AGFA HEALTHCARE DICOM Conformance Statement

HE/001359 Page 1 of 33 Document No. 001359 - Revision: 1.1 Livelink NodeID: 32940311 Agfa HealthCare AGFA HEALTHCARE DICOM Conformance Statement I...
Author: Matthew Blair
9 downloads 0 Views 349KB Size
HE/001359

Page 1 of 33 Document No. 001359 - Revision: 1.1 Livelink NodeID: 32940311

Agfa HealthCare

AGFA HEALTHCARE DICOM Conformance Statement

IMPAX EE Mammography

Document No. 001359 Revision: 1.1 Livelink NodeID: 32940311

When printed, this is NOT a controlled copy

HE/001359

Page 2 of 33 Document No. 001359 - Revision: 1.1 Livelink NodeID: 32940311

Agfa HealthCare

23 May, 2011

Document Information

Service-related contact information worldwide

Issued by: Agfa HealthCare SIV Connectivity Septestraat 27 B-2640 Mortsel Belgium

tel: +32 (0)3 444 7588 email: [email protected]

All service-related contact information is available on this URL

http://www.agfahealthcare.com/global/en/main/contact/index. jsp

Agfa shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance or use of this publication. Agfa reserves the right to revise this publication and to make changes to its content at any time, without obligation to notify any person or entity of such revisions and changes. This publication may only be used in connection with the promotion, sales, installation and use of Agfa equipment. Copyright  May, 11 Agfa HealthCare All rights reserved

HE/001359

Page 3 of 33 Document No. 001359 - Revision: 1.1 Livelink NodeID: 32940311

Agfa HealthCare

23 May, 2011

Conformance Statement Overview This DICOM conformance Statement refers to IMPAX EE Mammography. IMPAX EE Mammography conforms DICOM 3.0 2009 standard. IMPAX EE Mammography supports Standard Conformance to the SOP classes of Table 1.1-1. These SOP classes are included in DICOM Supplement 64. Table 1.1-1: Network Services Supported SOP Classes

Verification Computed Radiography Image Storage Digital Mammography X-Ray Image Storage – For Presentation Ultrasound Multi-frame Image Storage Secondary Capture Image Storage Multi-frame Grayscale Byte Secondary Capture Image Storage Multi-frame Grayscale Word Secondary Capture Image Storage Multi-frame True Color Secondary Capture Image Storage Grayscale Softcopy Presentation State Storage SOP Class Basic Text SR Storage Enhanced SR Storage Comprehensive SR Storage Key Object Selection Document Storage

User of Service (SCU)

Provider of Service (SCP)

Display

Yes

N/A

Yes

Yes

Yes

Yes

Yes

Yes Yes

Yes Yes

Yes Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes Yes Yes Yes

Yes Yes Yes Yes

No

N/A

No

N/A

No

N/A

No

N/A

No No No

N/A N/A N/A

Verification Yes Transfer Yes

Yes Yes Yes Yes Query/Retrieve

Patient Root Query/Retrieve Information Model – Yes FIND Patient Root Query/Retrieve Information Model – Yes MOVE Study Root Query/Retrieve Information Model – Yes FIND Study Root Query/Retrieve Information Model – Yes MOVE Workflow Management Modality Worklist Information Model – FIND Yes Storage Commitment Push Model SOP Class Yes Modality Performed Procedure Step SOP Class Yes

HE/001359

Page 4 of 33 Document No. 001359 - Revision: 1.1 Livelink NodeID: 32940311

Agfa HealthCare

23 May, 2011

Table 1.1-2: Media Service Supported Media Storage Application Profile Compact Disk - Recordable General Purpose CD-R

Write Files (FSC or FSU)

Read Files (FSR)

No

Yes

HE/001359

Page 5 of 33 Document No. 001359 - Revision: 1.1 Livelink NodeID: 32940311

Agfa HealthCare

23 May, 2011

Table of Contents 1 1.1 1.2 1.3 1.3.1 1.3.2 1.4 1.5

2

Introduction ............................................................................................7 Revision Record ................................................................................................... 7 Purpose and Intended Audience of this Document ................................................ 7 General Remarks ................................................................................................. 7 Integration and Validation Activities.................................................................. 7 Future Evolution .............................................................................................. 7 Acronyms and Abbreviations ................................................................................ 8 Related Documents .............................................................................................. 8

Networking .............................................................................................9

2.1 Implementation Model .......................................................................................... 9 2.1.1 Application Data Flow Diagram.......................................................................10 2.1.2 Functional Definitions of AE’s .........................................................................11 2.1.2.1 Functional Capability of IMPAX EE Mammography CLIENT.......................11 2.1.3 Sequencing of Real World Activities................................................................11 2.2 AE Specifications.................................................................................................12 2.2.1 IMPAX EE Mammography CLIENT Specification ............................................12 2.2.1.1 SOP Classes Supported ............................................................................12 2.2.1.2 Association Establishment Policies ............................................................13 2.2.1.2.1 General ................................................................................................13 2.2.1.2.2 Number of Associations ........................................................................13 2.2.1.2.3 Asynchronous Nature ...........................................................................13 2.2.1.2.4 Implementation Identifying Information..................................................13 2.2.1.3 Association Initiation Policies .....................................................................14 2.2.1.3.1 Verify Connection (SCU).......................................................................14 2.2.1.3.1.1 Description and Sequencing of Activity 14 2.2.1.3.1.2 Proposed Presentation Contexts 14 2.2.1.3.1.3 SOP Specific Conformance – Verification 14 2.2.1.3.2 Query images, KO, GSPS and SR ........................................................14 2.2.1.3.2.1 Description and Sequencing of Activity 14 2.2.1.3.2.2 Proposed Presentation Contexts 15 2.2.1.3.2.3 SOP Specific Conformance Statement for Q/R Study Root Information Model – FIND 16 2.2.1.3.2.4 SOP Specific Conformance Statement for Q/R Patient Root Information Model – FIND 17 2.2.1.3.3 Retrieve images, KO, GSPS and SR.....................................................17 2.2.1.3.3.1 Description and Sequencing of Activity 17 2.2.1.3.3.2 Proposed Presentation Contexts 17 2.2.1.3.4 Store images, GSPS, KO and SR .........................................................18 2.2.1.3.4.1 Description and Sequencing of Activity 18 2.2.1.3.4.2 Proposed Presentation Contexts 18 2.2.1.3.5 Storage Commitment (Request)............................................................18 2.2.1.3.5.1 Description and Sequencing of Activity 18 2.2.1.3.5.2 Proposed Presentation Contexts 18 2.2.1.3.6 Query Modality Worklist ........................................................................19 2.2.1.3.6.1 Description and Sequencing of Activity 19 2.2.1.3.6.2 Proposed Presentation Contexts 20 2.2.1.3.6.3 SOP Specific Conformance – Modality Worklist Information Model – FIND 20 2.2.1.3.7 Modality Performed Procedure Step (SCU)...........................................21 2.2.1.3.7.1 Description and Sequencing of Activity 21

