1. TR Processing Synthes Reusable Medical Devices Instruments, Instrument Trays and Cases

Page |1 TECH READY document Product DEPUY SYNTHES Reusable Medical Devices, Instruments, Instrument Trays and Cases and Non-Sterile Implants Combi...
Author: Tobias Lambert
2 downloads 0 Views 6MB Size
Page |1

TECH READY document Product

DEPUY SYNTHES

Reusable Medical Devices, Instruments, Instrument Trays and Cases and Non-Sterile Implants

Combined Tech Ready There are two distinct documents included in this Tech Ready document; both contain important information regarding the reprocessing of Depuy Synthes Reusable Medical Devices, Instruments, Instrument Trays and Cases and Non-Sterile Implants. Here are the documents: 1. TR Processing Synthes Reusable Medical Devices – Instruments, Instrument Trays and Cases Processing Synthes Reusable Medical Devices – Instruments, Instrument Trays and Cases

2. TR Processing Non-Sterile Implants Processing Non-Sterile Implants

Page |2

TECH READY document

Key for Colored Boxes – Sterilization Methods Steam ETO High-Level Disinfection STERRAD STERIS V-PRO STERIS SYSTEM-1E

Product

DEPUY SYNTHES

Reusable Medical Devices – Instruments, Instrument Trays and Cases

Point of Use

Source Document: Depuy_Synthes_Reusable_Medical_Devices_Instruments_Trays_Cases_DJ1305_rI_n_NonSterile_Implants_DJ1304_rJ_Rcv201407.pdf RH201412 DISCLAIMER: The information in this document is for reference only and includes only a portion of the manufacturer’s labeling/instructions. With the exception of the titles and header information, the content corresponds word for word with the Manufacturers document. It does not supersede or alter the manufacturer’s labeling/instructions. The manufacturer’s labeling/instructions contains other important information that must be considered in order to properly reprocess these items. Use this guide only after thoroughly reviewing, understanding, and complying with all of the manufacturer’s labeling/instructions requirements. If there is a conflict, always follow the manufacturer’s labeling/instructions. The user holds BPP blameless of any and all liability arising from the use of this document.

Page |3

TECH READY document

Key for Colored Boxes – Sterilization Methods Steam ETO High-Level Disinfection STERRAD STERIS V-PRO STERIS SYSTEM-1E

Disassembly

Cleaning / Decontamination

Source Document: Depuy_Synthes_Reusable_Medical_Devices_Instruments_Trays_Cases_DJ1305_rI_n_NonSterile_Implants_DJ1304_rJ_Rcv201407.pdf RH201412 DISCLAIMER: The information in this document is for reference only and includes only a portion of the manufacturer’s labeling/instructions. With the exception of the titles and header information, the content corresponds word for word with the Manufacturers document. It does not supersede or alter the manufacturer’s labeling/instructions. The manufacturer’s labeling/instructions contains other important information that must be considered in order to properly reprocess these items. Use this guide only after thoroughly reviewing, understanding, and complying with all of the manufacturer’s labeling/instructions requirements. If there is a conflict, always follow the manufacturer’s labeling/instructions. The user holds BPP blameless of any and all liability arising from the use of this document.

Page |4

TECH READY document

Key for Colored Boxes – Sterilization Methods Steam ETO High-Level Disinfection STERRAD STERIS V-PRO STERIS SYSTEM-1E

Cleaning / Decontamination (continued)

Source Document: Depuy_Synthes_Reusable_Medical_Devices_Instruments_Trays_Cases_DJ1305_rI_n_NonSterile_Implants_DJ1304_rJ_Rcv201407.pdf RH201412 DISCLAIMER: The information in this document is for reference only and includes only a portion of the manufacturer’s labeling/instructions. With the exception of the titles and header information, the content corresponds word for word with the Manufacturers document. It does not supersede or alter the manufacturer’s labeling/instructions. The manufacturer’s labeling/instructions contains other important information that must be considered in order to properly reprocess these items. Use this guide only after thoroughly reviewing, understanding, and complying with all of the manufacturer’s labeling/instructions requirements. If there is a conflict, always follow the manufacturer’s labeling/instructions. The user holds BPP blameless of any and all liability arising from the use of this document.

Page |5

TECH READY document

Key for Colored Boxes – Sterilization Methods Steam ETO High-Level Disinfection STERRAD STERIS V-PRO STERIS SYSTEM-1E

Cleaning / Decontamination (continued)

Cleaning / Decontamination continues on next page Source Document: Depuy_Synthes_Reusable_Medical_Devices_Instruments_Trays_Cases_DJ1305_rI_n_NonSterile_Implants_DJ1304_rJ_Rcv201407.pdf RH201412 DISCLAIMER: The information in this document is for reference only and includes only a portion of the manufacturer’s labeling/instructions. With the exception of the titles and header information, the content corresponds word for word with the Manufacturers document. It does not supersede or alter the manufacturer’s labeling/instructions. The manufacturer’s labeling/instructions contains other important information that must be considered in order to properly reprocess these items. Use this guide only after thoroughly reviewing, understanding, and complying with all of the manufacturer’s labeling/instructions requirements. If there is a conflict, always follow the manufacturer’s labeling/instructions. The user holds BPP blameless of any and all liability arising from the use of this document.

Page |6

TECH READY document

Key for Colored Boxes – Sterilization Methods Steam ETO High-Level Disinfection STERRAD STERIS V-PRO STERIS SYSTEM-1E

Cleaning / Decontamination (continued)

Cleaning / Decontamination continues on next page Source Document: Depuy_Synthes_Reusable_Medical_Devices_Instruments_Trays_Cases_DJ1305_rI_n_NonSterile_Implants_DJ1304_rJ_Rcv201407.pdf RH201412 DISCLAIMER: The information in this document is for reference only and includes only a portion of the manufacturer’s labeling/instructions. With the exception of the titles and header information, the content corresponds word for word with the Manufacturers document. It does not supersede or alter the manufacturer’s labeling/instructions. The manufacturer’s labeling/instructions contains other important information that must be considered in order to properly reprocess these items. Use this guide only after thoroughly reviewing, understanding, and complying with all of the manufacturer’s labeling/instructions requirements. If there is a conflict, always follow the manufacturer’s labeling/instructions. The user holds BPP blameless of any and all liability arising from the use of this document.

