Zenotec Scan S50 – Win 7 / 64 bit upgrade kit Installation guide

Contents 2

Disclaimer

3

Important to check before proceeding

3

Contents of the upgrade kit

3

Five steps of the installation procedure

11

Technical support

11

Manufacturer

3Series scanner upgrade kit installation guide All rights reserved. No part of this manual may be reproduced or transmitted, in any form or by any means, whether electronic, chemical, manual or otherwise, including but not limited to photocopying, recording, or information storage and retrieval systems, for any purpose without prior written permission from Dental Wings.

Disclaimer Dental Wings specifically disclaims all warranties, either express or implied, including, but not limited to, implied warranties of merchantability and fitness for a particular purpose with respect to this product and documentation. Updates to hardware and software components are made regularly. Therefore, some of the instructions, specifications and pictures in this documentation may differ slightly from your particular installation. All items described in this guide are for illustration purposes only and may not apply to your specific installation. No legal rights or entitlements may be obtained from descriptions made in this manual. Dental Wings reserves the right to revise, make changes or improvements to this product or documentation at any time without notice. Dental Wings shall not be liable for any consequential or incidental damages, including but not limited to the loss of business profits or any other commercial damages resulting from the use of this product.

2

Dental Wings Inc.

Important to check before proceeding Prior to performing the upgrade, verify you have the upgrade kit that corresponds to the serial number of your scanner. This kit and installation guide is for the 3Series with a serial number between 3S00233 and 3S01016 (model 1100-2 DDR3, MSI motherboard). If you own a 3Series with a serial number between 3S00001 and 3S00232 (model 1100-1 DDR-2, Gigabyte motherboard), another upgrade kit is applicable, please contact your distributor to obtain the correct kit.

You will find the serial number on the silver grey label found on the lower left side of the back panel of your scanner.

Contents of the upgrade kit The 3Series 64-bit upgrade kit contains the following items:

• 2 x 4 GB memory modules • USB flash drive that will enable your system to upgrade to Windows 7, 64 bits* • 12-unit multi-die plate • “Windows 7 embedded “ label * Windows 7, 64 bits is compatible with DWOS starting from release 3.1.0.

Five steps of the installation procedure A. Database backup*

B. Installation of the memory modules C. Retrieving the scanner’s information D. Upgrading to 64 bits

* Database backup is optional. The upgrade won’t automatically backup the existing DWOS data. All scanned cases, for example, won’t be saved unless the database is backed up.

E. Restoring the scanner’s information STI75-0106

3Series upgrade procedure

3

A. Database backup (optional) To backup your database, you must save two specific folders to an external hard drive. By default, the database is installed in C:\Program Files\Dental-Wings(3.0). Select both folders DentalWingsData and DW_MYSQL_DB and copy them to your external hard drive.

B. Installation of the memory modules 8 1

7

6

2

3

5 4

1. Turn off the scanner and unplug all

2. Unscrew beginning with screw #1,

devices from it (screen, keyboard, etc).

ending with screw #8, then carefully hold and remove the panel.

4

Dental Wings Inc.

memory modules

3. Insert the two new memory modules in the lower slots, according to the below instructions.

Insertion of memory modules The memory module has only one notch on the center and will only fit in the right orientation. To insert a memory module, push it in until the golden finger on the memory module is deeply inserted in the slot. The plastic clip at each side of the slot will automatically close when the memory module is properly seated.

STI75-0106

3Series upgrade procedure

5

Fan cable

4. Affix the “Windows 7 embedded” label over the existing Windows label.

5. Carefully replace the fan cable inside the scanner, ensuring it won’t be pinched by the back panel.

1

1

8

2

3

7

6

5

7. Fasten all screws lightly to seat the

6. Slide up the back panel, then screw

back panel, then tighten all screws to secure.

upper middle screw lightly. 6

4

Dental Wings Inc.