HE/001359

Page 6 of 33 Document No. 001359 - Revision: 1.1 Livelink NodeID: 32940311

Agfa HealthCare

23 May, 2011

2.2.1.3.7.2 Proposed Presentation Contexts 21 2.2.1.3.7.3 SOP Specific Conformance – MPPS 21 2.2.1.4 Association Acceptance Policies................................................................24 2.2.1.4.1 Verify Connection .................................................................................24 2.2.1.4.1.1 Description and Sequencing of Activity 24 2.2.1.4.1.2 Accepted Presentation Contexts 24 2.2.1.4.2 Receive images, GSPS, KO and SR.....................................................24 2.2.1.4.2.1 Description and Sequencing of Activity 24 2.2.1.4.2.2 Accepted Presentation Contexts 24 2.2.1.4.2.3 SOP Specific Conformance Statement for Storage SOP classes 25 2.2.1.4.3 Storage commitment (Report)...............................................................25 2.2.1.4.3.1 Description and Sequencing of Activity 25 2.2.1.4.3.2 Accepted Presentation Contexts 25 2.2.1.4.3.3 SOP Specific Conformance Statement for Storage Commitment (Report) 26 2.3 Network Interfaces...............................................................................................26 2.3.1 Physical Medium Support ...............................................................................26 2.4 Configuration.......................................................................................................27 2.4.1 AE Title/ Presentation Mapping.......................................................................27 2.4.1.1 Local AE Titles ..........................................................................................27 2.4.1.2 Remote AE Titles.......................................................................................27 2.4.2 Configuration Parameters...............................................................................27

3

Media Interchange................................................................................28

3.1 3.1.1 3.1.2 3.1.3 3.1.4 3.2 3.2.1 3.2.1.1

Implementation Model .........................................................................................28 Application Data Flow Diagram.......................................................................28 Functional Definition of IMPAX EE Mammography CLIENT AE.......................28 Sequencing of Real World Activities................................................................29 File Meta Information for Implementation Class and Version ...........................29 AE Specifications.................................................................................................29 IMPAX EE Mammography CLIENT AE Specification ......................................29 File Meta Information for Implementation IMPAX EE Mammography CLIENT 29 3.2.1.2 Real World Activities..................................................................................29 3.2.1.2.1 Real World Activity – Import DICOM objects from CD ...........................29 3.2.1.2.1.1 Media Storage Application Profile 29 3.2.1.2.2 Real World Activity – Load DICOM objects from CD..............................30 3.2.1.2.2.1 Media Storage Application Profile 30

4

Support for Extended Character Sets...................................................31

5

Security ................................................................................................32

5.1 5.2 5.3

6 6.1 6.1.1 6.2

Security Profiles ..................................................................................................32 Association Level Security ...................................................................................32 Application Level Security....................................................................................32

Annexes ...............................................................................................33 IOD Contents.......................................................................................................33 Coerced/ Modified Fields ................................................................................33 Grayscale Image Consistency .............................................................................33

HE/001359

Page 7 of 33 Document No. 001359 - Revision: 1.1 Livelink NodeID: 32940311

Agfa HealthCare

23 May, 2011

1

INTRODUCTION

1.1

Revision Record Dicom Conformance Statement IMPAX EE Mammography Revision Number 1.0 1.1

1.2

Date 2011-03-31 2011-05-12

Reason for Change Initial version Update document after review

Purpose and Intended Audience of this Document This document is a DICOM Conformance Statement for the DICOM Services of the IMPAX EE Mammography product. The user of this document is involved with system integration and/or software design. We assume that the reader is familiar with the terminology and concepts that are used in the DICOM 3.0 standard and the IHE Technical Framework. Readers not familiar with DICOM 3.0 terminology should first read the appropriate parts of the DICOM standard itself, prior to reading this conformance statement. Although the use of this conformance statement in conjunction with the DICOM 3.0 standard is intended to facilitate communication with Agfa IMPAX EE Mammography, it is not sufficient to guarantee, by itself, the inter-operation of the connection. The following issues need to be considered.

1.3

General Remarks

1.3.1

Integration and Validation Activities The integration of any device into a system of interconnected devices goes beyond the scope of the DICOM 3.0 standard and this conformance statement when interoperability is desired. The responsibility for analyzing the applications requirements and developing a solution that integrates the Agfa equipment with other vendors’ systems is the user’s responsibility and should not be underestimated. In some circumstances it might be necessary to perform a validation to make sure that functional interoperability between the Agfa equipment and non-Agfa devices works as expected. The user should ensure that any non-Agfa provider accepts responsibility for any validation required for their connection with the Agfa equipment.

1.3.2

Future Evolution As the DICOM 3.0 standard evolves to meet the user’s growing requirements and to incorporate new features and technologies, Agfa will follow the evolution of the standard. This evolution of the standard may require changes to devices that have implemented DICOM 3.0. The user should ensure that any non-Agfa provider, who connects with Agfa devices, also plans for future evolution of the DICOM standard. A refusal to do so may result in the loss of functionality and/or connectivity between the different products.

HE/001359

Page 8 of 33 Document No. 001359 - Revision: 1.1 Livelink NodeID: 32940311

Agfa HealthCare

1.4

23 May, 2011

Acronyms and Abbreviations Definitions, terms and abbreviations used in this document are defined within the different parts of the DICOM standard. Abbreviations and terms are as follows:

1.5

AE

DICOM Application Entity

CD-R

Compact Disk Recordable

DICOM

Digital Imaging and Communications in Medicine

DIMSE

DICOM Message Service Element

DVD

Digital Versatile Disc or Digital Video Disc

FSC

File-Set Creator

FSR

File-Set Reader

FSU

File-Set Updater

GSPS

Grayscale Softcopy Presentation State

IE

Information Entity

IOD

(DICOM) Information Object Definition

IP

Internet Protocol

ISO

International Standard Organization

KO

Key Object

MPPS

Modality Performed Procedure Step

PDU

DICOM Protocol Data Unit

SCP

DICOM Service Class Provider (DICOM server

SCU

DICOM Service Class User (DICOM client)

SMTP

Simple Mail Transfer Protocol

SOP

DICOM Service-Object Pair

SR

Structure Report

S/MIME

Secure/Multipurpose Internet Mail Extensions

TCP

Transmission Control Protocol

UID

Unique Identifier

VR

Value Representation

Related Documents 

ACR-NEMA Digital Imaging and Communications in Medicine (DICOM) V3.0.



