Design and Implementation of Sensing Shirt for Ambulatory Cardiopulmonary Monitoring

Journal of Medical and Biological Engineering, 31(3): 207-216 207 Design and Implementation of Sensing Shirt for Ambulatory Cardiopulmonary Monitori...
1 downloads 0 Views 1MB Size
Journal of Medical and Biological Engineering, 31(3): 207-216

207

Design and Implementation of Sensing Shirt for Ambulatory Cardiopulmonary Monitoring Zheng-Bo Zhang1

Yu-Hong Shen2

Bu-Qing Wang1 1

Wei-Dong Wang1,*

Jie-Wen Zheng3

Department of Biomedical Engineering, Chinese People’s Liberation Army General Hospital, Beijing 100853, China 2 The Quartermaster Research Institute, Beijing 100010, China 3 Institute of Medical Equipment, National Biological Protection Engineering Center, Tianjin 300161, China Received 31 May 2010; Accepted 11 Oct 2010; doi: 10.5405/jmbe.801

Abstract Ambulatory monitoring of cardiopulmonary parameters during wake, sleep, and activity states can expand our understanding of health and diseases. In this paper, we present a wearable cardiopulmonary monitoring system called the „Sensing Shirt‟. The Sensing Shirt consists of a T-shirt with sensors integrated for physiological monitoring, a data acquisition unit (DAQU), and a set of PC-based software. A number of vital signs such as electrocardiogram (ECG), rib cage (RC) and abdominal (AB) respiration, photoplethysmogram (PPG), SpO2, and posture/activities can be acquired by the shirt. The physiological data are stored in a microSD memory card, and then analyzed offline to extract parameters such as heart rate (HR), breathing rate, SpO2, tidal volume (TV), pulse wave transit time (PWTT), and respiratory sinus arrhythmia (RSA). In the Sensing Shirt system, the ECG is acquired by means of active electrodes. PPG and ECG are sampled by one high-speed 16-bit, analog-to-digital converter at 500 Hz for precise PWTT calculation. Technique of pulse amplitude modulation is used for signal conditioning both in respiratory inductive plethysmography and PPG circuits to reduce the power consumption. Powered by 2 AA batteries with 1600 mAh capacity, the whole system can work more than 48 h continuously without interruption. Basic performance tests demonstrate that prototypes of this system can capture ECG, RC and AB respiration, and PPG signals effectively, and extract parameters such as HR, breathing rate, SpO2, and TV accurately. This advanced ambulatory monitoring system may be used in both home healthcare and scientific research. Keywords: Wearable, Cardiopulmonary monitoring, Pulse amplitude modulation, Active electrode, Respiratory inductive plethysmography, Pulse wave transit time

1. Introduction Continuous monitoring of physiological and physical parameters is necessary for the assessment and management of personal health status [1]. It contributes significantly to the reduction of healthcare cost and ensuring those who need urgent care get it sooner. In recent years, there has been an increasing interest in wearable physiological monitoring systems, both in research and commercial areas. Many novel designs for wearable sensors and systems have been developed, such as AMON [2], SenseWear [3], LifeShirt [4], Smart Shirt [5], and Smart Vest [6]. Wearable health monitoring systems employ wearable sensors to acquire vital signs from human bodies such as heart rate (HR), respiration, body temperature, and blood pressure. Wrist, belt and smart clothes are * Corresponding author: Wei-Dong Wang Tel: +86-1-066-936921; Fax: +86-1-066-937921 E-mail: [email protected]; [email protected]

frequently used for placement of biomedical sensors and devices for ambulatory use. The field of smart textile-based wearable biomedical systems has been generating a lot of interest in the research and business communities [7,8], with the aims of getting the technology into the marketplace. Cardiopulmonary function monitoring is an essential part for a health-monitoring system. Chronic diseases such as chronic heart failure, atrial fibrillation, pulmonary emphysema, and chronic obstructive pulmonary disease (COPD) represent a considerable workload in primary care. There is also interest as to whether it would be feasible to conduct sleep disorder testing at home. LifeShirt is a currently commercially available monitoring system capable of cardiopulmonary physiological data capture [9,10]. The key technology of this system is its patented respiratory inductive plethysmography (RIP). A detailed quantification of volume, timing and shape parameters in the respiratory pattern waveform can be derived from RIP technology. LifeShirt opened a new era for comprehensive monitoring of cardiopulmonary physiology, but this system