Page |7

TECH READY document

Key for Colored Boxes – Sterilization Methods Steam ETO High-Level Disinfection STERRAD STERIS V-PRO STERIS SYSTEM-1E

Cleaning / Decontamination (continued)

Source Document: Depuy_Synthes_Reusable_Medical_Devices_Instruments_Trays_Cases_DJ1305_rI_n_NonSterile_Implants_DJ1304_rJ_Rcv201407.pdf RH201412 DISCLAIMER: The information in this document is for reference only and includes only a portion of the manufacturer’s labeling/instructions. With the exception of the titles and header information, the content corresponds word for word with the Manufacturers document. It does not supersede or alter the manufacturer’s labeling/instructions. The manufacturer’s labeling/instructions contains other important information that must be considered in order to properly reprocess these items. Use this guide only after thoroughly reviewing, understanding, and complying with all of the manufacturer’s labeling/instructions requirements. If there is a conflict, always follow the manufacturer’s labeling/instructions. The user holds BPP blameless of any and all liability arising from the use of this document.

Page |8

TECH READY document

Key for Colored Boxes – Sterilization Methods Steam ETO High-Level Disinfection STERRAD STERIS V-PRO STERIS SYSTEM-1E

Cleaning / Decontamination (continued)

Cleaning / Decontamination continues on next page Source Document: Depuy_Synthes_Reusable_Medical_Devices_Instruments_Trays_Cases_DJ1305_rI_n_NonSterile_Implants_DJ1304_rJ_Rcv201407.pdf RH201412 DISCLAIMER: The information in this document is for reference only and includes only a portion of the manufacturer’s labeling/instructions. With the exception of the titles and header information, the content corresponds word for word with the Manufacturers document. It does not supersede or alter the manufacturer’s labeling/instructions. The manufacturer’s labeling/instructions contains other important information that must be considered in order to properly reprocess these items. Use this guide only after thoroughly reviewing, understanding, and complying with all of the manufacturer’s labeling/instructions requirements. If there is a conflict, always follow the manufacturer’s labeling/instructions. The user holds BPP blameless of any and all liability arising from the use of this document.

Page |9

TECH READY document

Key for Colored Boxes – Sterilization Methods Steam ETO High-Level Disinfection STERRAD STERIS V-PRO STERIS SYSTEM-1E

Cleaning / Decontamination (continued)

Maintenance

Source Document: Depuy_Synthes_Reusable_Medical_Devices_Instruments_Trays_Cases_DJ1305_rI_n_NonSterile_Implants_DJ1304_rJ_Rcv201407.pdf RH201412 DISCLAIMER: The information in this document is for reference only and includes only a portion of the manufacturer’s labeling/instructions. With the exception of the titles and header information, the content corresponds word for word with the Manufacturers document. It does not supersede or alter the manufacturer’s labeling/instructions. The manufacturer’s labeling/instructions contains other important information that must be considered in order to properly reprocess these items. Use this guide only after thoroughly reviewing, understanding, and complying with all of the manufacturer’s labeling/instructions requirements. If there is a conflict, always follow the manufacturer’s labeling/instructions. The user holds BPP blameless of any and all liability arising from the use of this document.

P a g e | 10

TECH READY document

Key for Colored Boxes – Sterilization Methods Steam ETO High-Level Disinfection STERRAD STERIS V-PRO STERIS SYSTEM-1E

Maintenance (continued)

Assembly

Assembly continues on next page Source Document: Depuy_Synthes_Reusable_Medical_Devices_Instruments_Trays_Cases_DJ1305_rI_n_NonSterile_Implants_DJ1304_rJ_Rcv201407.pdf RH201412 DISCLAIMER: The information in this document is for reference only and includes only a portion of the manufacturer’s labeling/instructions. With the exception of the titles and header information, the content corresponds word for word with the Manufacturers document. It does not supersede or alter the manufacturer’s labeling/instructions. The manufacturer’s labeling/instructions contains other important information that must be considered in order to properly reprocess these items. Use this guide only after thoroughly reviewing, understanding, and complying with all of the manufacturer’s labeling/instructions requirements. If there is a conflict, always follow the manufacturer’s labeling/instructions. The user holds BPP blameless of any and all liability arising from the use of this document.

P a g e | 11

TECH READY document

Key for Colored Boxes – Sterilization Methods Steam ETO High-Level Disinfection STERRAD STERIS V-PRO STERIS SYSTEM-1E

Assembly (continued)

Source Document: Depuy_Synthes_Reusable_Medical_Devices_Instruments_Trays_Cases_DJ1305_rI_n_NonSterile_Implants_DJ1304_rJ_Rcv201407.pdf RH201412 DISCLAIMER: The information in this document is for reference only and includes only a portion of the manufacturer’s labeling/instructions. With the exception of the titles and header information, the content corresponds word for word with the Manufacturers document. It does not supersede or alter the manufacturer’s labeling/instructions. The manufacturer’s labeling/instructions contains other important information that must be considered in order to properly reprocess these items. Use this guide only after thoroughly reviewing, understanding, and complying with all of the manufacturer’s labeling/instructions requirements. If there is a conflict, always follow the manufacturer’s labeling/instructions. The user holds BPP blameless of any and all liability arising from the use of this document.

P a g e | 12

TECH READY document

Key for Colored Boxes – Sterilization Methods Steam ETO High-Level Disinfection STERRAD STERIS V-PRO STERIS SYSTEM-1E

Sterilization

Sterilization continues on next page Source Document: Depuy_Synthes_Reusable_Medical_Devices_Instruments_Trays_Cases_DJ1305_rI_n_NonSterile_Implants_DJ1304_rJ_Rcv201407.pdf RH201412 DISCLAIMER: The information in this document is for reference only and includes only a portion of the manufacturer’s labeling/instructions. With the exception of the titles and header information, the content corresponds word for word with the Manufacturers document. It does not supersede or alter the manufacturer’s labeling/instructions. The manufacturer’s labeling/instructions contains other important information that must be considered in order to properly reprocess these items. Use this guide only after thoroughly reviewing, understanding, and complying with all of the manufacturer’s labeling/instructions requirements. If there is a conflict, always follow the manufacturer’s labeling/instructions. The user holds BPP blameless of any and all liability arising from the use of this document.