IHE Radiology Technical Framework Revision 10 – Final Text, Feb 2011

HE/001359

Page 9 of 33 Document No. 001359 - Revision: 1.1 Livelink NodeID: 32940311

Agfa HealthCare

23 May, 2011

2

NETWORKING

2.1

Implementation Model All DICOM Services provided by IMPAX EE Mammography are implemented in one process, which is launched and terminated by the user. This process hosts a DICOM server listening on different ports that requires the assignation of an Application Entity Title to the application. That motivates to describe the DICOM functionality of IMPAX EE Mammography as provided an Application Entity: 

IMPAX EE Mammography CLIENT provides DICOM Services related with images and DICOM Structured Report Documents that includes Storage Services to receive images of MG, CR, US modalities, Grayscale Presentation States (GSPS), Key Object (KO) and Structure Report (SR) Documents. Selection Documents, but also the Storage Commitment Push Model Service as Service Class User (SCU), and Query/Retrieve Services, to query archives for available images, GSPS and KO objects and repositories for available SR documents. These Services also retrieve such DICOM information objects triggered by user interactions.

HE/001359

Page 10 of 33 Document No. 001359 - Revision: 1.1 Livelink NodeID: 32940311

Agfa HealthCare

2.1.1

23 May, 2011

Application Data Flow Diagram Dicom Interface

Verify Connection Verify Connection Load Images, KO GSPS and SR

Query KO,GSPS, Images and SR Retrieve KO,GSPS, Images and SR

Display Images, KO GSPS and SR

Receive KO, GSPS, Images and SR

IMPAX EE Mammography CLIENT

Find Objects Modality Performed Procedure Step

Move Objects

Storage Commitment

Query Modality Worklist

.

HE/001359

Page 11 of 33 Document No. 001359 - Revision: 1.1 Livelink NodeID: 32940311

Agfa HealthCare

23 May, 2011

2.1.2

Functional Definitions of AE’s

2.1.2.1

Functional Capability of IMPAX EE Mammography CLIENT     

2.1.3

Verify the connection to DICOM peer Application Entities (AEs) by acting as DICOM Verification SCU and SCP Query and retrieve images, SR, GSPS and KO from DICOM peer AEs by acting as DICOM Query/Retrieve SCU. Receive images, GSPS, KO and SR from DICOM peer AEs by acting as SCP for various Image Storage SOP Classes, DICOM Grayscale Presentation State Storage SOP Class, and KO Selection Document Storage SOP Class. Request storage commitment of image, KO, SR and GSPS Information Entities (IEs) to DICOM peer AEs by acting as DICOM Storage Commitment Push Model SCU. Send query to a Modality Worklist by acting as SCU and then sends a Modality Performed Procedure Step (MPPS) message to a remote AE.

Sequencing of Real World Activities Any retrieval of images by the IMPAX EE Mammography AE from a peer AE has preceded by – one or several – query requests to the peer AE, to get information about available images and values of study/series/image identifiers necessary for creating retrieve requests. Typically the receipt of images by the IMPAX EE Mammography AE from a peer AE is caused by an Image Retrieve Request to this peer AE by the IMPAX EE Mammography AE. The IMPAX EE Mammography AE creates derived images from source images. Image IEs have to be received by the IMPAX EE Mammography AE, before they can be stored to or printed at peer AEs.

HE/001359

Page 12 of 33 Document No. 001359 - Revision: 1.1 Livelink NodeID: 32940311

Agfa HealthCare

23 May, 2011

2.2

AE Specifications

2.2.1

IMPAX EE Mammography CLIENT Specification

2.2.1.1

SOP Classes Supported IMPAX EE Mammography implements DICOM Verification Service, Query/Retrieve Services, Storage, Services for Image, Key Objects, SR Documents, Grayscale Presentation State IEs, the Storage Commitment Push Model SOP Class. Table 2.2-1: SOP Class(es) for IMPAX EE Mammography CLIENT SOP Class Name

SOP Class UID

SCU

SCP

Yes

Yes

Yes Yes Yes Yes

Yes Yes Yes Yes

Yes

Yes

Yes Yes Yes Yes

Yes Yes Yes Yes

Yes Yes Yes Yes

Yes Yes Yes Yes

Yes Yes No

No No Yes

Yes Yes Yes Yes

No No No No

Verification Verification SOP Class

1.2.840.10008.1.1 Transfer

Computed Radiography Image Storage 1.2.840.10008.5.1.4.1.1.1 Digital Mammography X-ray Image Storage For Presentation 1.2.840.10008.5.1.4.1.1.1.2 Secondary Capture Image Storage 1.2.840.10008.5.1.4.1.1.7 Multi-frame Grayscale Byte Secondary Capture Image 1.2.840.10008.5.1.4.1.1.7.2 Storage Multi-frame Grayscale Word Secondary Capture Image 1.2.840.10008.5.1.4.1.1.7.3 Storage Multi-frame True Color Secondary Capture Image Storage 1.2.840.10008.5.1.4.1.1.7.4 Ultrasound Image Storage 1.2.840.10008.5.1.4.1.1.6.1 Ultrasound Multi-frame Image Storage 1.2.840.10008.5.1.4.1.1.3.1 Grayscale Softcopy Presentation 1.2.840.10008.5.1.4.1.1.11.1 State Storage SOP Class Key Object Selection Document Storage 1.2.840.10008.5.1.4.1.1.88.59 Basic Text SR Storage 1.2.840.10008.5.1.4.1.1.88.11 Enhanced SR Storage 1.2.840.10008.5.1.4.1.1.88.22 Comprehensive SR Storage 1.2.840.10008.5.1.4.1.1.88.33 Workflow Management Modality Worklist Information Model - FIND 1.2.840.10008.5.1.4.31 Modality Performed Procedure Step SOP Class 1.2.840.10008.3.1.2.3.3 Storage Commitment Push Model SOP Class 1.2.840.10008.1.20.1 Query Retrieve Patient Root Query/Retrieve Information Model – FIND 1.2.840.10008.5.1.4.1.2.1.1 Patient Root Query/Retrieve Information Model – MOVE 1.2.840.10008.5.1.4.1.2.1.2 Study Root Query/Retrieve Information Model – FIND 1.2.840.10008.5.1.4.1.2.2.1 Study Root Query/Retrieve Information Model – MOVE 1.2.840.10008.5.1.4.1.2.2.2

HE/001359

Page 13 of 33 Document No. 001359 - Revision: 1.1 Livelink NodeID: 32940311

Agfa HealthCare

23 May, 2011

2.2.1.2

Association Establishment Policies

2.2.1.2.1

General The DICOM standard application context name, which is always proposed, is listed in next Table 2.2-2. Table 2.2-2: DICOM Application context

Application Context Name

1.2.840.10008.3.1.1.1

The maximum PDU length is 16384 bytes. SOP Class extended negotiation is supported.

2.2.1.2.2

Number of Associations Typically the IMPAX EE Mammography AE initiates only one DICOM Association concurrently, but potentiality the number of parallel associations is only limited by the resources of the underlying operating system. Several DICOM Associations will be initiated concurrently, if the user launches a new Query/Retrieve, before an already performing Query/Retrieve operation was finished. Several open DICOM Associations may also occur during Storage Commitment, if the Storage Commitment SCP peer AE sends the Commitment Result in a separate Association – initiated by the peer AE and accepted by the IMPAX EE Mammography AE.

