Calibration Certificate

VAISALA Calibration Certificate Caltbrallon - Cer1Jficala No. 2D83.01 AS FOUND data only Company Pharmaceuticals Company Your Address, USA Ambient ...
Author: Gwen Tate
4 downloads 0 Views 2MB Size
VAISALA

Calibration Certificate

Caltbrallon - Cer1Jficala No. 2D83.01

AS FOUND data only Company Pharmaceuticals Company Your Address, USA Ambient Temperature(°C): Min: 20.5 Max: 23.7 0236624 VL-1016-22V Ambient Humidity(%RH): Min: 25.9 Max: 33.1 Calibration by comparison Method: Serial Number: 10021068 VCP1005 Procedures: 3694 Asset ID: The calibration(s) on this report are traceable to the United States of America National Institute ofStandards and Technology or to other recognized national or international standards or to accepted values of natural physical constants, and are accredited to ISO/IEC 17025. The laboratory meets the requirements of ANSI/NCSL 2540-1. Using methods detailed in the ISO "Guide to the Expression of Uncertainty in Measurement", reported uncertainties represent expanded uncertainties expressed at approximately the 95% confidence level using a coverage factor ofk=2. The calibrations were performed equivalently either with minimum test uncertainty ratios of 4: I using a coverage factor of k = 2, or with the statistical method of guard banding to reduce test limits. The results relate only to the item(s) calibrated. Certificate Number: Model Number·

CALIBRATION REFERENCE EQUIPMENT Fluke Black Stack SPRT Scanner Module Model 2560 Hart Scientific Secondary Standard PRT Probe Model 5626 CALIBRATION TEST RESULTS Chan Test Dcscri12tion Temperature Temperature Temperature Temperature Temperature

.!ln.i.!§

ac oc oc oc oc

Reference -89.92 -54.91 -19.93 15.04 50.02

Calibration Date Next � 12-Mav-15 16-May-16 27-0ct-16 27-0ct-15

Serial# 833423 3934 I Year Spec. 0.35 0.35 0.35 0.35 0.35

Measurement Uncertain!Y 0.02 0.02 0.02 0.02 0.02

As Found Diff. Result O.Q3 -89.89 0.00 -54.91 0,01 -19.92 14.97 0.07 O.Q3 50.05

p p p p p

As Left Result Diff. NA NA NA NA NA NA NA NA NA NA

u u u u u

P indicates Pass, F indicates Fail, 0 indicates overlaps guard band, U indicates tolerance unspecified Calibration Maintaining Calibration The electronic components in this data logger are ofthe highest quality. The unit has been designed to remain within its specifications. The length ofi11-eulib tion service can be affected by aging, temperature and shock. ·or those crs wi I critical needs such as accreditation demands, 1catio , or I. requirements, we recommend that the p rio · • asis.

Calibration Technician signature

Information on calibration services is available at the address below. This data logger was calibrated by: Vaisala Inc. I 0-D Gill Street Woburn, MA USA 01801 Toll Free: 1-800-683-8374, Email: [email protected], Website: www.vaisala.com

Calibration Date:

04-Feb-2016

Next Calibration: 04-Feb-2017

c:�LaboraQ.nager: Manager's signature This certificate is copyrighted by Vaisala Inc., and may not be reproduced, except in full, without written approval from Vaisala Inc.

VAISALA

Calibration Certificate

AS FOUND data only Company Pharmaceutical Company Your Address, USA Ambient Tempcrnturc(°C): Min: 21.1 Max: 22.0 crtiticatc Number: 0242281 Ambient Humidity(o/oRH): Min: 33.3 Max: 39.6 Model Number: VL-1416-44V Serial Number: 11071046 Calibration by comparison Method: Asset ID: Procedures: VCP I 005 The calibration(s) on this report are traceable to the United States of America National Institute of Standards and Technology or to other recognized national or international standards or to accepted values of natural physical constants, and are accredited to JSO/JEC 17025. The laboratory meets the requirements of ANSI/NCSL 2540-1. Using methods detailed in the ISO "Guide to the Expression of Uncertainty in Measurement", reported uncertainties represent expanded uncertainties expressed at approximately the 95% confidence level using a coverage factor of k=2. The calibrations were performed equivalently either with minimum test uncertainty ratios of 4: I using a coverage factor of k = 2, or with the statistical method of guard banding to reduce test limits. The results relate only to the item(s) calibrated. CALIBRATION REFERENCE EQUIPMENT Hart Scientific Black Stack SPRT Scanner Module Model 2560 Hart Scientific Secondary Standard PRT Probe Model 5626