208

J. Med. Biol. Eng., Vol. 31 No. 3 2011

acquires electrocardiogram (ECG) with the current passive Ag-AgCl electrodes, which has the disadvantages of baseline wander, power line interference, and movement artifact noise during ambulatory use [11]. As the pulse oximeter is integrated in the system by the serial port, it is hard to synchronize the photoplethysmogram (PPG) data received from the series port with the LifeShirt‟s intrinsic ECG signal to derive an accurate pulse wave transit time (PWTT), which is a useful parameter both for sleep apnea detection and continuous blood pressure estimation. Furthermore, as Life Shirt is a PDA-based data acquisition system, it is not convenient for users to carry this device in ambulatory condition. Battery life and cost of this system are also problems. In this paper, we present a microcontroller-based wearable, cardiopulmonary monitoring system called Sensing Shirt, which incorporates our research on wearable technology such as active electrodes for ambulatory ECG sensing, technique of pulse amplitude modulation (PAM) for RIP [12], and low-power PPG. The whole system consists of a T-shirt with sensors integrated, a data acquisition unit (DAQU), and PC-based software for data processing and display. The system integrates sensors and cables into an elastic, lightweight shirt, and sensors are distributed on several specific body locations to accurately measure the vital parameters. The DAQU is placed in a little pocket on the outer surface of the shirt, and a removable microSD card is incorporated in the DAQU for data recording. Vital parameters such as single-lead ECG, rib cage (RC) and abdominal (AB) respiration, tidal volume (TV), PPG, HR, SpO2, PWTT, respiratory sinus arrhythmia (RSA), and posture/activities can be acquired by this system. In the Sensing Shirt system, the PPG signal is sampled along with other cardiopulmonary parameters by one high-speed, 16-bit ADC. Synchronized AD conversion between PPG and ECG signals contributes to more accurate PWTT calculation. As the power consumption is a crucial factor for a wearable system powered by batteries, the technique of PAM is used both for RIP and PPG signal conditioning, with the microcontroller generating control signals, to reduce the power consumption dramatically in the analog section. Powered by two AA alkaline batteries with 1600 mAh capacity, the Sensing Shirt system can work continuously more than 48 h without interruption. Basic performance tests show that prototypes of this system can capture ECG, RC and AB respiration, and PPG signals effectively, and extract parameters such as HR, breathing rate, SpO2, and TV accurately.

2.1 Sensing shirt Basic cardiopulmonary parameters such as ECG, RC and AB respiration, PPG, and SpO2 can be acquired by the shirt. Figure 1 illustrates the Sensing Shirt with sensors integrated for collecting multiple cardiopulmonary parameters. A 3-axis accelerometer is embedded in the DAQU to record body posture and activities. ECG button electrodes and RIP inductive wires are integrated in the fabric of the Sensing Shirt and connected to the DAQU by wires, which are also stitched into the fabric. These sensors are placed at specific body locations and integrated into the shirt in a manner which makes them stay in good contact with the user‟s body. RIP sensors are woven into the shirt around the user‟s chest and abdomen. Two parallel sinusoidal arrays of insulated wires embedded in elastic bands are woven into the shirt. For each RIP sensing band, tightening devices of Velcro tape for individual adjustment of tightness of the band is equipped to prevent its longitudinal movement over the surface of the torso of body. Considering the convenience of handling and comfort of wearing, the ECG button electrodes are embedded at special locations (RA, RL, LL, as shown in Fig. 1) in the inner surface of the shirt, and wires are stitched into the shirt. Comfortable materials blending cotton and knitted lycra were chosen to fabricate the shirt. Wearable sensors integrated in the shirt are attached via one cable tree and a secure connector to the DAQU. That makes it easy for the ordinary user to plug and play, and avoid operation errors. PPG sensor placed at the index finger/ear lobe is plugged into the DAQU through another different cable and secure connector.

Figure 1. T-shirt with sensors integrated for wearable cardiopulmonary parameter monitoring.

2. Materials and methods 2.2 Data acquisition unit The Sensing Shirt system consists of three parts: a comfortable T-shirt with embedded sensors to capture the physiological signals, a DAQU for data acquisition and SpO2 calculation, and a set of PC-based software for offline parameter extraction. All data sampled is stored on a microSD memory card and transferred to PC with a USB card reader. Data analysis and display are performed on a PC with software developed on the LabVIEW 8.2 software platform (National Instruments, Austin, TX).

The DAQU is light-weight so as to be placed into a little pocket on the outer surface of the shirt, as shown in Fig. 1. Figure 2 illustrates the block diagram of DAQU, which includes the complete electronic components for signal conditioning, digitization, processing, and recording. The DAQU is designed to keep its size and power consumption to a minimum. A commercial microSD memory card with the capacity of 1G was chosen as the storage medium for its tiny dimension. An ARM Cortex-M0 based, low-power 32-bit

Sensing Shirt for Ambulatory Monitoring

Active ECG Electrodes

2*AA Batteries

Battery Monitor Circuit

DC-DC Converter (MAX1722)

3.4V

LDO Regulator (CAT6217-3.3) LDO Regulator (CAT6217-3.3)

Differential Amplifier

Band-pass Filter

Amplifier &LP

3.3V 3.3V

209

Analog Circuit Digital Circuit

ECG

LED Indication

X;Y;Z

Low-pass Filter

Triaxial Accelerometer

GPIO

16-bit ADC

RIP Sensors

Exciting in turn

Signal Conditioning

RC&AB

PPG Sensor

Exciting in turn

Signal Conditioning

PPG1&PPG2

PAM for RIP

SSP

32-bit ARM Cortex-M0

(LPC1114) SPI

microSD Card

PAM for PPG

CMOS Analog Switch

ARM GPIO

8-channel Analog Multiplexers

ARM GPIO

Figure 2. Block diagram of DAQU.