2.2.1.2.3

Asynchronous Nature Configurable via configurations file. At default, asynchronous mode is disabled.

2.2.1.2.4

Implementation Identifying Information

Table 2.2-3: DICOM implementation Class and Version for IMPAX EE Mammography CLIENT Implementation Class UID

1.2.40.0.13.1.1

Implementation Version Name

dcm4che-2.0

HE/001359

Page 14 of 33 Document No. 001359 - Revision: 1.1 Livelink NodeID: 32940311

Agfa HealthCare

23 May, 2011

2.2.1.3

Association Initiation Policies

2.2.1.3.1

Verify Connection (SCU)

2.2.1.3.1.1

Description and Sequencing of Activity The user of the IMPAX EE Mammography Application initiates a Verification Request by selecting the corresponding button in the AE Configuration Panel on the user interface.

2.2.1.3.1.2

Proposed Presentation Contexts

Table 2.2-4: Presentation Contexts Proposed by IMPAX EE Mammography CLIENT Presentation Context Table Abstract Syntax Name Verification SOP Class

2.2.1.3.1.3

UID 1.2.840.10008.1.1

Transfer Syntax Name List Implicit VR, Little Endian

UID List 1.2.840.10008.1.2

Role SCU

Extended Negotiation None

SOP Specific Conformance – Verification IMPAX EE Mammography CLIENT provides standard conformance to the DICOM Verification SOP Class as SCU.

2.2.1.3.2

Query images, KO, GSPS and SR

2.2.1.3.2.1

Description and Sequencing of Activity The user of the IMPAX EE Mammography Application initiates the Query Request by selecting the corresponding button or expanding a node in the Patient/Study Folder on the user interface. Wild card or specific information can be specified for various Patient, Study, Series and/or Image attributes. If the user triggers a query in sequence level, than the user also triggered a retrieve and a receive for KO and GSPS (if existing). The user of IMPAX EE Mammography Application initiates the Query Request by selecting an entry representing one Study in the Patient Folder

HE/001359

Page 15 of 33 Document No. 001359 - Revision: 1.1 Livelink NodeID: 32940311

Agfa HealthCare

23 May, 2011

2.2.1.3.2.2

Proposed Presentation Contexts The default behavior of IMPAX EE Mammography is to propose as SCU both Presentation Contexts. Table 2.2-5: Proposed Presentation Contexts Presentation Context Table Abstract Syntax

Transfer Syntax UID List

Role

Extended Negotiation

Name

UID

Name List

Query/Retrieve Patient Root Info. Model – FIND

1.2.840.10008.5.1.4.1.2.1.1

Implicit VR, Little Endian

1.2.840.10008.1.2

SCU

Yes

Query/Retrieve Study Root Info. Model – FIND

1.2.840.10008.5.1.4.1.2.2.1

Implicit VR, Little Endian

1.2.840.10008.1.2

SCU

Yes

If both presentation contexts are accepted by the peer AE, IMPAX EE Mammography will use the Query/Retrieve Study Root Info. Model – FIND SOP class for query requests. The used query level depends on the selection or the type of node at the Patient/Study Folder on the user interface which was expanded: Table 2.2-6: Expanded item for used query level selected/expanded item corresponds

used Query Level

None

PATIENT / STUDY (SOP specific)

Patient

STUDY

Study

STUDY

Series

SERIES

Image

IMAGE

HE/001359

Page 16 of 33 Document No. 001359 - Revision: 1.1 Livelink NodeID: 32940311

Agfa HealthCare

23 May, 2011

Following keys are supported by the SCU of both SOP classes: Table 2.2-7: Attributes for FIND Level Patient

Study

Series

Image

Attributes Name Patient Name Patient ID Issuer of Patient ID Patient’s Birth Date Patient’s Sex Study Date Study Time Study Description Accession Number Study ID Study Instance UID Modalities in Study Referring Physician’s Name Modality Body Part examined Requesting physician Requesting service Institution name Station Name Series Number Series Instance UID Request Attribute Sequence >Requested Procedure ID >Scheduled Procedure Step ID Performed Procedure Step Start Date Performed Procedure Step Start Time Institutional Department Name Instance Number Content Date Verification Flag SOP Instance UID

Tag (0010,0010) (0010,0020) (0010,0021) (0010,0030) (0010,0040) (0008,0020) (0008,0030) (0008,1030) (0008,0050) (0020,0010) (0020,000D) (0008,0061) (0008,0090) (0008,0060) (0018,0015) (0032,1032) (0032,1033) (0008,0080) (0008,1010) (0020,0011) (0020,000E) (0040,0275) (0040,1001) (0040,0009) (0040,0244) (0040,0245) (0008,1040) (0020,0013) (0008,0023) (0040,A493) (0008,0018)

Matching * S, * S R, S S R, S S, * S, * S, * L S * S S, * * S, * S, * S, * S, * L Sequence S S R, S S, * S, * R, S S L

The types of matching are: “*” is Wild Card matching, “R” is Range matching, “L” is List of UID Matching and Sequence matching, “S” is Single value matching.

2.2.1.3.2.3

SOP Specific Conformance Statement for Q/R Study Root Information Model – FIND Queries initiated by the user without selection of any Patient/Study entry are performed with query level STUDY. When querying SR objects, queries are initiated by the user selection of the Study entry (as indicated in Table 2.2-6 are performed with query level SERIES) and matching key Modality (0008,0060) = SR. If the Query Response contains entries for matching Series, a Retrieve for the SR Objects of this Series will be initiated.

HE/001359

Page 17 of 33 Document No. 001359 - Revision: 1.1 Livelink NodeID: 32940311

Agfa HealthCare

23 May, 2011

2.2.1.3.2.4

SOP Specific Conformance Statement for Q/R Patient Root Information Model – FIND By default, IMPAX EE Mammography Application uses Query/Retrieve Study Root Info. Model – FIND. In case that server does not support this SOP Class, IMPAX EE Mammography uses Query/Retrieve Patient Root Info. Model – FIND. Queries initiated by the user without selection of any Patient/Study entry are performed with query level STUDY. When querying SR objects, queries are initiated by the user selection of the Study entry (as indicated in Table 2.2-6 are performed with query level SERIES) and matching key Modality (0008,0060) = SR. If the Query Response contains entries for matching Series, a Retrieve for the SR Objects of this Series will be initiated.

2.2.1.3.3

Retrieve images, KO, GSPS and SR

2.2.1.3.3.1

Description and Sequencing of Activity The user of the IMPAX EE Mammography Application initiates the Retrieve Images, KO and GSPS Request by selecting the corresponding button in the Patient/Study Folder on the user interface, after selecting KO, studies/sequences with GSPS, single images, series or whole studies which shall be retrieved. The reception of images is reflected by a status change of the corresponding item in the Patient/Study Folder. Retrieve SR will be triggered by a preceding Query SR, with non empty Response.