CALIBRATION TEST RESULTS Chan Test Des�ri:ption Temperature Temperature Temperature Temperature I Temperature Temperature 2 Temperature 2 Temperature 2 Temperature 2 Temperature 2 Temperature 3 Temperature 3 3 Temperature 3 Temperature Temperature 3 Temperature 4 Temperature 4 Temperature 4

Units oc oc oc oc oc oc oc oc oc oc oc oc oc oc oc oc oc oc

Reference -89.91 -54.91 -19.93 15.12 50.13 -89.91 -54.91 -19.93 15.12 50.13 -89.91 -54.91 -19.93 15.12 50.13 -89.91 -54.91 -19.93

Calibration Date Next 21-May-16 21-May-17 27-0ct-15 27-0ct-16

Yfil

Serial# 802212 3934 1 Year Spec. 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35

Measurement Uncertainty 0.02 0.02 0.02 0.02 0.02 0.02 0.02 0.02 0.02 0.02 0.02 0.02 0.02 0.02 0.02 0.02 0.02 0.02

As Found Diff. Result -89.91 0.00 0.00 -54.91 0.01 -19.94 15.10 0.02 0.01 50.14 0.01 -89.92 0.01 -54.92 0.01 -19.94 15.10 0.02 50.13 0.00 -89.91 0.00 0.00 -54.91 -19.94 0.01 0.02 15.10 50.13 0.00 -89.91 0.00 0.01 -54.92 -19.95 0.02

p p p p p p

p p p p

p p p p p p p p

As Left Diff. Result NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA

u u u u

u u u u

u u u u u u u u u u

P indicates Pass, F indicates Fail, 0 indicates overlaps guard band, U indicates tolerance unspecified

Maintaining Calibration

Calibration

The electronic components in this data logger are of the highest quality. The unit has been designed to remain within its specifications. The length of in-calibro ·011 service can be affected by aging, temperature and shock. For those u r •ith critic I needs such as accreditation demands, I requirements, we recommend that the d. · osis.

Information on calibration services is available at the address below. This data Jogger was calibrated by: Vaisala Inc. 10-D Gill Street Woburn, MA USA 01801 Toll Free: 1-800-683-83 74, Email: [email protected], Website: www.vaisala.com

Calibration Date:

nlib

23-Aug-2016

Next Calibration: 23-Aug-2017

Page l of2 This certificate is copyrighted by Vaisala Inc., and may not be reproduced, except in full, without written approval from Vaisala Inc.

VAISALA

Calibration Certificate

� (icciiiotTED) Callbrallon • C•rtlflcate No. 20BJ.01

AS FOUND data only

Your Company

Certificate Number: 0242281 Ambient Tcmpcraturc(0C}: Min: 21.1 Max: 22.0 Model Number: VL-1416-44V Ambient Humidity(%RH): Min: 33.3 Max: 39.6 Calibration by comparison Serial Number: 11071046 Method: Asset ID: VCPI005 Procedures: The calibration(s) on this report are traceable to the United States of America National Institute of Standards and Technology or to other recognized national or international standards or to accepted values of natural physical constants, and are accredited to ISO/IEC 17025. The laboratory meets the requirements of ANSI/NCSL 2540-1. Using methods detailed in the ISO "Guide to the Expression of Uncertainty in Measurement", reported uncertainties represent expanded uncertainties expressed at approximately the 95% confidence level using a coverage factor of k=2. The calibrations were performed equivalently either with minimum test uncertainty ratios of 4:1 using a coverage factor of k = 2, or with the statistical method of guard banding to reduce test limits. The results relate only to the item(s) calibrated. 4 4

oc oc

Temperature Temperature

15.12 50.13

0.35 0.35

0.02 0.02

15.10 50.13

0.02 0.00

p p

NA NA

NA NA

u u

P indicates Pass, F indicates Fail, 0 indicates overlaps guard band, U indicates tolerance unspecified Maintaining Calibration Calibration

The electronic components in this data logger are of the highest quality. The unit has been designed to remain within its specifications. The length of in-calibratio service can be affected by aging, temperature and shock. For those user with critical ccds such as accreditation demands, govcrnme I spccifi · n , IS quircmcn(s, we recommend that the unit be c i rated 1 1 per· di • > ·i .