P a g e | 13

TECH READY document

Key for Colored Boxes – Sterilization Methods Steam ETO High-Level Disinfection STERRAD STERIS V-PRO STERIS SYSTEM-1E

Sterilization (continued)

Source Document: Depuy_Synthes_Reusable_Medical_Devices_Instruments_Trays_Cases_DJ1305_rI_n_NonSterile_Implants_DJ1304_rJ_Rcv201407.pdf RH201412 DISCLAIMER: The information in this document is for reference only and includes only a portion of the manufacturer’s labeling/instructions. With the exception of the titles and header information, the content corresponds word for word with the Manufacturers document. It does not supersede or alter the manufacturer’s labeling/instructions. The manufacturer’s labeling/instructions contains other important information that must be considered in order to properly reprocess these items. Use this guide only after thoroughly reviewing, understanding, and complying with all of the manufacturer’s labeling/instructions requirements. If there is a conflict, always follow the manufacturer’s labeling/instructions. The user holds BPP blameless of any and all liability arising from the use of this document.

P a g e | 14

TECH READY document

Key for Colored Boxes – Sterilization Methods Steam ETO High-Level Disinfection STERRAD STERIS V-PRO STERIS SYSTEM-1E

Product

DEPUY SYNTHES

Processing Non-Sterile Synthes Implants

Point of Use

Disassembly

No specific instructions given

Cleaning / Decontamination

Source Document: DePuy_Synthes_Reusable_Medical_Devices_Instruments_Trays_Cases_DJ1305_rI_n_Non_Sterile_Impants_DSUS_SPN_0115_0706_201501.pdf RH201504 DISCLAIMER: The information in this document is for reference only and includes only a portion of the manufacturer’s labeling/instructions. With the exception of the titles and header information, the content corresponds word for word with the Manufacturers document. It does not supersede or alter the manufacturer’s labeling/instructions. The manufacturer’s labeling/instructions contains other important information that must be considered in order to properly reprocess these items. Use this guide only after thoroughly reviewing, understanding, and complying with all of the manufacturer’s labeling/instructions requirements. If there is a conflict, always follow the manufacturer’s labeling/instructions. The user holds BPP blameless of any and all liability arising from the use of this document.

P a g e | 15

TECH READY document

Key for Colored Boxes – Sterilization Methods Steam ETO High-Level Disinfection STERRAD STERIS V-PRO STERIS SYSTEM-1E

Cleaning / Decontamination (continued)

Cleaning / Decontamination continues on next page Source Document: DePuy_Synthes_Reusable_Medical_Devices_Instruments_Trays_Cases_DJ1305_rI_n_Non_Sterile_Impants_DSUS_SPN_0115_0706_201501.pdf RH201504 DISCLAIMER: The information in this document is for reference only and includes only a portion of the manufacturer’s labeling/instructions. With the exception of the titles and header information, the content corresponds word for word with the Manufacturers document. It does not supersede or alter the manufacturer’s labeling/instructions. The manufacturer’s labeling/instructions contains other important information that must be considered in order to properly reprocess these items. Use this guide only after thoroughly reviewing, understanding, and complying with all of the manufacturer’s labeling/instructions requirements. If there is a conflict, always follow the manufacturer’s labeling/instructions. The user holds BPP blameless of any and all liability arising from the use of this document.

P a g e | 16

TECH READY document

Key for Colored Boxes – Sterilization Methods Steam ETO High-Level Disinfection STERRAD STERIS V-PRO STERIS SYSTEM-1E

Cleaning / Decontamination (continued)

Source Document: DePuy_Synthes_Reusable_Medical_Devices_Instruments_Trays_Cases_DJ1305_rI_n_Non_Sterile_Impants_DSUS_SPN_0115_0706_201501.pdf RH201504 DISCLAIMER: The information in this document is for reference only and includes only a portion of the manufacturer’s labeling/instructions. With the exception of the titles and header information, the content corresponds word for word with the Manufacturers document. It does not supersede or alter the manufacturer’s labeling/instructions. The manufacturer’s labeling/instructions contains other important information that must be considered in order to properly reprocess these items. Use this guide only after thoroughly reviewing, understanding, and complying with all of the manufacturer’s labeling/instructions requirements. If there is a conflict, always follow the manufacturer’s labeling/instructions. The user holds BPP blameless of any and all liability arising from the use of this document.

P a g e | 17

TECH READY document

Key for Colored Boxes – Sterilization Methods Steam ETO High-Level Disinfection STERRAD STERIS V-PRO STERIS SYSTEM-1E

Maintenance

Assembly

Source Document: DePuy_Synthes_Reusable_Medical_Devices_Instruments_Trays_Cases_DJ1305_rI_n_Non_Sterile_Impants_DSUS_SPN_0115_0706_201501.pdf RH201504 DISCLAIMER: The information in this document is for reference only and includes only a portion of the manufacturer’s labeling/instructions. With the exception of the titles and header information, the content corresponds word for word with the Manufacturers document. It does not supersede or alter the manufacturer’s labeling/instructions. The manufacturer’s labeling/instructions contains other important information that must be considered in order to properly reprocess these items. Use this guide only after thoroughly reviewing, understanding, and complying with all of the manufacturer’s labeling/instructions requirements. If there is a conflict, always follow the manufacturer’s labeling/instructions. The user holds BPP blameless of any and all liability arising from the use of this document.

P a g e | 18

TECH READY document

Key for Colored Boxes – Sterilization Methods Steam ETO High-Level Disinfection STERRAD STERIS V-PRO STERIS SYSTEM-1E

Sterilization

Source Document: DePuy_Synthes_Reusable_Medical_Devices_Instruments_Trays_Cases_DJ1305_rI_n_Non_Sterile_Impants_DSUS_SPN_0115_0706_201501.pdf RH201504 DISCLAIMER: The information in this document is for reference only and includes only a portion of the manufacturer’s labeling/instructions. With the exception of the titles and header information, the content corresponds word for word with the Manufacturers document. It does not supersede or alter the manufacturer’s labeling/instructions. The manufacturer’s labeling/instructions contains other important information that must be considered in order to properly reprocess these items. Use this guide only after thoroughly reviewing, understanding, and complying with all of the manufacturer’s labeling/instructions requirements. If there is a conflict, always follow the manufacturer’s labeling/instructions. The user holds BPP blameless of any and all liability arising from the use of this document.

_______________________________________________________________________________ Processing Synthes Reusable Medical Devices – Instruments, Instrument Trays and Cases

These recommendations are for processing Synthes reusable medical devices sold in North America. Synthes reusable medical devices include certain surgical instruments, instrument trays and cases. The information provided does not apply to Synthes implants. These recommendations are to be followed unless otherwise noted on specific product inserts.