2.2.1.3.3.2

Proposed Presentation Contexts The default behavior of IMPAX EE Mammography is to propose as SCU following Presentation Context. Table 2.2-8: Proposed Presentation Contexts Presentation Context Table Abstract Syntax

Transfer Syntax

Name

UID

Name List

Query/Retrieve Study Root Info. Model – MOVE

1.2.840.10008.5.1.4.1.2.2.2

Implicit VR, Little Endian

UID List 1.2.840.10008.1.2

Role SCU

Extended Negotiation None

If the presentation context is accepted by the peer AE, IMPAX EE Mammography will use the Query/Retrieve Study Root Info. Model – MOVE SOP class for retrieve requests. The used retrieve level depends on the selection at the Patient/Study Folder on the user: Table 2.2-9: Selected item for used Retrieve Level selected/expanded item corresponds None Patient Study Series Image

used Query Level PATIENT / STUDY (SOP specific) STUDY STUDY SERIES IMAGE

HE/001359

Page 18 of 33 Document No. 001359 - Revision: 1.1 Livelink NodeID: 32940311

Agfa HealthCare

23 May, 2011

2.2.1.3.4 Store images, GSPS, KO and SR 2.2.1.3.4.1

Description and Sequencing of Activity

The user of the IMPAX EE Mammography Application selects single images, series or complete studies and the destination AE in the local Patient/Study Folder before initiating the Image Storage by selecting the corresponding button on the user interface. Importing images, GSPS, KO and SR from external source or archive is the only possible way for storing such objects. At default, all Grayscale Presentation State Objects associated with selected images will be stored to the selected import destination. Also the selected Key Objects will be stored. The storage of images is reflected by a status change of the corresponding item in the Patient/Study Folder. If the destination AE also supports the Storage Commitment Push Model service - or a separate AE is associated with the destination AE to perform the Storage Commitment operation -, at default configuration, IMPAX EE Mammography sends a Storage Commitment request for the transmitted images and GSPS objects immediately after the storage operation was finished.

2.2.1.3.4.2

Proposed Presentation Contexts

The default behavior of IMPAX EE is to propose as SCU following Presentation Context: Table 2.2-10: Proposed Presentation Contexts Presentation Context Table Abstract Syntax Name

Transfer Syntax UID

All Storage SOP Classes listed in Table 2.2-1

Name List Implicit VR, Little Endian

UID List 1.2.840.10008.1.2

2.2.1.3.5

Storage Commitment (Request)

2.2.1.3.5.1

Description and Sequencing of Activity

Role SCU

Extended Negotiation None

At default configuration, IMPAX EE Mammography sends a Storage Commitment request for the transmitted images and GSPS objects immediately after the storage operations was finished to the destination AE - or to an associated AE responsible for performing the Storage Commitment operation. The commitment result is reflected by a status change of the corresponding item in the Patient/Study Folder.

2.2.1.3.5.2

Proposed Presentation Contexts

IMPAX EE Mammography will propose the single Presentation Context of next

Table 2.2-11.

HE/001359

Page 19 of 33 Document No. 001359 - Revision: 1.1 Livelink NodeID: 32940311

Agfa HealthCare

23 May, 2011

Table 2.2-11: Proposed Presentation Contexts Presentation Context Table Abstract Syntax Name

Transfer Syntax UID

Storage Commitment Push Model SOP Class

Name List

1.2.840.10008.1.20.1

Implicit VR, Little Endian

UID List 1.2.840.10008.1.2

Role SCU

Extended Negotiation None

The following elements are sent as part of the N-ACTION request: Table 2.2-12: Attributes for N-ACTION Attribute Transaction UID Referenced SOP Sequence >Referenced SOP Class UID >Referenced SOP Instance UID

Tag (0008,1195) (0008,1199) (0008,1150) (0008,1155)

At default, IMPAX EE Mammography releases the association immediately after it receives the N-ACTION RSP to the initial Storage Commitment N-ACTION Request. That forces the peer AE, to send the N-EVENT-REPORT request containing the commitment result in a separate association, which has to be initiated – and released – by the peer AE. The default behavior can be changed, that IMPAX EE Mammography keeps the initial association open until the corresponding N-EVENT-REPORT request was received and immediately confirmed by IMPAX EE Mammography – no matter if in the same or in a separate association. If a complete success of the requested Storage Commitment is notified by the N-EVENTREPORT, the status of the concerned image and presentation state objects changes to “committed”, which enables the user to delete these objects. If existing failure of the requested Storage Commitment is notified by the N-EVENTREPORT, the user will be asked, if the status of image and presentation state objects for which a successful commitment was reported, shall be changed to “committed”.

2.2.1.3.6

Query Modality Worklist

2.2.1.3.6.1

Description and Sequencing of Activity The user of IMPAX EE Mammography Application initiates a Query request to a modality worklist when selecting in the import dialog the option import via modality worklist. Wild card or specific information can be specified for the searched attributes.

HE/001359

Page 20 of 33 Document No. 001359 - Revision: 1.1 Livelink NodeID: 32940311

Agfa HealthCare

2.2.1.3.6.2

23 May, 2011

Proposed Presentation Contexts

Table 2.2-13: Presentation Context Proposed by IMPAX EE Mammography CLIENT Presentation Context Table Abstract Syntax Name

UID

Modality Worklist Information Model – FIND

2.2.1.3.6.3

Transfer Syntax Name List

1.2.840.10008.5.1.4.31

Implicit VR, Little Endian

Role

UID List 1.2.840.10008.1.2

SCU

Extended Negotiation None

SOP Specific Conformance – Modality Worklist Information Model – FIND DICOM Message Service Element (DIMSE) Service C-FIND Attributes: Table 2.2-14: Supported Attributes from Modality Worklist Information Model Attribute Accession Number Admission ID Admitting Date Admitting Diagnoses Description Current Patient Location Institution Name Issuer of patient ID Modality Patient’s Birth Date Patient’s Name Patient’s Sex Patient ID Performing Physician’s Name Referring Physician’s Name Requested Contrast Agent Requested Procedure Description Requested Procedure ID Requested Procedure Priority Requesting Physician Scheduled Procedure Step Start Date Scheduled Procedure Step Description Scheduled Procedure Step Status Scheduled Station AE Title Scheduled Procedure Step ID

Tag (0008,0050) (0038,0010) (0038,0020) (0008,1080) (0038,0300) (0008,0080) (0010,0021) (0008,0060) (0010,0030) (0010,0010) (0010,0040) (0010,0020) (0008,1050) (0008,0090) (0032,1070) (0032,1060) (0040,1001) (0040,1003) (0032,1032) (0040,0002) (0040,0007) (0040,0020) (0040,0001) (0040,0009)

HE/001359

Page 21 of 33 Document No. 001359 - Revision: 1.1 Livelink NodeID: 32940311

Agfa HealthCare

23 May, 2011

Attribute

Tag

Station name Study Instance UID Visit Status ID

(0008,1010) (0020,000D) (0038,0008)

2.2.1.3.7

Modality Performed Procedure Step (SCU)

2.2.1.3.7.1

Description and Sequencing of Activity IMPAX EE Mammography sends a DICOM MPPS message to the server after importing via Modality worklist if this option is active in the configuration panel of the user interface.

