll kinds of CAN recording, storage, analysis ƒ ƒONLINE monitoring and decoding of CAN messages ƒ ƒDisplay CAN information graphically ƒ ƒOnline mathematics and analysis of CAN channels ƒ ƒEasy dbc file decoding ƒ ƒOBDII support

DS-NET-CAN2

DEWESoft™ 7 – monitor with CAN, GPS, VIDEO

ƒ ƒFeatures DEWESoft™ X ƒ ƒGPS and video information

OBDII on CAN OBDII is a standardized vehicle interface which must be implemented in each vehicle sold in the USA, and California state environmental law requires a standard interface for measurement of basic parameters of the exhaust system. Parameters such as vehicle speed, engine RPM, throttle position and certain temperatures, can be read out of any vehicle which has the OBDII interface. DEWESoft™ only needs a CAN card and a special cable to connect to OBDII. DEWESoft™ 7 – dbc CAN File

DS-CAN2 CAN/USB Interface ƒ ƒ2 high speed CAN interfaces (isolated) ƒ ƒHW synchronization with all Dewesoft products ƒ ƒUp to 8 CAN interfaces per system ƒ ƒIncl. DEWESoft™ X-LT CAN monitoring/analysis program ƒ ƒ-40° .. +70° C operating temperature (fanless) ƒ ƒNo external power supply needed if CAN operation only

DEWESoft™ measurement innovation

59

Typical Configurations

Typical Configurations EPS and NVH test systems ƒ ƒ 8 analog inputs, multichannel (voltage, current, strain, potentiometer..) @ 5 kS/s ƒ ƒ 8 thermocouple inputs 100 S/s ƒ ƒ 8 analog inputs for NVH analysis with filters up to 200 kS/s ƒ ƒ 2 counters ƒ ƒ 1 CAN port

DS-NET system with CPU... ƒ ƒ 2 DS-NET BR4-D: 8 analog inputs, multichannel (voltage, current, strain, potentiometer..) @ 5 kS/s

... expanding with SIRIUS

SYNC

USB

ƒ ƒ 1 DS-NET TH8-C: 8 thermocouple inputs @ 100 S/s

ƒ ƒConnected via USB to the DS-NET CPU ƒ ƒ8 analog inputs for NVH analysis with filters up to 200 kS/s ƒ ƒ2 counters ƒ ƒ 1 CAN ports

DS-NET and DEWE-43A can be combined to ONE system or used separately SYSTEM 1 and 2 together ƒ ƒ 1 6 ICP inputs

ƒ ƒ 6 CAN ports (isolated)

NE

T

HE

R

ƒ ƒ 8 8 multiple inputs (strain, voltage , resistors..) ET

ETHE R N

ET

ƒ ƒ 2 4 smart - counter USB

USB

USB

Coating ƒ ƒ S ealed version for high temperature range and rugged applications ƒ ƒCondensing protection for low-temp ambient applications ƒ ƒAvailable on request

60