Information on calibration services is available at the address below. This data logger was calibrated by: Vaisala Inc. I 0-D Gill Street Woburn, MA USA 01801 Toll Free: 1-800-683-83 74, Email: [email protected], Website: www.vaisala.com

Calibration Date: 23-Aug-2016 Next Calibration: 23-Aug-2017 Calibration Laboratory Manager: Page 2 of2 This certificate is copyrighted by Yaisala Inc., and may not be reproduced, except in full, without written approval from Yaisala Inc.

VAISALA

ffl'

Calibration Certificate

(icciii.DITEJ>)

Callbratlon • Cartlflcala No. 2083.01

AS FOUND data only Your Company Your Address CcnilicateNumber:

0242280

Ambient Tcmpcraturc(OC): Min: 21.1 Max: 22.0 Model Number: VL-1416-44V Ambient Humidity(%Rl-1): Min: 33.3 Max: 39.6 Calibration by comparison 15111006 Serial Number: Method: Asset ID: Procedures: VCP1005 The calibration(s) on this report are traceable to the United States of America National Institute of Standards and Technology or to other recognized national or international standards or to accepted values of natural physical constants, and are accredited to ISO/JEC 17025. The laboratory meets the requirements of ANSI/NCSL 2540-1. Using methods detailed in the ISO "Guide to the Expression of Uncertainty in Measurement", reported uncertainties represent expanded uncertainties expressed at approximately the 95% confidence level using a coverage factor of k=2. The calibrations were performed equivalently either with minimum test uncertainty ratios of 4:1 using a coverage factor of k = 2, or with the statistical method of guard banding to reduce test limits. The results relate only to the item(s) calibrated. CALIBRATION REFERENCE EQUIPMENT

Hart Scientific Black Stack SPRT Scanner Module Model 2560 Hart Scientific Secondary Standard PRT Probe Model 5626 CALIBRATION TEST RESULTS

Chan T�! De:cription Temperature Temperature Temperature Temperature Temperature 2 Temperature Temperature 2 2 Temperature 2 Temperature 2 Temperature Temperature 3 Temperature 3 Temperature 3 Temperature 3 Temperature 3 4 Temperature 4 Temperature 4 Temperature

Units oc oc oc oc oc oc oc oc oc oc oc oc oc oc oc oc oc oc

Reference -89.91 -54.91 -19.93 15.12 50.13 -89.91 -54.91 -19.93 15.12 50.13 -89.91 -54.91 -19.93 15.12 50.13 -89.91 -54.91 -19.93

Calibration Date Last Next 21-May-16 21-May-17 27-0ct-16 27-0ct-15

Serial# 802212 3934 1 Year

� 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35

Measurement Uncertainty 0.02 0.02 0.02 0.02 0.02 0.02 0.02 0.02 0.02 0.02 0.02 0.02 0.02 0.02 0.02 0.02 0.02 0.02

As Found Diff. Result -89.91 0.00 0.00 -54.91 -19.95 0.02 15.08 0.04 50.11 0.02 0.01 -89.92 -54.93 0.02 O.Q3 -19.96 15.08 0.04 50.12 0.01 -89.91 0.00 0.01 -54.92 0.03 -19.96 15.07 0.05 50.11 0.02 -89.91 0.00 0.01 -54.92 0.03 -19.96

p p p p p p p p p p p p p p p p p p

As Left Result Diff. NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA

u u u u u u u u u u u u u u u u u u

P indicates Pass, F indicates Fail, 0 indicates overlaps guard band, U mdicates tolerance unspecified

