Scanning-Electron Microscope. Considerations in Microscopy. Magnification M Resolution

Scanning-Electron Microscope TEM SEM Scanning Electron Microscope SEM ƒ topographic contrast ƒ surface Transmission Electron Microscope TEM SEM-In...
Author: Vincent Perkins
0 downloads 2 Views 6MB Size
Scanning-Electron Microscope

TEM

SEM

Scanning Electron Microscope SEM

ƒ topographic contrast ƒ surface Transmission Electron Microscope TEM SEM-Introduction

1

SEM-Introduction

ƒ density contrast ƒ slices

2

Considerations in Microscopy ƒ Magnification M ƒ Resolution ƒ smallest separation of two points that are visible as distinct entities

ƒ high magnification without resultion is

JEOL 840

“nonsense” ƒ depth of focus or depth of field DOF ƒ damage ƒ preparation

Philips XL30 FEG

SEM-Introduction

LEO Supra 35

3

SEM-Introduction

4

History

for light: resolution in the visible ~ 200 nm with X-rays as small as 20 nm

ƒ 1878 Abbe ƒ 1932 de Broglie ƒ 1926 Busch ƒ ƒ ƒ ƒ

light diffraction limit electrons are waves can focus e’s with magnetic field 1932 Ruska TEM 1938 von Ardenne first SEM 1938 Siemens first commercial TEM 1965 first commercial SEM

SEM-Introduction

operate at large NA (NA = numerical aperture) NA = n sin(α) up to 1.4

for electrons: resultion ~ 1 Å operate at small NA NA = α ~ 10-2 -10-3 100 keV, yields λ = 0.04 Å

5

SEM-Introduction

Anatomy

Comparison light microscope vs SEM scanning optical microscope

optical microscope

SEM-Introduction

SEM

6

SEM

SEM

7

SEM-Introduction

8

Radiolarien (biogenes Sediment, hier aus Opal)

SEM-Introduction

Radiolarien (biogenes Sediment, hier aus Opal)

9

SEM-Introduction

magnified

SEM-Introduction

10

further magnified

11

SEM-Introduction

12

SEM-Introduction

13

SEM-Introduction

14

SEM components ƒ Vacuum System ƒ low-, high- and ultrahigh vacuum together

ƒ Microscope Column ƒ ƒ ƒ ƒ ƒ

Electron Gun Condenser to shape the beam Electromagnetic Lenses to focus the beam Scan Coils to deflect the beam Apertures to limit the beam

ƒ Sample Chamber ƒ Motorized Stage (x,y,z, tilt, rotation…) ƒ Detectors SEM-Introduction

15

SEM-Introduction

16

Zooming capability another example

SEM-Introduction

17

SEM-Introduction

18

SEM-Introduction

19

SEM-Introduction

20

SEM-Introduction

21

SEM-Introduction

Why Vacuum

Types of Pumps ƒ mechanical displacement

ƒ Electrons are scattered by gas molecules

ƒ rough-, rotary-, auxiliary pump

ƒ in ambient at 15 keV, mean-free path ∼ 10 cm ƒ it’s much worse for low-energy electrons ƒ need a mfp of order 1 m

ƒ oil diffusion pump ƒ diff pump

ƒ turbomolecular pump

ƒ Prevents beam induced chemical reactions ƒ Required for stable emission ƒ Required for some detectors and

ƒ turbo pump

ƒ sputter ion pump ƒ ion pum

ƒ Ti-sublimation pump ƒ cryopump

electrostatic lense

SEM-Vacuum

22

23

SEM-Vacuum

24

in an electron microscope

Rotary Pump ƒ initial pump ƒ e.g. to pump the chamber after sample change

ƒ good efficiency for high pressure ƒ 100 l/min and more

ƒ filled with special lubricant ƒ disadvantages: ƒ ƒ ƒ ƒ

SEM-Vacuum

25

vibrations! oil vapor exhausts oil vapor maintenance

SEM-Vacuum

26

Diffusion Pump ƒ invented by Gaede 1915 and Langmuir 1916 ƒ works by momentum transfer (not much to do with diffusion)

ƒ Considerations: ƒ ƒ ƒ ƒ ƒ

SEM-Vacuum

27

SEM-Vacuum

must be used together with another roughening pump very high pumping speed needs warm up and special oil, which is evaporated needs cooling, too! needs to be mounted vertically