MCU LPC1114 (NXP Semiconductors N.V., Eindhoven, The Netherlands) performs the AD conversion, control signal generation for PAM used in RIP and PPG analogy circuit, signal processing for SpO2 calculation, and data storage in the microSD card. An embedded real-time operating system (RTOS) UCOS-Ⅱ (version 2.86) runs on the ARM core for multiple tasks scheduling. Multiple analog signals are sampled by a 16-bit ADC ADS8320 (Texas Instruments, MN, USA) through the synchronous serial interface with the MCU. Eight-channel analog multiplexers ADG658 (Analog Devices, Inc., Mass., USA) act as channel selection controlled by the MCU. The system is powered by two AA batteries (LR6). The voltage of the battery is raised to 3.4 V by a step-up DC-DC converter MAX1722, and two LDO chips CAT6217-3.3 are used to generate two separate 3.3-V DC supplies feeding to analog and digital circuits, respectively. 2.3 Sensor unit All analog signal conditioning circuits are on one PCB board placed in the DAQU. Analog signals from the sensors are conditioned in the DAQU to levels suitable for digitization and processing. Though some special ECG sensors, such as silicon rubber electrodes with pure silver fillings, fabric electrodes, filled gold electrodes, have been developed [13], these kinds of electrodes are prone to contact impedance variation, baseline wander, and movement artifact noise when the user is in motion. In the Sensing Shirt system, conventional Ag-AgCl electrodes are used for ECG sensing for their high reliability. In order to improve the ECG signal quality for ambulatory use, active electrodes are incorporated in the Sensing Shirt. An active electrode with its biopotential amplifier embedded in the ECG button electrodes, as shown in Fig. 1, performs the signal amplification as close to the transducer as possible, contributing to a high-quality ECG

signal by its ultra-high input impedance and low output impedance. In our system, the gain of the amplifier circuit of the active electrode is set to one, and it actually acts as an emitter follower. The system can acquire one-lead ECG (standard lead-II) using three Ag-AgCl electrodes. On the analogy circuit board in the DAQU, a differential preamplifier with a gain of 8, a high-pass filter with 0.5 Hz cutoff frequency, a fourth-order Butterworth low-pass filter with 100-Hz cutoff frequency, and a main amplifier with a gain of 125 are used to acquire clear ECG signal. As the shirt is a battery-powered system and active electrodes have been used, 50-Hz notch filter in hardware is omitted. Digital signal processing technology is used in the PC-based software for digital filtering, scaling, etc. The ECG signal is sampled at 500 Hz. The HR is measured from the ECG signal, and R-waves are detected with 2-millisecond resolution. RIP sensors are integrated in the Sensing Shirt for respiration monitoring. Comparing to impedance pneumography, inductive plethysmography has more advantages, including greater accuracy, better sensitivity and better security for patients. The Sensing Shirt system incorporates our patented RIP with PAM technique to detect respiration with high signal quality and ultra-low power consumption [12]. RIP sensors are excited by one high constant-current source momentarily and periodically, and in a very short time, the respiration signals are detected and sampled. The RC and AB sensors are excited consecutively under the control of sequence signal to avoid crosstalk between each other. A low-voltage and low-on-resistance CMOS analog switch is employed for channel selection controlled by a digital gating signal applied to its control gate. In order to improve the integration and reduce the system power consumption, control signals for PAM are generated by the MCU through its general-purpose input and output ports

210

J. Med. Biol. Eng., Vol. 31 No. 3 2011

(GPIO). Thus, some digital components used by the original RIP module are omitted. In the Sensing Shirt system, a constant-current source generates a 15 mA peak-to-peak sinusoidal current at 400 kHz, and each sensor is switched on and off at a frequency of 50 Hz, with an excitation duration of 100 μs. Though the momentary current through the sensor is very high for improvement of signal-to-noise ratio (SNR) and sensitivity, its average current actually is only 1/10,000 of 15 mA. Thus, PAM can dramatically reduce the power consumption and still keep a high SNR. RIP signal is direct-current-coupled output and sampled by the ADS8320 at the end of each excitation. Thus, the sampling rate of each breathing channel is 50 Hz and the respiratory pattern waveform can be accurately acquired. The PPG sensor is placed at the index finger/ear lobe with the red light source at 660 nm and infrared light source at 940 nm. Red wavelength of 660 nm has higher absorption for reduced hemoglobin compared to oxyhemoglobin. Infrared 940 nm shows higher absorption for oxyhemoglobin. Two wavelength PPG signals are sampled and SpO2 is calculated in real time in the MCU based on Lambert-Beer Law [14]. As average change on PWTT derived by calculating the time intervals from ECG to finger PPG signals, pulse by pulse, is only about 7~23ms [15], synchronized and high speed of sampling on ECG and PPG signals is crucial for optimal determination of PWTT. In the Sensing Shirt system, ECG and PPG signals are sampled by the high-speed ADC ADS8320 at a sampling rate of 500 Hz to give a 2-millisecond resolution. Just as in the abovementioned RIP technology, PAM is also used in the PPG analog section to reduce the power consumption, and all control signals for PAM are generated by the MCU through its GPIO port. The red and infrared lights are turned on and off consecutively, at a frequency of 500 Hz. At the end of each excitation, PPG signals are sampled. A programmable gain amplifier is used for gain modification, controlled by the MCU to compensate for the individual users‟ differences in PPG amplitude. The accelerometer has been proposed as being suitable for monitoring human movements and has particular applicability to the monitoring of free-living subjects. A 3-axis accelerometer MMA7260 (Freescale Inc., TX, USA) is integrated in the DAQRU to capture posture and motility information. The MMA7260 features signal conditioning, low-pass filter, temperature compensation, and high sensitivity (800mV/[email protected]). Signals from the MMA7260 of each channel are sampled at a frequency of 20 Hz. In the PC-based software, these signals are processed further and integrated acceleration signal is used to detect periods of physical activity and rest. The posture signal can be used to adjust RC and AB channel gains of RIP for more accurate ambulatory TV estimations. 2.4 Signal processing Raw waveform signals stored on the microSD card are processed using software that runs on a PC. Data are subjected to automatic analyses including calibration, artifact screening, parameter extraction, making trends, and generating reports. The HR is calculated by analyzing the ECG waveform to