Cautions

Limits on reprocessing

Do not use steel wool or abrasive cleaners. Avoid solutions containing iodine or high chlorine content. Soiled or used Synthes devices should not be loaded in a case and cleaned in a mechanical washer. Synthes devices must be cleaned separately from Synthes instrument trays and Synthes cases. Soiled devices are devices that have blood, tissue and/or bodily fluid/matter in or on the surface of the devices. Long, narrow cannulations, blind holes and intricate parts require particular attention during cleaning. All devices must be thoroughly cleaned. Synthes instruments are critical devices and must be terminally sterilized prior to use. The sterilization parameters are only valid for devices that are adequately cleaned. Do not stack trays of instruments in a mechanical washer. See Sterilization section for Immediate-Use Steam Sterilization instructions. Immediate-Use Steam Sterilization is only intended for individual instruments. Synthes does not support or recommend sterilizing loaded cases or implants using this method. The following parameters are only valid for properly installed, maintained, calibrated and compliant reprocessing equipment. Cleaning agents with a pH within 7 – 9 are recommended. The recommended cleaning method for Power tools is manual. Do not place Power tools in an ultrasonic cleaner. Do not submerge Power tools in aqueous solutions. Refer to product specific literature for care of Power tools. The sterilization parameters cannot be used for the Synthes Power Drive Unit, PN: 530.100, Power Drive Set, PN: 105.957, Synthes Piezoelectric Handpiece, PN: 05.001.401 and the Synthes Piezoelectric System, PN: 68.001.400. For the Synthes Power Drive Unit and Power Drive Unit Set, the Synthes Piezoelectric Handpiece and Piezoelectric System, please refer to the User Manuals for Sterilization guidelines. Surgical patients identified as at-risk for Creutzfeldt-Jakob disease (CJD) and related infections should be treated with single-use instruments. Dispose of instruments used or suspected of use on a patient with CJD after surgery and/or follow current national recommendations. Repeated processing cycles that include ultrasonic, mechanical washing and sterilization have minimal effects on Synthes surgical instrumentation. End of life of a device is normally determined by wear and damage due to use. Evidence of damage and wear on a device may include but is not limited to corrosion (i.e. rust, pitting), discoloration, excessive scratches, flaking, wear and cracks. Improperly functioning devices, devices with unrecognizable markings, missing or removed (buffed off) part numbers, damaged and excessively worn devices should not be used.

Clinical Processing Instructions

Point of Use Care

Containment and Transportation

DJ1305 Rev I

Wipe blood and/or debris from device throughout surgical procedure to prevent it from drying onto the surface. Flush cannulated devices with sterile or purified water to prevent the drying of soil and/or debris to the inside. Soiled devices should be separated from non-contaminated devices to avoid contamination of personnel or surroundings. Devices should be covered with a towel dampened with sterile or purified water to prevent blood and/or debris from drying. Soiled devices should be transported separate from non-contaminated devices to avoid contamination.

Page 1 of 6

_______________________________________________________________________________ It is recommended that devices should be reprocessed as soon as is reasonably practical following use. Disassemble device, if device is able to be disassembled, prior to cleaning. Refer to technique guide or other supplemental information for specific device disassembly and/or reassembly instructions. Open devices with ratchets, box locks or hinges. Remove sharp devices for manual cleaning or place into a separate tray. Lumens/cannula of devices should be manually processed prior to cleaning. Lumens/cannula should first be cleared of debris. Lumens/cannula should be brushed thoroughly using appropriately sized soft-bristled brushes and twisting action. Brushes should be tight-fitting. Brush size should be approximately the same diameter of the lumen/cannulation to be cleaned. Using a brush that is too big or too small for the diameter of the lumen/cannulation may not effectively clean the surface of a lumen/cannulation. After brushing lumens/cannula, blow clean compressed air through lumen/cannulation to clear debris, if necessary. Soak and/or rinse heavily soiled devices or cannulated devices prior to cleaning to loosen any dried soil or debris. Use a neutral pH enzymatic soak or detergent to soak devices. Follow the enzymatic cleaner or detergent manufacturer’s instructions for use for correct exposure time, temperature, water quality and concentration. Use cold tap water to rinse devices. Synthes devices must be cleaned separately from Synthes instrument trays and Synthes cases. Lids should be removed from cases for the cleaning process, if applicable.

Preparation for Decontamination

Equipment: various sized soft-bristled brushes, lint-free cloths, syringes, pipettes and/or water jet, neutral enzymatic cleaner or neutral detergent with a pH between 7 and 9. 1. 2. 3. Cleaning – Manual Method

4. 5.

6. 7. 8. 9.

DJ1305 Rev I

Disassemble device, if device is able to be disassembled, prior to cleaning. Refer to technique guide or other supplemental information for specific device disassembly and/or reassembly instructions. Rinse soiled device under running cold tap water for a minimum of two minutes. Use a soft-bristled brush to assist in the removal of gross soil and debris. Soak device in a neutral pH enzymatic cleaner or detergent solution for a minimum of ten minutes. Follow the enzymatic cleaner or detergent manufacturer’s instructions for use for correct exposure time, temperature, water quality and concentration. Rinse device with cold water for a minimum of two minutes. Use a syringe, pipette, or water jet to flush lumens, channels and other hard to reach areas. Manually clean device for a minimum of five minutes in a freshly prepared neutral pH enzymatic cleaner or detergent solution. Use a soft-bristled brush to remove soil and debris. Actuate joints, handles and other movable device features to expose all areas to the detergent solution, if applicable. Clean device under water to prevent aerosolization of contaminants. Note: fresh solution is a newly- made, clean solution. Rinse device thoroughly with deionized (DI) or purified (PURW) water for a minimum of two minutes. Use a syringe, pipette or water jet to flush lumens and channels. Actuate joints, handles and other moveable device features in order to rinse thoroughly under running water, if applicable. Visually inspect device. Repeat the manual cleaning procedure (steps 2- 6) until no visible soil remains on device. Perform a final rinse on device using DI or PURW water. Dry device using a clean, soft, lint-free cloth or clean compressed air.

Page 2 of 6

_______________________________________________________________________________

Equipment: ultrasonic cleaner, various sized soft-bristled brushes, lint-free cloths, syringes, pipettes and/or water jet, neutral enzymatic cleaner or neutral detergent with a pH between 7 and 9.