Maintaining Calibration

Calibration

The electronic components in this data logger are of the highest quality. The unit has been designed to remain within its specifications. The length of in-calibration service can be affected by aging, temperature and shock. For those user. 'th critical needs such as accreditation demands, ISequircmcnts, O we recommend that the government specific i 11 ·, o unit be calib ed o. pcrio c ·is. �

Information on calibration services is available at the address below. This data logger was calibrated by: Vaisala Inc. 10-D Gill Street Woburn, MA USA 01801 Toll Free: 1-800-683-83 74, Email: [email protected], Website: www.vaisala.com Calibration Date:

23-Aug-2016

Next Calibration:

23-Aug-2017

Page I of2 This certificate is copyrighted by Vaisala Inc., and may not be reproduced, except in full, without written approval from Vaisala Inc.

VAISALA

Calibration Certificate

� ficciiiDITED) ci1nb,allon � c,ntflc.ait No. aoa:3,01

AS FOUND data only Your Company Your Address Ccrlificate Number;

0242280

Ambicn\ Tcmpcraturc(0 ): Min: 21.1 Max: 22.0 Model Number: VL-1416-44V Ambiclll Humidily(%RMl: Min: 33.3 Max: 39.6 Serial Number: 15111006 Calibration by comparison Method: Asset ID: Procedures: VCPI005 The calibration(s) on this report are traceable to the United States of America National Institute of Standards and Technology or to other recognized national or international standards or to accepted values of natural physical constants, and are accredited to ISO/JEC 17025. The laboratory meets the requirements of ANSI/NCSL Z540-1. Using methods detailed in the ISO "Guide to the Expression of Uncertainty in Measurement", reported uncertainties represent expanded uncertainties expressed at approximately the 95% confidence level using a coverage factor of k=2. The calibrations were performed equivalently either with minimum test uncertainty ratios of 4:1 using a coverage factor of k = 2, or with the statistical method of guard banding to reduce test limits. The results relate only to the item(s) calibrated. 4 4

oc oc

Temperature Temperature

15.12 50.13

0.35 0.35

0.02 0.02

15.08 50.12

0.04 0.01

p p

NA NA

NA NA

u u

P indicates Pass, F indicates Fail, 0 indicates overlaps guard band, U indicates tolerance unspecified

Maintaining Calibration

Calibration

The electronic components in this data logger are of the highest quality. The unit has been designed to remain within its specifications. The length of in-calibration service can be affected by aging, temperature and shock. For those users ith critical needs such as accreditation demands, government spccilica ons, o I quiremcnls, we recommend that the unit be calib� d O cri ic

Information on calibration services is available at the address below. This data logger was calibrated by: Vaisala Inc. I 0-D Gill Street Woburn, MA USA 01801 Toll Free: 1-800-683-83 74, Email: [email protected], Website: www.vaisala.com Calibration Date:

23-Aug-2016

Next Calibration:

23-Aug-2017

Page 2 of2 This certificate is copyrighted by Vaisala Inc., and may not be reproduced, except in full, without written approval from Vaisala Inc.

VAISALA

Calibration Certificate

� [icciiiOITED) Callbrallon - Certificate No. 2083.01

AS FOUND data only Your Company Your Address

Ambient Tcmpenuure(°C): Min: 2 I. I Max: 22.0 Ambient Humidity(%RH): Min: 33.3 Max: 39.6 Calibration by comparison Method: Procedures: VCPI005 The calibration(s) on this report are traceable to the United States of America National Institute of Standards and Technology or to other recognized national or international standards or to accepted values of natural physical constants, and are accredited to ISO/IEC 17025. The laboratory meets the requirements of ANSI/NCSL Z540-1. Using methods detailed in the ISO "Guide to the Expression of Uncertainty in Measurement", reported uncertainties represent expanded uncertainties expressed at approximately the 95% confidence level using a coverage factor of k=2. The calibrations were performed equivalently either with minimum test uncertainty ratios of 4: I using a coverage factor of k = 2, or with the statistical method of guard banding to reduce test limits. The results relate only to the item(s) calibrated.