determine the QRS complex beat by beat. Studies on automatic detection of QRS complex have been carried out for many years [16]. We detect the QRS complex by the first-derivative and threshold-comparison approach. Firstly, the power-line interference is removed by a subtraction procedure [17], if necessary. The procedure is simple and has good performance in coping with changes in amplitude of the power-line interference. Then the QRS complex is detected by a band-pass filter followed by a threshold comparison algorithm. Finally the R-wave position is located and saved, and the RR interval is calculated. As the sampling rate of the ECG signal is 500 Hz, it is easy to calculate the HR as follows: HR = 60/(RR interval / 500). When correctly calibrated, RIP allows the measurements of volume and time components of breathing cycles as well as the relative contribution of RC and AB. Some calibration techniques, based on the two-compartment model of the respiratory system model by Konno and Mead [18], have been developed, as shown in the following relationship: △Va = △Vrc + △Vab

(1)

The change of volume measured at the mouth (△Va) is the sum of an RC (△Vrc) and an AB (△Vab) contribution. These volume changes are obtained indirectly by measuring the variations of two representative thoracic and abdominal dimensions by RIP. Equation 1 can be rewritten as: △Va = △RC + △AB

(2)

where △RC and △AB are changes (△) in RIP ribcage and abdomen signals.  and  are calibration coefficients. Typically, RIP calibration involves collecting data to ascertain  and . The △Va, RC excursion (△RC), and AB excursion (△AB) are simultaneously recorded. These values are then used to set RC and AB channel amplifier gains (coefficients  and ) for implementing equation (2) during ventilation monitoring. After calibration procedure, the breath-by-breath TV can be estimated according to equation (2) by RC and △AB. Least squares method (LSQ) calibration [19] and qualitative diagnostic calibration (QDC) procedure [20] are commonly used for RIP calibration. We utilize the LSQ method for more accurate ambulatory TV estimations and the QDC method for relative TV estimations during sleeping periods. QDC calibration procedure does not require separate calibrations for upright and supine position of the torso, and can be finished during a 5-min quiet period of breathing. This calibration method is preferred during sleeping periods. The calculation of SpO2 is performed in real time in the MCU, and the SpO2 values are stored with other physiological parameters simultaneously. The approach of first-derivative and threshold-comparison is used to process each PPG signal to determine the occurrence of every pulse [21]. Then the peak-to-peak value of each PPG signal, pulse by pulse, is calculated by the peak detection method of finding the local maxima and minima of each pulse. After calibration, the normalized ratio between two wavelengths is used to predict SpO2 values by checking the conversion table, which converts the ratio of normalized absorbance into percent SpO2.

Sensing Shirt for Ambulatory Monitoring

2.5 Performance test A performance test was completed to study the consistency and reliability of the physiological data acquired by the Sensing Shirt. For ECG signal, the effectiveness of QRS detection algorithm was firstly evaluated. ECG signals from the MIT/BIH arrhythmia database were chosen as the testing input to the off-line analysis software for evaluation. We extracted ECG segment of 5 minutes‟ duration randomly from each file in the database for test. Then we used a patient simulator (Metrology Services, DNI Nevada ECG 200, USA) to generate typical 1-mV peak-to-peak normal ECG signals to the DAQRU through three Ag-AgCl electrodes. HR was changed manually from 30 to 240 BPM (beat per minute). Signals were recorded in the microSD memory card and an off-line analysis was performed to test the reliability and accuracy. Because the RIP measures the changes of cross-section area, there is no standardized apparatus to test its performance in clinical environment. We developed a relatively standard method for Sensing Shirt‟s RIP testing. A medical ventilator (Siemens, Servo-i, Germany) and a lung simulator (LS-2000® , BC Biomedical) were used to make the changes of cross-section area both in RC and AB positions. The ventilator mechanically pumped air into and out of the lung simulator, which was placed under the shirt. Constant 300-ml TV was generated by the ventilator periodically, so the volume of lung simulator changed under the control of the ventilator to generate a reproducible signal to the RIP sensors. Different breathing rates were set manually through the operation menu on the ventilator‟s screen. Data were recorded in the microSD memory card for an off-line breathing rate calculation. We used a SpO2 analyzer (Fluke Biomedical, daeg, Norway) to verify the accuracy of SpO2 measurement. A conversion table which converts the ratio of normalized absorbance into percent SpO2 was obtained from the simulator. After the preliminary check by the patient simulator, we carried out two human body experiments for ECG and RIP evaluation, respectively, with the standard measurement methods as reference. The accuracy of TV estimation by the Sensing Shirt was firstly evaluated. The experiment was conducted on 15 healthy male subjects aged from 20 to 40 from whom informed consent had been obtained. The validation of the device was carried out for about 30 min on each subject in supine position. RC and AB excursion and respiratory flow were simultaneously recorded by the Sensing Shirt and the ventilator tester (imtmedical, FlowAnalyser™ PF-300, Switzerland) respectively, with the subject wearing a face mask sealed around the nose and mouth. QDC calibration procedure was firstly used to determine the calibration coefficients for TV measurement, and then breath-by-breath TV was compared between the two systems. The absolute percentage error, correlation coefficient, and linear regression coefficient on breath-by-breath TV between the two systems were calculated to estimate the performance of the Sensing Shirt. We carried out treadmill exercise and cycle ergometer exercise test for dynamic ECG evaluation, with the American Thoracic Society/American College of Chest Physicians