2.2.1.3.7.2

Proposed Presentation Contexts

Table 2.2-15: Presentation Context Proposed by IMPAX EE Mammography CLIENT Presentation Context Table Abstract Syntax Name

UID

Modality Performed Procedure Step

2.2.1.3.7.3

Transfer Syntax Name List

1.2.840.10008.3.1.2.3.3

Implicit VR, Little Endian

UID List 1.2.840.10008.1.2

Role SCU

Extended Negotiation None

SOP Specific Conformance – MPPS IMPAX EE Mammography supports the next DIMSE Service N-CREATE MPPS Attributes: Table 2.2-16: Supported Attributes for MPPS N-CREATE Attribute Specific Character Set SOP Class UID SOP Instance UID Modality Procedure Code Sequence >Code Value >Coding Scheme Designator >Code Meaning Patient's Name Patient ID Issuer of Patient ID Patient's Birth Date Patient's Sex Patient's Mother's Birth Name Study ID Admission ID Performed Station AE Title Performed Station Name Performed Location Performed Procedure Step Start Date

Tag (0008,0005) (0008,0016) (0008,0018) (0008,0060) (0008,1032) (0008,0100) (0008,0102) (0008,0104) (0010,0010) (0010,0020) (0010,0021) (0010,0030) (0010,0040) (0010,1060) (0020,0010) (0038,0010) (0040,0241) (0040,0242) (0040,0243) (0040,0244)

HE/001359

Page 22 of 33 Document No. 001359 - Revision: 1.1 Livelink NodeID: 32940311

Agfa HealthCare

23 May, 2011

Attribute Performed Procedure Step Start Time Performed Procedure Step End Date Performed Procedure Step End Time Performed Procedure Step Status Performed Procedure Step ID Performed Procedure Step Description Performed Procedure Type Description Performed Protocol Code Sequence >Code Value >Coding Scheme Designator >Code Meaning Scheduled Step Attributes Sequence >Accession Number >Referenced Study Sequence >>Referenced SOP Class UID >>Referenced SOP Instance UID >Study Instance UID >Requested Procedure Description >Requested Procedure Code Sequence >>Code Value >>Coding Scheme Designator >>Code Meaning >Scheduled Procedure Step Description >Scheduled Protocol Code Sequence >>Code Value >>Coding Scheme Designator >>Code Meaning >Scheduled Procedure Step ID >Requested Procedure ID Performed Procedure Step Discontinuation Reason Code Sequence >Code Value >Coding Scheme Designator >Code Meaning Performed Series Sequence >Retrieve AE Title >Series Description >Performing Physician's Name >Operator's Name >Referenced Image Sequence >>Referenced SOP Class UID >>Referenced SOP Instance UID >Protocol Name >Series Instance UID >Referenced Non-Image Composite SOP Instance Sequence >>Referenced SOP Class UID

Tag (0040,0245) (0040,0250) (0040,0251) (0040,0252) (0040,0253) (0040,0254) (0040,0255) (0040,0260) (0008,0100) (0008,0102) (0008,0104) (0040,0270) (0008,0050) (0008,1110) (0008,1150) (0008,1155) (0020,000D) (0032,1060) (0032,1064) (0008,0100) (0008,0102) (0008,0104) (0040,0007) (0040,0008) (0008,0100) (0008,0102) (0008,0104) (0040,0009) (0040,1001) (0040,0281) (0008,0100) (0008,0102) (0008,0104) (0040,0340) (0008,0054) (0008,103E) (0008,1050) (0008,1070) (0008,1140) (0008,1150) (0008,1155) (0018,1030) (0020,000E) (0040,0220) (0008,1150)

HE/001359

Page 23 of 33 Document No. 001359 - Revision: 1.1 Livelink NodeID: 32940311

Agfa HealthCare

23 May, 2011

Attribute >>Referenced SOP Instance UID

Tag (0008,1155)

IMPAX EE Mammography supports the next DIMSE Service N-SET MPPS Attributes: Table 2.2-17: Supported Attributes for MPPS N-SET Attribute Procedure Code Sequence >Code Value >Coding Scheme Designator >Code Meaning Performed Procedure Step Start Date Performed Procedure Step Start Time Performed Procedure Step End Date Performed Procedure Step End Time Performed Procedure Step Status Performed Procedure Step Description Performed Procedure Type Description Performed Procedure Step Discontinuation Reason Code Sequence >Code Value >Coding Scheme Designator >Code Meaning Performed Series Sequence >Retrieve AE Title >Series Description >Performing Physician's Name >Operator's Name >Referenced Image Sequence >>Referenced SOP Class UID >>Referenced SOP Instance UID >Protocol Name >Series Instance UID >Referenced Non-Image Composite SOP Instance Sequence >>Referenced SOP Class UID >>Referenced SOP Instance UID

Tag (0008,1032) (0008,0100) (0008,0102) (0008,0104) (0040,0244) (0040,0245) (0040,0250) (0040,0251) (0040,0252) (0040,0254) (0040,0255) (0040,0281) (0008,0100) (0008,0102) (0008,0104) (0040,0340) (0008,0054) (0008,103E) (0008,1050) (0008,1070) (0008,1140) (0008,1150) (0008,1155) (0018,1030) (0020,000E) (0040,0220) (0008,1150) (0008,1155)

HE/001359

Page 24 of 33 Document No. 001359 - Revision: 1.1 Livelink NodeID: 32940311

Agfa HealthCare

2.2.1.4

23 May, 2011

Association Acceptance Policies The accepted calling and the called application entity titles – as the single TCP port number on which the application is listening for incoming association requests – can be configured in the AE configuration panel of the user interface. It accepts any association for which at least one presentation context is accepted.

2.2.1.4.1

Verify Connection

2.2.1.4.1.1

Description and Sequencing of Activity IMPAX EE Mammography CLIENT responds to a Verification request from a DICOM SCU.

2.2.1.4.1.2

Accepted Presentation Contexts

Table 2.2-18: Acceptable Presentation Context for IMPAX EE Mammography CLIENT Presentation Context Table Abstract Syntax Name Verification SOP Class

2.2.1.4.2

UID 1.2.840.10008.1.1

Transfer Syntax Name List Implicit VR, Little Endian

UID List 1.2.840.10008.1.1

Role SCP

Extended Negotiation None

Receive images, GSPS, KO and SR IMPAX EE Mammography CLIENT accepts an association when it receives an association request from a DICOM Storage SCU.

2.2.1.4.2.1

Description and Sequencing of Activity The receipt of an Image Storage Request is typically caused by a Retrieve Images Request issued by the IMPAX EE Mammography application entity to the remote system. When IMPAX EE Mammography accepts an association, it will receive any image, GSPS, KO or SR transmitted on that association and store the image locally and register it in the local database. The IMPAX EE Mammography application entity informs the user of the receipt of new objects.

