A Comparison of Laser Delivery Mode & Wavelength on Varicose Veins Monika G. Kiripolsky, MD; Susan P. Detwiler, MD and Mitchel P. Goldman, MD
Purpose: To determine the effects of endovenous laser wavelengths and delivery modes (continuous diode versus pulsed Nd:YAG) on isolated veins ex-vivo both macroscopically as well as microscopically. In addition, to determine whether protected-tip fibers are effective in reducing vein wall perforations.
Methods: An ex-vivo endovenous laser ablation model was used. Human veins harvested during ambulatory phlebectomy procedures were sutured closed at one end. Veins were then suspended by adding a weighting device, within a graduated column of warm saline at physiologic concentration. (See Figure 1).The lumen of each vein was filled with porcine blood, a fiber inserted into the lumen and the fiber attached to a mechanical pullback device (See Figure 2):
Human vein suspended within cylinder of physiologic saline (prior to insertion of laser fiber and subsequent laser treatment)
Vein has been filled with porcine blood & laser fiber has been inserted. Fiber is connected to automatic pullback device. The light shows where the laser tip is within the vein.
Figure 1
Figure 2
Laser and Fiber Matrix Laser:
1320nm Pulsed 1310nm CW 810nm CW 810nm CW 1470nm CW 2100nm Pulsed
No.
System Type
Power*
Pullback Rate
1 2 3 4 5 6
Nd:YAG Diode Diode Diode Diode Ho:YAG
7 watts 7 watts 7 watts 7 watts 7 watts 7 watts
1mm/sec 1mm/sec 1mm/sec 1mm/sec 1mm/sec 1mm/sec
Fiber Configuration
600um Standard Tip 600um Standard Tip 600um Protected Tip 365um Standard Tip 600um Standard Tip 600um Standard Tip
* Adjusted to 7 watts with a Molectron power meter CoolTouch Inc.
11_2011
7075-0132 Rev A
Macroscopic Results: •
1320 nm Nd:YAG pulsed (fiber #1) with two veins treated, yielded shrinkage of the vessel with no visible perforations, no coagulum on fiber tip.
Fiber Tip
Pre - Laser
•
Post - Laser
1310 nm diode CW (fiber #2) caused vein shrinkage, but, also caused dramatic perforations.
Fiber Tip
Pre-Laser
CoolTouch Inc.
Post-Laser
11_2011
7075-0132 Rev A
•
810 nm diode CW (fiber #3) protected tip fiber yielded vein shrinkage without perforations or coagulum formation on the tip
Fiber Tip
Pre - Laser
•
Post - Laser
810 nm diode CW (fiber#4) non-protected tip fiber yielded grossly discernable perforations and coagulum on the tip of the fiber.
Fiber Tip
Pre - Laser
CoolTouch Inc.
Post - Laser
11_2011
7075-0132 Rev A
•
1470 nm diode CW (fiber #5) loud popping sounds; fiber appeared to stick to vein wall, dragging it along by its site of initial contact via the rate set by the pull-back device. Also, coagulum formed on the fiber tip.
Pre - Laser •
Post - Laser
2100 nm diode CW (fiber #6) loud popping sounds; fiber appeared to stick to vein wall, dragging it along by its site of initial contact via the rate set by the pull-back device. Also, coagulum formed on the fiber tip
Pre - Laser CoolTouch Inc.
Post - Laser 11_2011
7075-0132 Rev A
Microscopic Results: A control (untreated vein) counterpart from the same patient was analyzed histologically. Cross-sections of vessels treated with each laser fiber combination (as well as corresponding non-treated controls) were viewed and interpreted by a histopathologist who remained blind as to the laser used. Laser
No.
Total # crosssections evaluated
Thermal damage*
Perforation or full thickness necrosis of a portion of the vein wall** (# sections where finding is observed/ total # crosssections)
1320nm pulsed
1310nm diode 810nm diode protected tip 810nm diode nonprotected tip
1
13 13
PT, FT, AD 0
2/13 (15%) 0 (0%)
2
11
PT, FT, AD
7/11 (64%)
11
0
0 (0%)
11
PT
0 (0%)
7
0
0 (0%)
3
PT, FT, AD 4
11
(AD focally near
5/11 (45%)
perforations)
1470nm diode
2100nm holmium
1320nm pulsed (redo)
5
6
1
8
0
0 (0%)
6
PT, FT, AD
4/6 (67%)
14
0
0 (0%)
10
PT, FT, AD
5/10 (50%)
10
0
0 (0%)
9
PT, FT, AD
1/9 (11%)
11
0
0 (0%)
*Thermal damage (amorphous amphophilic material; coagulation necrosis) - affecting the partial thickness of media with intervening muscle cell nuclei still visible ("PT"), affecting the full thickness of media with intervening muscle cell nuclei still visible ("FT"), affecting the advential dermis ("AD") - (# sections where finding is observed/ total # cross-sections evaluated) **Perforation (channel or absence of tissue in a portion of the vein wall) or full thickness necrosis of a portion of the vein wall (muscle cell nuclei are not visible and are replaced by coagulation necrosis and amorphous basophilic material) (# sections where finding is observed/ total # cross-sections evaluated)
CoolTouch Inc.
11_2011
7075-0132 Rev A
•
1320 nm Nd:YAG pulsed (fiber #1)
•
1310 nm Diode CW (fiber #2)
CoolTouch Inc.
11_2011
7075-0132 Rev A
•
810nm Diode CW Protected Tip (fiber #3)
CoolTouch Inc.
11_2011
7075-0132 Rev A
•
810nm Diode CW Non-Protected Tip (fiber #4)
•
1470nm Diode CW (fiber #5)
CoolTouch Inc.
11_2011
7075-0132 Rev A
•
2100nm Diode CW (fiber #6)
Conclusions: •
• •
The delivery mode, pulsed Nd:YAG vs. continuous wave diode may be just as important as the wavelength when lasers are used in an endovenous ablation procedure. Thus, the 1310 nm CW laser may NOT be equivalent to the 1320 nm pulsed laser Pulsing the laser output may dislodge any coagulum from forming on the fiber tip. This could prevent high tip temperatures, which are more likely to cause wall perforations Protected tip, 810 nm fibers may be less likely to yield wall perforations than their non-protected counterparts