CeriilicaLc umber: Model Number: Serial Number: Asset ID:

0242279 VL-1416-44V 13061270

CALIBRATION REFERENCE EQUIPMENT Hart Scientific Black Stack SPRT Scanner Module Model 2560 Hart Scientific Secondary Standard PRT Probe Model 5626 CALIBRATION TEST RESULTS Chan Test l)escri11tion Temperature Temperature Temperature Temperature Temperature Temperature 2 Temperature 2 Temperature 2 Temperature 2 Temperature 2 Temperature 3 Temperature 3 Temperature 3 Temperature 3 Temperature 3 Temperature 4 Temperature 4 Temperature 4

Units

oc oc oc oc oc oc oc oc oc oc oc oc oc oc oc oc oc oc

Reference -89.91 -54.91 -19.93 15.12 50.13 -89.91 -54.91 -19.93 15.12 50.13 -89.91 -54.91 -19.93 15.12 50.13 -89.91 -54.91 -19.93

Calibration Date Next !&§1 21-May-16 21-May-17 27-0ct-15 27-0ct-16

Serial# 802212 3934 I Year Spec. 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35

Measurement Uncertainty 0.02 0.02 0.02 0.02 0.02 0.02 0.02 0.02 0.02 0.02 0.02 0.02 0.02 0,02 0,02 0.02 0.02 0.02

As Found Diff. Result 0.01 -89.92 0.01 -54.92 0.02 -19.95 0.03 15.09 50.13 0.00 0.00 -89.91 0.01 -54.92 0.02 -19.95 15.08 0.04 0.01 50.12 -89.92 0.01 0.01 -54.92 0.02 -19.95 0,03 15.09 0.00 50.13 -89.92 0.01 0.01 -54.92 -19.95 0.02

p p p p p p p p p p p p p p p p p p

As Left Result Diff. NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA

u u

u u u

u u u

u u u

u

u u u

u

u u

P indicates Pass, F indicates Fail, 0 indicates overlaps guard band, U indicates tolerance unspecified Maintaining Calibration Calibration

The electronic components in this data logger are of the highest quality. The unit has been designed to remain within its specifications. The length of in-calibrati service can be affected by aging, temperature and shock. For those use with critical needs such as accreditation demands, gowrnmi:nl · ccifi ti ns, or I re iremcnts, we recommend that the unit be cali ra d a criodi basi .

Information on calibration services is available at the address below. This data logger was calibrated by: Vaisala Inc. I 0-D Gill Street Woburn, MA USA 01801 Toll Free: 1-800-683-8374, Email: [email protected], Website: www.vaisala.com Calibration Date:

23-Aug-2016

Next Calibration:

23-Aug-2017

Page l of2 This certificate is copyrighted by Vaisala Inc., and may not be reproduced, except in full, without written approval from Vaisala Inc.

VAISALA

Calibration Certificate

� (icciiiDITED) Callbratlon - CartlnCille No. 2083.01

AS FOUND data only Your Company Your Address

Certificate Number: Ambient empernture(0 '): Min: 24.4 Max: 27 .5 0242278 Model Number: Ambient Humidi1y(%RH}; Min: 36.2 Max: 54.0 VL-1416-44V Calibration by comparison 15111051 Serial Number: Method: VCPI005 Procedures: Asset ID: The calibration(s) on this report are traceable to the United States of America National Institute of Standards and Technology or to other recognized national or international standards or to accepted values of natural physical constants, and are accredited to ISO/IEC 17025. The laboratory meets the requirements of ANSI/NCSL Z540-1. Using methods detailed in the ISO "Guide to the Expression of Uncertainty in Measurement", reported uncertainties represent expanded uncertainties expressed at approximately the 95% confidence level using a coverage factor of k=2. The calibrations were performed equivalently either with minimum test uncertainty ratios of 4: I using a coverage factor of k = 2, or with the statistical method of guard banding to reduce test limits. The results relate only to the item(s) calibrated. CALIBRATION REFERENCE EQUIPMENT Hart Scientific Secondary Standard PRT Probe Model 5626 Hart Scientific Black Stack SPRT Scanner Module Model 2560 CALIBRATION TEST RESULTS Chan Test Qc crip1ion Temperature Temperature Temperature Temperature Temperature 2 Temperature 2 Temperature 2 Temperature Temperature 2 2 Temperature 3 Temperature 3 Temperature Temperature 3 Temperature 3 3 Temperature 4 Temperature 4 Temperature Temperature 4