2.2.1.4.2.2

Accepted Presentation Contexts IMPAX EE Mammography accepts only presentation contexts for supported SOP classes containing the DICOM default transfer syntax:

Table 2.2-19: Acceptable Presentation Context for IMPAX EE Mammography CLIENT Presentation Context Table Abstract Syntax Name

UID

All supported storage SOP classes listed in table 2.2-1

Transfer Syntax Name List Implicit VR, Little Endian

UID List 1.2.840.10008.1.2

Role SCP

Extended Negotiation no

HE/001359

Page 25 of 33 Document No. 001359 - Revision: 1.1 Livelink NodeID: 32940311

Agfa HealthCare

2.2.1.4.2.3

23 May, 2011

SOP Specific Conformance Statement for Storage SOP classes IMPAX EE Mammography implements Level2 (Full) conformance for the Storage SOP Class. IMPAX EE Mammography will issue a failure status if it is unable to store a DICOM object on disk. If the database contains already an entry referencing an image with the equal SOP Instance UID as the new received image, the image will not be stored. IMPAX EE Mammography will issue also a storage success status in this case. The following error/warning status codes can be sent by IMPAX EE Mammography in the context of a CSTORE- RSP message: Table 2.2-20: Response Status for Storage

Service Status Refused Error Error Success Warning

Further Meaning Out of Resources Data Set does not match SOP Class Cannot understand Data Set does not match SOP Class

Error Code

Reason

A700 A900

Out of Resources Data Set does not match SOP Class

C000 0000 B007

Data Set does not match SOP Class

2.2.1.4.3

Storage commitment (Report)

2.2.1.4.3.1

Description and Sequencing of Activity A Storage Commitment Report is caused by a former Storage Commitment Request issued by the IMPAX EE Mammography application entity to the remote system. At default, IMPAX EE Mammography releases the association immediately after it receives the N-ACTION RSP to the initial Storage Commitment N-ACTION Request. That forces the peer AE, to send the N-EVENT-REPORT request containing the commitment result in a separate association, which has to be initiated – and released – by the peer AE. The default behavior can be changed, that IMPAX EE Mammography keeps the initial association open until the corresponding N-EVENT-REPORT request was received and immediately confirmed by IMPAX EE Mammography – no matter if in the same or in a separate association.

2.2.1.4.3.2

Accepted Presentation Contexts

Table 2.2-21: Acceptable Presentation Contexts for IMPAX EE Mammography CLIENT Presentation Context Table Abstract Syntax Name Verification SOP Class Storage Commitment Push Model SOP Class

UID 1.2.840.10008.1.1 1.2.840.10008.1.20.1

Transfer Syntax Name List Implicit VR, Little Endian Implicit VR, Little Endian

UID List

Role

Extended Negotiation

1.2.840.10008.1.1

SCP

None

1.2.840.10008.1.1

SCU

None

HE/001359

Page 26 of 33 Document No. 001359 - Revision: 1.1 Livelink NodeID: 32940311

Agfa HealthCare

2.2.1.4.3.3

23 May, 2011

SOP Specific Conformance Statement for Storage Commitment (Report) If a complete success of the requested Storage Commitment is notified by the N-EVENTREPORT, the status of the concerned image and presentation state objects changes to “committed”, which enables the user to delete these objects. If existing failures of the requested Storage Commitment is notified by the N-EVENTREPORT, the user will be asked, if the status of image and presentation state objects for which a successful commitment was reported, shall be changed to “committed”.

2.3

Network Interfaces IMPAX EE Mammography provides DICOM V3.0 TCP/IP (Transmission Control Protocol / Internet Protocol) Network Communication Support as defined in PS 3.8 of the DICOM Standard. IMPAX EE Mammography inherits its TCP/IP stack from the computer system upon which it executes.

2.3.1

Physical Medium Support IMPAX EE Mammography is indifferent to the physical medium over which TCP/IP executes; it inherits the medium from the computer system upon which it is being executed.

HE/001359

Page 27 of 33 Document No. 001359 - Revision: 1.1 Livelink NodeID: 32940311

Agfa HealthCare

23 May, 2011

2.4

Configuration

2.4.1

AE Title/ Presentation Mapping The mapping of application entity titles to presentation addresses is configurable from IMPAX EE Mammography’s user interface in the configuration database.

2.4.1.1

Local AE Titles Table 2.4-1: AE Title Configuration Table

Application Entity

Default AE Title

IMPAX EE Mammography Client

Default TCP/IP Port

Hostname of computer

5515

The local AE Titles are configurable through web interface.

2.4.1.2

Remote AE Titles

2.4.2

Configuration Parameters Table 2.4-1: Configuration Parameters Table Parameter

Configurable (Yes/No)

Default Value

General Parameters Calling AE Title Port Read Timeout Default character Encoding Packet Size

Yes Yes Yes No No

Hostname of computer

Latin 1 16384

Modality Performed Procedure Step Specific Parameters Remote host Remote port MPPS called AE Title

Yes Yes Yes

HE/001359

Page 28 of 33 Document No. 001359 - Revision: 1.1 Livelink NodeID: 32940311

Agfa HealthCare

23 May, 2011

3

MEDIA INTERCHANGE

3.1

Implementation Model

3.1.1

Application Data Flow Diagram

Load Objects from CD Import Objects from CD

IMPAX EE Mammography CLIENT AE

Figure 3-1: Application Data Flow Diagram

3.1.2

Functional Definition of IMPAX EE Mammography CLIENT AE The selection of the option “Import CD/DVD” or “Import from CD” gives the possibility to import the selected patients and studies. These selected instances are transmitted to CLIENT AE. The user has the possibility to select any patient DICOM object contained in the CD and visualize it when selecting the CD drive or the path of the DICOMDIR object of the media as search node.

HE/001359

Page 29 of 33 Document No. 001359 - Revision: 1.1 Livelink NodeID: 32940311

Agfa HealthCare

3.1.3

23 May, 2011

Sequencing of Real World Activities When importing from CD, IMPAX EE Mammography CLIENT searches the desired patient and study in the DICOM DIR. Then, an import is executed via C-Store.

3.1.4

File Meta Information for Implementation Class and Version Table 3.1-1: File Meta Information for Implementation Class and Version File Meta Information Version Implementation Class UID Implementation Version Name

00x01 1.2.40.0.13.1.1 dcm4che-2.0

3.2

AE Specifications

3.2.1

IMPAX EE Mammography CLIENT AE Specification Table 3.2-1: AE Related Application Profiles, Real World Activities and Roles

Supported Application Profile STD-GEN-CD

3.2.1.1

Real-World Activity Import DICOM Objects to CD Load DICOM Objects to CD

Role FSR,FSU FSR

SC Option Interchange Interchange

File Meta Information for Implementation IMPAX EE Mammography CLIENT There is no File Meta Information that pertains to Application Entity IMPAX EE Mammography CLIENT.