Cleaning – Mechanical Method: Ultrasonic

Pre-clean method (Pre-clean method must be performed prior to ultrasonic mechanical method listed below.) 1. Disassemble device, if device is able to be disassembled, prior to cleaning. Refer to technique guide or other supplemental information for specific device disassembly and/or reassembly instructions. 2. Rinse soiled device under running cold tap water for a minimum of two minutes. Use a soft-bristled brush to assist in the removal of gross soil and debris. 3. Soak device in a neutral pH enzymatic cleaner or detergent solution for a minimum of ten minutes. Follow the enzymatic cleaner or detergent manufacturer’s instructions for use for correct exposure time, temperature, water quality and concentration. 4. Rinse device with cold water for a minimum of two minutes. Use a syringe, pipette, or water jet to flush lumens, channels and other hard to reach areas. 5. Manually clean device for a minimum of five minutes in a freshly prepared neutral pH enzymatic cleaner or detergent solution. Use a soft-bristled brush to remove soil and debris. Actuate joints, handles and other movable device features to expose all areas to the detergent solution, if applicable. Clean device under water to prevent aerosolization of contaminants. Note: fresh solution is a newly- made, clean solution. 6. Rinse device thoroughly using cold or warm tap water for a minimum of two minutes. Use a syringe, pipette or water jet to flush lumens and channels. Actuate joints, handles and other moveable device features in order to rinse thoroughly under running water, if applicable. 7. Visually inspect device. Repeat steps 2- 6 until no visible soil remains on device. Ultrasonic process: (Pre-cleaning steps 1 -7 should occur prior to this step.) 8. Prepare a fresh detergent solution using a neutral pH enzymatic cleaner or detergent. Follow the enzymatic cleaner or detergent manufacturer’s instructions for use for correct exposure time, temperature, water quality and concentration. Note: fresh solution is a newly- made, clean solution. 9. Clean Synthes device ultrasonically for a minimum of 15 minutes, using a minimum frequency of 40 KHz. 10. Rinse device thoroughly with deionized (DI) or purified (PURW) water for a minimum of two minutes. Use a syringe, pipette or water jet to flush lumens and channels. Actuate joints, handles and other moveable device features in order to rinse thoroughly under running water, if applicable. 11. Visually inspect device. Repeat steps 2- 10 until no visible soil remains on device. 12. Perform a final rinse on device using DI or PURW water for a minimum of 15 seconds. 13. Dry device using a clean, soft, lint-free cloth or clean compressed air.

DJ1305 Rev I

Page 3 of 6

_______________________________________________________________________________ Equipment: Ultrasonic cleaner, washer/disinfector, various sized soft-bristled brushes, lint-free cloths, syringes, pipettes and/or water jet, neutral enzymatic cleaner or neutral detergent with a pH between 7 and 9. Note: Ultrasonic cleaning may cause further damage to devices that have prior surface damage.

Cleaning -- Mechanical Method: Mechanical washer

Pre-clean method (Pre-clean method must be performed prior to mechanical washer method listed below.) 1. Disassemble device, if device is able to be disassembled, prior to cleaning. Refer to technique guide or other supplemental information for specific device disassembly and/or reassembly instructions. 2. Rinse soiled device under running cold tap water for a minimum of one minute. Remove gross soil using a soft bristled brush or soft, lint-free cloth. 3. Manually clean device for a minimum of two minutes in a freshly prepared neutral pH enzymatic or detergent solution. Follow the enzymatic cleaner or detergent manufacturer’s instructions for the correct dilution, temperature, water quality and exposure time. Use a soft-bristled brush to remove soil and debris. Actuate joints, handles and other movable device features to expose all areas to detergent solution, if applicable. Clean device under water to prevent aerosolization of contaminants. Note: fresh solution is a newly- made, clean solution. 4. Rinse device using cold to lukewarm running tap water for a minimum of one minute. Use a syringe, pipette or water jet to flush lumens and channels. Actuate joints, handles and other moveable device features in order to rinse thoroughly under running water, if applicable. 5. Prepare a fresh detergent solution using a neutral pH enzymatic cleaner or detergent. Follow the enzymatic cleaner or detergent manufacturer’s instructions for the correct dilution, temperature, water quality and exposure time. Note: fresh solution is a newly- made, clean solution. 6. Clean Synthes devices ultrasonically for a minimum of 15 minutes, using a minimum frequency of 40 KHz. 7. Rinse device using DI or PURW water for a minimum of two minutes. Use a syringe, pipette or water jet to flush lumens and channels. DI or PURW water must be used for final rinse. 8. Visually inspect device. Repeat steps 2 - 7 until no visible soil remains on device. Mechanical Washer process: (Pre-cleaning steps 1 -8 should occur prior to this step.) 9. .

Process device using the following cycle parameters:

Pre-wash

Minimum Time (minutes) 2

Minimum Temperature/Water Cold tap water

Wash I

2

Cold to warm tap water

Wash II

5

Warm tap water ( >40°C)

Rinse Dry

2 40

Warm DI or PURW (>40°C) 90°C

Cycle

Thermal disinfection

Drying

Inspection

DJ1305 Rev I

Type of Detergent N/A Neutral enzymatic pH between 7 and 9 Detergent with pH between 7 and 9 N/A N/A

For automated cleaning, thermal disinfect at 93°C for a minimum of 2 minutes and 30 seconds. For devices with cannula or lumens, orient the part such that the lumen or cannulation is in a vertical position. If this is not possible due to space limitations within the automated/mechanical washer, use an irrigating rack /load carrier with connections designed to ensure an adequate flow of process fluids to the lumen or cannulation of th e device if necessary. If a dry cycle is not included in the mechanical washer or if the device is not processed in a mechanical washer: Dry each device thoroughly inside and out to prevent rusting and malfunction. Use a clean, soft, lint-free cloth to avoid damage to the surface. Pay special attention to threads, ratchets and hinges or areas where fluid can accumulate. Open and close devices so that all areas are reached. Dry hollow parts using an air jet with clean compressed air. Synthes instruments should be inspected after processing, prior to sterilization, for: Cleanliness Damage, including but not limited to, corrosion (rust, pitting), discoloration, excessive scratches, flaking, cracks and wear Proper function, including but not limited to, sharpness of cutting tools, bending of flexible devices, movement of hinges/joints/box locks and moveable features such as handles, ratcheting and couplings and Missing or removed (buffed off ) part numbers Improperly functioning devices, devices with unrecognizable markings, missing or removed (buffed off) part numbers, damaged and worn devices should not be used. Disassembled devices should be reassembled prior to sterilization unless otherwise noted.