Units oc oc oc oc oc oc oc oc oc oc oc oc oc oc oc oc oc oc

Reference -89.92 -54.92 -19.95 IS.JO 50.12 -89.92 -54.92 -19.95 IS.JO 50.12 -89.92 -54.92 -19.95 15.10 50.12 -89.92 -54.92 -19.95

Calibration Date Next Last 28-Feb-l 7 29-Feb-16 21-May-16 21-Mav-17

Serial# 1609 802212 I Year Spec. 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35 0.35

Measurement Uncertain!)' 0.04 0.04 0.04 0.04 0.04 0.04 0.04 0.04 0.04 0.04 0.04 0.04 0.04 0.04 0.04 0.04 0.04 0.04

As Found Diff. Result -89.92 0.00 -54.93 0.01 -19.96 0.01 15.09 0.01 50.12 0.00 0.00 -89.92 -54.93 0.01 -19.95 0.00 15.08 0.02 50.11 0.01 O.QI -89.91 -54.92 0.00 0.01 -19.94 15.09 0.01 50.13 0.01 -89.91 0.01 0.00 -54.92 0.01 -19.94

p

p p p p p

p p p p

p p p p p

p p

p

As Left Result Diff. NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA

u u u u u u u u u u u u u u u u u u

P indicates Pass, F indicales Fail, 0 indicates overlaps guard band, U indica/es tolerance unspecified

Maintaining Calibration

Calibration

The electronic components in this data logger are of the highest quality. The unit has been designed to remain within its specifications. The length of in-calibration service can be affected by aging, temperature and shock. For those users •th critical needs such as accreditation demands, s or ISO requirements, we recommend that the ·riodic

Information on calibration services is available at the address below. This data logger was calibrated by: Vaisala Inc. 10-D Gill Street Woburn, MA USA 01801 Toll Free: 1-800-683-8374, Email: [email protected], Website: www.vaisala.com

c,::::;?ian

Calibration Date:

23-Aug-2016

Next Calibration:

23-Aug-2017

Calibration Laboratory Manager: Page l of2 This certificate is copyrighted by Vaisala Inc., and may not be reproduced, except in full, without written approval from Vaisala Inc.

VAISALA

Calibration Certificate

C11tbr11lon • Cutlffea1• No. 2012.01

AS FOUND data only Your Company Your Address

Ambient Tempcralurc(0 ): Min: 24.4 Max: 27.5 CertificaLc Number: 0242278 Ambient Humidity(%Rlll; Min: 36.2 Max: 54.0 Model Number: VL-1416-44V Calibration by comparison 15111051 Serial Number: Method: Asset ID: Procedures: VCPI005 The calibration(s) on this report are traceable to the United States of America National Institute of Standards and Technology or to other recognized national or international standards or to accepted values of natural physical constants, and are accredited to JSO/JEC 17025. The laboratory meets the requirements of ANSI/NCSL 2540-1. Using methods detailed in the ISO "Guide to the Expression of Uncertainty in Measurement", reported uncertainties represent expanded uncertainties expressed at approximately the 95% confidence level using a coverage factor ofk=2. The calibrations were performed equivalently either with minimum test uncertainty ratios of4:1 using a coverage factor of k = 2, or with the statistical method of guard banding to reduce test limits. The results relate only to the item(s) calibrated. 4 4

oc oc

Temperature Temperature

15.10 50.12

0.35 0.35

0.04 0.04

15.09 50.13

0.01 0.01

p p

NA NA

NA NA

u u

P indicates Pass, F indicates Fail, 0 indicates overlaps guard band, U indicates tolerance unspecified Calibration Maintaining Calibration

Information on calibration services is available at the address below. This data logger was calibrated by: Vaisala Inc. I 0-D Gill Street Woburn, MA USA 01801 Toll Free: 1-800-683-8374, Email: [email protected], Website: www.vaisala.com Calibration Date:

23-Aug-2016

Next Calibration:

23-Aug-2017

Calibration Laboratory Manager: Page 2 of2 This certificate is copyrighted by Vaisala Inc., and may not be reproduced, except in foll, without written approval from Vaisala Inc.