28

Diffusion Pump Principle

Diffusion Pump Advantages

Disadvantages

ƒ ƒ ƒ ƒ ƒ

ƒ oil vapor ƒ time to heat up and cool

simple cheap no moving parts pumps also light gases tolerant with particles

down

ƒ needs cooling water ƒ can overheat ƒ if roughening pump fails, oil may escape ƒ oil may be burnt or even distributed throughout the whole system ƒ can only work vertically

SEM-Vacuum

29

SEM-Vacuum

Turbomolecular Pump

30

Turbo Pump Construction

ƒ Becker, 1957 ƒ essentially a jet turbine, i.e. works also by momentum transfer ƒ multiple stages of rotating blades (rotor) spaced between fixed blades (stator)

ƒ Considerations: ƒ must be used together with another rough pump ƒ very high pumping speed (but not for light gases)

SEM-Vacuum

31

SEM-Vacuum

32

Turbo Pump Advantages

Disadvantages

ƒ ƒ ƒ ƒ ƒ

ƒ ƒ ƒ ƒ

clean no warm up high vacuum high pumping speeds can work in different orientation ƒ modern ones have magnetic bearings and are free of oil

Sputter Ion Pump ƒ can achieve ultrahigh-vacuum ƒ no moving parts ƒ ionization of gas by electrical fields and

relatively expensive not tolerant to particles can fail catastrophically high vacuum is pure hydrogen ƒ vibrations (for older ones) ƒ requires back-up pump

SEM-Vacuum

collisions ƒ ions sputter Ti, which reacts with residual gases and burries these under a film

33

SEM-Vacuum

Ion Pump Principle

Ion Pump Advantages

Disadvantages

ƒ clean ƒ no moving parts ƒ can measure pressure

ƒ ƒ ƒ ƒ

simultaneously

SEM-Vacuum

34

35

SEM-Vacuum

need to bake out not very efficient for water low capacity gasses not permanently removed

36

Types of Electron Guns ƒ ƒ ƒ ƒ

Ideal Emitter Characteristics ƒ ƒ ƒ ƒ ƒ ƒ ƒ ƒ ƒ

tungsten hairpin filament (thermionic emission) LaB6 cold field emission Schottky field emitter

SEM-Emitters

high current, i.e. low workfunction high melting point (stability), e.g. tungsten low outgasing chemically stable small source size (spatial coherence) high brightness “monochromatic” (temporal coherence) “infinite” life time “cheap”

37

SEM-Emitters

38

39

SEM-Emitters

40

Thermionic Emission ƒ current is passed through the emitter in order to heat it ƒ there is typically a large angular spread in emission and a broad distribution in energy ƒ have to shape the beam and stabilize it through the “Wehnelt Cylinder” (grid gap) ƒ grid gap reduces space-charge and increases the brightness (beam shaping)

SEM-Emitters

Wehnelt Cylinder

Crossover and Aperture works at 10-5 mbar

cross-over of source is demagnified in the lenses following below W-hairpin

Wehnelt cylinder

the angle α is called aperture (though we call the limitting pin-hole the aperture)

SEM-Emitters

41

SEM-Emitters

42

LaB6 Emitter Advantages

Disadvantages

ƒ brighter than W hairpin ƒ smaller source size ƒ energy spread is lower

ƒ ƒ ƒ ƒ ƒ ƒ ƒ

SEM-Emitters

needs < 10-6 mbar

reactive when hot higher vacuum required difficult to fabricate brittle must heat slowly must heat indirectly more expensive

43

SEM-Emitters

44

Field Emission

Field Emitter

ƒ proposed in 1954, achieved in 1966 ƒ needs UHV

ƒ single rystal wire of W, etched ƒ tip diameter 10-100 nm

ƒ large electrical field required, i.e. yielding 107 V/cm

SEM-Emitters

45

SEM-Emitters

Field Emitter

46

Electron Lenses

Advantages

Disadvantages

Electrostatic ƒ used in gun (only, few

ƒ cold source (no heating) ƒ very low spread ƒ very small “virtual” source

ƒ UHV (the higher the

exceptions) ƒ small ƒ fast response (e.g. beam blanker) ƒ it is conveniently uses in FIB’s

size ƒ very bright ƒ yields a large DOF

better) ƒ current is not very stable ƒ needs to be “flashed”

Electromagnetic ƒ solenoid ƒ high magnification possible ƒ needs cooling ƒ f∝H∝I ƒ rotates the image (need to be corrected)

