Cedex Cleaning & Maintenance SOP s QUALITY THAT COUNTS

Cedex Cleaning & Maintenance SOP’s Q U A L I T Y T H AT C O U N T S Daily Maintenance (1) Morning Start up • Check reagent containers, fill if need...
Author: Julius Hodges
0 downloads 2 Views 216KB Size
Cedex Cleaning & Maintenance SOP’s

Q U A L I T Y T H AT C O U N T S

Daily Maintenance (1) Morning Start up • Check reagent containers, fill if needed and be sure to rinse bottles prior to refilling • When filling Trypan Blue reagent container filter stock Trypan Blue prior to use • Empty waste container • In Liquid Management – perform Prime • Visually check syringe and all lines

E NOT

Please keep all reagent containers and sample cups covered and free of dust!

Cedex Cleaning & Maintenance SOP’s

Daily Maintenance (2) Runs throughout the day Before each run • Empty waste container if greater than 50% full • Check that system has been primed and there are no air pockets or bubbles in capillary tubes After each run (10-20 samples) • Perform Standard Clean (1.5 min) • Perform Chamber Clean 2 (0.5 min) • Empty waste container

Cedex Cleaning & Maintenance SOP’s

Daily Maintenance (3) End of Day Shutdown • Perform Standard Clean (1.5 min) • Perform Chamber Clean 2 (0.5 min) • Perform LM Shutdown • Empty waste container

Cedex Cleaning & Maintenance SOP’s

Weekly Maintenance (1) At the Start of the week (Monday) • Clean the reagent containers with distilled or de-ionized water and refill with new reagents. • Make sure to replace the Trypan Blue solution each week. • Clean external surface of capillary tubes, which reach into the sample vial & reagent containers, with a lint-free cloth. • Remove and clean syringe (using 70% ethanol and distilled or de-ionized water).

Cedex Cleaning & Maintenance SOP’s

Weekly Maintenance (2) At the End of the week (Friday) • Perform Standard Clean (1.5 min) • Perform Chamber Clean 2 (0.5 min) • Perform Intensive Clean (20 min) • Perform LM Shutdown • Empty waste container

Cedex Cleaning & Maintenance SOP’s

(Bi-)Monthly Maintenance Perform two Intensive Cleans with 0.1N to 1N HCl instead of Cedex Detergent Procedure • Replace Detergent in reservoir with 0.1N to 1N HCl • Run Prime • Start Intensive Clean • Replace HCl in reservoir with detergent • Run Prime • Run Standard Clean Use Syringe Kit to Backflush flow cell in reverse direction

Cedex Cleaning & Maintenance SOP’s

Syringe Backflush Routine • Detach the tubing and flangeless fittings from position 8 on the multi-way valve (far bottom right hand side of multi-way valve while looking at front of Cedex). • Attach tubing and flangeless fittings from position 8 to receiving port of tubing (supplied in kit). Place open end of tubing into empty beaker. • Unscrew flangeless nut and ferrule from syringe, supplied in kit. • Fill the syringe with DI, particle-free water. • Slide the flangeless nut over frosted tubing leading into waste container (In addition to the frosted tubing, there is also clear tubing leading into the waste container). Do not use this tubing with the nut threads facing the open end of the tubing. • Slip the flangeless ferrule over the tubing, with the tapered position of the ferrule facing toward the nut.

Cedex Cleaning & Maintenance SOP’s

Syringe Backflush Routine • Insert the frosted tubing with the flangeless nut and ferrule in place into the receiving port of the syringe filled with DI, particle-free water and tighten the nut finger tight. • Alternate pushing liquid and air from the syringe through the flowcell by rotating the syringe. Pull the plunger in the reverse direction to allow more air into the syringe. • Continue pushing liquid and air through the syringe. When syringe is empty, detach syringe from tubing and reattach tubing and flangeless fittings finger tight back into position 8. Perform a Standard Clean and check to make sure that there are no leaks coming from position number 8.

Cedex Cleaning & Maintenance SOP’s

Cedex Cleaning & Maintenance SOP’s Version 1.0, January, 2006 This Cleaning and Maintenance Standard Operation Procedure is valid for Cedex 1.3, 1.5, 1.6, 1.7 (Standard) and 1.10 (HiRes). Your individual innovatis contact persons: Service & Support:

Phone: Email:

Product Management:

Phone: Email:

Sales:

Phone: Email:

Cedex Cleaning & Maintenance SOP’s

Cedex Cleaning & Maintenance SOP’s innovatis AG Meisenstr. 96 33607 Bielefeld Germany Phone: +49 (0) 521 29 97 – 300 Fax: +49 (0) 521 29 97 – 285 innovatis Inc. 301 Lindenwood Drive, Suite 115 Malvern, PA 19355 USA Phone: 1-888-283 – 1564 Toll free Fax: 1-888-283 – 1631 Toll free Phone: 610-889 – 7319 http://www.innovatis.com

Cedex Cleaning & Maintenance SOP’s