211

recommendation for reference [22]. Ten healthy men (no known respiratory or cardiac disease) aged from 22 to 32 participated in this test, having given prior consent. The raw ECG data were recorded by the Sensing Shirt, and the beat-by-beat HR was stored simultaneously by an HR monitor (Polar Sport Tester, RS800, Finland) during two sequential conditions. Firstly, the participants completed an incremental treadmill test, which started in stationary standing state (0 mph) and continued in walking state (4 mph), jogging state (6 mph), and fast running state (8 mph). Each state lasted for 2 minutes. Secondly, they completed an incremental cycle ergometer exercise, which started in stationary sitting state on the stationary bicycle (0 rpm) and continued in slow pedaling state (20 rpm), medial pedaling state (30 rpm), and fast pedaling state (40 rpm). Each state lasted for 2 minutes. In the section of data processing, for each exercise state, the raw ECG data captured by the Sensing Shirt was firstly processed to derive HR, and then the mean HR of the two systems was calculated. The mean HR was analyzed statistically according to the method described by Bland–Altman [23] and plotted as the difference in measurements against the mean of the two measurements. The average of the differences between the two methods was used to calculate the overall bias and the standard deviation (SD) of the differences between the methods was used to calculate the precision. The bias and precision were analyzed to estimate the performance of the device.

3. Results The Sensing Shirt has been developed and tested for functionality and comfort levels for users. The DAQU weighs 104 g, with the three dimensions of 102 mm × 64 mm × 20 mm. Figure 3 illustrates the physiological data captured by Sensing Shirt during an incremental treadmill exercise. Figure 4 illustrates the screen shot of the off-line data analysis software with waveform of ECG, RC, AB, PPG, SpO2 and motions.

(a) Stationary standing stage

(b) Walking stage

(c) Jogging stage

(d) Fast running stage

Figure 3. Physiological data captured by the shirt during an incremental treadmill exercise.

J. Med. Biol. Eng., Vol. 31 No. 3 2011

212

Figure 4. Screen shot of the off-line data analysis software with waveform of ECG, RC, AB, PPG, SpO2 and motions.

Tables 1 and 2 show the test results of heat rate and respiratory rate evaluation, respectively. Table 3 shows the test results for the QRS detection algorithm, which produces 6 false-positive detections (FP), 4 false-negative detections (FN), and 7303 true-positive detections (TP) in the total 7407 QRS complexes (QT). This represented sensitivity (Se) of 99.95% and positive predictive accuracy (Ppa) of 99.92%. When SpO2 was predicted with the simulator probe in the range between 35% and 100%, the correlation coefficient was 0.999 and the SD of percent SpO2 was 0.15. The maximal absolute difference between the SpO2 module and the simulator during the whole test was 1%. For TV measurement comparison, sequential TV collected from the Sensing Shirt and PF-300 were aligned in adjacent columns of a data spreadsheet for each participant for evaluation. Of totally 11,437 validation breaths, 93.85% (10,734 breaths) of the breath-by-breath TV measured by Table 1. Results of heart rate test. Patient simulator 30 BPM (ECG 200) Sensing 30.0 ± 0 shirt

60 BPM

120 BPM

180 BPM

240 BPM

60.0 ± 0

120.0 ± 0.1 179.1 ± 0.3 240.0 ± 0.2

Table 3. Results of QRS detection with ECG signals from the MIT/BIH Arrhythmia Database. File name 100.dat 101.dat 102.dat 103.dat 104.dat 105.dat 107.dat 109.dat 112.dat 113.dat 115.dat 116.dat 117.dat 118.dat 119.dat 201.dat 202.dat 121.dat 122.dat 123.dat 124.dat

No. of QRS 371 340 366 354 377 418 352 432 428 289 316 395 251 362 326 440 265 303 421 249 252