ƒ there exists only converging lenses ƒ one cannot correct lens aberrations with compound lenses SEM-Emitters

47

SEM-Lens

48

Magnetic Lens (principle)

Einzellinse (electrostatic)

Davisson and Calbick 1931 Knoll 1932

SEM-Lens

49

SEM-Lens

50

Condenser and Objective Lens

SEM-Lens

51

condenser

objective

ƒ first lens (often two

ƒ large demagnification,

condensers) ƒ used to control the source size and beam current

e.g. short focal length ƒ two sets of deflection coils are included ƒ stigmator is built in too ƒ objective aperture (at the end)

SEM-Lens

52

Lens Aberrations ƒ ƒ ƒ ƒ ƒ

source size d0 (actual demagnified spot) spherical aberration (distance from axis) chromatic aberration (energy spread) diffraction (wavelength) astigmatism

astigmatism can be corrected (stigmator) aperture α is small in EM’s (unlike optical microscopes) at 10-30 keV (high voltage for an SEM), the limiting resolution is set by the source size and by spherical aberrations at 0.1-1 keV (low voltage for an SEM), the limiting resolution is set by diffraction and chromatic aberration

SEM-Lens

53

SEM-Lens

Linsenfehler

d 0 = C0 / α d S = C Sα 3 / 2

α ∝ D/B d C = CC

smallest spot (resolution limit)

where D is the pin-hole diameter and B the brightness

∆E α E

d d = 0.6λ / α

at high energy, C0 and CS dominate and the optimal spot size is given by:

SEM-Lens

54

(

d optimal ≈ C0 CS 3

)

1/ 4

55

SEM-Lens

56

Aperture (diamater) ƒ one or more used ƒ reduces the beam current ƒ lowers the angular spread and hence spherical aberration ƒ if too small, there is a lot of noise and the resolution is bad, because the virtual source size is increased ƒ must find a compromise!

SEM-Lens

57

SEM-Lens

58

same field of view with corrceted astigmatism

SEM-Lens

59

SEM-Lens

60

Depth of Field / Resolution ƒ Distance above and below plane of focus which appears to be in focus ƒ EM much larger than optical microscope ƒ DOF ∝ 1/α interrelated factors ƒ aperture ƒ working distance ƒ DOF

SEM-Parameters

ƒ resolution ƒ magnification

61

SEM-Parameters

62

Resolution vs WD

SEM-Parameters

63

small WD

large WD

ƒ ƒ ƒ ƒ

ƒ ƒ ƒ ƒ

small spot size large α small DOF high resolution

SEM-Parameters

large spot size small α large DOF low resolution

64

aperture 120 µm, too large !

SEM-Parameters

65

aperture 30 µm, good !

SEM-Parameters

high DOF due to small aperture of 10 µm

low DOF due to large aperture of 120 µm

SEM-Parameters

66

67

SEM-Parameters

68

high DOF due to small aperture of 10 µm

Signals and Interactions Electrons ƒ secondary (low energy) ƒ backscattered (high energy) ƒ transmitted ƒ Auger electrons (UHV) ƒ beam current Photons ƒ X-rays ƒ Cathodoluminescence

SEM-Parameters

69

SEM Advanced Signals and Interactions

Contrast

Everhart-Thornley

ƒ topographic ƒ compositional (elements) ƒ potential ƒ adsorbates (low energy) ƒ recombination radiation efficiency ƒ conductivity ƒ crystallographic

SEM Advanced Signals and Interactions

70

grid (usually at positive potential to attract the electrons)

71

SEM-Basic Detectors

72

ET-Detector

SEM-Basic Detectors

Origin of SE electrons

73

SEM-Basic Detectors

74

75

SEM Advanced Signals and Interactions

76

SE contrast with standard Everhart-Thornley detector

polished steel spheres from the workshop

in-lens contrast

SEM-Basic Detectors

Emission Zones

SEM Advanced Signals and Interactions

77

Backscattered vs Secondary esecondary electrons ƒ high resolution ƒ strongly topography sensitive ƒ little element sensitive ƒ sensitive to charging problems

SEM-Basic Detectors

78

Backscattered vs Secondary e-

backscattered electrons ƒ lower resolution ƒ atomic number contrast, in particular strong signal for heavy atoms ƒ less sensitive to charging problems

SE image

BSE image

a solder

SEM Advanced Signals and Interactions

79

SEM Advanced Signals and Interactions

80

Excitation Volume

flat polished surface which one is SE and BE ?