C. Retrieving the scanner’s information 1. Start the scanner. 2. Go to Control Panel > System: the installed memory (RAM) should now be 12.00 GB. 3. Insert the USB flash drive provided within your upgrade kit into any available USB port on the scanner. 4. Open the USB flash drive, then the “Scripts” folder.

5. Run the “SavePCInfo.bat” script by double clicking on that file. A Windows command line window will then open.

6. To continue the script execution, press any key on the keyboard. When prompted, enter the serial number of your scanner and press “Enter”.

STI75-0106

3Series upgrade procedure

7

7. You will then need to confirm that the serial number that you have entered is correct. Press “Y” if the right serial number is displayed. Press any other key to correct and re-type the serial number.

8. The script should finish its execution with a report. If everything was done properly, the script should indicate “Success” after every operation as shown.

9. If an operation has failed, please contact your distributor.

8

Dental Wings Inc.

D. Upgrading to 64 bits Attention: This operation will erase your database if it wasn’t backed-up within step A. 1. Note the serial number of your scanner (on lower left of back panel), as you will need it further within the procedure. 2. Restart the computer. 3. At the first screen, press “F11” key on the keyboard to enter boot menu. 4. Select the USB : Kingston Data Traveler as shown and press “Enter”.

5. Press “Enter” as soon as you see this screen, otherwise it will continue as the “Automatic boot” countdown reaches 0 seconds. It will then load for a few seconds before the next screen.

6. You will then be prompted twice to confirm that you really want to proceed with the restoration. Both times, confirm by pressing “y”, then “Enter”. It will then start restoring for about an hour. 7. Once the restoration is done, press “Enter” to continue. 8. Press “1” to restart the computer, then remove the USB flash drive as the computer is booting.

Important: The scanner will only restart properly if the USB flash drive has been removed.

STI75-0106

3Series upgrade procedure

9

E. Restoring the scanner’s information 1. Reinsert the USB flash drive into the scanner. 2. Go back to the “Scripts” folder on the flash drive.

3. Run the “RestorePCInfo.bat” script by double clicking on that file. A Windows command line window will open.

4. To continue the script execution, press any key on the keyboard. The script should finish its execution by a report. If everything was done properly, the script should indicate “Success” after every operation. The upgrade process is now finished.

5. If an operation has failed, please contact your distributor.

Database backup: If you have done a backup of your database (step A) it is now time to copy it on your system. To do, copy the same two folders from your external hard drive into the folder D:\Dental-Wings(3.0)\DWData.

10

Dental Wings Inc.

Technical support For complete software instructions, refer to DWOS® User Guide found on D:\Dental-Wings\ DWDocumentation where D:\Dental-Wings is the default installation folder. If needed, please contact your distributor for technical support.

Manufacturer Dental Wings Inc. 2030 Pie-IX, Suite 219 Montreal, QC CANADA H1V 2C8 Telephone +1 514 807 8485 Fax +1 514 223 3981 Email [email protected] Website dentalwings.com

STI75-0106

3Series upgrade procedure

11

D

E

N

T A

L

EXPECT THE DIFFERENCE! BY WIELAND. As a major supplier of dental system solutions, Wieland embodies both tradition and progress in matters of dental products and technology. Since our company was founded in 1871, we have stayed true to our corporate philosophy of combining tradition, innovation and quality with the best in customer care. Today, our core competencies and key strengths lie in the forward-looking integration of technologies and materials for dental prosthetics. This ensures that patients can trust in the quality of their restorations, and our partners in dental practices and laboratories can continue with confidence on the path to digitalisation and greater competitiveness.

www.wieland-dental.de Wieland Dental + Technik GmbH & Co. KG Schwenninger Straße 13, 75179 Pforzheim, Germany Fon +49 72 31 / 37 05 - 0, Fax +49 72 31 / 35 79 59

547004e.00.06/12

Wieland offers a wide range of products and services from CAD/CAM technologies and dental alloys to veneering ceramics and electroforming. Thanks to our worldwide presence and international network, Wieland is never far away, and your contact person can always be located via the Internet.