3.2.1.2

Real World Activities

3.2.1.2.1

Real World Activity – Import DICOM objects from CD IMPAX EE Mammography CLIENT AE works as FSR during the import from CD. It obtains the CD information data from the DICOMDIR structure and once selected the desired patient and study(s) it imports via DICOM C-Store. Optionally, it can act as FSU when user chooses the option to change some DICOM attributes of the imported data.

3.2.1.2.1.1

Media Storage Application Profile IMPAX EE Mammography CLIENT AE supports STD-GEN-CD Application Profile. The SOP classes supported are listed in Table 1.1-1: Network Services Supported. The used Transfer Syntax is Explicit VR Little Endian with Transfer Syntax UID 1.2.840.10008.1.2.1.

HE/001359

Page 30 of 33 Document No. 001359 - Revision: 1.1 Livelink NodeID: 32940311

Agfa HealthCare

23 May, 2011

3.2.1.2.1.1.1

Options User of IMPAX EE Mammography has three options to import: Direct Import, Ad Hoc and Import via Modality Worklist. When user chooses Ad Hoc or Via Modality Worklist some DICOM attributes from the imported image are changed, then, IMPAX EE Mammography is working as FSU.

3.2.1.2.1.1.1.1

Direct Import: All DICOM Attributes will remain as original data from CD.

3.2.1.2.1.1.1.2

Ad Hoc: Some DICOM Attributes will be substituted by the User of IMPAX Mammography. These attributes are: Table 3.2-2: Attributes modified by the user Attribute Name Patient Name Patient ID Patient’s Birth Date Patient’s Sex Accession Number

3.2.1.2.1.1.1.3

Tag (0010,0010) (0010,0020) (0010,0030) (0010,0040) (0008,0050)

Import via Modality Worklist: DICOM Attributes will be substituted by the obtained attributes from a modality worklist query. These attributes are listed in Table 2.2-14: Supported Attributes from Modality Worklist Information Model.

3.2.1.2.2

Real World Activity – Load DICOM objects from CD IMPAX EE Mammography CLIENT AE works as FSR when loading data from a CD. It obtains the CD information data from the DICOMDIR structure and the user can choose which information is loaded.

3.2.1.2.2.1

Media Storage Application Profile IMPAX EE Mammography CLIENT AE supports STD-GEN-CD Application Profile. The SOP classes supported are listed in Table 1.1-1: Network Services Supported. The used Transfer Syntax is Explicit VR Little Endian with Transfer Syntax UID 1.2.840.10008.1.2.1.

HE/001359

Page 31 of 33 Document No. 001359 - Revision: 1.1 Livelink NodeID: 32940311

Agfa HealthCare

4

23 May, 2011

SUPPORT FOR EXTENDED CHARACTER SETS IMPAX EE Mammography supports the following character sets: Character Set Description

Defined Term

Latin alphabet No. 1 Thai Cyrillic Arabic Korean Greek Hebrew

ISO-IR 100 ISO-IR 166 ISO-IR 144 ISO-IR 127 ISO-IR 149 ISO-IR 126 ISO-IR 138

UTF-8 JIS X 0208 Japanese Kanji JIS X 0201 Japanese Katakana Chinese

ISO-IR 192 ISO-IR 87 ISO-IR 13 GB18030

Character Set Description Defined Term Basic G0 Set ISO-IR 6 ISO 8859-1 Latin Alphabet No. 1 ISO-IR 100 (default) Latin Alphabet No. 2 ISO-IR 101

HE/001359

Page 32 of 33 Document No. 001359 - Revision: 1.1 Livelink NodeID: 32940311

Agfa HealthCare

23 May, 2011

5

SECURITY

5.1

Security Profiles IMPAX EE Mammography supports secure DICOM communication in conformance with the Secure Profiles: 

Basic TLS Secure Transport Connection Profile User of IMPAX EE Mammography can activate the use of TLS security.



AES TLS Secure Transport Connection Profile User of IMPAX EE Mammography can activate the use of AET TLS security complying with TLS_RSA_WITH_AES_128_CBC_SHA profile.



Basic User Identity Association Profile IMPAX EE Mammography can be configured to use Basic User Identity Association Profile.



User Identity plus Pass code Association Profile IMPAX EE Mammography can be configured to use User Identity plus Pass code Association Profile.



Basic Network Address Management Profile IMPAX EE Mammography supports Basic Network Address Management Profile acting as DNS Client and DHCP Client



Basic Time Synchronization Profile IMPAX EE Mammography supports Basic Time Synchronization Profile acting as NTP client.

5.2

Association Level Security IMPAX EE Mammography does not support any Association Level Security.

5.3

Application Level Security IMPAX EE Mammography application access requires a valid user and password in order to login.

HE/001359

Page 33 of 33 Document No. 001359 - Revision: 1.1 Livelink NodeID: 32940311

Agfa HealthCare

23 May, 2011

6

ANNEXES

6.1

IOD Contents

6.1.1

Coerced/ Modified Fields As specified in Chapter 3.2 IMPAX EE Mammography acting as SCU can modify attributes when importing. These attributes are listed in Table 3.2-2: Attributes modified by the user.

6.2

Grayscale Image Consistency The high resolution display monitor attached to the product can be calibrated according to the Grayscale Standard Display Function (GSDF).

Details as of PDF Creation Date Document Metadata Title:

001359_IMPAX EE Mammography DICOM Conformance Statement

Livelink ID:

32940311

Version#:

3

Version Date:

2011/05/16 10:43 AM CET

Status:

Approved on 2011/05/23 10:01 AM CET

Owner:

Peter Buytaert (awabr)

Created By:

Peter Buytaert (awabr)

Created Date:

2011/04/06 10:11 AM CET

PDF Creation Date:

2011/05/23 10:03 AM CET

This document was approved by: Signatures: 1. 2. 3. 4.

Bruno Laffin (nawyv) on 2011/05/16 04:42 PM CET Peter Steiger (awpqz) on 2011/05/17 04:12 PM CET Sabine Kolics (awppx) on 2011/05/17 07:35 PM CET Sebastian Zambal (axdhg) on 2011/05/23 09:34 AM CET

Detailed Approver History: 

Approval Workflow started on 2011/05/16 10:47 AM CET  Approval task originally assigned to and completed by Peter Steiger (awpqz) on 2011/05/17 04:12 PM CET  Approval task originally assigned to and completed by Bruno Laffin (nawyv) on 2011/05/16 04:42 PM CET  Approval task originally assigned to and completed by Sebastian Zambal (axdhg) on 2011/05/23 09:34 AM CET  Approval task originally assigned to and completed by Sabine Kolics (awppx) on 2011/05/17 07:35 PM CET

Version & Status History Version# Date Created 3

2011/05/16 10:43 AM CET

Status Approved - 2011/05/23

2

2011/04/11 03:05 PM CET

1

2011/04/06 10:11 AM CET

Rejected - 2011/04/26

Suggest Documents