No. of correct No. of false detection positives 371 0 338 1 366 1 354 0 375 3 418 1 352 0 432 0 428 0 289 0 316 0 395 0 251 0 362 0 326 0 440 0 265 0 303 0 421 0 249 0 252 0

% QRS detected 100% 99.4% 100% 100% 99.5% 100% 100% 100% 100% 100% 100% 100% 100% 100% 100% 100% 100% 100% 100% 100% 100%

Table 2. Results of breathing rate test. Ventilator (SERVO-i) Sensing shirt Ventilator (SERVO-i) Sensing shirt Ventilator (SERVO-i) Sensing shirt

4 BPM

6 BPM

10 BPM

15 BPM

3.9 ± 0.1

5.9 ± 0.2

10.0 ± 0.1 15.0 ± 0

20.0 ± 0.2

25 BPM

30 BPM

40 BPM

60 BPM

50 BPM

25.0 ± 0.2 30.0 ± 0.7 39.9 ± 0.5 51.2 ± 3.8 70 BPM

80 BPM

90 BPM

100 BPM

69.5 ± 3.0 79.7 ± 2.6 89.5 ± 2.1 100.5 ± 4.3

20 BPM

59.8 ± 5.1

Sensing Shirt were shown to be within ±10% of simultaneous PF-300 measurements, 98.10% (11,220 breaths) within 15%, and 99.03% (11326 breaths) within ±20%. The result was comparable with others‟ report [24]. Table 4 shows the test results of breath-by-breath TV comparison between two systems. The Bland–Altman plot for the HR detection during incremental exercises is illustrated in Fig. 5. The bias for HR detection during exercises was 0.24, and precision was 1.54. There existed systematic difference on the mean HR

Sensing Shirt for Ambulatory Monitoring

213

Table 4. Results of breath-by-breath TV comparison between two systems. Participant 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 Total

No. of breaths 700 647 811 864 921 858 741 1073 742 915 584 519 692 952 418 11437

Error > ± 10% 4.78% 5.87% 10.73% 1.62% 3.48% 9.91% 10.95% 5.23% 3.10% 7.76% 10.27% 5.97% 4.64% 1.94% 9.70% 6.15%

Error > ± 15% 0.72% 2.47% 3.33% 0.46% 1.20% 1.75% 3.92% 1.59% 0.54% 2.51% 3.25% 3.08% 2.31% 0.21% 3.15% 1.90%

Figure 5. Bland-Altman plots for HR comparison from 10 subjects during incremental treadmill and cycle ergometer exercises. The limits of agreement defined by 2SD above and below the bias are shown by dashed lines.

measurements between Sensing Shirt and the Polar Sport Tester RS800, but the mean difference between systems was small (0.24). The HR during incremental exercises resided within the mean ± 2SD range, which signifies good stability in HR detection during exercises.

4. Discussion This work contributes to the state of the art for wearable, ambulatory healthcare monitoring, by integration of multiple sensors into the shirt. The Sensing Shirt system is a logical extension of ECG monitors, with more physiological parameters, more patient comfort and more ease of use. Our wearable sensor unit has been carefully designed to decrease the power consumption and improve the signal quality for ambulatory use. Sensing Shirt represents our initial efforts in developing wearable technology for non-intrusive cardiopulmonary monitoring. Performance tests on these basic parameters demonstrated the accuracy and reliability of this prototype system.

Error > ± 20% 0.29% 1.70% 1.23% 0.35% 0.76% 0.35% 1.89% 0.84% 0.27% 1.64% 1.02% 2.31% 1.30% 0.11% 1.69% 0.97%

Correlation coefficient Regression coefficient 0.94 1.01 0.87 1.00 0.96 1.01 0.88 1.00 0.94 1.01 0.85 1.00 0.90 1.01 0.84 1.00 0.85 1.02 0.87 1.01 0.88 0.98 0.85 1.00 0.96 0.98 0.91 1.00 0.98 1.00 0. 90 1.00

Some of the extracted parameters such as HR, breathing rate, TV, and SpO2 have been validated by the preliminary check, and the reliability of HR detection during exercises was also tested. Figure 3 shows the physiological data captured by the Sensing Shirt at different movement stages during an incremental treadmill exercise. Though more movement artifact noise was imposed on the physiological data with the increment of exercise intensity (which can be affirmed by the signals of X, Y, and Z from the 3-axis accelerometer), the Sensing Shirt system could capture the physiological data and extract basic parameters effectively. For HR detection, though statistically, there is systematic difference in mean HR between the Sensing Shirt system and the Polar Sport Tester RS800, the mean difference is very small (0.24). This result is comparable with others‟ report with respect to the validity and reliability of cardiorespiratory measurements recorded by the Life Shirt during exercise [25]. A comparison of HR detection between the Life Shirt and Polar Sport Tester also demonstrated there was significant difference between systems, and the mean difference was also small. The Bland–Altman plot in Fig. 5 shows all the HR were within the mean ± 2SD range, and the limits of agreement (-2.8 and 3.3) are small enough for Sensing Shirt system to make an accurate and reliable HR detection during exercise. For RIP signals during movement, digital signal processing methods such as adaptive and wavelet filtering can be used for improving the accuracy of respiratory measurement [26]. Body acceleration signals can be used as a reference for signal processing. As certain parameters may only be reliably obtainable during a standardized sitting or supine resting period, it is still a challenge for us to minimize confounding effects from physical activity and movement artifacts in ambulatory monitoring. In follow-on research, clinical validation of cardiopulmonary parameters in ambulatory conditions is needed. Healthy volunteers and patients with cardiac and respiratory diseases should be involved. Meanwhile, as there is no standardized test procedure or simulator to evaluate PWTT and RSA, more future work needs to be done on the methodology, algorithms and clinical validation regarding