Page 4 of 6

_______________________________________________________________________________ Only legally marketed, FDA-cleared sterilization barriers (e.g. wraps, pouches or containers) should be used by the end-user for packaging terminally sterilized devices.

Packaging

Rigid Sterilization Container Use Instructions and Considerations In order to ensure proper sterilization of Synthes’ devices when using a rigid sterilization container, the following must be taken into consideration: o Only FDA-cleared rigid sterilization containers may be used with Synthes’ devices and loaded graphic cases (a graphic case with all or part of its assigned contents). o The rigid sterilization container manufacturer’s instructions for use are to be followed. If questions arise regarding the use of the rigid sterilization container, Synthes recommends contacting the manufacturer of that specific container for guidance. o The options in using rigid sterilization containers with Synthes’ devices and loaded graphic cases are as follows:  No more than one (1) fully loaded graphic case can be placed directly into a rigid sterilization container.  Instrument trays from no more than one (1) loaded graphic case can be placed in the rigid sterilization container.  Stand-alone modules/racks or single devices must be placed, without stacking, in a container basket to ensure optimal ventilation. o Rigid sterilization container must have a maximum volume to vent ratio of no greater than 3 2 127in /in . For any questions related to the volume to vent ratio, please contact the container manufacturer. o Only rigid sterilization containers approved for pre-vacuum steam sterilization can be used with Synthes’ devices and loaded graphic cases following the sterilization parameters provided by Synthes. o End-users should follow ANSI/AAMI ST79 for additional information concerning the use of rigid sterilization containers.

The following are the recommendations for the sterilization of Synthes devices:

Sterilization

Cycle Type

Minimum Sterilization Exposure Time (minutes)

Minimum Sterilization Exposure Temperature

Minimum Dry Time*

Prevacuum

4

132°C (270°F)

20 minutes

*When applying dry times to Synthes cases and their accessories, dry times outside the standard healthcare prevacuum parameters may be required. This is especially important for polymer-based (plastic) cases/trays used in conjunction with heavy duty nonwoven sterilization wraps. The current recommended dry times for Synthes cases can range from a standard 20 minutes to an extended time of 60 minutes. The dry time is most often influenced by the presence of polymer based (plastic) materials; therefore, changes such as elimination of silicone mats and/or change in sterile barrier system (e.g. heavy grade to light grade wrap or the use of rigid sterilization containers) can reduce the necessary dry time. Dry times may be highly variable due to differences in packaging materials (e.g. nonwoven wraps), environmental conditions, steam quality, device materials, total mass, sterilizer performance and varying cool down time. The user should employ verifiable methods (e.g. visual inspections) to confirm adequate drying. The autoclave manufacturer’s operating instructions and recommended guidelines for maximum sterilization load should be followed. The autoclave must be properly installed, maintained, and calibrated. Only legally marketed, FDA-cleared sterilization barriers (e.g. wraps, pouches or containers) should be used by the end-user for packaging terminally sterilized devices. Most devices are designed to be sterilized assembled unless: o Etched with "Disassemble for Sterilization"; o The graphic case is not configured for the assembled device or o According to instructions in product specific information. Immediate-Use Steam Sterilization is only intended for individual instruments. Synthes does not support sterilizing loaded graphic cases or implants using this method. The following parameters are to be used when using Immediate-use steam sterilization: o Unwrapped instrument o 4 (four) minute exposure o A minimum 3 (three) pulse fractionated prevacuum cycle o 132 °C (270 °F) Exceptions listed in the Cautions section of this document are not intended for Immediate-Use Steam Sterilization. Refer to the technique guide or other supplemental instruction to determine if device needs to be disassembled for sterilization. Immediate-Use Steam Sterilization should be performed in accordance with current AORN and AAMI recommendations.

DJ1305 Rev I

Page 5 of 6

_______________________________________________________________________________ Storage

Additional Information

Manufacturer Contact

Sterilized products should be stored in a dry, clean environment, protected from direct sunlight, pests, and extremes of temperature and humidity. Cleaning Agent Information: Synthes used the following cleaning agents during validation of these reprocessing recommendations. These cleaning agents are not listed in preference to other available cleaning agents which may perform satisfactorily – neutral pH enzymatic detergents (e.g. Prolystica 2X Concentrate Enzymatic Cleaner, Enzol, Endozime and Neodisher Medizym) and neutral pH detergents (e.g. Prolystica 2X Neutral Detergent). The cleaning and sterilization information is provided in accordance with ANSI/AAMI ST81, ISO 17664, AAMI TIR 12, ANSI/AAMI/ISO 17665-1, ANSI/AAMI ST79 and AAMI ST77. The recommendations provided above have been validated by the medical device manufacturer as being capable of preparing a non-sterile Synthes medical device. It remains the responsibility of the processor to ensure that the processing is actually performed, using equipment, materials and personnel in the reprocessing facility, and achieves the desired result. This requires validation and routine monitoring of the process. Likewise, any deviation by the processor from the recommendations provided should be properly evaluated for effectiveness and potential adverse consequences. All users should be qualified personnel with documented expertise, competency and training. Users should be trained on hospital policies and procedures along with current applicable guidelines and standards. Users should don appropriate personal protective equipment (PPE) when processing devices in accordance with the Department of Environmental and Occupational Health and Safety’s (OSHA) bloodborne pathogen guidelines. For further information, contact DePuy Synthes Customer Service Department at 1.800.523.0322.

DePuy Synthes, 1301 Goshen Parkway, West Chester, PA 19380

DJ1305 Rev I

Page 6 of 6

DePuySynthes ~~™ COMPANIES OF

~•gotMM'OK

Processing Non-sterile Synthes Implants These recommendations are for processing non-sterile Synthes implants that are sold in North America. The information provided applies to Synthes implants only. Explanted Synthes implants must never be reprocessed and should be handled according to hospital protocol upon removal. Any implant that has not been used , but has become soiled, should be handled according to hospital protocols. Refer to implant's product insert for any specific processing, cleaning, and sterilization instructions for that implant.

