Blood Flow Measurement Techniques

04.12.2014 Blood Flow Measurement Techniques !  not only blood pressure but blood flow important information !  describes the supply of blood in org...
Author: Dwight Jordan
16 downloads 1 Views 2MB Size
04.12.2014

Blood Flow Measurement Techniques !  not only blood pressure but blood flow important

information !  describes the supply of blood in organs and tissue !  abnormal changes in the blood flow or blood velocity gives

rise to malformatuon of vessels !  blood flow Q: volume of blood per time [ml/min] !  perfusion: blood supply in capillary bed [ml/min/100g]

PD Dr. Zöllner I Folie 155 I 9/9/2014

Blood Flow Values Body

Typical values for blood flow [cm/s]: Aorta 100 – 250 Abdominal 100 Carotis commuis 80 – 150 Vena Cava 5 – 40

PD Dr. Zöllner I Folie 156 I 9/9/2014

1

04.12.2014

Nicols & O‘Rourke, McDonald‘s Blood Flow in Arteries

Blood Flow and Pressure Waves Abdominal Vasculature

PD Dr. Zöllner I Folie 157 I 9/9/2014

Electromagnetic flow meter !  based on Faraday‘s law of induction !  blood is a conductor of electricity !  can be ‚aligned‘ by external magnet field

Nicols & O‘Rourke, McDonald‘s Blood Flow in Arteries PD Dr. Zöllner I Folie 158 I 9/9/2014

2

04.12.2014

Blood velocity measurement using US

PD Dr. Zöllner I Folie 159 I 9/9/2014

Typical Values

speed of sound: velocity of blood: transmitted US-frequency:

Doppler-frequency " Audio Range PD Dr. Zöllner I Folie 160 I 9/9/2014

3

04.12.2014

Continuous Wave (CW) - US sender

fS

sender mixer IM

bandpass

base wave

receiver fE

flow signal fS-fE

display

echo

product

Doppler

PD Dr. Zöllner I Folie 161 I 9/9/2014

Pulse Wave (PW) - US

PD Dr. Zöllner I Folie 162 I 9/9/2014

4

04.12.2014

US Doppler Probes - Invasive

Nicols & O‘Rourke, McDonald‘s Blood Flow in Arteries

PD Dr. Zöllner I Folie 163 I 9/9/2014

MR Blood Flow Measurements (1)

#  spins moving along a magnetic field gradient acquire a difference in the phase (right) #  amount of phase difference proportional to the velocity of the moving spins PD Dr. Zöllner I Folie 164 I 9/9/2014

Lotz et al. , RadioGraphics 2002

5

04.12.2014

MR Blood Flow Measurements (2)

#  flow encoding by means of difference in the bipolar gradients #  acquisition of two images Lotz et al. , RadioGraphics 2002 #  phase difference by substraction PD Dr. Zöllner I Folie 165 I 9/9/2014

MR Blood Flow Measurements (3) Phase difference expressed in formulars:

γ Δm ν

gyromagnetic ratio difference of the first moment of the gradient time curve velocity PD Dr. Zöllner I Folie 166 I 9/9/2014

6

04.12.2014

MR Blood Flow Measurements (4)

#  ΔM relates to the velocity encoding (VENC) #  gives sensitivity of measurable velocities #  relates to the gradient strength PD Dr. Zöllner I Folie 167 I 9/9/2014

Diagram of MR PC sequence

PD Dr. Zöllner I Folie 168 I 9/9/2014

7

04.12.2014

PC-MR Images #  receive 2 images #  magnitude image with velocity #  phase image with velocity and direction Flow encoding needs an operator defined value: VENC [cm/sec] -VENC < v < VENC PD Dr. Zöllner I Folie 169 I 9/9/2014

3D PC MRI #  flow encoding is directional #  depends on the gradient axis applied #  therefore: 3D flow measurements possible in each voxel

PD Dr. Zöllner I Folie 170 I 9/9/2014

8

04.12.2014

Examples

Courtesey of J Henning, Freiburg

PD Dr. Zöllner I Folie 171 I 9/9/2014

MIP Projection – Renal Artery

PD Dr. Zöllner I Folie 173 I 9/9/2014

9

04.12.2014

RAS, Secondary Hypertension, etc. Parameters to detect status: !  Flow reduction !  Flow velocity !  Kidney size !  Overall and split renal

function

!  RAS leading cause for

secondary hypertension " CKD !  Reduced flow triggers autoregulation of systemic circulation !  Loss of renal parenchyma in the stenosed kidney !  Microangiopathy in the contralateral kidney

Allow for stenosis grading and better selection of patients for PTA PD Dr. Zöllner I Folie 174 I 9/9/2014

Example - Acquisition Protocol ECG gated PC cine MR sequence with: !  TE=4ms, TR=37ms, FA=30° !  VENC=100 cm/s !  ca. 20 images / cardiac cycle !  Resolution 0.5 x0.5 mm !  Slice thickness 6mm

1.5T Siemens Symphony Vision

PD Dr. Zöllner I Folie 175 I 9/9/2014

10

04.12.2014

Example – Slice Positioning

a.  Abdominal aorta b.  Images acquired from MR angiography c.  Image slice through the vessel d.  Imaging plane for PC-MRI (perpendicular to c)

Schönberg et al. , Radiology 1997

PD Dr. Zöllner I Folie 176 I 9/9/2014

Clinical Interpretation of Flow Profile

#  4 section within RR interval #  Systolic peak (I) #  Incision (II) followed by midsystolic peak (III) #  Diastolic phase (IV)

PD Dr. Zöllner I Folie 177 I 9/9/2014

Schönberg et al. , Radiology 1997

11

04.12.2014

Flow Profile in Stenoic Patient

#  First systolic peak and incision disappear with increasing degree of stenosis

Schönberg et al. , Radiology 1997 PD Dr. Zöllner I Folie 178 I 9/9/2014

Flow Measurement in Stenosis Model

measurement setup

stenosis model at different flow rates

Scheuer et al. Z Med Phys 2010 PD Dr. Zöllner I Folie 179 I 9/9/2014

12

04.12.2014

Flow Measurement in Stenosis Model

flow profiles from 3D PC-MRI at different locations along the stenosis model

Scheuer et al. Z Med Phys 2010 PD Dr. Zöllner I Folie 180 I 9/9/2014

Summary !  Basic laws to describe pressure and flow: Bernoulli & Poiseuille !  Blood flow in normal regimes laminar flow # 

Reynolds number

# 

Parabolic shape of flow in tubes

!  Blood pressure measurement techniques # 

Korotkof sounds

# 

Cuff & stetoscope

!  Blood flow measurement techniques # 

EM flow meters

# 

Doppler US

# 

PC-MRI PD Dr. Zöllner I Folie 181 I 9/9/2014

13

Suggest Documents