214

J. Med. Biol. Eng., Vol. 31 No. 3 2011

these two parameters [27,28]. In clinical practice, the use of PSG recording is a standard evaluation procedure for sleep-related breathing disorders. However, PSG is not suitable for long-term monitoring in the home environment. As electroencephalogram (EEG) is not integrated in most ambulatory monitoring system, PWTT will improve the detection of sleep respiratory events and microarousals during sleep testing at home [29]. A number of rigidly controlled experiments will be carried out between the PSG and the Sensing Shirt to evaluate their sensitivity and specificity for sleep-event detection. The Sensing Shirt is able to simultaneously monitor several physiological systems, thus allowing for assessments of the interactions among various physiological parameters. This system offers us a useful tool to acquire circadian cardiopulmonary physiological parameters. The cardiorespiratory interaction and its changes during the diurnal cycle will provide detailed information on the autonomic nerve system status [30]. Data mining from these long-time physiological parameters to give explicit explanation about health status is another challenge for us. In the Sensing Shirt‟s current implementation, raw data is off-line processed on a PC platform but not in real time. This may limit the possibilities for detailed inspection of the data by the researcher and physicians. In the next step, a Bluetoothand Smartphone-based telemonitoring system will be developed. With the Bluetooth module integrated in the Sensing Shirt, the telemonitoring application will work on any mobile phone that can run Java, and data can be transmitted either over a wireless or cellular network. As the Sensing Shirt prototypes have demonstrated the feasibility of the concept and solutions to the key technologies utilized in wearable sensors and systems, mobile phone-based tele-monitoring system will be relatively easy to be developed.

5. Conclusions A wearable cardiopulmonary monitoring system called Sensing Shirt has been developed with wearable sensors integrated in an elastic T-shirt. A number of vital signs, including ECG, RC and AB respiration, PPG, SpO2, and posture/activities, can be acquired by this system. The technique of PAM used in the RIP and PPG sections can reduce the power consumption effectively, and synchronized AD conversion between ECG and PPG contributes to precise PWTT calculation. Powered by 2 AA batteries with 1600-mAh capacity, the whole system can work more than 48 h continuously without interruption. Performance tests on these basic parameters demonstrated the accuracy and reliability of this prototype system. This advanced ambulatory monitoring system may be used in both home healthcare and scientific research.

the Beijing Natural Science Foundation (Grant Number: 3102028). The authors are thankful to clinical engineer Juan Zhou at the PLA General Hospital, Beijing for her help in performance test.

References [1] [2]

[3]

[4]

[5]

[6]

[7]

[8]

[9]

[10]

[11]

[12]

[13]

[14] [15]

[16]

[17]

[18]

Acknowledgements [19]

This project was supported in part by the National Natural Science Foundation of China (Grant Number: 60801009) and