• • • • • • • • •

Cautions

• • •

Limits on reprocessing



Synthes does not recommend processing used implants . A used implant is an implant that has come into direct contact with a patient. Synthes implants should not be lubricated . Do not use a Synthes implant if the surface has been damaged . Do not use steel wool or abrasive cleaners on Synthes implants . In accordance with the CDC, AORN , and AAMI guidelines, Synthes does not recommend or support immediateuse steam sterilization of implants. Cleaning agents with a pH within 7-9 are recommended . Synthes implants are critical devices and must be terminally sterilized prior to use . The sterilization parameters are only valid for devices that are adequate ly cleaned . The following parameters are only val id for properly installed, maintained, calibrated, and compliant reprocessing equipment. Surgical patients identified as at-risk for Creutzfeldt-Jakob disease (CJD) and related infections should be treated with single-use instruments. Properly dispose of instruments used, or suspected of use, on a patient with CJD after surgery and/or follow current national recommendations. Repeated processing cycles that include ultrasonic, mechanical washing, and sterilization have minimal effects on Synthes implants. Synthes implants should be inspected for corrosion, damage such as scratches and notches, debris, discoloration, or residue . Any implant with corrosion, scratches, notches, residue, or debris should be discarded .

Processing Instructions

• Point of Use Care Containment and Transportation Preparation for Processing

• • •

Synthes does not recommend processing used implants. A used implant is an implant that has come into direct contact with a patient. Equipment: various sized soft-bristled brushes, lint-free cloths, syringes, pipettes and/or water jet, neutral enzymatic cleaner or neutral detergent with a pH between 7 and 9. I 2.

3. Processing- Manual

4.

Method

5. 6. 7. 8.

DSUS/SPN/O115/0706

Implants should remain covered until needed to avoid becoming soi led or contaminated with blood, tissue, and/or bodily fluids/matter. Only those to be implanted should be handled. Minimal handling of implants is necessary to prevent damage to the surface . Avoid cross contamination of implants with soiled instruments during transport.

Rinse device under running cold tap water for a minimum of two minutes. Use a soft-bristled brush to clean the device. Soak device in a neutral pH enzymatic cleaner or detergent solution for a minimum often minutes. Follow the enzymatic cleaner or detergent manufacturer' s instructions for use for correct exposure time, temperature, water quality, and concentration. Rinse device with cold water for a minimum of two minutes. Use a syringe, pipette, or water jet to flush lumens, channels, and other hard to reach areas. Manually clean device for a minimum of five minutes in a freshly prepared neutral pH enzymatic cleaner or detergent solution using a soft-bristled brush. Clean device under water to prevent aerosolization of contaminants. Note: freshly prepared solution is a newly-made, clean solution. Rinse device thoroughly with deionized (DI) or purified (PURW) water for a minimum of two minutes. Use a syringe, pipette, or water jet to flush lumens and channels. Visually inspect device. Perform a final rinse on device using DI or PURW water. Dry device using a clean, soft, lint-free cloth or clean compressed air.

January 2015

Page 1of4

Equipment: Ultrasonic cleaner, various sized soft-bristled brushes, lint-free cloths, syringes, pipettes and/or water jet, neutral enzymatic cleaner or neutral detergent with pH between 7 and 9.

ProcessingMechanical MethodUltrasonic

Pre-clean method (Pre-clean method must be performed prior to ultrasonic mechanical method listed below.) I. Rinse device under running cold tap water for a minimum of two minutes. Use a soft-bristled brush to clean the device. 2. Soak device in a neutral pH enzymatic cleaner or detergent solution for a minimum often minutes. Follow the enzymatic cleaner or detergent manufacturer's instructions for use for correct exposure time, temperature, water quality and concentration. 3. Rinse device with cold water for a minimum of two minutes. Use a syringe, pipette, or water jet to flush lumens, channels, and other hard to reach areas. 4. Manually clean device for a minimum of five minutes in a fresh ly prepared neutral pH enzymatic cleaner or detergent solution using a soft-bristled brush. Clean device under water to prevent aerosolization of contaminants. Note: freshly prepared solution is a newly made, clean solution. 5. Rinse device thoroughly using co ld or warm tap water for a minimum of two minutes. Use a syringe, pipette, or water jet to flush lumens and channels. 6. Visually inspect dev ice. Ultrason ic process: (Pre-c leaning steps 1-7 should occur prior to this step.) 7.

8. 9. 10. 11. 12.

Prepare a fresh detergent solution using a neutral pH enzymatic cleaner or detergent Foll ow the enzymatic cleaner or detergent manufacturer 's instructions for use for correct exposure time, temperature, water quality, and concentration. Note: freshly prepared solution is a newly made, clean solution. Clean Synthes device ultrasonically for a minimum of 15 minutes. Rinse device thoroughly with deionized (DI) or purified (PURW) water for a minimum of two minutes. Use a syringe, pipette, or water jet to flush lumens and channels. Visually inspect device. Perform a final rinse on dev ice using DI or PURW water for a minimum of 15 seconds. Dry device using a clean soft, lint-free cloth or clean compressed air.

Equipment: Ultrason ic cleaner, washer/disinfector, various sized soft-bristled brushes, lint-free cloths, syringes, pipettes, and/or water jet, neutral enzymatic cleaner or neutral detergent with a pH between 7 and 9.

ProcessingMec hanical MethodMec hanical Washer

Pre-clean method (Pre-clean method must be performed prior to mechanical washer method li sted below.) 1 Rinse device under running cold tap water for a minimum of one minute. 2. Manually clean device for a minimum of two minutes in a fres hly prepared neutral pH enzymatic or detergent solution using a soft-bristled brush. Follow the enzymatic cleaner or detergent manufacturer ' s instructions for the correct dilution, temperature, water quality, and exposure time. Clean device under water to prevent aerosolization of contaminants. Note: freshly prepared solution is a newly made, clean solution. 3. Rinse device using cold to lukewarm running tap water for a minimum of one minute. Use a syringe, pipette, or water jet to flush lumens and channels. 4. Prepare a fresh detergent solution using a neutral pH enzymatic cleaner or detergent Foll ow the enzymatic cleaner or detergent manufacturer's instructions for the correct dilution, temperature, water quality, and exposure time. Note: freshly prepared solution is a newly made, clean solution. 5. Clean Synthes devices ultrasonically for a minimum of 15 minutes. 6. Rinse device using DI or PURW water for a minimum of two minutes. Use a syringe, pipette, or water jet to flush lumens and channels. DI or PURW water must be used for final rinse. 7. Visually inspect device. Mechan ical Washer process: (Pre-cl ean ing steps 1-8 should occur prior to this step.) 8.