ƒ Electron penetration volume and depth depends on: ƒ Electron beam energy

• •

more energy, deeper penetration more energy, less surface sensitive

ƒ Atomic number of specimen



higher Z, lower pepentration and smaller volume

origin of topographic contrast in SE detection

SEM Advanced Signals and Interactions

81

SEM Advanced Signals and Interactions

82

SEM Advanced Signals and Interactions

83

SEM Advanced Signals and Interactions

84

polished steel surface at 5 keV

SEM Advanced Signals and Interactions

85

SEM Advanced Signals and Interactions

polished steel surface at 1 keV

86

X-ray Detection ƒ X-ray intensity ∝ Z ƒ wavlength dispersive detector ƒ energy dispersive detector

SEM Advanced Signals and Interactions

87

SEM Advanced Signals and Interactions

88

SEM Advanced Signals and Interactions

89

SEM Advanced Signals and Interactions

90

91

SEM Advanced Signals and Interactions

92

Wavelength Dispersive System (WDS)

SEM Advanced Signals and Interactions

SEM Advanced Signals and Interactions

93

Potential Contrast on a Chip

SEM Advanced Signals and Interactions

94

Crystallographic Contrast Electron Channeling Contrast Imaging (ECCI)

flat surface, small beam divergence BSE detector

SEM Advanced Signals and Interactions

95

SEM Advanced Signals and Interactions

96

Cathodoluminescence

SEM Advanced Signals and Interactions

Cathodoluminescence

97

SEM Advanced Signals and Interactions

Signal Processing

Dynamic focusing & Tilt correction ƒ When sample is tilted a very large DOF would

ƒ signal averaging (by pixels, lines and ƒ ƒ ƒ ƒ

98

be required to maintain the entire sample in focus ƒ Dynamic focusing corrects for this

frames) dynamic focusing tilt correction gamma control etc.

ƒ Tilting the sample results in distortion ƒ Correct the picture accordingly so that length are proper SEM-Lens

99

SEM-Lens

100

Tilt Correction Example

SEM-Lens

large WD and small aperture Î small beam current

101

SEM-Lens

102

large WD and small aperture Î small beam current need to integrate over long time

50 µm SEM-Lens

103

SEM-Lens

104

Preparation and Problems ƒ Preparation is an art, all kind of artefacts are possible ƒ charging ƒ contamination ƒ damage

SEM-Lens

105

10 kV

SEM-Basic Detectors

SEM-Basic Detectors

106

10 kV

107

SEM-Basic Detectors

108

2 kV

SEM-Basic Detectors

0.5 kV

109

SEM-Basic Detectors

110

Charging Effects

0.5 kV and twice WD

ƒ deflection of SE’s ƒ change in emission pattern ƒ periodic burst ƒ moving images ƒ jumps ƒ depends on beam energy

SEM-Basic Detectors

111

SEM-Basic Detectors

112

Metal Coating ƒ sputter coating ƒ Ar ions strike target ƒ quite homogeneous even if not in line of sight ƒ good adhesion

ƒ thermal evaporation ƒ target is heated ƒ shadow effect ƒ “cleaner”, but more grainy

SEM-Basic Detectors

Typical SEM sputter coater

ƒ often C, Au, AuPd ƒ artefacts in pictures: ƒ thermal damage ƒ etching damage ƒ roughness and grain of added film is imaged

113

SEM-Basic Detectors

114

Sample Preparation (organic) ƒ Fixation ƒ preserve fine structure by e.g. cross-linking

• formalin, chromic acid, glutaraldehyde, paraformaldehyde, Acrolein, OsO4

ƒ Dehydration

ƒ do it before and not in the column of the SEM:

• air drying (not prefered, because of surface tension) • ethanol, aceton, ether, chloroform… • freeze drying and critical-point drying ƒ Mounting ƒ Coating SEM-Basic Detectors

115

SEM-Basic Detectors

116

Critical Point Drying ƒ no heat of vaporization or rapid ƒ ƒ ƒ ƒ ƒ ƒ

SEM-Basic Detectors

117

SEM-Basic Detectors

Phase Diagram

SEM-Basic Detectors

density change no surface tension fix samples first dehydrate in e.g. ethanol mix with liquid transitional fluid pressurize move around the critical point

118

Transitional Fluids

119

SEM-Basic Detectors

120

CPD Maschine

SEM-Basic Detectors

121

Suggest Documents