P. Bonato, “Clinical applications of wearable technology,” Proc. Conf. 2009 IEEE Eng. Med. Biol. Soc., 6580-6583, 2009. U. Anliker, J. A. Ward, P. Lukowicz, G. Troster, F. Dolveck, M. Baer, F. Keita, E. B. Schenker, F. Catarsi, L. Coluccini, A. Belardinelli, D. Shklarski, M. Alon, E. Hirt, R. Schmid and M. Vuskovic, “AMON: a wearable multiparameter medical monitoring and alert system,” IEEE Trans. Inf. Technol. Biomed., 8: 415-427, 2004. M. L. Fruin and J. W. Rankin, “Validity of a multi-sensor armband in estimating rest and exercise energy expenditure,” Med. Sci. Sports Exerc., 36: 1063-1069, 2004. P. Grossman, “The LifeShirt: a multi-function ambulatory system monitoring health, disease, and medical intervention in the real world,” Stud. Health Technol. Inform., 108: 133-141, 2004. [5] S. Park and S. Jayaraman, “e-Health and quality of life: the role of the Wearable Motherboard,” Stud. Health Technol. Inform., 108: 239-252, 2004. P. S. Pandian, K. Mohanavelu, K. P. Safeer, T. M. Kotresh, D. T. Shakunthala, P. Gopal and V. C. Padaki, “Smart Vest: wearable multi-parameter remote physiological monitoring system,” Med. Eng. Phys., 30: 466-477, 2008. A. Lymberis and R. Paradiso, “Smart fabrics and interactive textile enabling wearable personal applications: R&D state of the art and future challenges,” Proc. Conf. 2008 IEEE Eng. Med. Biol. Soc., 5270-5273, 2008. A. Pantelopoulos and N. Bourbakis, “A survey on wearable biosensor systems for health monitoring,” Proc. Conf. 2008 IEEE Eng. Med. Biol. Soc., 4887-4890, 2008. F. H. Wilhelm, W. T. Roth and M. A. Sackner, “The LifeShirt: an advanced system for ambulatory measurement of respiratory and cardiac function,” Behav. Modif., 27: 671-691, 2003. J. D. Witt, J. R. Fisher, J. A. Guenette, K. A. Cheong, B. J. Wilson and A. W. Sheel, “Measurement of exercise ventilation by a portable respiratory inductive plethysmograph,” Respir. Physiol. Neurobiol., 154: 389-95, 2006. N. Halin, M. Junnila, P. Loula and P. Aarnio, “The LifeShirt system for wireless patient monitoring in the operating room,” J. Telemed. Telecare, 11: 41-43, 2005. Z. B. Zhang, M. S. Yu, R. X. Li, T. H. Wu and J. L. Wu, “Design of a wearable respiratory inductive plethysmograph and its applications,” Space Med. Med. Eng. (Beijing), 19: 377-381, 2006. L. Rattfält, M. Lindén, P. Hult, L. Berglin and P. Ask, “Electrical characteristics of conductive yarns and textile electrodes for medical applications,” Med. Biol. Eng. Comput., 45: 1251-1257, 2007. Y. Mendelson, “Pulse oximetry: theory and applications for noninvasive monitoring,” Clin. Chem., 38: 1601-1607, 1992. M. J. Drinnan, J. Allen, and A. Murray, “Relation between heart rate and pulse transit time during paced respiration,” Physiol. Meas., 22: 425-32, 2001. G. M. Friesen, T. C. Jannett, M. A. Jadallah, S. L. Yates, S. R. Quint and H. T. Nagle, “A comparison of the noise sensitivity of nine QRS detection algorithms,” IEEE Trans. Biomed. Eng., 37: 85-98, 1990. I. Dotsinsky and T. Stoyanov, “Power-line interference cancellation in ECG signals,” Biomed. Instrum. Technol., 39:155-162, 2005. K. Konno and J. Mead, “Measurement of the separate volume changes of rib cage and abdomen during breathing,” J. Appl. Physiol., 22: 407-422, 1967. R. Sartene, C. Dartus, J. L. Bernard. M. Mathieu and M. D. Goldman, “Comparison of thoracoabdominal calibration methods in normal human subjects,” J. Appl. Physiol., 75: 2142-2150, 1993.

Sensing Shirt for Ambulatory Monitoring

[20] J. R. Thompson, “Qualitative diagnostic calibration technique,” J. Appl. Physiol., 87: 869-872, 1999. [21] H. Hsiu, P. T. Chao, W. R. Chiang, C. L. Hsu, M. Y. Jan, W. K. Wang and Y. Y. Lin Wang, “Effect of pulse number on the pulse-based synchronized-averaging analysis in laser Doppler signal,” Biomed. Eng., Appl., Basis, and Commun., 19: 179-185, 2007. [22] R. M. Ross, “ATS/ACCP Statement on cardiopulmonary exercise testing,” Am. J. Respir. Crit. Care Med., 167: 1451-1452, 2003. [23] J. M. Bland and D. G. Altman, “Statistical methods for assessing agreement between two methods of clinical measurements,” Lancet, 1: 307-310, 1986. [24] T. S. Chadha, H. Watson, S. Birch, G. A. Jenouri, A. W. Schneider, M. A. Cohn and M. A. Sackner, “Validation of respiratory inductive plethysmography using different calibration procedures,” Am. Rev. Respir. Dis., 125: 644-649, 1982. [25] L. Kent, B. O‟Neill, G. Davison, A. Nevill, J. S. Elborn and J. M. Bradley, “Validity and reliability of cardiorespiratory measurements recorded by the LifeShirt during exercise tests,” Respir. Physiol. Neurobiol., 167: 162-167, 2009.

215

[26] D. B. Keenan and F. H. Wilhelm, “Adaptive and wavelet filtering methods for improving accuracy of respiratory measurement,” Biomed. Sci. Instrum., 41: 37-42, 2005. [27] Y. Chen, C. Y. Wen, G. C. Tao. M. Bi and G. Li, “Continuous and noninvasive blood pressure measurement: a novel modeling methodology of the relationship between blood pressure and pulse wave velocity,” Ann. Biomed. Eng., 37: 2222-2233, 2009. [28] J. W. Denver, S. F. Reed and S. W. Porges, “Methodological issues in the quantification of respiratory sinus arrhythmia,” Biol. Psychol., 74: 286-294, 2007. [29] J. L. Pépin, N. Delavie, I. Pin, C. Deschaux, J. Argod, M. Bost and P. Levy, “Pulse transit time improves detection of sleep respiratory events and microarousals in children,” Chest, 127: 722-730, 2005. [30] D. Cysarz, D. von Bonin, P. Brachmann, S. Buetler, F. Edelhäuser, K. Laederach-Hofmann and P. Heusser, “Day-to-night time differences in the relationship between cardiorespiratory coordination and heart rate variability,” Physiol. Meas., 29: 1281-1291, 2008.

216

J. Med. Biol. Eng., Vol. 31 No. 3 2011