DSUS/SPN/O11 5/0706

Process device using the fo llowing cycle parameters: Cycle

Minimum Time (minutes)

Minimum Temperature/\Nater

Pre-wash

2

Cold tap water

Wash I

2

Cold to warm tap water

Wash II Rinse Dry

5 2 40

Warm tap water (>40°C) Warm DI or PURW (>40°C) 90°C

January 2015

Type of Detergent

NIA Neutral enzymatic pH between 7 and 9 Detergent with pH between 7 and 9

NIA NIA

Page 2 of 4

Thermal disinfection Inspection

Packaging

For automated cleaning, thermal disinfect at 93°C for minimum of2 minutes and 30 seconds.

• • •

Synthes implants should be inspected after processing, prior to steri lization . Any implant with corrosion, discoloration, scratches, fl aws, residue or debris should be discarded . Only legally marketed, FDA-cleared sterilization barriers (e.g. wraps, pouches, or containers) should be used by the end-user for packaging terminally sterilized devices.



Ri2id Sterilization Container Use Instructions and Considerations In order to ensure proper steri lization of Synthes ' devices when using a rigid steri lization container, the following must be taken into cons ideration: Only FDA-cleared rigid sterilization containers may be used with Synthes' devices and loaded graphic 0 cases (a graphic case with all or part of its assigned contents). The rigid steri lization container manufacturer' s instructions for use are to be fo ll owed. If questions arise 0 regarding the use of the rigid steri lization container, Synthes recommends contacting the manufacturer of that specific container for gu idance. The options in using rigid steri lization containers with Synthes ' devices and loaded graphic cases are as 0 follows: No more than one (1) fully loaded graphic case can be placed into a rigid steri lization container. Instrument trays from no more than one (1) loaded graphic case can be placed in the rigid sterilization container. Stand-alone modules/racks or single devices must be placed, without stacking, in a container basket to ensure optimal venti lation. Ri2id sterilization conta iner must have a maximum volume to vent ratio of no 2reater than 127 0 in 3 /in 2 • For any guestions related to the volume to vent ratio glease contact the container manufacturer. Only rigid steri lization containers approved for pre-vacuum steam steri lization can be used with 0 Synthes ' devices and loaded graph ic cases following the sterilization parameters provided by Synthes. End-users shou ld follow ANSI/AAMI ST79 for additional information concern ing the use of rigid 0 sterilization containers.

. . .

The following are the recommendations for the sterilization of Synthes' implants:

Sterilization

DSUS/SPN/O11 5/0706

Cycle Type

Minimum Sterilization Exposure Time (minutes)

Minimum Sterilization Exposure Temperature

Minimum Dry Time*

Prevacuum

4

l 32°C (270°F)

20 minutes

*When applying dry times to Synthes cases and their accessories, dry times outside the standard healthcare prevacuum parameters may be required. Th is is especially important for polymer-based (plastic) cases/trays used in conjunction with heavy duty nonwoven steri lization wraps. The current recommended dry times for Synthes cases can range from a standard 20 minutes to an extended time of 60 minutes. The dry time is most often influenced by the presence of polymer based (plastic) materials; therefore, changes such as elimination of si licone mats and/or change in steril e barrier system (e .g. heavy grade to light grade wrap or the use ofrigid sterilization containers) can reduce the necessary dry time. Dry times may be highly variable due to differences in packaging materials (e.g. nonwoven wraps), environmental conditions, steam quality, implant materials, total mass, steri lizer performance, and varying cool down time . The user shou ld employ verifiable methods (e.g. visual inspections) to confirm adequate drying. • The autoclave manufacturer's operating instructions and recommended guidelines for maximum sterilization load should be followed . The autoc lave must be properly installed, maintained, and calibrated. Only legally marketed, FDA-cleared steri lization barriers (e.g. wraps, pouches, or containers) shou ld be used by the end-user for packaging terminally sterilized devices. • For product sold steri le, refer to device specific insert for limitations regarding resterilization .

January 2015

Page 3 of 4



Add itional Information

Manufacturer Contact

Sterile product that has exceeded the labeled expiration date and has intact and undamaged original packaging may be sterilized using the above sterilization parameters unless product is coated, i.e HA, plasma, etc. or resterilization is not permitted. Refer to device specific insert for limitations regarding resterilization. • Cleaning Agent Information: Synthes used the following cleaning agents during validation of these reprocessing recommendations. These cleaning agents are not listed in preference to other available cleaning agents which may perform satisfactori ly- neutral pH enzymatic detergents (e.g. Prolystica 2X Concentrate Enzymatic Cleaner, Enzol, Endozime, and Neodisher Medizym) and neutral pH detergents (e.g. Prolystica 2X Neutral Detergent). • The cleaning and sterilization information is provided in accordance with ANSI/AAMI ST8 l , ISO 17664, AAMI TIR 12, ANSI/AAMI/ISO 17665-1 , ANSI/AAMI ST79, and AAMI ST77. • The recommendations provided above have been validated by the medical device manufacturer as being capable of preparing a non-sterile Synthes medical implant It remains the responsibility of the processor to ensure that the processing is actually performed, using equipment, materials, and personnel in the reprocessing faci lity, and achieves the desired result. This requires validation and routine monitoring of the process. Likewise, any deviation by the processor from the recommendations provided should be properly evaluated for effectiveness and potential adverse consequences. • All users should be qualified personnel with documented expertise, competency, and training. Users should be trained on hospital policies and procedures along with current applicable guidelines and standards. • Users should don appropriate personal protective equipment (PPE) when processing implants in accordance with the Department of Environment and Occupational Health and Safety ' s (OSHA) bloodborne pathogen gu idelines For further information, contact DePuv Synthes Customer Service Department at 1.800.523 .0322.

Manufactured by (United States) Synthes USA Products, LLC 1302 Wrights Lane East West Chester, PA 19380 Telephone (610) 719-5000 To order (800) 523-0322 Fax: (610) 251-9056 Legal manufacturer (Canada): Synthes (Canada) Ltd. 2566 Meadowpine Blvd. Mississauga, Ontario L5N 6P9 Telephone (905) 567-0440 To order (800) 668-111 9 Fax: (905) 567-3185

DSUS/SPN/O11 5/0706

January 2015

Page 4 of 4